      "status": "Terminated",
      "city": "Fajardo",
      "state": "PR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71721",
      "recalling_firm": "Customed, Inc",
      "address_1": "Calle Igualdad Final #7",
      "address_2": "N/A",
      "postal_code": "00738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed Only in Puerto Rico.",
      "recall_number": "Z-2677-2015",
      "product_description": "LABOR & DELIVERY 4/CS  Used by medical staff to perform surgical procedures and  patient care procedures such as wound healing, suture removals, irrigate wounds, etc.  .",
      "product_quantity": "300",
      "reason_for_recall": "Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.",
      "recall_initiation_date": "20150710",
      "center_classification_date": "20150903",
      "termination_date": "20170412",
      "report_date": "20150909",
      "code_info": "Code:900419  Lots:  140613836 exp. 1/31/16  140814953 exp. 10/31/15  150116537 exp. 1/31/16  150116535 exp. 1/31/16  141216339 exp. 12/31/15  141216339 exp. 12/31/15  150418410 exp. 4/30/16  150418411 exp. 4/30/16  150518802 exp. 5/31/16  150518802 exp. 5/31/16"
    }
  ]
}