      "country": "United States",
      "city": "Reading",
      "address_1": "2400 Bernville Rd",
      "reason_for_recall": "Arrow is recalling additional lots that were identified as part of an active recall. Arrow is recalling the affected product due to the possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter.",
      "address_2": "",
      "product_quantity": "9,037 units in the U.S. and 4,500 units internationally",
      "code_info": "Device Listing # D149129  Material # ASK-45552-RWJ4  PR-45552-HPHNL",
      "center_classification_date": "20170707",
      "distribution_pattern": "Distributed to SC, AL, NJ, IN, MA, GA, CA, PA, AZ, VA , WA and Bangkok",
      "state": "PA",
      "product_description": "Arrow EVOLUTION Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology",
      "report_date": "20170719",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Arrow International Inc",
      "recall_number": "Z-2698-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77604",
      "termination_date": "20180906",
      "more_code_info": "",
      "recall_initiation_date": "20170612",
      "postal_code": "19605-9607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}