      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85853",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2669-2020",
      "product_description": "SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI 00801741066214), LH-0032YN (UDI 00801741066238), LH-0033YN (UDI 00801741066252), LH-0034YN (UDI 00801741066276), LH-0035YN (UDI 00801741066290 ), LH-0036YN (UDI 00801741066313), LH-0037YN (UDI 00801741066337), LH-0038YN (UDI 00801741066351), LH-0039YN (UDI 00801741066313), LH-0040YN (UDI 00801741066399)",
      "product_quantity": "5,296,132 total",
      "reason_for_recall": "Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.",
      "recall_initiation_date": "20200619",
      "center_classification_date": "20200718",
      "termination_date": "20220729",
      "report_date": "20200729",
      "code_info": "Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI 00801741066214), LH-0032YN (UDI 00801741066238), LH-0033YN (UDI 00801741066252), LH-0034YN (UDI 00801741066276), LH-0035YN (UDI 00801741066290 ), LH-0036YN (UDI 00801741066313), LH-0037YN (UDI 00801741066337), LH-0038YN (UDI 00801741066351), LH-0039YN (UDI 00801741066313), LH-0040YN (UDI 00801741066399).      Lot Numbers:  ASDQF068  ASDRF001  ASDRF002  ASDRF003  ASDRF005  ASEPF018  ASEPF019  ASEPF020  ASEPF026  ASEQF008  ASEQF025  ASEQF030  ASEQF039  ASEQF041  ASCYF038  ASCYF097  ASDQF065  ASDQF066  ASDQF067  ASDWF061  ASDWF062  ASDWF063  ASDWF064  ASDWF065  ASEQF045  ASEQF046  ASCYF039  ASCYF065  ASCYF093  ASCZF007  ASCZF008  ASCZF018  ASCZF019  ASCZF041  ASCZF042  ASDNF026  ASDQF027  ASDQF029  ASDQF037  ASDQF043  ASDQF051  ASDQF061  ASDRF037  ASDYF011  ASDYF016  ASDYF024  ASDYF034  ASDYF045  ASDYF055  ASDYF064  ASDYF066  ASDYF075  ASDYF078  ASDYF087  ASDZF073  ASDZF077  ASDZF087  ASDZF094  ASENF004  ASENF009  ASENF020  ASENF040  ASENF056  ASENF071  ASENF077  ASENF078  ASENF086  ASENF087  ASENF088  ASENF089  ASEPF001  ASEPF002  ASEPF003  ASEPF004  ASEPF005  ASCYF025  ASCYF026  ASCYF027  ASCZF028  ASDPF030  ASDQF016  ASDQF019  ASDQF024  ASDRF036  ASDZF014  ASDZF019  ASDZF023  ASDZF034  ASDZF040  ASDZF044  ASDZF052  ASDZF059  ASDZF067  ASEQF056  ASEQF069  ASEQF075  ASEQF076  ASCXF016  ASCXF017  ASDQF007  ASDQF010  ASDQF013  ASERF021  ASCXF014  ASCXF015  ASDPF038  ASDPF039  ASEQF108  ASEQF109  ASEQF111  ASEQF117  ASEQF121  ASEQF126  ASCYF096  ASDWF083  ASDWF084  ASEQF044  ASDRF007  ASDRF008  ASEQF042  ASCYF008  ASCYF009  ASCYF095  ASCZF030  ASDPF031  ASDPF032  ASCYF003  ASCYF004  ASDPF036  ASDPF037  ASEQF085  ASEQF093  ASEQF094  ASEQF095  ASERF041  ASCYF002  ASCYF094  ASDPF040  ASDPF041  ASCXF011  ASDNF044  ASDNF045  ASEQF100  ASEQF101  ASEQF102  ASEQF103  ASEQF104    Additional Lot as of 9/11/20:  ASEQF034",