  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carpinteria",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69191",
      "recalling_firm": "Dako North America Inc.",
      "address_1": "6392 Via Real",
      "address_2": "N/A",
      "postal_code": "93013-2921",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AT, AU, BE, CA, CN, DE, DK, ES, Fl, FR, GB, IE, IT, NL, NO, SE, US, DZ, GR, IS, OM, SG, UY.",
      "recall_number": "Z-2708-2014",
      "product_description": "Dako Autostainer Link 48 with software version Dako Link 4.0.3, Model No. AS480    Product Usage:  For in vitro diagnostic use. The Dako Autostainer Link 48 is an automated slide processing system compatible with reagents for the staining of paraffin-embedded and frozen tissue sections, cytospins, cell smears and fine needle aspirates. The system is designed to automate manual staining methods routinely used in immunohistochemistry and cytochemistry, enabling the transfer of established protocols from the bench to the Autostainer Link 48 automated workstations operating either singly or in networked configurations.",
      "product_quantity": "191 units total (57 units in US)",
      "reason_for_recall": "Dako is recalling the Dako Link 4.0.3 software on the Dako Autostainer Link 48 because Dako has determined that an upgrade to this version of the software may cause premature rejection of reagent bottles or inadequate aspiration of reagent if the instrument has also had a preventative maintenance or waste pump replacement, and the baseplate is not completely aligned.",
      "recall_initiation_date": "20140829",
      "center_classification_date": "20140925",
      "termination_date": "20150303",
      "report_date": "20141001",
      "code_info": "Serial Numbers:    AS1101D0902  AS1722D1009  AS2404D1203  AS2710D1211  AS1200D0904  AS1298D0906  AS1420D0908  AS1606D1004  AS1608D1004  AS1610D1004  AS1406D0908  AS1498D0910  AS1511D0910  AS1542D1001  AS1566D1002  AS1684D1008  AS1685D1008  AS1708D1008  AS1767D1010  AS1779D1010  AS1802D1010  AS1865D1011  AS2002D1105  AS2039D1106  AS2040D1106  AS2041D1106  AS2042D1106  AS2044D1106  AS2052D1106  AS2057D1107  AS2058D1107  AS2071D1107  AS2701D1211  AS2707D1211  AS48-0363-01  AS1251D0905  AS1149D0903  AS1216D0904  AS1270D0905  AS1271D0905  AS1311D0906  AS1314D0906  AS1356D0907  AS1363D0907  AS1691D1008  AS1883D1101  AS1934D1103  AS2109D1108  AS2111D1108  AS2141D1108  AS2158D1109  AS2159D1109  AS2170D1109  AS2182D1109  AS2192D1110  AS2300D1111  AS2358D1202  AS2359D1202  AS2361D1202  AS2416D1204  AS2504D1206  AS2516D1206  AS2678D1210  AS48-0196-01  AS48-0233-01  AS48-0234-01  AS48-0236-01  AS48-0237-01  AS48-0306-01  AS48-0332-01  AS48-0344-01  AS1025D0809  AS1027D0809  AS1029D0809  AS1030D0809  AS1032D0809  AS1072D0811  AS1185D0904  AS1186D0904  AS1187D0904  AS1418D0908  AS1220D0904  AS1292D0906  AS1354D0907  AS1472D0909  AS1473D0909  AS1556D1001  AS1817D1011  AS1904D1102  AS2036D1106  AS2353D1202  AS2357D1202  AS48-0273-01  AS48-0337-01  AS1266D0905  AS1624D1005  AS1695D1008  AS1560D1001  AS1603D1004  AS1636D1005  AS1941D1103  AS2230D1110  AS2234D1110  AS2235D1110  AS2236D1110  AS2308D1111  AS2542D1207  AS1342D0907  AS1446D0909  AS48-0114-01  AS1293D0906  AS2110D1108  AS1161D0903  AS1162D0903  AS1427D0909  AS2147D1109  AS2148D1109  AS2418D1204  AS1431D0909  AS1014D0809  AS1093D0901  AS1190D0904  AS1210D0904  AS48-0309-01  AS48-0310-01  AS48-0317-01  AS48-0318-01  AS48-0326-01  AS1052D0811  AS1094D0901  AS1109D0902  AS1123D0903  AS1127D0903  AS1138D0903  AS1140D0903  AS1238D0904  AS1239D0904  AS1240D0905  AS1301D0906  AS1320D0906  AS1322D0906  AS1331D0907  AS1332D0907  AS1376D0907  AS1508D0910  AS1520D0910  AS1528D0911  AS1551D1001  AS1637D1006  AS1663D1007  AS1681D1007  AS1682D1007  AS1712D1008  AS1732D1009  AS1749D1009  AS1765D1010  AS1775D1010  AS1778D1010  AS1800D1010  AS1837D1011  AS1839D1011  AS1848D1011  AS1850D1011  AS1855D1011  AS1866D1012  AS1867D1012  AS1880D1101  AS1975D1104  AS1993D1105  AS2030D1106  AS2055D1106  AS48-0150-01  AS48-0152-01  AS48-0156-01  AS48-0166-01  AS48-0167-01  AS48-0177-01  AS48-0192-01  AS48-0205-01  AS48-0207-01  AS48-0213-01  AS48-0248-01  AS48-0259-01  AS48-0296-01  AS48-0315-01    AS2761D1301  AS48-0129-01  AS2169D1109  AS1820D1011  AS1584D1003  AS2014D1105"
    }
  ]
}