  "results": [
    {
      "status": "Terminated",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85871",
      "recalling_firm": "Steris Corporation",
      "address_1": "5960 Heisley Rd",
      "address_2": "N/A",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, and United Arab Emirates.",
      "recall_number": "Z-2700-2020",
      "product_description": "InnoWave Sonic Irrigator, There are currently six models (PCF, Pro, Ultra, Ultra+, Unity, and ECO) - Product Usage: STERIS InnoWave Sonic Irrigators are intended to thoroughly clean (remove tissue, blood, and other contaminants from) a variety of reusable surgical instruments. The equipment provides sonic irrigation (ultrasonic cavitation) to both interior and exterior of lumened instruments.",
      "product_quantity": "1283",
      "reason_for_recall": "A recent FDA inspection at the firm identified that the foil test performed during installation of the device does not have clear acceptance criteria.",
      "recall_initiation_date": "20200626",
      "center_classification_date": "20200724",
      "termination_date": "20201222",
      "report_date": "20200805",
      "code_info": "Serial numbers   11167539801  11167539802  11167539803  01167539804  12167539806  11167440102  01177735903  11167539811  01177735902  01177735901  01177735906  01177735904  01177735905  12167539808  01167539810  02177975001  12167539809  01177736803  01177736804  01177736805  01177736806  01177736808  02177975002  01177736807  02177956701  02177956702  02177956703  02177956704  02177956705  03177956706  02177956710  03177956707  03177956708  02177956711  03177956709  03177956901  03177956902  02177956907  03177956903  03177956904  03177956905  03177956906  03177957111  04177957105  04177957106  04177957107  04177957108  04177957109  04177957110  05178203702  05178203703  05178203704  05178203705  05178203701  05178203706  05178203707  05178203708  05178203709  05178203710  06178204001  06178204002  06178204003  06178204004  06178204006  06178204007  06178204009  01177736802  06178204010  04177957112  04178203711  06178460901  06178460902  06178460903  06178460904  06178460905  06178460906  06178460907  0817 87702 01  0817 87702 02  06178460908  06178460909  06178460910  06178460911  06178460912  06178460913  06178460914  06178460915  05178203713  08178478301  06178204011  08178478302  08178478303  06178204012  08178478304  08178478305  08178478306  08178478307  08178478308  06178460917  08178478309  08178478310  08178478311  08178478312  08178478314  08178478315  0917 88031 01  0917 88031 02  0917 88031 03  0917 88031 04  0917 88031 05  0817 87702 04  0817 87702 03  07178460918  09178427701  09178427702  09178427703  09178427704  09178427705  06178204005  09178803106  09178803107  09178803108  09178803109  09178803110  09178427706  09178427707  09178427708  09178427709  09178427710  09178803111  1017 88034 01  08178478317  1017 88034 02  0917 87702 06  09178478318  1017 88034 03  1017 88034 05  1017 88034 06  1017 88034 07  10178427712  0917 87702 07  1017 88034 08  1017 88034 09  1017 88034 10  1017 88034 11  1017 88034 12  1017 88034 13  1017 88034 14  08178478313  1017 88034 15  1217 88033 01  1217 88033 02  1217 88033 03  1217 88033 04  1217 88033 05  1217 88033 06  01188948301  01188948302  10178427713  01188948303  0917 87702 08  01188948304  01188948305  01188948306  01188948307  01188948308  01188948310  1217 88033 07  1117 88034 16  1117 88034 17  1117 88034 18  1217 88033 08  1217 88033 09  1217 88033 11  1217 88033 12  1217 88033 13  1217 88033 14  0917 87702 10  09178803112  01189087701  12178803316  12178803317  01189087702  01189087703  01189087704  02189087724  02189087723  01189087705  01189087706  02189087721  1017 88035 02  1017 88035 03  1017 88035 04  01189087707  1017 88035 05  01189087708  01189087709  01189087710  01189087711  01189087712  01189087713  1017 88035 06  01189087714  01189087715  01189087716  1017 88036 01  01189087718  01189087719  01189087720  03189220501  03189220502  01189087717  03189220503  03189220504  03189220516  03189220517  1017 88036 02  03189220511  03189220512  03189220513  03189220514  03189220515  04189322801  04189322802  04189322803  03189220505  03189220520  03189220519  03189322821  03189220506  03189220508  03189322822  03189220509  03189220510  04189322804  04189322805  04189322806  04189322807  04189322808  04189322809  03189322825  04189322810  03189322824  04189322811  03189322823  04189322812  04189322813  