      "status": "Terminated",
      "city": "Fort Worth",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74824",
      "recalling_firm": "Alcon Research, Ltd.",
      "address_1": "6201 South Fwy",
      "address_2": "N/A",
      "postal_code": "76134-2099",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Armenia, Belgium, Bulgaria, Croatia, Finland, France, Germany, Hungary, Italy, Lithuania, Malaysia, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Slovenia, Spain, Sweden, Switzerland, Thailand, United Kingdom (Great Britain and Ireland), and United Arab Emirates. *** While this product is available in the US, these lots were not distributed to US consignees ***",
      "recall_number": "Z-2708-2016",
      "product_description": "Alcon AcrySof IQ IOL with ULTRASERT Delivery System    Product Usage:  The Alcon AcrySof IQ Intraocular Lens (IOL) is an acrylic foldable singlepiece posterior chamber lens for the replacement of the human crystalline lens in the visual correction of aphakia in adult patients following cataract surgery. This material is capable of being folded prior to insertion. The lens gently unfolds to a fullsize lens body following implantation. The lens has a biconvex optic with supporting haptics The AcrySof IQ IOLs are provided in the ULTRASERT Preloaded Delivery System for a convenient, controlled means to reliably place these lenses into the capsular bag.",
      "product_quantity": "371 units",
      "reason_for_recall": "The ULTRASERT Delivery System from certain lots have an inferior surface characteristic that could result in the IOL becoming lodged in the Delivery System",
      "recall_initiation_date": "20160728",
      "center_classification_date": "20160901",
      "termination_date": "20170607",
      "report_date": "20160907",
      "code_info": "Lot numbers: 12407022, 12407085, 12407087, 12409013, 12409024, 12409026, 12409028"
    }
  ]
}