  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75081",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AZ, CA, CT, FL, IL, MA, MS, MO, NC, PR, TX, WI.    OUS: Algeria, Australia, Bangladesh, Belgium, Brazil, Canada, Colombia, Dominican republic, Egypt, France, Germany, Ghana, Guatemala, HONG KONG, Hungary, India, Indonesia, Iran, Italy, Jamaica, Japan, Korea, Kuwait, LEBANON, Lithuania, Maldives, Martinique, Mexico, MOROCCO, Netherlands, New Zealand, OMAN, Pakistan, Philippines, Poland, Portugal, Romania, SAUDI ARABIA, Singapore, South Africa, Spain,Sweden, Tanzania, Thailand,   Turkey, United Arab Emirates, United Kingdom, Vietnam.",
      "recall_number": "Z-2731-2016",
      "product_description": "GE Healthcare SIGNA Creator, SIGNA Explorer.  Diagnostic Imaging Device.",
      "product_quantity": "383 (US = 24;OUS = 359)",
      "reason_for_recall": "GE Healthcare is conducting a recall dealing with  an incorrect RF Power Monitor Function  involving  SIGNA CREATOR and SIGNA EXPLORER which  could result in higher than expected thermal dose to the patient and higher than expected localized heating. There have been no failures of this type reported, and no injuries reported as a result of this issue.",
      "recall_initiation_date": "20160831",
      "center_classification_date": "20160902",
      "termination_date": "20180731",
      "report_date": "20160914",
      "code_info": "Mfg Lot or Serial # System ID  MNEXP1600002MR 480425MR7  001709TA5                    520ASRMR  003175TA7                    916RAFMR  002448TA9                    559449SPMR  002674TA0                    559449SPMR  002027TA1                   203337MREXP  003012TA2                  GON: 4371049  002974TA4                 239458MR1  003399TA3                239292MR1  002380TA4                305983MR3  002753TA2                727869MR  003356TA3                954473MR1  003176TA5              773843SCMR  000753TA4               574206MR  003013TA0               508746JMR2  003019TA7               417269MRS2  003371TA2               601853MR1  001027TA2               704384MINT  001059TA5 7            04384MINT  002453TA9              513569MR1  002741TA7             513584MR  001063TA7            254965CTOMR1  002879TA5            920560EXP  002831TA6            GON: 1532602  002023TA0           DZ1035MR03  002081TA8   DZ1863MR01  003302TA7           GON: 4399202  002742TA5   GON: 4374136  003076TA7   DZ1149MR02  003731TA7   DZ2014MR01  003079TA1   DZ1908MR01  003078TA3   DZ1174MR01  003565TA9   GON: 4345190  003134TA4   DZ1947MR01  002883TA7   83147183761  003174TA0   83140192745  003508TA9   83151589682  003403TA3   083027600486516  001232TA8   190073MR02  003439TA7   2626380  003441TA3   MRR10842  002687TA2   MRR10744  003440TA5   MRR10833  002090TA9   MRR10671  003793TA7   GON: 4393875  002949TA6   604988MRI  002219TA4   MRR10679  003212TA8   MRR10843  002876TA1   MRR10762  003177TA3   MRR10799  003754TA9   MRR10892  003359TA7   MRR5954  002885TA2    EG1826MR02  002450TA5                     EG1990MR01  002886TA0   EG2266MR01  001502TA4   EG1112MR02  003213TA6   EG1709MR01  003083TA3   EG1522MR01  003133TA6   EG1237MR01  003354TA8   EG1505MR02  002446TA3   EG1196MR01  002940TA5   EG1998MR01  002743TA3   EG1245MR03  003267TA2   EG1069MR02  003172TA4   GP1022MR02  001405TA0   X227040904  001124TA7   A1039630  001231TA0   X2881410  000898TA7   M4174549  