{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77351",
      "recalling_firm": "Shimadzu Medical Systems Usa Com",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Nationwide distribution to AK, AR, AZ, CA, CT, FL, IL, IN, KS, LA, MI,  MO, NE, NJ, NM, NV, NY,OH, OK, PA, RI, SC, TN, TX, WA, WI, and Hawaii  Foreign:  Canada and Japan",
      "recall_number": "Z-2654-2017",
      "product_description": "SHIMADZU RADspeed Pro Catalog Number: RADspeed Pro;  Stationary x-ray system",
      "product_quantity": "1673 units total",
      "reason_for_recall": "Cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.",
      "recall_initiation_date": "20170523",
      "center_classification_date": "20170629",
      "termination_date": "20191001",
      "report_date": "20170705",
      "code_info": "Serial No. 0162M4501 thru 0562M45103, 3ZC5C2A0C002 thru 3ZC5C1734001, D362M42902 thru 0562M41101, 41E733E69003, MP95A8F5A001 thru MP95A9F6B001"
    }
  ]
}