  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85853",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2670-2020",
      "product_description": "PowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336),   0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350), 0132015 (UDI 00801741047367), 0132075 (UDI 00801741047374), 0132210 (UDI 00801741047381), 0132215 (UDI 00801741047398), 0132275 (UDI 00801741047404), 0141910 (UDI 00801741047411), 0141915 (UDI 00801741047428), 0141975 (UDI 00801741047435), 0142010 (UDI 00801741047442), 0142015 (UDI 00801741047459), 0142075 (UDI 00801741047466), 0142210 (UDI 00801741047473), 0142215 (UDI 00801741047480), 0142275 (UDI 00801741047497)",
      "product_quantity": "5,296,132 total",
      "reason_for_recall": "Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.",
      "recall_initiation_date": "20200619",
      "center_classification_date": "20200718",
      "termination_date": "20220729",
      "report_date": "20200729",
      "code_info": "Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336),   0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350), 0132015 (UDI 00801741047367), 0132075 (UDI 00801741047374), 0132210 (UDI 00801741047381), 0132215 (UDI 00801741047398), 0132275 (UDI 00801741047404), 0141910 (UDI 00801741047411), 0141915 (UDI 00801741047428), 0141975 (UDI 00801741047435), 0142010 (UDI 00801741047442), 0142015 (UDI 00801741047459), 0142075 (UDI 00801741047466), 0142210 (UDI 00801741047473), 0142215 (UDI 00801741047480), 0142275 (UDI 00801741047497).    Lot Numbers:  ASCXF007  ASCXF019  ASDRF074  ASDRF075  ASDRF079  ASDRF084  ASDVF021  ASCXF001  ASDNF040  ASDNF043  ASDVF031  ASDWF118  ASCXF006  ASCXF018  ASDRF053  ASDRF058  ASDRF060  ASDRF071  ASDVF029  ASDWF117  ASCYF028  ASCYF029  ASCZF025  ASDPF029  ASDQF015  ASDQF039  ASDQF041  ASDRF065  ASDTF043  ASDTF054  ASDTF055  ASDTF061  ASDVF018  ASCYF023  ASCYF024  ASDTF031  ASDTF032  ASDTF033  ASDTF037  ASDTF038  ASCYF040  ASCYF076  ASCZF055  ASDNF001  ASDNF002  ASDNF004  ASDNF018  ASDQF036  ASDTF075  ASDTF096  ASDTF104  ASDTF105  ASDTF115  ASDUF006  ASDUF019  ASDUF026  ASDUF033  ASDUF042  ASDUF059  ASDUF063  ASDUF078  ASDUF097  ASDUF102  ASDUF105  ASDUF111  ASDUF119  ASDVF005  ASDVF014  ASDWF070  ASDWF081  ASDWF086  ASDWF102  ASDWF114  ASDTF023  ASDTF024  ASDTF025  ASDTF034  ASDTF035  ASCYF020  ASDTF027  ASDTF028  ASDTF036  ASCYF021  ASCYF022  ASDTF016  ASDTF017  ASDTF018  ASDTF019  ASDWF106  ASCXF009  ASCXF010  ASDRF088  ASDRF089  ASDRF092  ASDRF095  ASDRF096  ASDXF071  ASDXF073  ASENF043  ASENF047  ASENF048  ASCXF002  ASDNF042  ASDXF083  ASDXF084  ASCXF005  ASCXF008  ASDSF028  ASDSF032  ASDSF044  ASDSF051  ASDSF055  ASDSF056  ASENF061  ASENF075  ASEQF074  ASEQF086  ASEQF099  ASDNF027  ASDNF031  ASDNF035  ASDNF038  ASDQF021  ASDQF025  ASDQF030  ASDVF088  ASDWF003  ASDWF009  ASDWF012  ASDWF019  ASDWF072  ASERF011  ASDQF002  ASDQF003  ASDQF004  ASDQF006  ASDTF101  ASDVF084  ASDXF036  ASDXF037  ASCZF056  ASCZF062  ASDNF003  ASDNF005  ASDNF029  ASDNF032  ASDNF037  ASDNF046  ASDNF052  ASDTF040  ASDTF044  ASDTF059  ASDTF076  ASDTF098  ASDTF106  ASDUF003  ASDUF009  ASDUF021  ASDUF028  ASDUF043  ASDUF060  ASDUF077  ASDUF095  ASDUF108  ASDUF112  ASDWF073  ASDWF082  ASDWF091  ASDWF092  ASDWF113  ASDWF120  ASDXF002  ASDXF015  ASDZF021  ASDZF029  ASDZF037  ASDZF047  ASENF007  ASENF069  ASENF079  ASENF083  ASEPF007  ASCYF018  ASCYF019  ASCZF026  ASDRF047  ASDRF048  ASDRF049  ASDRF050  ASDXF049  ASDXF050  ASENF026  ASENF027  ASENF028  ASERF006  ASERF007  ASCYF007  ASDRF063  ASCYF005  ASCYF006  ASDRF009  ASDRF015  ASDRF016  ASDRF028  ASDVF068  ASDZF060  ASDZF063  ASDZF066  ASDZF068  ASERF026  ASERF028",
      "more_code_info": ""
    }
  ]
}