      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "Carestation 620, 650 and 650c has Potential for Elevated FiCO2 and Unexpected System Malfunction",
      "address_2": "",
      "product_quantity": "3,448 (334 US; 3,114 OUS)",
      "code_info": "Carestation 620 A1:    SM616460036MA  SM616460037MA  SM616400001MA  SM616250007MA  SM616340023MA  SM616460048MA  SM616460049MA  SM616460022MA  SM616250005MA  SM616400004MA  SM616400005MA  SM617110011MA  SM617110012MA  SM617110013MA  SM617110014MA  SM616430024MA  SM616200020MA  SM616200019MA  SM616200021MA  SM616200022MA  SM616200023MA  SM616200024MA  SM617010108MA  SM616270021MA  SM616270022MA  SM617020019MA  SM617020020MA  SM617020021MA  SM616460047MA  SM616340001MA  SM616340002MA  SM616340039MA  SM616340025MA  SM616460052MA  SM616460053MA  SM616460054MA  SM616460055MA  SM616460056MA  SM616460057MA  SM616250006MA  SM616460035MA  SM617010114MA  SM617010115MA  SM617010116MA  SM617010117MA  SM617010118MA  SM616340027MA  SM616410061MA  SM616340018MA  SM616340019MA  SM616340020MA  SM616460030MA  SM616460028MA  SM616460029MA  SM616410043MA  SM617020010MA  SM617020011MA  SM617020012MA  SM617020013MA  SM617020014MA  SM617020015MA  SM617020016MA  SM617020017MA  SM617020018MA  SM617060084MA  SM617100056MA  SM617100057MA  SM616400016MA  SM616400017MA  SM616170021MA  SM616170022MA  SM616200012MA  SM616340028MA  SM616340029MA  SM616340030MA  SM617100058MA  SM616340021MA  SM616430034MA  SM616430035MA  SM617100157MA  SM616250001MA  SM616250002MA  SM616250003MA  SM616250004MA  SM617100019MA  SM617100020MA  SM617100021MA  SM617100022MA  SM617100023MA  SM617100024MA  SM617100025MA  SM617020026MA  SM616100020MA  SM616100021MA  SM616100022MA  SM616100023MA  SM616100024MA  SM616200013MA  SM616200016MA  SM616200014MA  SM616200015MA  SM617020007MA  SM617020008MA  SM617020009MA  SM615500007WA  SM616150007WA  SM616220004WA  SM616220005WA  SM616220006WA  SM616370002WA  SM616370003WA  SM616370005WA  SM616370006WA  SM616450003WA  SM617030015WA  SM617030016WA  SM617030017WA  SM617040018WA  SM617040019WA  SM617040020WA  SM617040021WA  SM617040022WA  SM617040023WA  SM617060001WA  SM617060002WA  SM617060003WA  SM617060004WA  SM617120006WA  SM617120007WA  SM617080015WA  SM617080016WA  SM617080017WA  SM617080018WA  SM617080019WA  SM617080020WA  SM617080021WA  SM617080022WA  SM616400019MA  SM616400020MA  SM616460023MA  SM616460024MA  SM616460025MA  SM616460026MA  SM616460027MA  SM617050039MA  SM616330011WA  SM616330012WA  SM617040001WA  SM615520001WA  SM615520002WA  SM615520003WA  SM616240002WA  SM616340049WA  SM616340050WA  SM616240001WA  SM615390002WA  SM615390003WA  SM615390004WA  SM616430045WA  SM616430046WA  SM616270025WA  SM616270026WA  SM616220009WA  SM616220010WA  SM616220011WA  SM616340043WA  SM615430004WA  SM616140005WA  SM616140006WA  SM616280020WA  SM616280021WA  SM615390001WA  SM616330002WA  SM616330003WA  SM616330004WA  SM616280005WA  SM616280006WA  SM616280007WA  SM616280008WA  SM616280009WA  SM616280010WA  SM616280011WA  SM616280012WA  SM616280013WA  SM616280014WA  SM616280015WA  SM616280016WA  SM616280017WA  SM616280018WA  SM616280019WA  SM616400021WA  SM616400022WA  SM616400023WA  SM616190005WA  SM615500005WA  SM616210004WA  SM616460005WA  SM616460006WA  SM616460007WA  SM616460008WA  SM616460009WA  SM616320002MA  SM616320003MA  SM616410012MA  SM616410013MA  SM616410014MA  SM616440003WA  SM616500002WA  SM616500003WA  SM616500004WA  SM616510001WA  SM616510002WA  SM616510003WA  SM616510004WA  SM616170018WA  SM616290001WA  SM616030001WA  SM616030002WA  SM616030003WA  SM616030004WA  SM616030005WA  SM616170007WA  SM616440006WA  SM616170008WA  SM616150014WA  SM616350001WA  SM616350003WA  SM616410025WA  SM616410026WA  SM617030018WA  SM617030019WA  SM617030020WA  SM617030021WA  SM617030022WA  SM617030023WA  SM617030024WA  SM616150013WA  SM616290002WA  SM616290003WA  SM616290004WA  SM616290005WA  SM616290006WA  SM616150010WA  SM616170010WA  SM616170011WA  SM616170012WA  SM616170013WA  SM616410031WA  SM616410039WA  SM616410040WA  SM616150011WA  SM616150012WA  SM617040011WA  SM617040013WA  SM616290007WA  SM616350004WA  SM616410002WA  SM616410023WA  SM616410027WA  SM616410028WA  SM616410029WA  SM616410036WA  SM616420001WA  SM616430041WA  SM616430060WA  SM616500001WA  SM616510005WA  SM616510006WA  SM616510007WA  SM616510008WA  SM616510009WA  SM616510010WA  SM617010120WA  SM617010121WA  SM617010122WA  SM617010123WA  SM617030038WA  SM617030039WA  SM617080001WA  SM617080002WA  SM617080003WA  SM616160007WA  SM616160008WA  SM617030033WA  SM617030034WA  SM616290008WA  SM616410037WA  SM616420002WA  SM616150015WA  SM616160005WA  SM616160006WA  SM616020006WA  SM616020007WA  SM616020008WA  SM616410030WA  SM616410035WA  SM616290015WA  SM616290016WA  SM616290017WA  SM616290019WA  SM616290020WA  SM616330015WA  SM616330016WA  SM615490002WA  SM615490003WA  SM615490007WA  SM615490008WA  SM615490009WA  SM615490011WA  SM616020009WA  SM616020010WA  SM616020012WA  SM616170006WA  SM616170016WA  SM616430052WA  SM616430053WA  SM616430054WA  SM616430055WA  SM616430056WA  SM616430057WA  SM616430058WA  SM616430059WA  SM616430062WA  SM615500002WA  SM616050005WA  SM616170019WA  SM616170020WA  SM616430061WA  SM616350002WA  SM616410024WA  SM616300005WA  SM616300006WA  SM616330005WA  SM616430063WA  SM616430064WA  SM617060140WA  SM616440001WA  SM616440002WA  SM616020011WA  SM616170014WA  SM617040008WA  SM616410038WA  SM616020003WA  SM616410033WA  SM616410034WA  SM617040014WA  SM617040016WA  SM617040017WA  SM616430051WA  SM616510011WA  SM617020028WA  SM617020029WA  SM617020030WA  SM617020031WA  SM617020032WA  SM617020033WA  SM617020034WA  SM617030025WA  SM617030026WA  SM617030027WA  SM617030028WA  SM617030029WA  SM617030030WA  SM617030031WA  SM617030032WA  SM617030037WA  SM617030040WA  SM617040009WA  SM617040010WA  SM617040012WA  SM617040015WA  SM617060128WA  SM617060129WA  SM617060130WA  SM617060131WA  SM617060132WA  SM617060133WA  SM617060134WA  SM617060135WA  SM617060136WA  SM617060137WA  SM617060138WA  SM617060139WA  SM617070010WA  SM617070011WA  SM617070012WA  SM617070013WA  SM617070014WA  SM617070015WA  SM617070016WA  SM617070017WA  SM617070018WA  SM617070019WA  SM617070020WA  SM617070021WA  SM617070023WA  SM617070024WA  SM617070025WA  SM617070026WA  SM617070027WA  SM617070028WA  SM617070029WA  SM617070030WA  SM617070031WA  SM617070032WA  SM617070033WA  SM617070034WA  SM617070035WA  SM617070036WA  SM617070037WA  SM617070038WA  SM617070039WA  SM617070040WA  SM617070041WA  SM617070042WA  SM617070043WA  SM617070044WA  SM617070045WA  SM617070046WA  SM617070047WA  SM617070048WA  SM617070049WA  SM617080004WA  SM617080005WA  SM617080006WA  SM617080007WA  SM617080008WA  SM617080009WA  