001307TA8   M23084330  003307TA6   M1265006  003511TA3 M1265008  001029TA8 X102860306  001171TA8 M185296206  001169TA2 X756087020  003265TA6 M2569125  000973TA8 M8321415  001028TA0 X116370509  001503TA2 M2269855  001233TA6 M2533228  001170TA0 M2748222  001936TA4  M2322255  001238TA5 M8317617  001938TA0 M215631205  000331TA9 M386271501  001123TA9 M2718163  001237TA7 X751964008  000965TA4 M203097404  002022TA2 M4213545  001251TA8 M2316318  001026TA4 RE1017MR01  002975TA1 X868131014  003725TA9 A1039633  003604TA6 M2485629  003783TA8 L2202007  003266TA4 GP1052MR03  003563TA4 M1974139  003730TA9 A6023417  003306TA8 X453783201  003600TA4 B1081916  003728TA3 X2266305  003566TA7 X1020807  003218TA5 RE1029MR01  003478TA5 RE1044MR02  003739TA0 396399MR02  002755TA7 HC0285MR01  001982TA8 PC1401MR01  000971TA2 387462MR01  003729TA1 PC7479MR04  001234TA4 527556MR02  003178TA1 410725MR01  003435TA5 HC1711MR01  001981TA0 570231MR02  003722TA6  487692MR01  000149TA5 \"PC0345MR02  \"  003475TA1 TP0009MR01  003301TA9 GON: 4389578  001411TA8 MRR1051  002218TA6 MRR3881  002739TA1 0828270047  001616TA2 0828270045  003507TA1 0828270050  003512TA1 828270049  001932TA3 HU1099MR02  003023TA9 HU1243MR02  002976TA9 083027848946716  002970TA2 083027607457016  003024TA7  083027108785616  003602TA0 083027108786816  001230TA2 083027000633815  001478TA7 083027210062849  002579TA1 083027245187515  003397TA7 083027247170216  003513TA9 083027241541616  002375TA4 083027242389615  002444TA8 083027243696615  003510TA5 083027240320216  001979TA4 083027062669515  003443TA9 083027000026516  003077TA5 083027828631016  001604TA8 083027861954815  003014TA8 083027867369616  001599TA0 083027869960115  001058TA7 083027280565115  001476TA1 083027282542615  002443TA0 083027201528815  002880TA3 083027207170016  003438TA9 083027200144616  001600TA6 083027209389215  003298TA7 083027209980616  003401TA7 083027600155416  001602TA2 083027601581615  002082TA6 083027064149415  002452TA1 083027045280015  002084TA2 083027803313615  001504TA0 083027803032115  003605TA3 083027801497116  002576TA7 083027804757116  003437TA1 083027809326316  002824TA1 083027847477816  003016TA3 083027848271716  003561TA8 083027840049116  002679TA9 083027845987715  001603TA0 083027021866615  003481TA9 083027024406116  002882TA9 083049026692016  000368TA1 083027607006215  003724TA2 083027609969016  002972TA8 5512MR0007  003504TA8 GON: 4410359  003480TA1 GON: 4409805  002677TA3 A5390311  001931TA5 A402682101  001407TA6 A339043902  000160TA2 C208415204  001771TA5 A5194602  003173TA2 AE1557MR01  003782TA0 A5388205  003210TA2 A344729901  003667TA3 A456623203  002683TA1 A5697716  003778TA8 A5820152  003214TA4 B5738501  001235TA1 A389746501  003215TA1 A5978807  002942TA1 MRR10827  002971TA0 PC0815  003217TA7 YM1450  002024TA8 PC0013  001479TA5 PC1007  002270TA7  PC0904  002382TA0 YM3057  001710TA3 PC1012  003080TA9 PC0813  000159TA4 \"PC0801  \"  001408TA4 PC1010  001937TA2 PC1013  002091TA7 PC1015  002085TA9 PC6050  002384TA6 PC0804  003514TA7 PC1026  001636TA0  PC1008  002826TA6 PC0814  003303TA5 PC1027  002083TA4 PC0803  002756TA5 PC1018  002878TA7 PC1023  003017TA1 PC0816  001475TA3 PC0802  001635TA2 PC0805  002622TA9 PC0809  002676TA5 PC1017  002685TA6 PC1022  003209TA4 PC0909  001406TA8 PC1004  002376TA2 YM0834  001253TA4 PC1002  002877TA9 PC0812  003355TA5 YM1462  001477TA9 PC0902  002829TA0 YM1745  002828TA2 PC1019  002086TA7 YM0534  