  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Barrington",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64796",
      "recalling_firm": "GE Healthcare It",
      "address_1": "540 W Northwest Hwy",
      "address_2": "N/A",
      "postal_code": "60010-3051",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution to GA, OH, NY, TX, TN ,MN and Internationally  to India, Australia, Canada, England, Qatar, and Scotland.",
      "recall_number": "Z-2671-2016",
      "product_description": "Centricity Laboratory Core Lab System 4.1    The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network.",
      "product_quantity": "17",
      "reason_for_recall": "The Centricity Laboratory removes free-text notes on a master panel when the user deletes the slave panel. The deletion of a slave panel will (auto slash) the  corresponding test (item) results on the master panel. Additionally, the displayed and HL7 results show neither the appropriate abnormal flag value nor the reference range values, when using large_num.g >7 to report numeric results.",
      "recall_initiation_date": "20121120",
      "center_classification_date": "20160826",
      "termination_date": "20171229",
      "report_date": "20160907",
      "code_info": "Versions 4.1, 4.0, 3.3.2 & 3.3Almp"
    }
  ]
}