SM617080010WA  SM617080011WA  SM617080012WA  SM617080013WA  SM617080014WA  SM617080023WA  SM617080024WA  SM617080025WA  SM617080026WA  SM617080027WA  SM617080028WA  SM617080029WA  SM617080030WA  SM617080031WA  SM617090020WA  SM617090021WA  SM617090022WA  SM617090023WA  SM617090024WA  SM617090025WA  SM617090026WA  SM617090027WA  SM617090028WA  SM617090029WA  SM617090030WA  SM617090031WA  SM616170009WA  SM616290018WA  SM615500003WA  SM616440005WA  SM617030035WA  SM617030036WA  SM616020004WA  SM616440004WA  SM617020001WA  SM617020002WA  SM617020003WA  SM617020004WA  SM617020005WA  SM616160003WA  SM616160004WA  SM616330013WA  SM616330014WA  SM617100026MA  SM617100027MA  SM616430026MA  SM616200018MA  SM616430025MA  SM617100155MA  SM617100018MA  SM616310012MA  SM616310029MA  SM616140003WA  SM616140004WA  SM616280022MA  SM616280023MA  SM616280024MA  SM616280025MA  SM616400007MA  SM616400008MA  SM616400009MA  SM616400010MA  SM616400011MA  SM617010109MA  SM617010110MA  SM617010111MA  SM617010112MA  SM617010113MA  SM617100158MA  SM617100159MA  SM617100160MA  SM616050004WA  SM616070003WA  SM616170001WA  SM616170003WA  SM616170005WA  SM617010124WA  SM617010125WA  SM617010126WA  SM617010127WA  SM617010128WA  SM617010132WA  SM617010133WA  SM617010134WA  SM617010135WA  SM617060126WA  SM617060127WA  SM616340014MA  SM616340015MA  SM616340016MA  SM616400018MA  SM616410044MA  SM616410045MA  SM616410046MA  SM616410047MA  SM617010106MA  SM617010107MA  SM617010119MA  SM617100151MA  SM617100152MA  SM616150001WA  SM616410032WA  SM616270002WA  SM617040004WA  SM617020022MA  SM617020023MA  SM616100005MA  SM616100006MA  SM617060098MA  SM617060099MA  SM617060100MA  SM616280028MA  SM616280029MA  SM616280030MA  SM616280031MA  SM616300009MA  SM616300010MA  SM616300011MA  SM616300012MA  SM616300014MA  SM616300015MA  SM616300016MA  SM616300017MA  SM616300018MA  SM616310001MA  SM616310002MA  SM616310003MA  SM616310004MA  SM616310005MA  SM616310006MA  SM616310007MA  SM616310008MA  SM616310009MA  SM616310010MA  SM616310011MA  SM617050009MA  SM617050010MA  SM617050011MA  SM617050012MA  SM617050013MA  SM617050014MA  SM617050015MA  SM617050016MA  SM617050017MA  SM617050018MA  SM617050019MA  SM617050020MA  SM617050021MA  SM617050022MA  SM617050023MA  SM617050024MA  SM617050025MA  SM617050026MA  SM617050027MA  SM617050028MA  SM617050029MA  SM617050030MA  SM617050031MA  SM617050032MA  SM617050033MA  SM617050034MA  SM617050035MA  SM617050036MA  SM617050037MA  SM617050038MA  SM617060006MA  SM617060007MA  SM617060008MA  SM617060009MA  SM617060010MA  SM617060011MA  SM617060012MA  SM617060013MA  SM617060014MA  SM617060015MA  SM617060016MA  SM617060017MA  SM617060018MA  SM617060019MA  SM617060020MA  SM617060021MA  SM617060022MA  SM617060023MA  SM617060024MA  SM617060025MA  SM617060026MA  SM617060027MA  SM617060028MA  SM617060029MA  SM617060030MA  SM617060031MA  SM617060032MA  SM617060033MA  SM617060034MA  SM617060035MA  SM617060039MA  SM617060040MA  SM617060041MA  SM617060042MA  SM617060043MA  SM617060044MA  SM617060045MA  SM617060046MA  SM617060047MA  SM617060048MA  SM617060049MA  SM617060050MA  SM617060051MA  SM617060052MA  SM617060053MA  SM617060054MA  SM617060055MA  SM617060056MA  SM617060057MA  SM617060058MA  SM617060059MA  SM617060060MA  SM617060061MA  SM617060062MA  SM617060063MA  SM617060064MA  SM617060065MA  SM617060066MA  SM617060067MA  SM617060068MA  SM617060069MA  SM617060070MA  SM617060071MA  SM617060072MA  SM617060073MA  SM617060074MA  SM617060075MA  SM617060076MA  SM617060077MA  SM617060078MA  SM617060079MA  SM617060080MA  SM617060081MA  SM617060082MA  SM617060083MA  SM617100028MA  SM617100029MA  SM617100030MA  SM617100031MA  SM617100032MA  SM617100033MA  SM617100034MA  SM617100035MA  SM617100036MA  SM617100037MA  SM617100038MA  SM617100039MA  SM617100040MA  SM617100041MA  SM617100042MA  SM617100043MA  SM617100044MA  SM617100045MA  SM617100046MA  SM617100047MA  SM617100048MA  SM617100049MA  SM617100050MA  SM617100051MA  SM617100052MA  SM617100053MA  SM617100054MA  SM617100055MA  SM617100071MA  SM617100072MA  SM617100073MA  SM617100074MA  SM617100075MA  SM617100076MA  SM617100077MA  SM617100078MA  SM617100079MA  SM617100080MA  SM617100081MA  SM617100082MA  SM617100083MA  SM617100084MA  SM617100085MA  SM617100086MA  SM617100087MA  SM617100088MA  SM617100089MA  SM617100090MA  SM617100091MA  SM617100092MA  SM617100093MA  SM617100094MA  SM617100095MA  SM617100096MA  SM617100097MA  SM617100098MA  SM617100099MA  SM617100100MA  SM617100101MA  SM617100102MA  SM617100103MA  SM617100104MA  SM617100105MA  SM617100106MA  SM617100107MA  SM617100108MA  SM617100109MA  SM617100110MA  SM617100111MA  SM617100112MA  SM617100113MA  SM617100114MA  SM617100115MA  SM617100116MA  SM617100117MA  SM617100118MA  SM617100119MA  SM617100120MA  SM617100121MA  SM617100122MA  SM617100123MA  SM617100124MA  SM617100125MA  SM617100126MA  SM617100127MA  SM617100128MA  SM617100131MA  SM617100132MA  SM616400012MA  SM616400013MA  SM615500008WA  SM616480018WA  SM616480019WA  SM616450006WA  SM616320024WA  SM617020035WA  SM617020036WA  SM615030014WA  SM616340022MA  SM616410056MA  SM616410057MA  SM616460031MA  SM616460032MA  SM616460033MA  SM616460034MA  SM616400014MA  SM616280032WA  SM617110001WA  SM617110002WA  SM617110003WA  SM615370001WA  SM616440025WA  SM616100033WA  SM615350082WA  SM615350083WA  SM615350084WA  SM615350085WA  SM615370003WA  SM615370020WA  SM615370021WA  SM615370018WA  SM615370019WA  SM616430042WA  SM616430044WA  SM615440001WA  SM615350088WA  SM615350089WA  SM617110009WA  SM616200002WA  SM615440003WA  SM616210001WA  SM616210002WA  SM616210003WA  SM617030011WA  SM617040005WA  SM617040006WA  SM617100069WA  SM615510004WA  SM615510005WA  SM615510006WA  SM616220020WA  SM616220019WA  SM616360005WA  SM615370022WA  SM615500004WA  SM615500009WA  SM615510001WA  SM615510002WA  SM616190014WA  SM616190016WA  SM616220007WA  SM616220008WA  SM616220012WA  SM616220013WA  SM616360022WA  SM616440024WA  SM616220001WA  SM617070009WA  SM617100014WA  SM615510003WA  SM616340044WA  SM615370006WA  SM615370008WA  SM615370010WA  SM615370011WA  SM615370012WA  SM615370005WA  SM616370001WA  SM616220021WA  SM616220022WA  SM616220025WA  SM616240005WA  SM616240006WA  SM616320022WA  SM616320023WA  SM616430043WA  SM615340070WA  SM615370009WA  SM615430001WA  SM615370007WA  SM616220023WA  SM616220024WA  SM615360005WA  SM615440002WA  SM616100038WA  SM616150004WA  SM616150005WA  SM616150006WA  SM616160009WA  SM616330001WA  SM616440023WA  SM616440026WA  SM617030012WA  SM617030013WA  SM617030014WA  SM615500001WA  SM615500016WA  SM616340045WA  SM615330048WA  SM615480007WA  SM615480009WA  SM616330006WA  SM616330007WA  SM616330008WA  SM616330009WA  SM616330010WA  SM617100061WA  