  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.",
      "address_2": "",
      "product_quantity": "33",
      "code_info": "728243 \u0013 Brilliance BigBore Oncology CT         7116 7611 7354 7063 7309 7169 7205 7016 7019 7043 7069 7073 7091 7112 7132 7163 7141 7194 7226 7325 7294 7350 7316 7480 7114 7210 7310 7438 7530 7186 7119 7042",
      "center_classification_date": "20160825",
      "distribution_pattern": "US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,  MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM,  NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY",
      "state": "OH",
      "product_description": "Brilliance CT Big Bore Oncology, Model number 728243    Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-2653-2016",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "74820",
      "termination_date": "20180702",
      "more_code_info": "",
      "recall_initiation_date": "20160720",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}