  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Bloomington",
      "address_1": "750 N Daniels Way",
      "reason_for_recall": "Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation.  degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.",
      "address_2": "",
      "product_quantity": "148,334",
      "code_info": "All Lots.  Catalog prefixes / suffixes NPAS-/-SST    Catalog/product number; Global/order number  NPAS-100-HC-NT-U-SST G52963  NPAS-100-HC-U-SST    G52962  NPAS-101-HC-NT-U-SST G23067  NPAS-101-HC-U-SST    G23066  NPAS-104-HC-NT-JWG-U-SST G52966  NPAS-104-HC-NT-U-SST G52965  NPAS-104-HC-U-SST    G52964  NPAS-105-HC-NT-U-SST G52967  NPAS-108-HC-NT-U-SST G25516  NPAS-120-HC-NT-U-SST G25517  NSSW-4.0-18-NPAS-100-HC-SST G34253",
      "center_classification_date": "20160822",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, \"Palestinian Territory, Occupied\", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.",
      "state": "IN",
      "product_description": "Aprima Access Nonvascular Introducer Set  Accessories, Catheter  Intended for single-puncture percutaneous access to facilitate placement of an .038 inch (0.97 mm) diameter working wire guide for interventional radiology procedures",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Cook Inc.",
      "recall_number": "Z-2620-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74010",
      "termination_date": "20170804",
      "more_code_info": "",
      "recall_initiation_date": "20160415",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}