SM616440022WA  SM615480006WA  SM616150008WA  SM615500012WA  SM615490010WA  SM615500013WA  SM616240009WA  SM616330018WA  SM616100032WA  SM616100034WA  SM616100036WA  SM616100037WA  SM615440004WA  SM617090008WA  SM617040007WA  SM616300003WA  SM616300004WA  SM617100067WA  SM617100068WA  SM615370002WA  SM615370004WA  SM615370014WA  SM615370015WA  SM615370016WA  SM615370017WA  SM615480008WA  SM615480010WA  SM615500011WA  SM615500015WA  SM616440011WA  SM616440012WA  SM616440013WA  SM617110005WA  SM617110006WA  SM615440005WA  SM615440006WA  SM615440007WA  SM616080005WA  SM616100031WA  SM616320025WA  SM616420009WA  SM615340071WA  SM615350072WA  SM615350075WA  SM615350077WA  SM615350078WA  SM616080004WA  SM616080006WA  SM616190013WA  SM615260034WA  SM616160010WA  SM616190015WA  SM615370013WA  SM615430005WA  SM615500014WA  SM616240003WA  SM616240004WA  SM616190002WA  SM616190003WA  SM616190004WA  SM616230011WA  SM616230012WA  SM616230013WA  SM616230014WA  SM616230015WA  SM616230016WA  SM616230017WA  SM616230018WA  SM616230019WA  SM616230020WA  SM616410015MA  SM616410016MA  SM616410017MA  SM616410018MA  SM616410019MA  SM616410020MA  SM616410021MA  SM616410022MA  SM616430033MA  SM616340034MA  SM616340033MA  SM616230004WA  SM616230005WA  SM617070007WA  SM615460002WA  SM616450005WA  SM616010007WA  SM616400025WA  SM616280003WA  SM616230007WA  SM616280002WA  SM616500005WA  SM616500006WA  SM616500007WA  SM616500008WA  SM616500009WA  SM617110004WA  SM616300007WA  SM616300008WA  SM616080001WA  SM616080002WA  SM616290026WA  SM616290027WA  SM616510012WA  SM615480002WA  SM616360016WA  SM616420017WA  SM616470009WA  SM616480009WA  SM617090003WA  SM616470013WA  SM617010129WA  SM617070006WA  SM616280001WA  SM616360012WA  SM616460063WA  SM616470011WA  SM617070002WA  SM616340048WA  SM616340046WA  SM616450011WA  SM617090009WA  SM616360013WA  SM617070022WA  SM616270023WA  SM616360015WA  SM616420003WA  SM616420004WA  SM616430050WA  SM617070001WA  SM616100002WA  SM616140001WA  SM616460012WA  SM617100009WA  SM616270024WA  SM616400024WA  SM617090007WA  SM616480008WA  SM616230009WA  SM616480011WA  SM616230001WA  SM716340041WA  SM716340042WA  SM616480010WA  SM616470010WA  SM616470005WA  SM616470006WA  SM616360007WA  SM617100010WA  SM616360014WA  SM616320001WA  SM616230010WA  SM616470002WA  SM616470008WA  SM617100008WA  SM616470003WA  SM616290025WA  SM617090005WA  SM616160001WA  SM616360006WA  SM616340047WA  SM616340051WA  SM617070008WA  SM616420007WA  SM616470004WA  SM616470007WA  SM617010130WA  SM617070005WA  SM617090010WA  SM617090018WA  SM617090019WA  SM617100011WA  SM617100012WA  SM617100013WA  SM616140002WA  SM616230006WA  SM616270001WA  SM616290024WA  SM616450012WA  SM616460013WA  SM616480012WA  SM617010131WA  SM617100062WA  SM616420005WA  SM616420006WA  SM617090006WA  SM616470012WA  SM616200011MA  SM616340017MA  SM616410010MA  SM616410011MA  SM617060101MA  SM617060102MA  SM617060103MA  SM615340062WA  SM616200001WA  SM616480020WA  SM616470001WA  SM617100063WA  SM617100064WA  SM617060005WA  SM617110007WA  SM617110008WA  SM617110015WA  SM616230002WA  SM616230003WA  SM616480016WA  SM617090001WA  SM617090002WA  SM617090004WA  SM617090017WA  SM617060125MA  SM616190017WA  SM616510013WA  SM616510014WA  SM616510015WA  SM616510016WA  SM616510017WA  SM616510018WA  SM616510019WA  SM616330017WA  SM616460003WA  SM617110010WA  SM616010008WA  SM616480017WA  SM616010002WA  SM616010003WA  SM616010004WA  SM616010005WA  SM616010006WA  SM616160002WA  SM616170015WA  SM616290009WA  SM616290010WA  SM616290011WA  SM616290012WA  SM616290013WA  SM616360001WA  SM616360002WA  SM616480013WA  SM616480014WA  SM616480015WA  SM617040002WA  SM617040003WA  SM615490004WA  SM616410042MA  SM615520005WA  SM615520006WA  SM615520007WA  SM616490002WA  SM617060121MA  SM617060122MA  SM617060123MA  SM617060124MA  SM616410059MA  SM616410060MA  SM617060036MA  SM617060037MA  SM617060038MA  SM617100017MA  SM616460040MA  SM616460041MA  SM616460042MA  SM616460043MA  SM616460038MA  SM616460039MA  SM616460050MA  SM616460051MA  SM617060085MA  SM617060086MA  SM617060087MA  SM616050001WA  SM616100001WA  SM616160011WA  SM616160012WA  SM616190012WA  SM616450001WA  SM616450002WA  SM616460061WA  SM616460062WA  SM616020005WA  SM617060092MA  SM616310030MA  SM616320004MA  SM616320005MA  SM616320006MA  SM616320007MA  SM616320008MA  SM617010100MA  SM617100153MA  SM616170002WA  SM616170004WA  SM616400015MA  SM617010101MA  SM617010102MA  SM615490005WA  SM615490006WA  SM615500006WA  SM616050003WA  SM616020001WA  SM616290014WA  SM615420001WA  SM616410048MA  SM616010001WA  SM616280027MA  SM616250008MA  SM616250009MA  SM616270003MA  SM616270004MA  SM616270005MA  SM616270006MA  SM616270007MA  SM616270008MA  SM616270009MA  SM616270010MA  SM616270011MA  SM616270012MA  SM616270013MA  SM616270014MA  SM616270015MA  SM616270016MA  SM616270017MA  SM616270018MA  SM616270019MA  SM616270020MA  SM616410003MA  SM616410004MA  SM616410005MA  SM616410006MA  SM616410007MA  SM616410008MA  SM616410009MA  SM616250010MA  SM616280026MA  SM616340036MA  SM616340037MA  SM616340038MA  SM616400006MA  SM616430027MA  SM616430028MA  SM615430002WA  SM615430003WA  SM616100035WA  SM616100039WA  SM616100040WA  SM616190006WA  SM616190007WA  SM616190008WA  SM616190009WA  SM616190010WA  SM616190011WA  SM616210023WA  SM616210024WA  SM616320027WA  SM616320028WA  SM616320029WA  SM616320030WA  SM616320031WA  SM616320032WA  SM616320033WA  SM616320034WA  SM616320035WA  SM616320036WA  SM616430047WA  SM616430048WA  SM616430049WA  SM616440007WA  SM616440008WA  SM616440009WA  SM616440010WA  SM617030001WA  SM617030002WA  SM617030003WA  SM617030004WA  SM617030005WA  SM617030006WA  SM617030007WA  SM617030008WA  SM617030009WA  SM617030010WA  SM617020024MA  SM617020025MA  SM616050002WA  SM616150003WA  SM616200003WA  SM616200004WA  SM616380001WA  SM616380002WA  SM616380003WA  SM616380004WA  SM617020027WA  SM616110004WA  SM616110002WA  SM616110003WA  SM616440015WA  SM615460001WA  SM616110001WA  SM616410001WA  SM616460010WA  SM616360021WA  SM616360019WA  SM616360020WA  SM616380005WA  SM616360023WA  SM616480004WA  SM616070004WA  SM616360017WA  SM616360018WA  SM616440014WA  SM616230021WA  SM616280004WA  SM616450004WA  SM616460011WA  SM616480005WA  SM616480006WA  SM616480007WA  SM616480001WA  SM616170017WA  SM617100154MA  SM615480001WA  SM616490003WA  SM616490004WA  SM615380001WA  SM615380002WA  SM616290021WA  SM616240010WA  SM616510020WA  SM616510021WA  SM616100041WA  SM616460014WA  SM616460015WA  SM616460016WA  SM616460017WA  SM616450010WA  SM616070001WA  SM616070002WA  SM616420008WA  SM617120001WA  SM617120002WA  SM617120003WA  SM615480003WA  SM615480004WA  SM615490001WA  SM615500010WA  SM616100003WA  SM616100004WA  SM616240008WA  