  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "KOMORNIKI",
      "state": "N/A",
      "country": "Poland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68594",
      "recalling_firm": "ARJOHUNTLEIGH POLSKA Sp. z.o.o.",
      "address_1": "UI. KS. PIOTRA WAWRZYNIAKA 2",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -  US Nationwide in the state of MA and the countries of Australia, Germany, and Italy.",
      "recall_number": "Z-2693-2014",
      "product_description": "Minstrel (with scale)    Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to  safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair.",
      "product_quantity": "18 Lifts",
      "reason_for_recall": "ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp",
      "recall_initiation_date": "20140728",
      "center_classification_date": "20140923",
      "termination_date": "20160913",
      "report_date": "20141001",
      "code_info": "Model Numbers: HMA0002 and HMB002-US;   Serial Numbers: MPLAT0409W1177 ;  MPLAT0410W1213 ;  MPLAT0410W1228 ;  MPLAT0410W1229 ;  MPLAT0410W1231 ;  MPLAT0410W1232 ;  MPLAT0412W1290 ;  MPLAT0412W1296 ;  MPLAT0412W1297 ;  MPLAT0502W1403 ;  MPLAT0502W1404 ;  MPLAT0503W1485 ;  MPLAT0503W1486 ;  MPLAT0504W1518 ;  MPLAT0504W1519 ;  MPLAT0504W1520 ;  MPLAT0504W1531"
    }
  ]
}