      "country": "United States",
      "city": "Middleton",
      "address_1": "3150 Pleasant View Rd",
      "reason_for_recall": "Some of the Elite 2 and 3 Obstetric probes and 5 MHz Vascular Probes may exhibit excessive background noise that may mask the user\u0019s ability to clearly hear heart rate and/or blood flow.",
      "address_2": "",
      "product_quantity": "2,335 units total",
      "code_info": "Serial numbers: CAQ0100  CAQ0101  CAQ0102  CAQ0103  CAQ0104  CAQ0105  CAQ0106  CAQ0107  CAQ0108  CAQ0109  CAQ0110  CAQ0111  CAQ0112  CAQ0113  CAQ0114  CAQ0115  CAQ0116  CAQ0117  CAQ0118  CAQ0119  CAQ0120  CAQ0121  CAQ0122  CAQ0123  CAQ0124  CAQ0125  CAQ0126  CAQ0127  CAQ0128  CAQ0129  CAQ0130  CAQ0131  CAQ0132  CAQ0133  CAQ0134  CAQ0135  CAQ0136  CAQ0137  CAQ0138  CAQ0139  CAQ0140  CAQ0141  CAQ0142  CAQ0143  CAQ0144  CAQ0145  CAQ0146  CAQ0147  CAQ0148  CAQ0149  CAQ0150  CAQ0151  CAQ0152  CAQ0153  CAQ0154  CAQ0155  CAQ0156  CAQ0157  CAQ0158  CAQ0159  CAQ0160  CAQ0161  CAQ0162  CAQ0163  CAQ0164  CAQ0165  CAQ0166  CAQ0167  CAQ0168  CAQ0169  CAQ0170  CAQ0171  CAQ0172  CAQ0173  CAQ0174  CAQ0175  CAQ0176  CAQ0177  CAQ0178  CAQ0179  CAQ0180  CAQ0181  CAQ0182  CAQ0183  CAQ0184  CAQ0185  CAQ0186  CAQ0187  CAQ0188  CAQ0189  CAQ0190  CAQ0191  CAQ0192  CAQ0193  CAQ0194  CAQ0195  CAQ0196  CAQ0197  CAQ0198  CAQ0199  CAQ0200  CAQ0201  CAQ0202  CAQ0203  CAQ0204  CAQ0205  CAQ0206  CAQ0207  CAQ0208  CAQ0209  CAQ0210  CAQ0211  CAQ0212  CAQ0213  CAQ0214  CAQ0215  CAQ0216  CAQ0217  CAQ0218  CAQ0219  CAQ0220  CAQ0221  CAQ0222  CAQ0223  CAQ0224  CAQ0225  CAQ0226  CAQ0227  CAQ0228  CAQ0229  CAQ0230  CAQ0231  CAQ0232  CAQ0233  CAQ0234  CAQ0235  CAQ0236  CAQ0237  CAQ0238  CAQ0239  CAQ0240  CAQ0241  CAQ0242  CAQ0243  CAQ0244  CAQ0245  CAQ0246  CAQ0247  CAQ0248  CAQ0249  CAQ0250  CAQ0251  CAQ0252  CAQ0253  CAQ0254  CAQ0255  CAQ0256  CAQ0257  CAQ0258  CAQ0259  CAQ0260  CAQ0261  CAQ0262  CAQ0263  CAQ0264  CAQ0265  CAQ0266  CAQ0267  CAQ0268  CAQ0269  CAQ0270  CAQ0271  CAQ0272  CAQ0273  CAQ0274  CAQ0275  CAQ0276  CAQ0277  CAQ0278  CAQ0279  CAQ0280  CAQ0281  CAQ0282  CAQ0283  CAQ0284  CAQ0285  CAQ0286  CAQ0287  CAQ0288  CAQ0289  CAQ0290  CAQ0291  CAQ0292  CAQ0293  CAQ0294  CAQ0295  CAQ0296  CAQ0297  CAQ0298  CAQ0299  CAQ0300  CAQ0301  CAQ0302  CAQ0303  CAQ0304  CAQ0305  CAQ0306  CAQ0307  CAQ0308  CAQ0309  CAQ0310  CAQ0311  CAQ0312  CAQ0313  CAQ0314  CAQ0315  CAQ0316  CAQ0317  CAQ0318  CAQ0319  CAQ0320  