SM616490001WA  SM616360003WA  SM616370004WA  SM615480005WA  SM616020002WA  SM616240007WA  SM616200017MA  SM616310019MA  SM616340024MA  SM616340026MA  SM616430029MA  SM616430030MA  SM616430031MA  SM616430032MA  SM616460018MA  SM616460019MA  SM617010103MA  SM617010104MA  SM617050005MA  SM617050006MA  SM617050007MA  SM617050008MA  SM617050040MA  SM617060090MA  SM617060091MA  SM617060093MA  SM617060094MA  SM617060095MA  SM617060096MA  SM617060097MA  SM617060104MA  SM617060105MA  SM617100156MA  SM616200010MA  SM616430040MA  SM616320009MA  SM616320010MA  SM616320011MA  SM617060088MA  SM617060089MA  SM616400002MA  SM616400003MA  SM617050001MA  SM617050002MA  SM617050003MA  SM617050004MA  SM616440016WA  SM616440019WA  SM616440020WA  SM616440021WA  SM616450013WA  SM616460001WA  SM616460002WA  SM617090011WA  SM617090012WA  SM617090013WA  SM617090014WA  SM617090015WA  SM617090016WA  SM616290022WA  SM616290023WA  SM616420010WA  SM616190001WA  SM616200025WA  SM616200026WA  SM616200027WA  SM616200028WA  SM616200029WA  SM616200030WA  SM616450009WA  SM616360004WA  SM616360008WA  SM616360009WA  SM616360010WA  SM616360011WA  SM616220014WA  SM616220015WA  SM616220016WA  SM616220017WA  SM616220018WA  SM616220026WA  SM616220027WA  SM616230008WA  SM616480002WA  SM616480003WA  SM617120004WA  SM617120005WA  SM617070003WA  SM617070004WA  SM616440017WA  SM616440018WA  SM616320026WA  SM717100065WA  SM717100066WA  SM616210005WA  SM616210006WA  SM616210007WA  SM616210008WA  SM616210009WA  SM616210010WA  SM616210011WA  SM616210012WA  SM616210013WA  SM616210014WA  SM616210015WA  SM616210016WA  SM616210017WA  SM616210018WA  SM616210019WA  SM616210020WA  SM616210021WA  SM616210022WA  SM616220002WA  SM616220003WA  SM616150009WA  SM616370007WA  SM616420011WA  SM616420012WA  SM616420013WA  SM616420014WA  SM616420015WA  SM616420016WA  SM617100015MA  SM617100016MA  SM615020004WA  SM616170023MA  SM616170024MA  SM616170025MA  SM616170026MA  SM616170027MA  SM616170028MA  SM616170029MA  SM616170030MA  SM616200005MA  SM616200006MA  SM616200007MA  SM616200008MA  SM616200009MA  SM616310013MA  SM616310014MA  SM616310015MA  SM616310016MA  SM616310017MA  SM616310018MA  SM616310020MA  SM616310021MA  SM616310022MA  SM616310023MA  SM616310024MA  SM616310025MA  SM616310026MA  SM616310027MA  SM616310028MA  SM616320012MA  SM616320013MA  SM616320014MA  SM616320015MA  SM616320016MA  SM616320017MA  SM616320018MA  SM616320019MA  SM616320020MA  SM616320021MA  SM616340003MA  SM616340004MA  SM616340005MA  SM616340006MA  SM616340007MA  SM616340008MA  SM616340009MA  SM616340010MA  SM616340011MA  SM616340012MA  SM616340013MA  SM616410049MA  SM616410050MA  SM616410051MA  SM616410052MA  SM616410053MA  SM616410054MA  SM616410055MA  SM616410058MA  SM616430001MA  SM616430002MA  SM616430003MA  SM616430004MA  SM616430005MA  SM616430006MA  SM616430007MA  SM616430008MA  SM616430009MA  SM616430010MA  SM616430011MA  SM616430012MA  SM616430013MA  SM616430014MA  SM616430015MA  SM616430016MA  SM616430017MA  SM616430018MA  SM616430019MA  SM616430020MA  SM616430021MA  SM616430022MA  SM616430023MA  SM616430036MA  SM616430037MA  SM616430038MA  SM616430039MA  SM616460020MA  SM616460021MA  SM617010105MA  SM617060106MA  SM617060107MA  SM617060108MA  SM617060109MA  SM617060110MA  SM617060111MA  SM617060112MA  SM617060113MA  SM617060114MA  SM617060115MA  SM617060116MA  SM617060117MA  SM617060118MA  SM617060119MA  SM617060120MA    Carestation 650 A1:    SM717020018MA  SM717020019MA  SM717020020MA  SM717020021MA  SM717020022MA  SM717010103MA  SM716300040MA  SM716300041MA  SM716300031MA  SM716480059MA  SM716480060MA  SM716510010MA  SM716510011MA  SM716510012MA  SM716510013MA  SM716510014MA  SM716380022MA  SM716380023MA  SM716380024MA  SM716480021MA  SM716480022MA  SM716480023MA  SM716480024MA  SM716480025MA  SM716480026MA  SM716480027MA  SM716480028MA  SM717020026MA  SM717020027MA  SM717020028MA  SM717020029MA  SM717020030MA  SM716510015MA  SM716510016MA  SM716510017MA  SM717020040MA  SM716300024MA  SM716300025MA  SM716300026MA  SM716430052MA  SM716430053MA  SM716430054MA  SM716430055MA  SM716480057MA  SM716480029MA  SM716480030MA  SM716480031MA  SM716480032MA  SM716480033MA  SM716270006MA  SM716270007MA  SM717010115MA  SM717010117MA  SM717010118MA  SM717010120MA  SM717010121MA  SM717010122MA  SM717010123MA  SM717010124MA  SM717010125MA  SM717010126MA  SM717010127MA  SM717010128MA  SM717010129MA  SM717010130MA  SM717010131MA  SM717010132MA  SM717010133MA  SM717010134MA  SM717010135MA  SM717010136MA  SM717010137MA  SM717010138MA  SM717010139MA  SM717010140MA  SM717020039MA  SM717100043MA  SM717100044MA  SM717100045MA  SM717010116MA  SM717010119MA  SM716300027MA  SM716480058MA  SM717100047MA  SM717100048MA  SM717100049MA  SM716480055MA  SM717100042MA  SM716200046MA  SM716200047MA  SM716430046MA  SM717010109MA  SM716310001MA  SM716430066MA  SM716430067MA  SM716430068MA  SM716390011MA  SM716390012MA  SM716270001MA  SM717020047MA  SM717020048MA  SM716200033MA  SM717010104MA  SM716270008MA  SM716250022MA  SM716250023MA  SM716430063MA  SM716430064MA  SM716430065MA  SM716300032MA  SM716300033MA  SM716300034MA  SM716300035MA  SM716300036MA  SM716300037MA  SM716300038MA  SM716430047MA  SM717100039MA  SM717100040MA  SM716200048MA  SM716200049MA  SM716200050MA  SM716250016MA  SM716250017MA  SM716250018MA  SM717100041MA  SM716100021MA  SM716480056MA  SM717020036MA  SM716380017MA  SM717020057MA  SM716380015MA  SM717080055MA  SM716480045MA  SM716480046MA  SM716480047MA  SM716480048MA  SM716480049MA  SM716480050MA  SM716430050MA  SM716430051MA  SM716460006MA  SM716460007MA  SM716390006MA  SM716390007MA  SM716390008MA  SM717020023MA  SM717020024MA  SM717020025MA  SM716300044MA  SM716300045MA  SM716460023MA  SM716460024MA  SM716460025MA  SM716460026MA  SM716460027MA  SM716460028MA  SM716460029MA  SM716460030MA  SM716460031MA  SM716460032MA  SM716460033MA  SM716460034MA  SM716460035MA  SM716460036MA  SM716460037MA  SM716460038MA  SM716460039MA  SM716460040MA  SM716460041MA  SM716460042MA  SM716460063MA  SM716460064MA  SM716460065MA  SM716460066MA  SM716460067MA  SM716460068MA  SM716480034MA  SM716480035MA  SM716480036MA  SM716300039MA  SM717100046MA  SM716460070MA  SM717080060MA  SM716100022MA  SM716100023MA  SM716100025MA  SM716100026MA  SM716100029MA  SM716100030MA  SM716100031MA  SM716100032MA  SM716100033MA  SM716100034MA  SM716100035MA  SM716200042MA  SM716380018MA  SM716380019MA  SM716380020MA  SM716100027MA  SM716100028MA  SM716200032MA  SM716200034MA  SM716200035MA  SM716200036MA  SM716200037MA  SM716200038MA  SM716200039MA  SM716200040MA  SM716200041MA  SM716200043MA  SM716200044MA  SM716250019MA  SM716250020MA  SM716430031MA  SM716430033MA  SM716430034MA  SM716430035MA  