  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69102",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including MO, TX, MD, AR, VA, KY, RI, NJ, MA, NY, LA, PA, VT, SC, CA, OH, FL, MN, MI and Internationally to  Netherlands, Canada, Finland, Austria, France, Malaysia, Denmark, Norway, UK, Australia, Belgium, New Zealand.",
      "recall_number": "Z-2678-2014",
      "product_description": "No packaging.  The device is a computed tomography scanner.      Intended to provide corss sectional images of the human body and visualization of the internal organs of the body.",
      "product_quantity": "61",
      "reason_for_recall": "When the Extended Display Firled ov View reconstruction is used, the CT numbers and anatomy in teh extended region are not accurate.",
      "recall_initiation_date": "20070109",
      "center_classification_date": "20140922",
      "termination_date": "20151106",
      "report_date": "20141001",
      "code_info": "Model 728243 Software version 2.2.1 and 2.2.2"
    }
  ]
}