  "results": [
    {
      "country": "United States",
      "city": "Bloomington",
      "address_1": "750 N Daniels Way",
      "reason_for_recall": "Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation.  degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.",
      "address_2": "",
      "product_quantity": "26,076",
      "code_info": "All Lots.  Catalog/product number; Global/order number  023565-BT  G18074",
      "center_classification_date": "20160822",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, \"Palestinian Territory, Occupied\", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.",
      "state": "IN",
      "product_description": "Kumpe Access Catheter  Cathether, Ureteral, General & Plastic Surgery  Used in combination with a HiWire, Bentson, or other flexible-tipped wire guide to gain difficult ureteral access beyond a redundant or tortuous ureteral segment",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Cook Inc.",
      "recall_number": "Z-2617-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74010",
      "termination_date": "20170804",
      "more_code_info": "",
      "recall_initiation_date": "20160415",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",