SM717100034MA  SM716100024MA  SM716430032MA  SM716250021MA  SM716300020MA  SM716460010MA  SM716460011MA  SM716460012MA  SM716460013MA  SM716460014MA  SM715500017WA  SM716200002WA  SM716200014WA  SM716380016MA  SM716390009MA  SM716390010MA  SM717020052MA  SM717020053MA  SM716350005WA  SM716370037WA  SM716230001WA  SM716230002WA  SM716230003WA  SM716250001WA  SM716250002WA  SM716250003WA  SM715440002WA  SM715390003WA  SM716420007WA  SM716070001WA  SM716320009WA  SM716320010WA  SM716290019WA  SM716190026WA  SM716450029WA  SM716450030WA  SM716450031WA  SM716240001WA  SM717060010WA  SM716230012WA  SM716420006WA  SM716420008WA  SM716460073WA  SM716350004WA  SM716290020WA  SM715480019WA  SM715480020WA  SM715490012WA  SM715490013WA  SM715490014WA  SM715490015WA  SM716050005WA  SM716330020WA  SM716330021WA  SM716340012WA  SM716470006WA  SM716470007WA  SM716470008WA  SM716470009WA  SM716470010WA  SM716470011WA  SM716470012WA  SM716490006WA  SM716490007WA  SM716490008WA  SM716490009WA  SM717020059WA  SM717120001WA  SM715360006WA  SM715490019WA  SM715490020WA  SM716010009WA  SM716030002WA  SM716040005WA  SM716040006WA  SM716040007WA  SM716040008WA  SM716040009WA  SM716040010WA  SM716160001WA  SM716240004WA  SM716240005WA  SM716240006WA  SM716240007WA  SM716240008WA  SM716290014WA  SM716290015WA  SM716370010WA  SM716440029WA  SM716460020WA  SM716460071WA  SM716460072WA  SM716490002WA  SM716490003WA  SM716490004WA  SM717010145WA  SM717100067WA  SM717110023WA  SM717110024WA  SM717110025WA  SM715360014WA  SM715480021WA  SM715480022WA  SM715490001WA  SM716010004WA  SM716010005WA  SM716160002WA  SM716160003WA  SM716160005WA  SM716300028MA  SM716300029MA  SM716300030MA  SM716430056MA  SM716430057MA  SM716430058MA  SM716430059MA  SM716430060MA  SM715250029WA  SM716360036WA  SM716360014WA  SM716290002WA  SM716360015WA  SM716360016WA  SM716410011WA  SM717020017WA  SM717020058WA  SM717120003WA  SM716080021WA  SM716290001WA  SM716190024WA  SM716280003WA  SM716290003WA  SM716360013WA  SM716410001WA  SM716410002WA  SM716410003WA  SM716410004WA  SM716410005WA  SM716410006WA  SM716410007WA  SM716410008WA  SM716410009WA  SM716410010WA  SM716300054WA  SM717060014WA  SM717060015WA  SM715490022WA  SM715490023WA  SM715380007WA  SM717100036MA  SM717080056MA  SM717080057MA  SM717080058MA  SM717080059MA  SM717100033MA  SM716200045MA  SM717020049MA  SM716430070MA  SM716270004MA  SM717060030MA  SM717060031MA  SM717060032MA  SM717060033MA  SM717060034MA  SM717060035MA  SM717060036MA  SM717060037MA  SM716050013WA  SM716080005WA  SM716080007WA  SM716160006WA  SM716160007WA  SM716160008WA  SM716300042MA  SM716300043MA  SM716080014WA  SM715350090WA  SM715460011WA  SM715460012WA  SM716160009WA  SM715360008WA  SM715370002WA  SM715370003WA  SM715440003WA  SM716050003WA  SM716050004WA  SM716050007WA  SM716050009WA  SM716050010WA  SM716050011WA  SM716190017WA  SM716240020WA  SM716240021WA  SM716240022WA  SM716240023WA  SM716240024WA  SM716280019WA  SM716320014WA  SM716330009WA  SM716360009WA  SM716360017WA  SM716370012WA  SM716370013WA  SM716370014WA  SM716370016WA  SM716370017WA  SM716370018WA  SM716370021WA  SM716370022WA  SM716370023WA  SM716370024WA  SM716440019WA  SM716440021WA  SM717110026WA  SM717110027WA  SM717110028WA  SM717110029WA  SM717110030WA  SM717110031WA  SM717110032WA  SM717110033WA  SM717110034WA  SM717110035WA  SM715330050WA  SM716050002WA  SM716100036MA  SM716080006WA  SM715490018WA  SM715490021WA  SM716080002WA  SM716080003WA  SM716080004WA  SM716210004WA  SM716440002WA  SM716350003WA  SM716490001WA  SM715420004WA  SM716490020WA  SM716450027WA  SM716280020WA  SM716280021WA  SM715470001WA  SM715470002WA  SM716070004WA  SM716070005WA  SM717080064WA  SM716280014WA  SM716280017WA  SM716280018WA  SM717110008WA  SM715430005WA  SM715470020WA  SM715480015WA  SM716460017WA  SM716220006WA  SM716220005WA  SM715420009WA  SM716200017WA  SM716340013WA  SM716420013WA  SM717060009WA  SM717060019WA  SM717020037MA  SM717020038MA  SM717060017WA  SM717060018WA  SM716450016WA  SM716020008WA  SM716290004WA  SM716290005WA  SM716460016WA  SM716470001WA  SM716460018WA  SM716460019WA  SM716090001WA  SM716140004WA  SM716190020WA  SM716210002WA  SM716210003WA  SM716240044WA  SM716240045WA  SM716290021WA  SM716300001WA  SM717060016WA  SM716370039WA  SM715430006WA  SM716190025WA  SM715340056WA  SM716050006WA  SM716090002WA  SM716140003WA  SM716420014WA  SM716420015WA  SM716420016WA  SM715420010WA  SM715420011WA  SM716160011WA  SM716200015WA  SM716200016WA  SM716440032WA  SM717040001WA  SM717040002WA  SM717040003WA  SM717090007WA  SM715500014WA  SM716070008WA  SM716200051WA  SM716200052WA  SM716200053WA  SM716250013WA  SM716250014WA  SM716250015WA  SM715270035WA  SM716480064WA  SM716440031WA  SM715350073WA  SM715350074WA  SM716040001WA  SM716040002WA  SM716040003WA  SM716040004WA  SM716080008WA  SM716260001WA  SM716260002WA  SM716080017WA  SM716080018WA  SM716080019WA  SM716080020WA  SM716460076WA  SM716460077WA  SM716010001WA  SM716010002WA  SM716190028WA  SM715030009WA  SM715280039WA  SM715290047WA  SM715330052WA  SM715490027WA  SM715500015WA  SM715500016WA  SM716070009WA  SM716070010WA  SM716070011WA  SM716080001WA  SM716240042WA  SM716360007WA  SM717010154WA  SM717020008WA  SM717020009WA  SM717020010WA  SM717090001WA  SM716020004WA  SM715270037WA  SM715290046WA  SM716190011WA  SM716190012WA  SM716500001WA  SM716500002WA  SM716500003WA  SM716500004WA  SM716500005WA  SM716500006WA  SM716500007WA  SM716500008WA  SM716500009WA  SM716500012WA  SM716500013WA  SM716320011WA  SM716320012WA  SM716320013WA  SM716240041WA  SM716360010WA  SM716300052WA  SM716300053WA  SM717090009WA  SM715260032WA  SM715280040WA  SM716270011WA  SM716410013WA  SM716410014WA  SM715260031WA  SM715270036WA  SM717060024WA  SM716070007WA  SM717010142WA  SM715470004WA  SM715470005WA  SM715470006WA  SM715470007WA  SM715470008WA  SM715470009WA  SM715470010WA  SM715470011WA  SM715470012WA  SM715470013WA  SM715470014WA  SM715470015WA  SM715470016WA  SM715470017WA  SM715470018WA  SM715470019WA  SM715280038WA  SM716070012WA  SM",
      "center_classification_date": "20170707",
      "distribution_pattern": "US Distribution including PR and to the states of :  AL, AR, AZ, CA, CO, CT, FL GA, ID,IL, IN, KY, LA, , MA, MD, MI, MO, MS, MT,  MN, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX,  VA, WA, WI, and UT.      OUS:     Algeria  Argentina  Australia  Austria  Belgium  Bolivia  Bosnia and Herzegovina  Brazil  Bulgaria  Cambodia  Canada  Chile  Colombia  Costa Rica  Cote d'Ivoire  Croatia  Czech Republic  Ecuador  Egypt  El Salvador  Estonia  Finland  France  Germany  Greece  Honduras  Hong Kong  Hungary  India  Indonesia  Iraq  Ireland  Israel  Italy  Japan  Jordan  Kenya  Korea (Republic Of)  Kuwait  Latvia  Lebanon  Lithuania  Malaysia  Maldives  Malta  Mauritius  Mexico  Moldova  Morocco  Myanmar  Netherlands  New Zealand    Nigeria  Norway  Oman  Pakistan  Paraguay  Peru  Philippines  Poland  Portugal  Romania  Russia  Saudi Arabia  Senegal  Serbia  Singapore  Slovakia  Slovenia  South Africa  Spain  Sweden  Switzerland  Taiwan  Tanzania  Thailand  Turkey  United Arab Emirates  United Kingdom  Uruguay  Vietnam  Zimbabwe",
