  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79315",
      "recalling_firm": "Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US Nationwide.",
      "recall_number": "Z-2663-2018",
      "product_description": "Lateral Assembly, Radial Implant, Size 1 Orthopedic implant for replacement of the proximal end of the radius in the Stryker rHead Radial Head and Uni-Elbow system.",
      "product_quantity": "211",
      "reason_for_recall": "Revision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.",
      "recall_initiation_date": "20171024",
      "center_classification_date": "20180808",
      "termination_date": "20230518",
      "report_date": "20180815",
      "code_info": "310-0000"
    }
  ]
}