  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77467",
      "recalling_firm": "Medacta Usa Inc",
      "address_1": "3973 Delp St",
      "address_2": "N/A",
      "postal_code": "38118-6110",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "CA, AZ, NC, Israel, France",
      "recall_number": "Z-2661-2017",
      "product_description": "Medacta GMK HINGE, Tibial Insert Fixed, SIZE 2, HEIGHT 23 mm, REF 02.09.0223H, STERILE EO",
      "product_quantity": "4 units",
      "reason_for_recall": "One lot of product does not include a screw for attachment to the tibial base  plate.",
      "recall_initiation_date": "20170519",
      "center_classification_date": "20170703",
      "termination_date": "20180529",
      "report_date": "20170712",
      "code_info": "Lot # 163646"
    }
  ]
}