  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St NE",
      "reason_for_recall": "78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.",
      "address_2": "",
      "product_quantity": "78 (39 US, 39 OUS)",
      "code_info": "Product Name: Evera XT DR lCD.  Models: DDBB1D4, DDBB1D1, DDBB2D4, DDBB2D1.  Serial Numbers:   BWC202738H, BWC202754H, BWB202998H, BWB203157H, BWB203167H, BWB203173H, BWB203186H, BWE601558S, BWE601571S, BWE601578S, BWE601579S, BWE601581S, BWE601589S, BWE601591S, BWE601594S, BWE601600S, BWE601605S, BWD602122S.    Product Name: Evera S DR lCD.  Models: DDBC3D1, DDBC3D4.  Serial Numbers:   BWG600597S, BWF600969S, BWF600970S, BWF600972S, BWF600973S, BWF600975S, BWF600977S, BWF600978S, BWF600979S, BWF600983S, BWF600984S, BWF600985S, BWF600987S, BWF600989S, BWF600991S, BWF600992S, BWF600996S, BWF601001S.    Product Name: Evera XT VR lCD.  Models: DVBB1D1, DVBB2D4.  Serial Numbers:   BWI201423H, BWI201436H, BW1201440H, BWI201451H, BWI201454H, BWI201462H, BWI201473H, BWJ601102S, BWJ601108S, BWJ601112S, BWJ601479S, BWJ601101S, BWJ601103S, BWJ601106S.",
      "center_classification_date": "20160824",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.",
      "state": "MN",
      "product_description": "Medtronic, Evera Implantable Cardioverter Defibrillators:    Product       Model  EVERA XT   DDBB1D1, DDBB1D4, DDBB2D1, DDBB2D4, DVBB2D4, DVBB1D1.  EVERA S     DDBC3D1, DDBC3D4    Product Usage:  The Medtronic Evera DR dual chamber and Evera VR single chamber, implantable cardioverter defibrillator (lCD) are a family of multiprogrammable cardiac devices that monitor and regulate the patients heart rate by providing dual or single chamber rate-responsive bradycardia pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies (dual-chamber devices only).",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure",
      "recall_number": "Z-2651-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74923",
      "termination_date": "20170120",
      "more_code_info": "",
      "recall_initiation_date": "20160812",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}