{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Northbrook",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69034",
      "recalling_firm": "Nanosphere, Inc.",
      "address_1": "4088 Commercial Ave",
      "address_2": "N/A",
      "postal_code": "60062-1829",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including  AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.",
      "recall_number": "Z-2683-2014",
      "product_description": "The Verigene Gram-Positive Blood Culture Nucleic Acid Test (BC-UP) performed using the sample-to-result Verigene System is a qualitative, multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause bloodstream infection (BSI). BC-UP is performed directly on blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria. BC-UP detects and identifies the following bacterial genera and species: Staphylococcus spp., Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus lugdunensis, Enterococcus faecalis, Enterococcus faecium, Streptococcus spp., Streptococcus pneumoniae, Streptococcus pyogenes, Streptococcus agalactiae, Streptococcus anginosus, group and Listeria spp.   In addition, BC-UP detects the mnecA resistance marker, inferring mecA-mediated, methicillin resistance, and the vanA and vanB resistance markers, inferring vanA/vanB-mediated vancomycin resistance. In mixed growth, BC-UP does not specifically attribute van-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA-mediated methicillin resistance to either S. aureus or S. epidermidis.   BC-UP is indicated for use in conjunction with other clinical and laboratory findings to, aid in the diagnosis of bacterial bloodstream infections; however, is not to be used to, monitor these infections. Sub-culturing of positive blood cultures is necessary to recover, organisms for susceptibility testing, identification of organisms not detected by BC-UP, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.",
      "product_quantity": "4,440 units",
      "reason_for_recall": "Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.",
      "recall_initiation_date": "20131010",
      "center_classification_date": "20140923",
      "termination_date": "20150616",
      "report_date": "20141001",
      "code_info": "Product Number: 20-009-018; Lot Numbers and Expiration Dates: 1) Lot: 090613018B, Expiration Date: 03/05/14; 2) Lot: 091413018B, Expiration Date: 03/13/14; 3) Lot: 091713018C, Expiration Date: 03/16/14; 4) Lot: 091813018B, Expiration Date: 03/17/14; 5) Lot: 092113018C, Expiration Date: 03/20/14; 6) Lot: 100213018B, Expiration Date: 03/31/14"
    }
  ]
}