      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69045",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US distribution in OH and countries of: Australia, China, Belgium, Israel, and Sweden.",
      "recall_number": "Z-2664-2014",
      "product_description": "The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.  These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.",
      "product_quantity": "6",
      "reason_for_recall": "During a Field Test (a customer external evaluation period), the customer detected a problem in the CIRS 4.0 Beta 2 software.  When using the CIRS 4.0 Beta 2 software, during the recon during ready stage of reconstruction, images may be overlapped with, or superimposed on other images.  Philips has disabled this feature at all sites evaluating 4.0 Beta 2 software.  Philips will issue a software up",
      "recall_initiation_date": "20110901",
      "center_classification_date": "20140918",
      "termination_date": "20150701",
      "report_date": "20140924",
      "code_info": "91003, 30018, 95504, 300004, 300030, 300010, 95130, 95157",
      "more_code_info": ""
    }
  ]
}