{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "40 Liberty Blvd",
      "reason_for_recall": "Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved.  This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.",
      "address_2": "",
      "product_quantity": "1 unit",
      "code_info": "Model # 10430603  Serial #  74192",
      "center_classification_date": "20160823",
      "distribution_pattern": "Distributed to: MI, NY, CA, KY, ND, NE",
      "state": "PA",
      "product_description": "SOMATOM Definition AS with software version VA48A-SP2;   Model # 10430603, computed tomography x-ray system",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-2624-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74840",
      "termination_date": "20170124",
      "more_code_info": "",
      "recall_initiation_date": "20160713",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}