  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85853",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2661-2020",
      "product_description": "PowerLoc MAX Power-Injectable Infusion Set myPICK Port Access Kit; Catalog Numbers CPA00001 (UDI 00801741098345), CPA00002 (UDI 00801741098352), CPA00028 (UDI 00801741129728), CPA00032 (UDI 00801741137501), CPA00033 (UDI 00801741137501), CPA00036 (UDI 00801741138911), CPA00051 (UDI 00801741175718), CPA00055 UDI 00801741182181), CPA00056 (UDI 00801741182907)",
      "product_quantity": "5,296,132 total",
      "reason_for_recall": "Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.",
      "recall_initiation_date": "20200619",
      "center_classification_date": "20200718",
      "termination_date": "20220729",
      "report_date": "20200729",
      "code_info": "Catalog Numbers CPA00001 (UDI 00801741098345), CPA00002 (UDI 00801741098352), CPA00028 (UDI 00801741129728), CPA00032 (UDI 00801741137501), CPA00033 (UDI 00801741137501), CPA00036 (UDI 00801741138911), CPA00051 (UDI 00801741175718), CPA00055 UDI 00801741182181), CPA00056 (UDI 00801741182907).    Lot Numbers:  REEN0212  REEP2685  REEP3660  REEQ3563  REER1968  REDZ0705  REEP3313  REEP3547  REER0577  REEN0214  REEQ3569  REDY3552  REEN1934  REEP3280  REEP3552  REER1115  REEN0218  REEP3659  REEP1436  REER0298  REER1335  REDY0992  REEN0278  REEP1437  REEQ0232",
      "more_code_info": ""
    }
  ]
}