      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St NE",
      "reason_for_recall": "78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.",
      "address_2": "",
      "product_quantity": "78 (39 US, 39 OUS)",
      "code_info": "Product Name:  Viva XT CRT -D .  Models:  DTBA1D4, DTBA1D1.  Serial Numbers:   BLF203128H, BLF204746H, BLE202888H, BLE202889H, BLE202890H, BLE202901H, BLE202941H, BLE202947H, BLE202954H, BLE202958H, BLE202961H, BLE202962H, BLE202964H, BLE202981H, BLE202987H, BLE202989H, BLE202990H, BLE202991H, BLE203019H, BLE203026H, BLE203027H, BLE203029H, BLE203032H, BLE203046H, BLE203052H, BLE203073H    Product Name: Viva S CRT-D.  Models:  DTBB1D4, DTBB1D1.  Serial Numbers: BLO202272H, BLN202206H.",
      "center_classification_date": "20160824",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.",
      "state": "MN",
      "product_description": "Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators:  Product                 Model  VIVAXTCRT-D     DTBA1D1, DTBA1D4  VIVASCRT-D       DTBB1D1, DTBB1D4    Product Usage:  The Medtronic Viva CRT -D dual chamber, implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT -D) is a multiprogrammable cardiac device that monitors and regulates the patient\u0019s heart rate by providing single or dual chamber, rate responsive bradycardia pacing; sequential biventricular pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies.",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure",
      "recall_number": "Z-2650-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74923",
      "termination_date": "20170120",
      "more_code_info": "",
      "recall_initiation_date": "20160812",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}