  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85853",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2666-2020",
      "product_description": "SafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482), PA-0029YN (UDI 00801741085499), PA-0030 (UDI 00801741085505), PA-0030YN (UDI 00801741085512), PA-0031 (UDI 00801741085529), PA-0031YN (UDI 00801741085536), PA-0032 (UDI 00801741085543), PA-0032YN (UDI 00801741085550), PA-0033 (UDI 00801741085567), PA-0033YN (UDI 00801741085574), PA-0034 (UDI 00801741085581), PA-0034YN (UDI 00801741085598), PA-0036 (UDI 00801741085604), PA-0038 (UDI 00801741085628), PA-0038YN (UDI 00801741085635)",
      "product_quantity": "5,296,132 total",
      "reason_for_recall": "Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.",
      "recall_initiation_date": "20200619",
      "center_classification_date": "20200718",
      "termination_date": "20220729",
      "report_date": "20200729",
      "code_info": "Catalog Numbers PA-0029 (UDI 00801741085482), PA-0029YN (UDI 00801741085499), PA-0030 (UDI 00801741085505), PA-0030YN (UDI 00801741085512), PA-0031 (UDI 00801741085529), PA-0031YN (UDI 00801741085536), PA-0032 (UDI 00801741085543), PA-0032YN (UDI 00801741085550), PA-0033 (UDI 00801741085567), PA-0033YN (UDI 00801741085574), PA-0034 (UDI 00801741085581), PA-0034YN (UDI 00801741085598), PA-0036 (UDI 00801741085604), PA-0038 (UDI 00801741085628), PA-0038YN (UDI 00801741085635).    Lot Numbers:  REDX2834  REDX4856  REDY1792  REDY2773  REEN2075  REEN4559  REEP1004  REEP3058  REEQ4021  REER0466  REER1235  REER2399  REDW3688  REDZ3026  REEN1109  REEP2916  REEQ2690  REEQ4063  REDS1399  REDS1875  REDS3371  REDS4425  REDT1836  REDU4512  REDX2861  REDY0778  REDY3542  REDZ3102  REEN1044  REEP3066  REEQ2648  REER2400  REDR1897  REDR2695  REDR2796  REDS1288  REDS1400  REDS2408  REDS4621  REDT0697  REDT1645  REDT1837  REDT2924  REDT4207  REDU2262  REDU2347  REDU3328  REDV1094  REDV1239  REDV2772  REDV2983  REDW0413  REDX3752  REDY1723  REDY3731  REDZ0146  REDZ1859  REDZ3027  REEN2076  REEN2144  REEN3136  REEP0964  REER2366  REER2367  REER2401  REDN3937  REDQ0732  REDQ1532  REDQ3957  REDS1850  REDS2265  REDT0698  REDT1646  REDT2747  REDT4188  REDU0781  REDU2263  REDV2200  REDV3476  REDW0549  REDX1497  REDX3753  REDZ1056  REDZ1761  REEN1110  REEN2794  REEN4560  REEP3956  REDY3732  REDZ1860  REEN2774  REEN4491  REEP1005  REEQ1672  REER1236  REDQ1535  REDR2696  REDR2795  REDS1857  REDS4422  REDT1838  REDT2744  REDZ1762  REEN1045  REEN2786  REEN4561  REEP2919  REEP4051  REEQ4065  REER1273  REDX0146  REDY0779  REEN2792  REEP0965  REDZ3028  REEN4492  REEP0983  REEQ0701  REEQ1660  REEQ4024  REER0454  REDW1196  REDY0780  REDZ3104  REER2354  REDX1373  REEP0984  REEQ4066  REER2388  REEN1048  REEN2077  REEQ0671  REDX1374  REDZ1065  REEN4562  REEP0942",
      "more_code_info": ""
    }
  ]
}