CAQ0321  CAQ0322  CAQ0323  CAQ0324  CAQ0325  CAQ0326  CAQ0327  CAQ0328  CAQ0329  CAQ0330  CAQ0331  CAQ0332  CAQ0333  CAQ0334  CAQ0335  CAQ0336  CAQ0337  CAQ0338  CAQ0339  CAQ0340  CAQ0341  CAQ0342  CAQ0343  CAQ0344  CAQ0345  CAQ0346  CAQ0347  CAQ0348  CAQ0349  CAQ0350  CAQ0351  CAQ0352  CAQ0353  CAQ0354  CAQ0355  CAQ0356  CAQ0357  CAQ0358  CAQ0359  CAQ0360  CAQ0361  CAQ0362  CAQ0363  CAQ0364  CAQ0365  CAQ0366  CAQ0367  CAQ0368  CAQ0369  CAQ0370  CAQ0371  CAQ0372  CAQ0373  CAQ0374  CAQ0375  CAQ0376  CAQ0377  CAQ0378  CAQ0379  CAQ0380  CAQ0381  CAQ0382  CAQ0383  CAQ0384  CAQ0385  CAQ0386  CAQ0387  CAQ0388  CAQ0389  CAQ0390  CAQ0391  CAQ0392  CAQ0393  CAQ0394  CAQ0395  CAQ0396  CAQ0397  CAQ0398  CAQ0399  CAQ0400  CAQ0401  CAQ0402  CAQ0403  CAQ0404  CAQ0405  CAQ0406  CAQ0407  CAQ0408  CAQ0409  CAQ0410  CAQ0411  CAQ0412  CAQ0413  CAQ0414  CAQ0415  CAQ0416  CAQ0417  CAQ0418  CAQ0419  CAQ0420  CAQ0421  CAQ0422  CAQ0423  CAQ0424  CAQ0425  CAQ0426  CAQ0427  CAQ0428  CAQ0429  CAQ0430  CAQ0431  CAQ0432  CAQ0433  CAQ0434  CAQ0435  CAQ0436  CAQ0437  CAQ0438  CAQ0439  CAQ0440  CAQ0441  CAQ0442  CAQ0443  CAQ0444  CAQ0445  CAQ0446  CAQ0447  CAQ0448  CAQ0449  CAQ0450  CAQ0451  CAQ0452  CAQ0453  CAQ0454  CAQ0455  CAQ0456  CAQ0457  CAQ0458  CAQ0459  CAQ0460  CAQ0461  CAQ0462  CAQ0463  CAQ0464  CAQ0465  CAQ0466  CAQ0467  CAQ0468  CAQ0469  CAQ0470  CAQ0471  CAQ0472  CAQ0473  CAQ0474  CAQ0475  CAQ0476  CAQ0477  CAQ0478  CAQ0479  CAQ0480  CAQ0481  CAQ0482  CAQ0483  CAQ0484  CAQ0485  CAQ0486  CAQ0487  CAQ0488  CAQ0489  CAQ0490  CAQ0491  CAQ0492  CAQ0493  CAQ0494  CAQ0495  CAQ0496  CAQ0497  CAQ0498  CAQ0499  CAQ0500  CAQ0501  CAQ0502  CAQ0503  CAQ0504  CAQ0505  CAQ0506  CAQ0507  CAQ0508  CAQ0509  CAQ0510  CAQ0511  CAQ0512  CAQ0513  CAQ0514  CAQ0515  CAQ0516  CAQ0517  CAQ0518  CAQ0519  CAQ0520  CAQ0521  CAQ0522  CAQ0523  CAQ0524  CAQ0525  CAQ0526  CAQ0527  CAQ0528  CAQ0529  CAQ0530  CAQ0531  CAQ0532  CAQ0533  CAQ0534  CAQ0535  CAQ0536  CAQ0537  CAQ0538  CAQ0539  CAQ0540  CAQ0541  CAQ0542  CAQ0543  CAQ0544  CAQ0545  CAQ0546  CAQ0547  CAQ0548  CAQ0549  CAQ0550  CAQ0551  CAQ0552  CAQ0553  CAQ0554  CAQ0555  CAQ0556  CAQ0557  CAQ0558  CAQ0559  CAQ0560  CAQ0561  CAQ0562  CAQ0563  CAQ0564  CAQ0565  CAQ0566  CAQ0567  CAQ0568  CAQ0569  CAQ0570  CAQ0571  CAQ0572  CAQ0573  CAQ0574  CAQ0575  CAQ0576  CAQ0577  CAQ0578  CAQ0579  CAQ0580  CAQ0581  CAQ0582  CAQ0583  CAQ0584  CAQ0585  CAQ0586  CAQ0587  CAQ0588  