04189322815  04189322816  03189220521  04189322826  04189322817  04189322819  1017 88036 03  05189437921  04189322820  05189437923  05189437901  05189437902  05189437903  05189437906  05189437907  05189437908  05189437909  06189437924  05189437904  05189437905  05189437910  1017 88036 05  05189437911  05189437912  06189437925  05189437913  05189437914  05189437915  05189437916  05189437917  05189437918  05189437920  07189716601  07189716602  07189716608  06189558701  06189558706  06189558703  07189716610  06189558704  07189716609  06189558707  06189558705  07189686504  07189686502  07189686503  07189686505  07189716603  08189716612  07189716604  07189716605  06189558709  08189873406  07189716606  07189716607  08189686506  08189686507  06189558710  08189686508  08189686509  08189686510  07189686511  08189873408  08189843301  08189843302  08189843313  08189843314  08189843303  07189686512  08189843304  09189843305  09189843316  09189843306  09189843307  05189437922  08189686611  09189843308  09189843309  09189843311  09189843310  08178478316  09189843312  09189843101  08189686612  10189686601  10189686602  10189686603  10189686604  10189686606  10189686607  10189686608  10189686605  10189686614  10189686615  10189686609  10189686610  08189686613  10189855505  10189855507  10189855509  10189850401  10189855510  10189855511  10189855513  09189855501  10189855514  11189855515  11189855516  11189855517  08189873409  08189873410  11189855518  11189855519  09189855502  11189855605  10189898801  11189855606  11189855607  11189850406  12189928803  12189928804  10189898802  10189898803  09189855503  11189855609  11189855520  11189855521  12189855610  12189855611  12189855612  09189855504  10189898804  1017 88036 04  10189855601  10189898805  12189855613  12189855614  12189855616  11189855522  12189855617  12189855618  12189855619  10189855602  10189855603  0119 98703 09  12189855523  10189855604  12189855525  01199870311  12189855524  06189558708  01199870313  01199870314  08189873401  01199870310  01199870312  01199870315  01199870316  01199870317  01199870319  01199944403  01199961203  01199870318  01199870320  01199870321  01199855622  12189855620  12189855621  01199870322  01199855624  01199944401  01199961206  01199870323  01199899105  01199899106  01199961201  01199961202  01199961204  01199961205  01199855625  08189873402  11189870324  01199870301  11189870325  01199899107  01199899108  01199899109  01199899110  01199855623  01199870302  02199962401  02199962402  02199962403  02199962404  02199899114  02199899116  02199899117  02199899118  02199899119  02199899112  02199899113  02199962405  02199962406  02199899115  01199870303  01199870304  11189870327  08189873403  02199914101  02199914102  02199914103  02199914106  02199914107  02199870305  02199870307  11189870326  02199914108  02199870306  02199114112  02199914109  02199914111  02199914113  02199914115  02199914116  02199870308  02199899120  08189873404  08189873405  01199914201  02199962407  02199966303  02199966304  02199966305  02199966306  02199966307  02199914114  02199914117  01188948309  02199942614  02199942615  12189928802  02199966308  02199966302  10189855506  02199914118  02199942613  02199899121  02199899122  01199914202  03199942504  03199942505  03199942506  02199899124  1218 993 8204  03199942616  03199942617  02199899125  02199899126  03199989201  03199942618  03199942620  03199942621  01199914203  03199952505  03199952506  03199942619  03199942622  03199899128  01199914204  02199899123  03199942623  03199942624  03199942627  03199942626  03199942629  03199899127  03199899129  03199914002  03199942628  10189850404  03199942630  03199942631  03199942632  04199952605  03199914001  03199914003  03199914004  04199988803  04199988804  04199988805  04199988806  03199952503  04199952603  04199952604  04199952606  04199952607  04199952608  04199952609  12189938202  04199988807  03199914005  03199914006  03199952504  04199952610  04199952611  04199952612  04199952613  04199952614  04199952616  01199914205  04199914008  04199914009  03199914007  04199952615  04199914011  12189855615  04199952602  03199942501  04199988801  04199914012  03199942502  04199952617  04199952618  04199952619  04199942602  04199952620  04199952621  05199964313  05199964314  04199942603  04199942604  04199952622  05199964318  04199964205  04199964206  04199988903  04199988904  04199988905  05190013501  05190013502  05190013503  05190013504  04199914010  05199964317  05199964323  