      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85908",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.",
      "recall_number": "Z-2696-2020",
      "product_description": "IntelliVue MX800 patient monitor, model no. 865240",
      "product_quantity": "28",
      "reason_for_recall": "Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.",
      "recall_initiation_date": "20200630",
      "center_classification_date": "20200723",
      "termination_date": "20240125",
      "report_date": "20200729",
      "code_info": "Serial numbers    DE50887321 DE50887434 DE50887128 DE50887025  DE50887196 DE50887429 DE50887125 DE50887043  DE50887303 DE50887420 DE50887106 DE50887164  DE50887322 DE50887607 DE50887061 DE50887179  DE50887425 DE50887427 DE50887142   DE50887446 DE50887152 DE50887064   DE50887445 DE50887137 DE50887063",
      "more_code_info": ""
    }
  ]
}