  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Hartland",
      "address_1": "900 Walnut Ridge Dr",
      "reason_for_recall": "Potential for RadSuite AV viewer to skip image slices when the user presses page up or page down when the viewer displays multiple viewports and the user has mouse focused on any viewport other than the first viewport.",
      "address_2": "",
      "product_quantity": "25 sites potentially have the affected versions",
      "code_info": "Versions: 5.30.7, 5.30.7 HF0429E, 5.30.7 HF0441E, 5.30.7 HF0446E, 5.30.7 HF0470E, 5.30.7 HF0473E, 5.30.8, 5.30.8 HF0477E, 5.30.8 HF0490E, 5.30.8 HF0495 SB0508E, 5.30.8 HF0495E, 5.30.8 HF0495E, SB504E, 5.30.8 HF0550, 5.30.8 SB0485E, 5.30.8 SB0486E, 5.30.X SB0381E, 5.35.1, 5.35.1 HF0487E, 5.35.1 HF0489E, 5.35.2, 5.35.3, 5.35.4, and 5.35.4 HF0513",
      "center_classification_date": "20160823",
      "distribution_pattern": "Distribution was made to AL, AZ, CA, IL, MD, MA, MI, MO, NV, NJ, NC, OH, PA, TN, and VA.  Government distribution was made to OK.  There was no foreign/military distribution.",
      "state": "WI",
      "product_description": "RadSuite software.  The firm name on the label is Merge Healthcare, Hartland, WI.",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Merge Healthcare, Inc.",
      "recall_number": "Z-2627-2016",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "74919",
      "termination_date": "20170210",
      "more_code_info": "",
      "recall_initiation_date": "20160131",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}