CAQ0589  CAQ0590  CAQ0591  CAQ0592  CAQ0593  CAQ0594  CAQ0595  CAQ0596  CAQ0597  CAQ0598  CAQ0599  CAQ0600  CAQ0601  CAQ0602  CAQ0603  CAQ0604  CAQ0605  CAQ0606  CAQ0607  CAQ0608  CAQ0609  CAQ0610  CAQ0611  CAQ0612  CAQ0613  CAQ0614  CAQ0615  CAQ0616  CAQ0617  CAQ0618  CAQ0619  CAQ0620  CAQ0621  CAQ0622  CAQ0623  CAQ0624  CAQ0625  CAQ0626  CAQ0627  CAQ0628  CAQ0629  CAQ0630  CAQ0631  CAQ0632  CAQ0633  CAQ0634  CAQ0635  CAQ0636  CAQ0637  CAQ0638  CAQ0639  CAQ0640  CAQ0641  CAQ0642  CAQ0643  CAQ0644  CAQ0645  CAQ0646  CAQ0647  CAQ0648  CAQ0649  CAQ0650  CAQ0651  CAQ0652  CAQ0653  CAQ0654  CAQ0655  CAQ0656  CAQ0657  CAQ0658  CAQ0659  CAQ0660  CAQ0661  CAQ0662  CAQ0663  CAQ0664  CAQ0665  CAQ0666  CAQ0667  CAQ0668  CAQ0669  CAQ0670  CAQ0671  CAQ0672  CAQ0673  CAQ0674  CAQ0675  CAQ0676  CAQ0677  CAQ0678  CAQ0679  CAQ0680  CAQ0681  CAQ0682  CAQ0683  CAQ0684  CAQ0685  CAQ0686  CAQ0687  CAQ0688  CAQ0689  CAQ0690  CAQ0691  CAQ0692  CAQ0693  CAQ0694  CAQ0695  CAQ0696  CAQ0697  CAQ0698  CAQ0699  CAQ0700  CAQ0701  CAQ0702  CAQ0703  CAQ0704  CAQ0705  CAQ0706  CAQ0707  CAQ0708  CAQ0709  CAQ0710  CAQ0711  CAQ0712  CAQ0713  CAQ0714  CAQ0715  CAQ0716  CAQ0717  CAQ0718  CAQ0719  CAQ0720  CAQ0721  CAQ0722  CAQ0723  CAQ0724  CAQ0725  CAQ0726  CAQ0727  CAQ0728  CAQ0729  CAQ0730  CAQ0731  CAQ0732  CAQ0733  CAQ0734  CAQ0735  CAQ0736  CAQ0737  CAQ0738  CAQ0739  CAQ0740  CAQ0741  CAQ0742  CAQ0743  CAQ0744  CAQ0745  CAQ0746  CAQ0747  CAQ0748  CAQ0749  CAQ0750  CAQ0751  CAQ0752  CAQ0753  CAQ0754  CAQ0755  CAQ0756  CAQ0757  CAQ0758  CAQ0759  CAQ0760  CAQ0761  CAQ0762  CAQ0763  CAQ0764  CAQ0765  CAQ0766  CAQ0767  CAQ0768  CAQ0769  CAQ0770  CAQ0771  CAQ0772  CAQ0773  CAQ0774  CAQ0775  CAQ0776  CAQ0777  CAQ0778  CAQ0779  CAQ0780  CAQ0781  CAQ0782  CAQ0783  CAQ0784  CAQ0785  CAQ0786  CAQ0787  CAQ0788  CAQ0789  CAQ0790  CAQ0791  CAQ0792  CAQ0793  CAQ0794  CAQ0795  CAQ0796  CAQ0797  CAQ0798  CAQ0799  CAQ0800  CAQ0801  CAQ0802  CAQ0803  CAQ0804  CAQ0805  CAQ0806  CAQ0807  CAQ0808  CAQ0809  CAQ0810  CAQ0811  CAQ0812  CAQ0813  CAQ0814  CAQ0815  CAQ0816  CAQ0817  CAQ0818  CAQ0819  CAQ0820  CAQ0821  CAQ0822  CAQ0823  CAQ0824  CAQ0825  CAQ0826  CAQ0827  CAQ0828  CAQ0829  CAQ0830  CAQ0831  CAQ0832  CAQ0833  CAQ0834  CAQ0835  CAQ0836  CAQ0837  CAQ0838  CAQ0839  CAQ0840  CAQ0841  CAQ0842  CAQ0843  CAQ0844  CAQ0845  CAQ0846  CAQ0847  CAQ0848  CAQ0849  CAQ0850  CAQ0851  CAQ0852  CAQ0853  CAQ0854  CAQ0855  CAQ0856  CAQ0857  