      "state": "WI",
      "product_description": "Carestation 620 A1, Carestation 650 A1, and Carestation 650c A1",
      "report_date": "20170719",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-2703-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77244",
      "termination_date": "20180213",
      "more_code_info": "716320019WA  SM716320020WA  SM716320021WA  SM716320022WA  SM716420029WA  SM716420030WA  SM716420031WA  SM716430001WA  SM716430002WA  SM716430003WA  SM716430004WA  SM716430005WA  SM716430006WA  SM716430007WA  SM716430008WA  SM716430009WA  SM716430010WA  SM716430011WA  SM716430014WA  SM716430015WA  SM716430016WA  SM716430017WA  SM716430018WA  SM716430019WA  SM716430020WA  SM716430021WA  SM716430022WA  SM716430023WA  SM716430024WA  SM716430025WA  SM716440012WA  SM716440013WA  SM716440014WA  SM716440015WA  SM716440016WA  SM716440017WA  SM716440018WA  SM716010010WA  SM716300003WA  SM716320002WA  SM716460008MA  SM716460009MA  SM716460069MA  SM716480037MA  SM716480038MA  SM716480039MA  SM716480040MA  SM716480041MA  SM716480042MA  SM716480043MA  SM717080053MA  SM717080054MA  SM716300023MA  SM716270009MA  SM716270010MA  SM715290045WA  SM715420008WA  SM715470021WA  SM715470022WA  SM715480001WA  SM715480002WA  SM715480010WA  SM715480011WA  SM715480012WA  SM716100048WA  SM716100049WA  SM716300019WA  SM716150011WA  SM716010006WA  SM716020001WA  SM716020002WA  SM716020003WA  SM716020005WA  SM716020006WA  SM716020007WA  SM716240043WA  SM716320008WA  SM717080073WA  SM717080074WA  SM717080075WA  SM717080076WA  SM717090008WA  SM715450003WA  SM716240026WA  SM716240027WA  SM716240028WA  SM716240029WA  SM717080070WA  SM715420001WA  SM715420002WA  SM715420003WA  SM716330010WA  SM715490002WA  SM716300006WA  SM716420017WA  SM716360006WA  SM715260033WA  SM715330049WA  SM715340064WA  SM716220002WA  SM716510001WA  SM716510003WA  SM716510004WA  SM717110015WA  SM715510005WA  SM715510006WA  SM716360012WA  SM717030001WA  SM717080071WA  SM717100003WA  SM717100004WA  SM716230009WA  SM716230010WA  SM716330017WA  SM716370009WA  SM715350076WA  SM716360008WA  SM717060026WA  SM716510002WA  SM715480008WA  SM715500018WA  SM715500019WA  SM715500020WA  SM715500021WA  SM715500022WA  SM715510001WA  SM715510002WA  SM715510003WA  SM715510004WA  SM716240018WA  SM716240019WA  SM716440020WA  SM717080002WA  SM717110036WA  SM716100051WA  SM716100052WA  SM716240003WA  SM716440005WA  SM715490003WA  SM715490004WA  SM717100017WA  SM717100018WA  SM717100059WA  SM717100060WA  SM717100061WA  SM717100062WA  SM717100063WA  SM717100064WA  SM717110006WA  SM717110007WA  SM715500009WA  SM715500010WA  SM715500011WA  SM715500012WA  SM715500013WA  SM717080041MA  SM717080042MA  SM717080043MA  SM716190002WA  SM715480003WA  SM716150002WA  SM716510005WA  SM716510006WA  SM716510007WA  SM716510008WA  SM716510009WA  SM716510030WA  SM716510031WA  SM716510032WA  SM716510033WA  SM716510034WA  SM716520001WA  SM716520002WA  SM716520003WA  SM716520004WA  SM716520005WA  SM716520006WA  SM716520007WA  SM716520008WA  SM715360005WA  SM715360093WA  SM715500005WA  SM715500006WA  SM716420028WA  SM717020007WA  SM716470002WA  SM716470003WA  SM717090002WA  SM716300055WA  SM716440001WA  SM717010105MA  SM717010106MA  SM717010107MA  SM717100013WA  SM717100014WA  SM717100015WA  SM717060007WA  SM717060008WA  SM716100007WA  SM716150020WA  SM716150021WA  SM716150022WA  SM716150023WA  SM716490005WA  SM715280041WA  SM715380003WA  SM716020009WA  SM716140005WA  SM716400001WA  SM717080003WA  SM717080004WA  SM717100005WA  SM717100006WA  SM717100007WA  SM717100008WA  SM716300011WA  SM716300012WA  SM716300013WA  SM716300014WA  SM716300015WA  SM716300016WA  SM716300017WA  SM716300018WA  SM716300051WA  SM716100005WA  SM716100006WA  SM716280004WA  SM716280005WA  SM716480066WA  SM716480067WA  SM716480068WA  SM716480069WA  SM716480070WA  SM717040004WA  SM716010011WA  SM716090017WA  SM716450026WA  SM715480013WA  SM715480014WA  SM716400002WA  SM717010147WA  SM717010148WA  SM717090024WA  SM715520007WA  SM716040018WA  SM716470016WA  SM717020005WA  SM716110015WA  SM716270015WA  SM716340002WA  SM716470015WA  SM717010150WA  SM716480002WA  SM716340015WA  SM716110016WA  SM716470014WA  SM717070015WA  SM716330019WA  SM717010149WA  SM717020002WA  SM716480017WA  SM717020011WA  SM717090011WA  SM716040020WA  SM716360005WA  SM716440003WA  SM716440004WA  SM716450011WA  SM716450024WA  SM716480013WA  SM716480020WA  SM717020004WA  SM716290018WA  SM717010153WA  SM716110005WA  SM716470013WA  SM717080072WA  SM717090010WA  SM716370025WA  SM716450033WA  SM717070011WA  SM716480019WA  SM717090018WA  SM716410018WA  SM716460005WA  SM716040012WA  SM716120001WA  SM717020006WA  SM716460074WA  SM717090014WA  SM717090017WA  SM717100001WA  SM716150008WA  SM717090016WA  SM717100002WA  SM717060025WA  SM717020003WA  SM716320023WA  SM716340007WA  SM716220007WA  SM716380010WA  SM715520006WA  SM716420003WA  SM716350001WA  SM716340003WA  SM717090022WA  SM715520002WA  SM716110023WA  SM716330005WA  SM717090013WA  SM717100011WA  SM716110004WA  SM716220001WA  SM716290017WA  SM716320025WA  SM717090026WA  SM715520003WA  SM716340006WA  SM716450032WA  SM716110014WA  SM716290011WA  SM716040016WA  SM715520004WA  SM715520005WA  SM716480001WA  SM716100047WA  SM716290012WA  SM716290013WA  SM717090015WA  SM716460021WA  SM716460022WA  SM716150006WA  SM716230004WA  SM716360019WA  SM716460075WA  SM716410016WA  SM716450017WA  SM716340017WA  SM717100010WA  SM716430026WA  SM716480007WA  SM716340004WA  SM716480008WA  SM716450025WA  SM716320001WA  SM717020012WA  SM715510011WA  SM716150007WA  SM716150009WA  SM716440006WA  SM716040019WA  SM716290010WA  SM716420023WA  SM716420024WA  SM716450012WA  SM716470004WA  SM716470005WA  SM716480009WA  SM717010151WA  SM717090019WA  SM717090025WA  SM717090027WA  SM717090029WA  SM715510012WA  SM715520001WA  SM715520008WA  SM716040011WA  SM716040013WA  SM716040014WA  SM716040015WA  SM716040017WA  SM716120002WA  SM716120003WA  SM716130001WA  SM716150010WA  SM716280001WA  SM716280002WA  SM716290016WA  SM716340001WA  SM716340005WA  SM716340014WA  SM716340016WA  SM716420004WA  SM716420025WA  SM716450020WA  SM716480004WA  SM716480005WA  SM716480006WA  SM716480014WA  SM716480015WA  SM717010152WA  SM717060020WA  SM717090020WA  