  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79315",
      "recalling_firm": "Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US Nationwide.",
      "recall_number": "Z-2674-2018",
      "product_description": "Lateral Assembly, Radial Head Impl Assembly, Size 2. Orthopedic implant for replacement of the proximal end of the radius in the Stryker rHead Radial Head and Uni-Elbow system.",
      "product_quantity": "150",
      "reason_for_recall": "Revision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.",
      "recall_initiation_date": "20171024",
      "center_classification_date": "20180808",
      "termination_date": "20230518",
      "report_date": "20180815",
      "code_info": "310-00011"
    }
  ]
}