CAQ0858  CAQ0859  CAQ0860  CAQ0861  CAQ0862  CAQ0863  CAQ0864  CAQ0865  CAQ0866  CAQ0867  CAQ0868  CAQ0869  CAQ0870  CAQ0871  CAQ0872  CAQ0873  CAQ0874  CAQ0875  CAQ0876  CAQ0877  CAQ0878  CAQ0879  CAQ0880  CAQ0881  CAQ0882  CAQ0883  CAQ0884  CAQ0885  CAQ0886  CAQ0887  CAQ0888  CAQ0889  CAQ0890  CAQ0891  CAQ0892  CAQ0893  CAQ0894  CAQ0895  CAQ0896  CAQ0897  CAQ0898  CAQ0899  CAQ0900  CAQ0901  CAQ0902  CAQ0903  CAQ0904  CAQ0905  CAQ0906  CAQ0907  CAQ0908  CAQ0909  CAQ0910  CAQ0911  CAQ0912  CAQ0913  CAQ0914  CAQ0915  CAQ0916  CAQ0917  CAQ0918  CAQ0919  CAQ0920  CAQ0921  CAQ0922  CAQ0923  CAQ0924  CAQ0925  CAQ0926  CAQ0927  CAQ0928  CAQ0929  CAQ0930  CAQ0931  CAQ0932  CAQ0933  CAQ0934  CAQ0935  CAQ0936  CAQ0937  CAQ0938  CAQ0939  CAQ0940  CAQ0941  CAQ0942  CAQ0943  CAQ0944  CAQ0945  CAQ0946  CAQ0947  CAQ0950  CAQ0951  CAQ0953  CAQ0955  CAQ0956  CAQ0957  CAQ0958  CAQ0959  CAQ0960  CAQ0961  CAQ0962  CAQ0963  CAQ0964  CAQ0965  CAQ0966  CAQ0967  CAQ0968  CAQ0969  CAQ0970  CAQ0971  CAQ0972  CAQ0973  CAQ0974  CAQ0975  CAQ0976  CAQ0977  CAQ0978  CAQ0979  CAQ0980  CAQ0981  CAQ0982  CAQ0983  CAQ0984  CAQ0985  CAQ0986  CAQ0987  CAQ0988  CAQ0989  CAQ1157  CAQ1350  CAQ1351  CAQ1352  CAQ1353  CAQ1355  CAQ1356  CAQ1357  CAQ1358  CAQ1359  CAQ1360  CAQ1361  CAQ1362  CAQ1363  CAQ1364  CAQ1365  CAQ1366  CAQ1367  CAQ1368  CAQ1369  CAQ1370  CAQ1371  CAQ1372  CAQ1373  CAQ1374  CAQ1375  CAQ1376  CAQ1377  CAQ1378  CAQ1379  CAQ1380  CAQ1381  CAQ1382  CAQ1383  CAQ1384  CAQ1385  CAQ1386  CAQ1387  CAQ1388  CAQ1389  CAQ1390  CAQ1391  CAQ1392  CAQ1393  CAQ1394  CAQ1395  CAQ1396  CAQ1397  CAQ1398  CAQ1399  CAQ1400  CAQ1401  CAQ1402  CAQ1403  CAQ1404  CAQ1405  CAQ1406  CAQ1407  CAQ1408  CAQ1409  CAQ1410  CAQ1411  CAQ1412  CAQ1413  CAQ1414  CAQ1415  CAQ1416  CAQ1417  CAQ1418  CAQ1419  CAQ1420  CAQ1422  CAQ1426  CAQ1428  CAQ1429  CAQ1430  CAQ1431  CAQ1432  CAQ1433  CAQ1434  CAQ1435  CAQ1436  CAQ1437  CAQ1438  CAQ1439  CAQ1440  CAQ1441  CAQ1442  CAQ1443  CAQ1444  CAQ1445  CAQ1446  CAQ1447  CAQ1448  CAQ1449  CAQ1450  CAQ1451  CAQ1452  CAQ1453  CAQ1454  CAQ1455  CAQ1456  CAQ1457  CAQ1458  CAQ1459  CAQ1460  CAQ1461  CAQ1462  CAQ1463  CAQ1464  CAQ1465  CAQ1466  CAQ1467  CAQ1468  CAQ1469  CAQ1470  CAQ1471  CAQ1472  CAQ1473  CAQ1474  CAQ1475  CAQ1476  CAQ1477  CAQ1478  CAQ1479  CAQ1480  CAQ1481  CAQ1482  CAQ1483  CAQ1484  CAQ1485  CAQ1486  CAQ1487  CAQ1488  CAQ1489  CAQ1490  CAQ1491  CAQ1492  CAQ1493  CAQ1494  CAQ1495  