  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "HIROSHIMA",
      "state": "N/A",
      "country": "Japan",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68909",
      "recalling_firm": "MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK",
      "address_1": "6-22, 4-CHOME, KAN-ON-SHIN-MACHI",
      "address_2": "NISHI-KU",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed in the states of NY, FL, and TX.",
      "recall_number": "Z-2653-2014",
      "product_description": "VERO  Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator.    MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.",
      "product_quantity": "3 units in use in the US",
      "reason_for_recall": "Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status.  Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment",
      "recall_initiation_date": "20140705",
      "center_classification_date": "20140916",
      "termination_date": "20150401",
      "report_date": "20140924",
      "code_info": "Software Version 3.0.0 and after.  Serial Numbers 201902, 203901, and 203919",
      "more_code_info": ""
    }
  ]
}