  "results": [
    {
      "status": "Terminated",
      "city": "Bridgend",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68974",
      "recalling_firm": "ORTHO-CLINICAL DIAGNOSTICS",
      "address_1": "FELINDRE MEADOWS",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "NY, NJ, CT, MA, PA, VA, MD, FL and GA",
      "recall_number": "Z-2671-2014",
      "product_description": "VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody to hepatitis C virus (anti-HCV) in human serum and plasma. REF 680 2454, IVD -- Ortho-Clinical Diagnostics High Wycombe UK. Note that one sales unit has 5 Reagent Packs contained within it.  Each Reagent Pack can test for 100 samples (patient, Quality Control, etc.).",
      "product_quantity": "463 sales units",
      "reason_for_recall": "VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in unexpected negative test results being obtained for reactive sample fluids (quality control and patient samples).",
      "recall_initiation_date": "20140806",
      "center_classification_date": "20140919",
      "termination_date": "20170414",
      "report_date": "20141001",
      "code_info": "Lot 9090 (exp. December 15, 2014)"
    }
  ]
}