04199942601  03199942625  05199964315  05199964316  04199942605  04199942606  O5190013504  05199964319  05199964320  05199964321  05199964322  05199942607  03199942503  04199988802  05199964324  03199952501  04190012803  05199988901  05190010001  05190010002  05190010004  05190010005  05199964328  05199964329  05199942608  05190010003  12189938205  05199964326  05199964327  05199942611  03199952502  05199982509  05199942612  05199964330  05199942609  05199942610  05199982505  05199982506  05199964331  05199964332  05199952623  02199966301  05199982507  05199964333  05199964334  04199964207  05199952624  05199952625  05199952626  05199983203  05199983202  05199983204  05199983205  05199983206  04190012805  04199964208  06199983208  05199983201  04190012804  05199982508  06190024201  06190024202  06190024203  06190024204  05199983207  06199983209  06199983211  06199983212  05199952627  06190024205  06199983213  05199952629  04190012801  06199983210  06199983215  05199964201  03199942508  06190024301  06199983214  06199983217  06190001205  06199983218  05199952630  06199952631  05199964202  06199983219  06199983220  06199983221  06190024302  06199983222  06199983223  06199952632  06199952633  06199952634  06199983224  06199964302  06199964303  06199964304  06190024303  06190024305  06190024304  06199983225  07190001016  07199983226  07199983227  07199983228  06199964305  06199964306  06190024307  07190001017  05199964203  04190012802  07190001018  07190001019  07190001020  07190001022  07199983229  06199964307  06190024309  07190001021  07199983230  07199964308  07199964309  06190024308  09189843315  07190001207  07190001023  07190001024  07199964310  10189850402  07190025401  07190025402  07190025403  07190025404  07190025405  05199988902  07190001208  07190001209  07199964311  07199982701  01289938201  07190001210  05199982501  07190001025  07190001026  07190001028  07199982702  07199982703  07199982704  05199982502  07190001027  07190040301  07190040302  07190038101  07190038102  07190040305  08190001029  08190001030  07199982705  05199982504  07190009751  08190009753  08190009754  07190009752  05199982503  08190009755  08190001031  08190001032  08190001033  08190001034  07199982706  07199982707  06190001204  07190009746  03199942510  06190001202  08190009756  08190001038  06190001203  08190025801  08190009757  08189686513  08199982708  08190001035  08190001037  08199982709  07190009747  12189899101  08190001036  08190001039  06190001201  07190040303  07190040304  08190001040  08190001041  08190001042  08199982710  08190025804  08190009758  08199982711  08190025803  08190025805  08190001043  08190001044  08199982712  07190001212  08190009716  08199982713  08199982714  08190009759  08190009717  08199982715  08190058501  08190001045  08190009719  08199982716  08190009718  08199982717  03199942509  07190001213  07190001214  08190058504  08190009720  09190009721  08199982718  09190009725  08199982719  09190025501  08190025504  08190025505  09190009727  08199982720  08190025503  09190025506  09190009722  09190009723  09190001001  09190001002  09190025507  09190025508  09190009724  09190009726  09190025509  09190009728  09190009729  09190025510  08190058503  09190025511  09190009730  09190009731  06190024310  09190025512  09190009732  09190009733  09190025513  09190065201  09190065202  08190058502  09190039208  09190009734  09190025515  09190025516  09190009735  09190009737  09190009738  09190025517  09190009736  09190039503  09190039504  09190039505  09190039506  09190009739  10190009740  10190009741  09190001004  07190009749  09190061501  09190061502  09190061504  09190039507  10190009742  07190009748  09190001005  09190039508  09190039501  09190039502  09190039509  10190075501  09190039206  09190061503  10190009743  09190039510  09190001006  09190039210  09190001007  09190001008  10190054401  10190054402  10190009744  10190009745  10190025601  10190025602  08190058505  10190025606  10190025607  07190009750  09190039511  09190039512  10190075502  09190039207  09190039513  10190025603  10190025604  10190025608  09190001009  10190075503  09190039201  09190001010  09190039514  10190025605  09190025502  09190039515  09190039516  09190001011  10190054404  09190001012  10190025609  09190001013  09190039202  09190039517  10190025610  10190062604  10190054407  10190025611  10190025612  10190025614  09190039203  11190062606  10190054406  10190025616  11190025617  10190001014  10190054504  10190025613  10190009701  10190009702  