  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77389",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of TX, NY,FL, IL, OH, and countries of France, Ireland, Japan, and Taiwan",
      "recall_number": "Z-2630-2017",
      "product_description": "da Vinci Xi Surgical System    Product Usage:  The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon in the accurate control of surgical instruments such as endoscopes, scissors, forceps, and retractors during endoscopic surgery. The system has four universal manipulators (arms) which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures in the pelvic, abdominal, and thoracic cavities.",
      "product_quantity": "10 units",
      "reason_for_recall": "da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially under-torqued brake screws in a section of the system arms.",
      "recall_initiation_date": "20170601",
      "center_classification_date": "20170626",
      "termination_date": "20171002",
      "report_date": "20170705",
      "code_info": "Serial Numbers:  SK0822, SK1154, SK1001, SK1170, SK1040, SK1176, SK1082, SK1180, SK1146, SK1190"
    }
  ]
}