{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Concord",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69113",
      "recalling_firm": "Motion Concepts",
      "address_1": "84 Citation Drive",
      "address_2": "Units 1-7",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.",
      "recall_number": "Z-2657-2014",
      "product_description": "TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.",
      "product_quantity": "US: 59 units",
      "reason_for_recall": "Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System  (TRx Vertical Positioner).",
      "recall_initiation_date": "20140828",
      "center_classification_date": "20140917",
      "termination_date": "20150603",
      "report_date": "20140924",
      "code_info": "All units manufactured between June 2011 and May 2014: Serial Numbers 28583, 28597, 28604, 28607, 28629, 30334, 028380, U001050, U001083, U001544, U002020, U002326, U002386, U002403, U002411, U002489, U002502, U002505, U002509, U002510, U002519, U002716, U002922, U003577, U004011, U004018, U004073, U004079, U004145, U004214, U004400, U004404, U004472, U004485, U006042, U006234, U006235, U006260, U006295, U006296, U006353, U006378, U007126, U007138, U007163, U007239, U007245, U007279, U007288, U007407, U008556, U008631, U008698, U008840, U008896, U009548, U009569, U009579, U009707, U009718.",
      "more_code_info": ""
    }
  ]
}