SM717090023WA  SM716480016WA  SM716300005WA  SM717090021WA  SM716210001WA  SM717020001WA  SM717090012WA  SM716320024WA  SM716410019WA  SM716480003WA  SM716320026WA  SM716350002WA  SM717090003WA  SM717010108MA  SM716190027WA  SM716230005WA  SM716480010WA  SM716480011WA  SM716480012WA  SM717080028MA  SM717080029MA  SM717080030MA  SM716080013WA  SM716110018WA  SM716110019WA  SM716110020WA  SM716110021WA  SM716110022WA  SM716160010WA  SM716190004WA  SM716360029WA  SM716360030WA  SM716360031WA  SM716360032WA  SM716360033WA  SM716360034WA  SM716360035WA  SM717020013WA  SM717020014WA  SM717020015WA  SM716460052MA  SM716460053MA  SM716460054MA  SM716460055MA  SM716460056MA  SM716460057MA  SM716460058MA  SM716460059MA  SM716460060MA  SM716460061MA  SM716460062MA  SM716190022WA  SM716190023WA  SM716420005WA  SM716420009WA  SM716420010WA  SM716480065WA  SM716200031MA  SM717020045MA  SM717020046MA  SM717080018MA  SM717080019MA  SM717080020MA  SM717080021MA  SM717080022MA  SM717080023MA  SM717080024MA  SM717080025MA  SM717080026MA  SM717080027MA  SM715380005WA  SM716080009WA  SM716370011WA  SM716370029WA  SM716370030WA  SM716370031WA  SM716370032WA  SM716370033WA  SM716370034WA  SM716370035WA  SM716370036WA  SM716370040WA  SM716380001WA  SM716380002WA  SM716380003WA  SM716380004WA  SM716380005WA  SM715380006WA  SM715420012WA  SM715420013WA  SM715340057WA  SM715360016WA  SM715420005WA  SM715420006WA  SM715420007WA  SM716240017WA  SM716300002WA  SM716300004WA  SM716340018WA  SM716340019WA  SM716350006WA  SM716350007WA  SM716350008WA  SM716350009WA  SM716350010WA  SM716350011WA  SM716350012WA  SM716350013WA  SM716380014WA  SM716420026WA  SM716420027WA  SM716450013WA  SM716450014WA  SM716450015WA  SM717100016WA  SM717100068WA  SM717100069WA  SM717100070WA  SM717080051MA  SM717080052MA  SM717080039MA  SM717080040MA  SM716430061MA  SM716430062MA  SM716480054MA  SM716430048MA  SM716430049MA  SM717130001WA  SM715400001WA  SM715400001WA  SM716070006WA  SM716160004WA  SM716070003WA  SM716300022MA  SM716300047MA  SM715330051WA  SM715360009WA  SM715380008WA  SM715390001WA  SM716100020WA  SM716100041WA  SM716100042WA  SM716240025WA  SM716490017WA  SM716490018WA  SM716490019WA  SM715280043WA  SM715280044WA  SM715350087WA  SM716190021WA  SM716330018WA  SM716340009WA  SM716340010WA  SM716340011WA  SM715390002WA  SM715480009WA  SM716430028WA  SM716150012WA  SM716150013WA  SM716150014WA  SM716150015WA  SM717010143WA  SM716190018WA  SM716190019WA  SM716220003WA  SM715480016WA  SM715480017WA  SM715480018WA  SM716320017WA  SM716320018WA  SM716070002WA  SM716150001WA  SM715440001WA  SM716260003WA  SM716390001WA  SM716100050WA  SM716200018WA  SM716390004WA  SM716390005WA  SM716050014WA  SM717090028WA  SM715280042WA  SM716150019WA  SM716290007WA  SM716450028WA  SM717010141WA  SM715510007WA  SM715430004WA  SM716300021MA  SM715350079WA  SM715480007WA  SM716050001WA  SM716100043WA  SM716210005WA  SM716320027WA  SM716320028WA  SM716320029WA  SM716320030WA  SM716320031WA  SM716420018WA  SM716450018WA  SM716450019WA  SM716450021WA  SM716450022WA  SM716450023WA  SM717020041MA  SM717020042MA  SM717020043MA  SM717080044MA  SM717080045MA  SM717080046MA  SM717080047MA  SM717080048MA  SM715330054WA  SM715380004WA  SM716420001WA  SM716440030WA  SM716480018WA  SM717080001WA  SM716150018WA  SM716150016WA  SM715340055WA  SM716240040WA  SM716020011WA  SM716460001WA  SM716460002WA  SM715330053WA  SM716020010WA  SM716020012WA  SM715450005WA  SM716320005WA  SM716230011WA  SM716360018WA  SM716190003WA  SM716440026WA  SM716150017WA  SM716320003WA  SM716320004WA  SM716320006WA  SM716320007WA  SM716440022WA  SM716440023WA  SM716440024WA  SM716440027WA  SM716440028WA  SM716460003WA  SM716460004WA  SM716200019WA  SM716320015WA  SM716320016WA  SM716370004WA  SM715360010WA  SM715370004WA  SM715380001WA  SM715380002WA  SM715510008WA  SM715510009WA  SM715510010WA  SM716150004WA  SM716150005WA  SM716240009WA  SM716240010WA  SM716240011WA  SM716240012WA  SM716370001WA  SM716370002WA  SM716370003WA  SM716490012WA  SM716490013WA  SM716490014WA  SM716490015WA  SM716490016WA  SM715480006WA  SM717020032MA  SM716250024MA  SM716250025MA  SM716270005MA  SM716300048MA  SM716300049MA  SM716380021MA  SM717020031MA  SM716430012MA  SM716430013MA  SM716100037MA  SM716100038MA  SM716180008MA  SM716270003MA  SM717010110MA  SM717010111MA  SM717010112MA  SM717010113MA  SM717010114MA  SM716500010WA  SM716500011WA  SM716110024WA  SM716370038WA  SM716190007WA  SM716190008WA  SM716210006WA  SM716210007WA  SM716280015WA  SM716280016WA  SM716330022WA  SM716400003WA  SM715360007WA  SM716010003WA  SM716290008WA  SM716290009WA  SM716380006WA  SM716380007WA  SM716380008WA  SM716490010WA  SM716490011WA  SM716500014WA  SM716500015WA  SM716500016WA  SM717110044WA  SM717110045WA  SM717110046WA  SM715460010WA  SM717020033MA  SM717020034MA  SM717020035MA  SM717020054MA  SM717080036MA  SM717080037MA  SM717090005WA  SM717090006WA  SM717090004WA  SM717080031MA  SM717080032MA  SM715480005WA  SM717080033MA  SM717080034MA  SM717080035MA  SM716340008WA  SM717020055MA  SM717020056MA  SM717080049MA  SM717080050MA  SM716380011WA  SM716440011WA  SM716440007WA  SM716440008WA  SM716440009WA  SM716440010WA  SM716420011WA  SM716420012WA  SM716410012WA  SM717080038MA  SM716110026WA  SM716110027WA  SM716240002WA  SM717060011WA  SM717110001WA  SM717110002WA  SM716520009WA  SM716100004WA  SM715340060WA  SM715340061WA  SM715340063WA  SM715450001WA  SM716110006WA  SM716110007WA  SM716110008WA  SM716110009WA  SM716110010WA  SM716110011WA  SM716110012WA  SM716110013WA  SM716440033WA  SM716440034WA  SM716440035WA  SM716440036WA  SM716440037WA  SM716440038WA  SM716440039WA  SM716440040WA  SM716440041WA  SM716440042WA  SM716450001WA  SM716450002WA  SM716450003WA  SM716450004WA  SM716450005WA  SM716450006WA  SM716450007WA  SM716450008WA  SM716450009WA  SM716450010WA  SM716490023WA  SM716490024WA  SM716490025WA  SM716490026WA  SM716490027WA  SM716490028WA  SM716490029WA  SM717070005WA  SM717070006WA  SM717120002WA  SM716490033WA  SM715490005WA  SM715490006WA  SM716160012WA  SM716160013WA  SM716160014WA  SM716160015WA  SM716160016WA  SM716160017WA  SM716160018WA  SM716160020WA  SM716160022WA  SM716160023WA  SM716160024WA  SM716160025WA  SM716160026WA  SM716160027WA  SM716160028WA  SM716160029WA  SM716160030WA  SM716160031WA  SM716160032WA  SM716160033WA  SM715460013WA  SM716090018WA  SM716090019WA  SM716090020WA  SM716090021WA  SM716090022WA  SM716090023WA  SM716090024WA  SM716090025WA  SM716090026WA  SM716090027WA  SM716420002WA  SM716410017WA  SM715370001WA  SM715460014WA  SM715460015WA  SM716110017WA  SM715450002WA  SM716150024WA  SM716140002WA  SM717100012WA  SM717110037WA  SM716080022WA  SM715340058WA  