  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink&quot; Data Management System, software version 14.  The issue may cause the software to stop executing commands, including uploading validated results to the LIS.  The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentraLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.",
      "address_2": "PO BOX 6101",
      "product_quantity": "858",
      "code_info": "Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474",
      "center_classification_date": "20140918",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Lebanon, Lithuania, Luxemburg, Malaysia, Mexico, Netherland, New Zealand Norway, Pakistan, Portugal, Peru, Qatar, Russia, Saudi Arabia, Singapore Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand United Arab Emirates, United Kingdom, Uruguay, Zambia.",
      "state": "DE",
      "product_description": "Siemens CentraLink Data Management System, software version 14.      The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.",
      "report_date": "20140924",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2665-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69129",
      "termination_date": "20170512",
      "more_code_info": "",
      "recall_initiation_date": "20140819",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}