  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69178",
      "recalling_firm": "Synthes, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution including the states of IL, SC, VA, TX, MA, NC, LA, CA, AL, CO, NE, OR and KS., and Internationally to Canada.",
      "recall_number": "Z-2674-2014",
      "product_description": "DePuy Synthes TI Vectra Plates:    TI Vectra-One\" Plate 2 Level / 34 mm\t  part 04.613.184\t  3337672    TI Vectra\" Plate 3 Level / 57 mm\t  part 04.613.257\t  lot 3693878    TI Vectra\" Plate 4 Level / 64 mm\t  part 04.613.364\t  lot 8143905",
      "product_quantity": "16",
      "reason_for_recall": "DePuy Synthes is initiating a voluntary medical device recall of certain lots of the TI Vectra Plates, which is a part of the DePuy Synthes Vectra Anterior Cervical Plate Family of Systems.  The Vectra Systems are intended for anterior screw fixation to the cervical spine (C2-C7). It was discovered in certain lots, that the clips are potentially missing from the TI Vectra Plates resulting in the i",
      "recall_initiation_date": "20140902",
      "center_classification_date": "20140922",
      "termination_date": "20150902",
      "report_date": "20141001",
      "code_info": "TI Vectra-One\" Plate 2 Level / 34 mm   part 04.613.184   3337672    TI Vectra\" Plate 3 Level / 57 mm   part 04.613.257   lot 3693878    TI Vectra\" Plate 4 Level / 64 mm   part 04.613.364   lot 8143905"
    }
  ]
}