  "results": [
    {
      "status": "Terminated",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74967",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "N/A",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of LA, PA, WA, and the countries of: Australia, Denmark, France, India, Italy, and Japan.",
      "recall_number": "Z-2691-2016",
      "product_description": "HexaPOD evo RT System    Product Usage:  The intended use of the device is the control of accurate patient positioning with assistance of a 3D Tracking System in a radiotherapy environment.",
      "product_quantity": "13",
      "reason_for_recall": "Potentially unrecognized incorrect position of the treatment couch in 3D workflow, i.e. the HexaPOD has not moved fully to the 3D position.",
      "recall_initiation_date": "20160818",
      "center_classification_date": "20160830",
      "termination_date": "20170523",
      "report_date": "20160907",
      "code_info": "iGuide 2.2.0"
    }
  ]
}