  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69148",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-2680-2014",
      "product_description": "The Ysio Systems with software version VC10    The Ysio enables radiographic and tomographic exposures of the whole Body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.",
      "product_quantity": "62",
      "reason_for_recall": "Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image.  If these artifacts appear in the region of interest, the examination may need to be repeated.",
      "recall_initiation_date": "20140827",
      "center_classification_date": "20140923",
      "termination_date": "20150330",
      "report_date": "20141001",
      "code_info": "Model numbers: 10281013, 10281163."
    }
  ]
}