  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77001",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico  Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore,  Spain, Switzerland, The Netherlands and United Kingdom",
      "recall_number": "Z-2662-2017",
      "product_description": "Logic Fit Tibial Tamp Head",
      "product_quantity": "573",
      "reason_for_recall": "The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.",
      "recall_initiation_date": "20170525",
      "center_classification_date": "20170703",
      "termination_date": "20190726",
      "report_date": "20170712",
      "code_info": "Catalog #: 213-75-01  Lot #'s: 37166-002, 44944-004, 44944-005, 44944-006, 48073-010, 48073-011, 48073-012, 49036-017, 51180-004, 51180-005, 51180-006, 51180-008,  51754-031, 51754-032, 51754-033, 51754-034, 59743-001, 61872-001,and   61872-002."
    }
  ]
}