002654TA2 YM4111  001934TA9 PC1009  002738TA3 PC1021  002943TA9 PC0006  000969TA6 PC1001  001499TA3 PC1011  001500TA8 YM0847  002442TA2 PC0807  003179TA9 PC1025  000970TA4 PC0901  003562TA6 PC1028  003483TA5 PC1100  001412TA6 PC0808  003219TA3 YM1705  001104TA9 PC1003  003398TA5 PC0912  003473TA6 PC0913  002379TA6 PC0806  002827TA4 PC0811  002884TA5  PC1020  002968TA6 PC1024  001252TA6 PC0888  002678TA1 YM0668  002445TA5 PC1014  001236TA9 PC1005  002026TA3 PC0810  002945TA4 PC0908  002675TA7 YM0659  003018TA9 YM0662  002476TA0 PC0906  002596TA5 PC0907  002944TA7 PC0911  003085TA8 PC0910  001250TA0 PC1006  002623TA7 PC1016  000966TA2 PC0905  003723TA4 PC1036  003131TA0 PC1030  001935TA6 PC0903  002977TA7 0850270494  003601TA2 0850270476  002088TA3 0850270483  002087TA5 0850270474  002624TA5 0850270479  002969TA4 0850270493  003360TA5 0850270496  003264TA9 0850270495  003560TA0 0850270497  002830TA8 0850270208  001249TA2 0850270472  002441TA4 0850270485  003779TA6 0850270141  001598TA2 0850270480  001498TA5 0850270473  001933TA1 0850270484  003304TA3 KW1013MR02  003299TA5 LB1005MR01  003361TA3 LB1220MR01  001254TA2 LT4028MR01  003506TA3 LT4022MR02  001501TA6 083027001828115  001229TA4 MQ1038MR03  002754TA0 This order is cancelled  001708TA7  MA1465MR01  001497TA7 MA1581MR01  001410TA0 MA1574MR01  002328TA3 MA1148MR01  002948TA8 MA1527MR01  003396TA9  MA1516MR01  003436TA3 MA1540MR01  003780TA4 GON: 4374894  002832TA4 RTD0123  003081TA7 85147575570  002973TA6 OM1087MR01  003082TA5 OM1121MR01  003211TA0 PK1248MR01  001775TA6 PK1242MR01  003358TA9 GON: 4395723  003784TA6 0843270007  000972TA0 PL3500MR01  002025TA5 PL3544MR02  001409TA2 PL3548MR01  002680TA7 PL3623MR01  002740TA9 PL3696MR01  001308TA6 PL3547MR01  002825TA8 PL3594MR01  001601TA4 PL1063MR01  000366TA5 \"PL3362MR01  \"  002757TA3 PL3697MR01  002267TA3 PL1724MR01  003795TA2 PL3810MR01  002268TA1 PL3595MR01  002684TA9 PL1778MR01  002595TA7 PL3596MR01  003075TA9 PPR58104  001496TA9 MRR1448  002875TA3 MRR3155  003084TA1 MRR2728  003479TA3 GON: 4420297  002447TA1 RO1321MR01  002381TA2 RO1311MR01  003400TA9 RO1090MR02  003022TA1 RO1200MR01  003020TA5 RO1252MR01  003402TA5 GON: 4381507  003021TA3 RO1038MR01  002221TA0 SA2330MR01  002089TA1  SA2468MR01  002383TA8 SA2167MR01  003753TA1 GON: 4398438  003305TA0 SA2118MR01  001980TA2  6073MR0008  002681TA5 1573MR0004  002220TA2 6073MR0009  002682TA3 0847270018  003505TA5 6073MR0012  002271TA5 ZA2222MR02  002440TA6 YP170A26  001770TA7 CP749713  002266TA5 FPG40900  002377TA0 FPG97802  003777TA0 DPN29025  003726TA7 GON: 4424831  002451TA3 YP602101  002378TA8 DEL71943  003132TA8 NPV71300  003308TA4 GON: 4399370  003353TA0 GON: 4399371  002575TA9 N006MR01  002881TA1 A013MR02  003357TA1 TZ1052MR01  003509TA7 1573MR0008  003603TA8 1573MR0007  003482TA7 1573MR0009  002655TA9 1573MR0003  002686TA4 1573MR0005  003433TA0 02008MRS01  001141TA1 06167MRS01  002449TA7 06169MRS01  003434TA8 16003MRS01  003442TA1 25008MRS01  002439TA8 34416MRS01  002021TA4 46008MRS01  003015TA5 06172MRS01  003622TA8 33044MRS02  003781TA2 GON: 4410001  002947TA0 34419MRS01  003477TA7 34433MRS01  003474TA4 52013MRS01  002946TA2 53012MRS01  003794TA5 09018MRS01  003130TA2 GON: 4294813  001769TA9 IE1590MR01  001404TA3 10718MRS01  001140TA3 10719MRS01  002941TA3 70001MRM01  002269TA9 10724MRS01  003300TA1 10079MRM83  003216TA9 5973MR0009  003476TA9 5973MR0010",
      "more_code_info": ""
    }
  ]
}