10190009703  10190054405  09190039204  10190054501  10190025615  11190025618  11190025619  11190025620  11190025621  10190009704  10190054408  10190054409  11190025622  10190054410  11190025623  10190062602  10191154505  02199914104  10190009705  10190009706  12189899102  11190063601  11190063602  11190063603  11190063604  11190093801  11190093802  02199914105  10190009707  10190054411  11190025624  11190025625  11190025626  11190009708  11190009710  11190025627  11190025629  11190025628  11190025630  11190039601  09190061505  10190054412  10190054413  11190009711  09190039209  11190062609  10190054414  11190039602  11190039603  10190054608  10190062603  11190039604  11190009712  11190039605  11190039606  11190039607  11190039608  11190039609  11190039610  01190054601  11190063701  11190054506  11190054507  11190054508  11190093803  10190054415  12190039612  08190025802  10190054416  12190039611  12190039613  12190039614  12190039615  11190063702  11190090403  11190063703  10190054417  01190054602  10190054418  12190039616  12190039617  12190039618  01190054603  10190054502  10190054419  01190054605  12190039619  11190090404  10190054503  10190062601  11190054604  01200039623  12190039620  11190009713  01200077301  01200077302  01200039625  12190039621  10190062605  11190062607  11190062608  11190063705  11190063704  11190063606  0120 0063707  11190063605  01200039622  01200063708  01200039624  01200039626  01200039627  11190054606  11190063607  11190063608  01200039628  09190039205  01200063709  01200039629  01200039630  01200054311  01200054313  01200054314  01200054315  01200054316  01200054317  11190009715  10190054607  01200063710  01200108203  01200108204  01200108205  01200108206  12190108201  12190108202  11190063609  11190063610  11190009714  01200063711  01200096712  11190063706  01200054318  11190064001  01200054312  01200054319  11190063611  01200054320  01200077602  11190063614  11190063616  11190063613  11190062601  00190039508  11190063615  01200054321  01200054322  01200077601  01200077303  11190063617  11190025701  11190025702  11190025703  11190025704  11190063619  01200054323  01200054324  01200054326  12190025705  12190025706  12190077305  12190025707  12190025708  01200077502  01200077505  01200054325  12190077306  01200054327  12190025709  12190077307  12190077308  01200054328  01200054329  02200077606  12190077309  01200054330  02200054331  02200054333  01200063713  01200063714  01200077304  02200054332  02200054334  12190025710  12190025711  01200077603  02200054335  02200054336  12190025712  01200077501  01200063715  01200077605  02200054337  02200054338  12190025713  01200077503  02200077607  02200077608  02200077609  02200077612  02200054340  02200077614  02200064116  03200064117  01200077604  02200077613  02200077615  02200064111  02200077610  01200091343  02200064113  03200064118  03200064119  03200064120  02200064112  02200064114  02200064115  01200039101  12190077310  02200054339  03200064121  01200039102  10189855512  01200077504  11190090402  O6190024305  11190064002  03200100605  03200100606  03200100607  O9190025513  01200077508  02200077611  02200100604  03200100608  03200100609  11190064003  01200091344  03200064122  03200064123  03200064124  03200064125  03200064128  03200064129  01200039103  01200039104  01200039105  03200064126  03200064130  03200064131  03200064132  01200039106  02200039109  02200091358  03200100610  03200100611  03200100612  03200100613  03200064133  03200064134  02200039114  03200064127  02200039115  04200128902  03200064135  03200064136  03200064137  03200064138  03200064139  03200064140  04200091311  02200100601  03200100614  03200100615  03200100616  04200091312  04200091314  03200077617  03200077619  04200077623  04200091315  03200077618  04200077622  03200100617  02200091359  03200077506  03200077507  04200077621  04200091313  04200091316  04200091317  04200091318  02200100701  03200100618  04200091328  02200100702  04200077624  04200091363  04200091365  03200100619  04200077311  04200077312  04200077313  04200077315  02200100703  02200100704  03200077620  02200100602  04200077314  02200100603  04200077316  04200091319  04200091320  03200091336  04200091322  04200091335  05200077413  03200091337  05200091367  04200091321  04200091323  04200091327  04200091324  04200091325  04200091326  05200077411  05200077412  03200077401  03200091338  05200091368  04200077317  04200077625  04200077318  03200077402  04200091361  04200077319  04200091362"
    }
  ]
}