CAQ1496  CAQ1497  CAQ1498  CAQ1499  CAQ1500  CAQ1501  CAQ1502  CAQ1503  CAQ1504  CAQ1505  CAQ1506  CAQ1507  CAQ1508  CAQ1509  CAQ1511  CAQ1512  CAQ1513  CAQ1514  CAQ1515  CAQ1516  CAQ1517  CAQ1518  CAQ1519  CAQ1520  CAQ1521  CAQ1522  CAQ1523  CAQ1524  CAQ1525  CAQ1526  CAQ1527  CAQ1528  CAQ1529  CAQ1530  CAQ1531  CAQ1532  CAQ1533  CAQ1534  CAQ1535  CAQ1536  CAQ1537  CAQ1538  CAQ1539  CAQ1540  CAQ1541  CAQ1542  CAQ1543  CAQ1544  CAQ1545  CAQ1546  CAQ1547  CAQ1548  CAQ1549  CAQ1550  CAQ1551  CAQ1552  CAQ1553  CAQ1554  CAQ1555  CAQ1556  CAQ1557  CAQ1558  CAQ1559  CAQ1560  CAQ1561  CAQ1562  CAQ1563  CAQ1564  CAQ1565  CAQ1566  CAQ1567  CAQ1568  CAQ1569  CAQ1570  CAQ1571  CAQ1572  CAQ1573  CAQ1574  CAQ1575  CAQ1576  CAQ1577  CAQ1578  CAQ1579  CAQ1580  CAQ1581  CAQ1582  CAQ1583  CAQ1584  CAQ1585  CAQ1586  CAQ1587  CAQ1588  CAQ1589  CAQ1590  CAQ1591  CAQ1592  CAQ1593  CAQ1594  CAQ1595  CAQ1596  CAQ1597  CAQ1598  CAQ1599  CAQ1600  CAQ1601  CAQ1602  CAQ1603  CAQ1604  CAQ1605  CAQ1606  CAQ1607  CAQ1608  CAQ1609  CAQ1610  CAQ1611  CAQ1612  CAQ1613  CAQ1614  CAQ1615  CAQ1616  CAQ1617  CAQ1618  CAQ1619  CAQ1620  CAQ1621  CAQ1622  CAQ1623  CAQ1624  CAQ1625  CAQ1626  CAQ1627  CAQ1628  CAQ1629  CAQ1630  CAQ1631  CAQ1632  CAQ1633  CAQ1634  CAQ1635  CAQ1636  CAQ1637  CAQ1638  CAQ1639  CAQ1640  CAQ1641  CAQ1642  CAQ1643  CAQ1644  CAQ1645  CAQ1646  CAQ1647  CAQ1648  CAQ1649  CAQ1650  CAQ1651  CAQ1652  CAQ1653  CAQ1654  CAQ1655  CAQ1656  CAQ1657  CAQ1658  CAQ1659  CAQ1660  CAQ1661  CAQ1662  CAQ1663  CAQ1664  CAQ1665  CAQ1666  CAQ1667  CAQ1668  CAQ1669  CAQ1670  CAQ1671  CAQ1672  CAQ1673  CAQ1674  CAQ1675  CAQ1676  CAQ1677  CAQ1678  CAQ1679  CAQ1680  CAQ1681  CAQ1682  CAQ1683  CAQ1684  CAQ1685  CAQ1686  CAQ1687  CAQ1688  CAQ1689  CAQ1690  CAQ1691  CAQ1692  CAQ1693  CAQ1694  CAQ1695  CAQ1696  CAQ1697  CAQ1698  CAQ1699  CAQ1700  CAQ1701  CAQ1702  CAQ1703  CAQ1704  CAQ1705  CAQ1706  CAQ1707  CAQ1708  CAQ1709  CAR0100  CAR0101  CAR0102  CAR0103  CAR0104  CAR0105  CAR0106  CAR0107  CAR0108  CAR0109  CAR0110  CAR0111  CAR0112  CAR0113  CAR0114  CAR0115  CAR0116  CAR0117  CAR0118  CAR0119  CAR0120  CAR0121  CAR0122  CAR0123  CAR0124  CAR0125  CAR0126  CAR0127  CAR0128  CAR0129  CAR0130  CAR0131  CAR0132  CAR0133  CAR0134  CAR0135  CAR0136  CAR0137  CAR0138  CAR0139  CAR0140  CAR0141  CAR0142  CAR0143  CAR0144  CAR0145  CAR0146  CAR0147  CAR0148  CAR0149  CAR0150  CAR0151  CAR0152  CAR0153  CAR0154  