{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Oro Valley",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94985",
      "recalling_firm": "Ventana Medical Systems, Inc.",
      "address_1": "1910 E Innovation Park Dr",
      "address_2": "",
      "postal_code": "85755-1962",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, \tCT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, \tNH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Albania, Argentina, Armenia, Australia, Azerbaijan, Bahrain, Belarus, Brazil, Cameroon, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Georgia, Ghana, Guatemala, Honduras, Hong Kong, India, Iran, Iraq, Israel, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mali, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, New Zealand, Nicaragua, Nigeria, North Macedonia, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Russian Fed., Rwanda, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Sri Lanka, Taiwan, Tanzania, Thailand, T¿rkiye, Uganda, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.",
      "recall_number": "Z-2644-2024",
      "product_description": "Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD",
      "product_quantity": "22,839 kits",
      "reason_for_recall": "Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result  in false positive results.",
      "recall_initiation_date": "20240627",
      "center_classification_date": "20240816",
      "report_date": "20240828",
      "code_info": "UDI: 04015630972579/J04613, J11853, J17541, J25047, J30286, K00982, K06239, K09880, K14266, K19784, K26461, and M00669",
      "more_code_info": ""
    }
  ]
}