SM716140001WA  SM716370027WA  SM716280006WA  SM716280007WA  SM716280008WA  SM715410001WA  SM715410002WA  SM715410003WA  SM716200008WA  SM716200009WA  SM716520010WA  SM716520011WA  SM717100009WA  SM716370026WA  SM715360013WA  SM716010007WA  SM715260030WA  SM715350080WA  SM715350081WA  SM715450004WA  SM716010008WA  SM716090028WA  SM716090029WA  SM716100001WA  SM716100002WA  SM716100003WA  SM716110025WA  SM716170001WA  SM716240013WA  SM716240014WA  SM716240015WA  SM716240016WA  SM716270016WA  SM716270017WA  SM716280009WA  SM716280010WA  SM716280011WA  SM716280012WA  SM716280013WA  SM716360024WA  SM716360025WA  SM716360026WA  SM716360027WA  SM716360028WA  SM716390002WA  SM716390003WA  SM716430029WA  SM716440025WA  SM716490021WA  SM716490022WA  SM716200030MA  SM716430036MA  SM716430037MA  SM716430038MA  SM716430039MA  SM716430040MA  SM716430041MA  SM716430042MA  SM716430043MA  SM716430044MA  SM716430045MA  SM717100019MA  SM717100020MA  SM717100021MA  SM717100022MA  SM717100023MA  SM717100024MA  SM717100025MA  SM717100026MA  SM717100027MA  SM717100028MA  SM717100029MA  SM717100030MA  SM717100031MA  SM717100032MA  SM716050008WA  SM716050012WA  SM716180003MA  SM716180004MA  SM716180005MA  SM716180006MA  SM716180007MA  SM716180010MA  SM716200020MA  SM716200021MA  SM716200022MA  SM716200023MA  SM716200024MA  SM716200025MA  SM716200026MA  SM716200027MA  SM716200028MA  SM715340059WA  SM716220008WA  SM716220009WA  SM716270002MA  SM716460015MA  SM715500001WA  SM716080010WA  SM716100008WA  SM716100009WA  SM716100010WA  SM716100011WA  SM716100012WA  SM716100013WA  SM716100014WA  SM716100015WA  SM716100016WA  SM716100017WA  SM716100018WA  SM716100019WA  SM716200012WA  SM716330011WA  SM716330012WA  SM716330013WA  SM716330014WA  SM716330015WA  SM716330016WA  SM716370005WA  SM716370006WA  SM716370007WA  SM716370008WA  SM716370019WA  SM716370020WA  SM716420019WA  SM716420020WA  SM717110042WA  SM717110043WA  SM715490008WA  SM715490017WA  SM716080015WA  SM716110001WA  SM716110002WA  SM716110003WA  SM716150025WA  SM716150026WA  SM716230006WA  SM716230007WA  SM716230008WA  SM716240030WA  SM716240031WA  SM716240032WA  SM716240033WA  SM716240034WA  SM716240036WA  SM716240037WA  SM716240038WA  SM716240039WA  SM716420021WA  SM716430027WA  SM717010146WA  SM715020005WA  SM715030006WA  SM715450006WA  SM715450007WA  SM715450008WA  SM715490016WA  SM715490024WA  SM716270012WA  SM716270013WA  SM716270014WA  SM716330001WA  SM716330002WA  SM716330003WA  SM716330006WA  SM716330007WA  SM716330008WA  SM716360020WA  SM716360021WA  SM716360022WA  SM716360023WA  SM716420022WA  SM717060021WA  SM717060022WA  SM717060023WA  SM717110003WA  SM717110004WA  SM717110005WA  SM717110009WA  SM717110010WA  SM717110011WA  SM717110012WA  SM717110013WA  SM716250004WA  SM716250005WA  SM716250006WA  SM716250007WA  SM716250008WA  SM716250009WA  SM716250010WA  SM716250011WA  SM716250012WA  SM715480004WA  SM715350086WA  SM715500003WA  SM715500004WA  SM716090003WA  SM716090004WA  SM716190029WA  SM716190030WA  SM716200001WA  SM716300007WA  SM716300008WA  SM716300009WA  SM716300010WA  SM717070001WA  SM717070002WA  SM717070003WA  SM717070004WA  SM717070007WA  SM717070008WA  SM717070009WA  SM717070010WA  SM716370028WA  SM716400004WA  SM716400005WA  SM717020016WA  SM715430003WA  SM715420014WA  SM715420015WA  SM715420016WA  SM715450009WA  SM715450010WA  SM715450012WA  SM715450013WA  SM715460001WA  SM715460002WA  SM715490009WA  SM715490010WA  SM715490011WA  SM715490028WA  SM716380012WA  SM716380013WA  SM716410015WA  SM716410020WA  SM716480061WA  SM716480062WA  SM716480063WA  SM717110014WA  SM717110016WA  SM717110017WA  SM717110018WA  SM717110019WA  SM717110020WA  SM717110021WA  SM717110022WA  SM717110038WA  SM717110039WA  SM717110040WA  SM717110041WA  SM716090005WA  SM716090006WA  SM716090007WA  SM716090008WA  SM716090009WA  SM716090010WA  SM716090011WA  SM716090012WA  SM716090013WA  SM716090014WA  SM716090015WA  SM716090016WA  SM717070012WA  SM717070013WA  SM717070014WA  SM717080005WA  SM717080006WA  SM717080007WA  SM717080008WA  SM717080009WA  SM717080010WA  SM717080011WA  SM717080012WA  SM717080013WA  SM717080014WA  SM717080015WA  SM717080016WA  SM717080017WA  SM717080061WA  SM717080062WA  SM717080063WA  SM717080065WA  SM717080066WA  SM717080067WA  SM717080068WA  SM717080069WA  SM715340067WA  SM715340069WA  SM715460004WA  SM715460005WA  SM715460006WA  SM715460007WA  SM715460008WA  SM715460009WA  SM715500007WA  SM715500008WA  SM716100044WA  SM716100045WA  SM716100046WA  SM716200010WA  SM716330004WA  SM717060027WA  SM717060028WA  SM717060029WA  SM717060038WA  SM716150003WA  SM717120004WA  SM717120005WA  SM717120006WA  SM715430001WA  SM715430002WA  SM715450011WA  SM715460003WA  SM715500002WA  SM716140006WA  SM715490007WA  SM716490030WA  SM716490031WA  SM716490032WA  SM715340065WA  SM715490025WA  SM715490026WA  SM716080011WA  SM716080012WA  SM716080016WA  SM716190005WA  SM716190006WA  SM716190009WA  SM716190010WA  SM716190013WA  SM716190014WA  SM716190015WA  SM716190016WA  SM716200011WA  SM716200013WA  SM716290006WA  SM716360001WA  SM716360002WA  SM716360003WA  SM716360004WA  SM716380009WA  SM717060001WA  SM717060002WA  SM717060003WA  SM717060004WA  SM717060005WA  SM717060006WA  SM717060012WA  SM717060013WA  SM716430069MA  SM717020044MA  SM716180001MA  SM716180002MA  SM716180009MA  SM716200029MA  SM717010100MA  SM717010101MA  SM717010102MA  SM717020050MA  SM717020051MA  SM716220004WA      Carestation 650C A1:    SM817080012MA  SM816350012MA  SM816350008MA  SM816350009MA  SM816350010MA  SM816350011MA  SM816240001WA  SM816240002WA  SM816240003WA  SM815490001WA  SM815490002WA  SM815500001WA  SM816160001WA  SM816160002WA  SM816160003WA  SM816160004WA  SM816160005WA  SM816210001WA  SM815480001WA  SM815480002WA  SM817010001WA  SM817100001WA  SM817100002WA  SM815460001WA  SM815460002WA  SM816240004WA  SM816240005WA  SM816200001WA  SM816200002WA  SM816200003WA  SM816200004WA  SM816200005WA  SM816230001WA  SM816270001WA  SM816270002WA  SM816350001WA  SM816350002WA  SM816350003WA  SM816350004WA  SM816350005WA  SM816350006WA  SM816350007WA  SM816360001WA  SM816380001WA  SM816380002WA  SM816380003WA  SM816080001WA  SM816100001WA  SM817080001MA  SM817080002MA  SM817080003MA  SM817080004MA  SM817080005MA  SM817080006MA  SM817080007MA  SM817080008MA  SM817080009MA  SM817080010MA  SM817080011MA  SM815360005WA  SM815360006WA  SM815360007WA  SM815360008WA  SM815360009WA  SM816490001WA  SM817110001WA  SM817110002WA  SM817110003WA  SM815020001WA  SM815050028WA  SM815340066WA  SM815340068WA  SM815360091WA  SM815360092WA  SM815420001WA  SM815420002WA  SM816240006WA",
      "recall_initiation_date": "20170502",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}