CAR0155  CAR0156  CAR0157  CAR0158  CAR0159  CAR0160  CAR0161  CAR0162  CAR0163  CAR0164  CAR0165  CAR0166  CAR0167  CAR0168  CAR0169  CAR0170  CAR0171  CAR0172  CAR0173  CAR0174  CAR0175  CAR0176  CAR0177  CAR0178  CAR0179  CAR0180  CAR0181  CAR0182  CAR0183  CAR0184  CAR0185  CAR0186  CAR0187  CAR0188  CAR0189  CAR0190  CAR0191  CAR0192  CAR0193  CAR0194  CAR0195  CAR0196  CAR0197  CAR0198  CAR0199  CAR0200  CAR0201  CAR0202  CAR0203  CAR0204  CAR0205  CAR0206  CAR0207  CAR0208  CAR0209  CAR0210  CAR0211  CAR0212  CAR0213  CAR0214  CAR0215  CAR0216  CAR0217  CAR0218  CAR0219  CAR0220  CAR0221  CAR0222  CAR0223  CAR0224  CAR0225  CAR0226  CAR0227  CAR0228  CAR0229  CAR0230  CAR0231  CAR0232  CAR0233  CAR0234  CAR0235  CAR0236  CAR0237  CAR0238  CAR0239  CAR0240  CAR0241  CAR0242  CAR0243  CAR0244  CAR0245  CAR0246  CAR0247  CAR0248  CAR0249  CAR0250  CAR0252  CAR0253  CAR0254  CAR0255  CAR0256  CAR0257  CAR0258  CAR0259  CAR0260  CAR0261  CAR0262  CAR0263  CAR0264  CAR0265  CAR0266  CAR0267  CAR0268  CAR0269  CAR0270  CAR0271  CAR0272  CAR0273  CAR0274  CAR0275  CAR0276  CAR0277  CAR0278  CAR0279  CAR0280  CAR0281  CAR0282  CAR0283  CAR0284  CAR0285  CAR0286  CAR0287  CAR0288  CAR0289  CAR0290  CAR0291  CAR0292  CAR0293  CAR0294  CAR0295  CAR0296  CAR0297  CAR0298  CAR0299  CAR0300  CAR0310  CAR0311  CAR0312  CAR0313  CAR0314  CAR0315  CAR0316  CAR0317  CAR0318  CAR0319  CAR0320  CAR0321  CAR0322  CAR0323  CAR0324  CAR0325  CAR0326  CAR0327  CAR0328  CAR0329  CAR0330  CAR0331  CAR0332  CAR0333  CAR0334  CAR0335  CAR0336  CAR0337  CAR0338  CAR0339  CAR0340  CAR0341  CAR0342  CAR0343  CAR0344  CAR0345  CAR0346  CAR0347  CAR0348  CAR0349  CAR0350  CAR0351  CAR0352  CAR0353  CAR0354  CAR0355  CAR0356  CAR0357  CAR0358  CAR0359  CAR0360  CAR0361  CAR0362  CAR0363  CAR0364  CAR0365  CAR0366  CAR0367  CAR0368  CAR0369  CAR0370  CAR0371  CAR0372  CAR0373  CAR0374  CAR0375  CAR0376  CAR0377  CAR0378  CAR0379  CAR0380  CAR0381  CAR0382  CAR0383  CAR0384  CAR0385  CAR0386  CAR0387  CAR0388  CAR0389  CAR0390  CAR0391  CAR0392  CAR0393  CAR0394  CAR0395  CAR0396  CAR0397  CAR0398  CAR0399  CAR0400  CAR0401  CAR0402  CAR0403  CAR0404  CAR0405  CAR0406  CAR0410  CAR0411  CAR0412  CAR0413  CAR0414  CAR0415  CAR0416  CAR0417  CAR0418  CAR0419  CAR0420  CAR0421  CAR0422  CAR0423  CAR0424  CAR0425  CAR0426  CAR0427  CAR0428  CAR0429  CAR0430  CAR0431  CAR0432  CAR0433  CAR0434  CAR0435  CAR0436  