      "status": "Terminated",
      "city": "Exeter",
      "state": "NH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69116",
      "recalling_firm": "Vapotherm, Inc.",
      "address_1": "22 Industrial Dr Ste 1",
      "address_2": "N/A",
      "postal_code": "03833-4557",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Canada, France, Swizerland, Korea, Sweden, Germany , Belgium, Spain, Mexico, Scotland, UAE, Australia, Israel, Hong Kong, Ireland, kuwait, and Italy.",
      "recall_number": "Z-2706-2014",
      "product_description": "Vapotherm Precision Flow,Disposable Patient Circuit  PF-NODPC-Low.    For use with the Precision Flow system and Ikaria INOmax DSIR  ¿ System.  This DPC allows clinicians to administer nitric oxide non-invasively via high  flow nasal cannula. In the Precision Flow High Flow Therapy application, a  special disposable patient circuit re-routes the gas to the INOmax injector  module then back into the Precision Flow system",
      "product_quantity": "207 units",
      "reason_for_recall": "Water may leak into the center gas lumen of the  Patient Circuit",
      "recall_initiation_date": "20140813",
      "center_classification_date": "20140925",
      "termination_date": "20151002",
      "report_date": "20141001",
      "code_info": "Lot Numbers 1402021 through 1406010, excluding lots 1402024 through 1402028 and 1404003. These numbers are applicable regardless of the prefix preceding them (i.e. HF, HFE, HFH, HFN, LF, LFE, LFH or LFN)."
    }
  ]
}