CAR0437  CAR0438  CAR0439  CAR0440  CAR0441  CAR0442  CAR0443  CAR0444  CAR0445  CAR0446  CAR0447  CAR0448  CAR0449  CAR0450  CAR0451  CAR0452  CAR0453  CAR0454  CAR0455  CAR0456  CAR0457  CAR0458  CAR0459  CAR0460  CAR0461  CAR0462  CAR0463  CAR0464  CAR0465  CAR0466  CAR0467  CAR0468  CAR0469  CAR0471  CAR0472  CAR0473  CAR0474  CAR0475  CAR0476  CAR0477  CAR0478  CAR0479  CAR0480  CAR0481  CAR0482  CAR0483  CAR0484  CAR0485  CAR0486  CAR0487  CAR0488  CAR0489  CAR0490  CAR0491  CAR0492  CAR0493  CAR0494  CAR0495  CAR0496  CAR0497  CAR0498  CAR0499  CAR0500  CAR0501  CAR0502  CAR0503  CAR0504  CAR0505  CAR0506  CAR0507  CAR0508  CAR0509  CAR0510  CAR0511  CAR0512  CAR0513  CAR0514  CAR0515  CAR0516  CAR0517  CAR0518  CAR0519  CAR0520  CAR0521  CAR0522  CAR0523  CAR0524  CAR0525  CAR0526  CAR0527  CAR0528  CAR0529  CAR0530  CAR0531  CAR0532  CAR0533  CAR0534  CAR0535  CAR0536  CAR0537  CAR0538  CAR0539  CAR0540  CAR0541  CAR0542  CAR0543  CAR0544  CAR0545  CAR0546  CAR0547  CAR0548  CAR0549  CAR0550  CAR0551  CAR0552  CAR0553  CAR0554  CAR0555  CAR0556  CAR0557  CAR0558  CAR0559  CAR0560  CAR0561  CAR0562  CAR0563  CAR0564  CAR0565  CAR0566  CAR0567  CAR0568  CAR0569  CAR0570  CAR0571  CAR0572  CAR0573  CAR0574  CAR0575  CAR0576  CAR0577  CAR0578  CAR0579  CAR0580, and   CAR0600.",
      "center_classification_date": "20170713",
      "distribution_pattern": "Worldwide Distribution -- US, to the following states: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH , OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV and, to the countries of Australia, Canada, Chile, Columbia, Germany, India, Japan, Jordan, Kuwait, Netherlands, Papua Guinea, Philippines, Qatar, Saudi Arabia, Singapore, Sweden, Switzerland, Thailand, and United Kingdom.",
      "state": "WI",
      "product_description": "Nicolet¿ Elite¿ 3 MHz OB Probe, Model #: X1L001, Catalog Code: N300.    Intended for non-invasive support, i.e. early detection of fetal life, monitor blood flow before and after surgery, etc.",
      "report_date": "20170719",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Natus Neurology Inc",
      "recall_number": "Z-2728-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77569",
      "termination_date": "20200722",
      "more_code_info": "",
      "recall_initiation_date": "20170609",
      "postal_code": "53562-4800",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}