{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sylmar",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92923",
      "recalling_firm": "Abbott Medical",
      "address_1": "15900 Valley View Ct",
      "address_2": "",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: Nation-wide  O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Nepal, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom.  Latin America - Pending",
      "recall_number": "Z-2614-2023",
      "product_description": "Gallant VR Implantable Cardioverter Defibrillator:  VR ICD REF CDVRA500Q  Gallant DR Implantable Cardioverter Defibrillator:  DR ICD REF CDDRA500Q    Neutrino VR Implantable Cardioverter Defibrillator:  NxT VR ICD REF CDVRA600Q  Neutrino DR Implantable Cardioverter Defibrillator:  NxT DR ICD REF CDDRA600Q    Entrant VR Implantable Cardioverter Defibrillator:  VR ICD REF CDVRA300Q  Entrant DR Implantable Cardioverter Defibrillator:  DR ICD REF CDDRA300Q    The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted through the transvenous technique with the distal tips positioned in the RA (dual chamber) and RV.  Cardiac Resynchronization Therapy Defibrillator (CRT-D) systems consist of a right atrial (RA) pacing lead, a right ventricular (RV) defibrillation lead, as well as a third lead placed in the coronary venous system overlying the left ventricle (LV) to improve ventricular resynchronization through biventricular pacing for treatment of heart failure. CRT-D devices, more generally known as an implantable pulse generator (IPG), are typically implanted in a subcutaneous device pocket located in the pectoral region and are attached to the pacing/defibrillator leads that are implanted through transvenous technique.",
      "product_quantity": "38,033 devices",
      "reason_for_recall": "Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication.  In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.",
      "recall_initiation_date": "20230816",
      "center_classification_date": "20230919",
      "report_date": "20230927",
      "code_info": "Gallant: Model Number: VR ICD REF CDVRA500Q UDI-DI Code: 05415067031990 Serial Numbers: 111033498 810000696 810002524 111031208 111025117 111030959 111006327 111028481 111014296 111033775 810021414 810001946 810001996 810002011 810001942 810007459 111036536 810019608 810019393 810019581 111042257 810026722 810028531 810028585 810027003 111011936 810021423 111025530 111015227 111036960 810026790 111036957 111036933 111036964 111007506 810021413 810021445 810021449 111033604 111000553 111000558 111000567 111000472 111000706 111000667 111006793 111000664 111000678 111018664 111018644 111018802 111012460 111006783 111006791 111006789 111000786 111000582 111000550 111000721 111001035 111001025 111014950 111021456 111021461 111021703 111021722 111021450 111036417 111012272 111011597 810001013 810001011 810006759 810001046 810001021 810019014 111024417 810002439 111024415 810006774 111024454 810022318 111036914 810013066 810015963 111002002 810021662 810013192 810011364 111003380 810011481 111020960 111011436 111020992 810015820 810015794 111009835 111004271 111003046 111024175 111022969 111009840 810019313 810018954 111021002 111009844 810008620 111010818 810018999 111021710 810016110 810019437 810008498 111024550 810008497 810019426 810011359 810019424 810011511 810011353 810011328 111011181 111011372 111021007 111021712 111024183 810013436 810019054 810019546 810022117 810027515 810013071 810007469 111020996 111009733 111003861 111009782 810011335 810016103 111021851 810004570 111008584 111009245 111003267 810008499 111021988 111012318 111003040 111010800 111003332 111022768 810016526 810027041 111009747 810013248 111011899 111010457 810021351 810021705 810015780 111018656 111010771 111021437 810005318 111010761 111022974 111021994 810007454 810008483 810011285 810011458 810005317 111008582 111003173 111008588 111003555 111018654 810027111 810019552 810016459 111009820 810008557 810026959 111002712 111021707 111006776 111006779 111010814 111017230 810021576 810016112 111003506 810027040 810027014 810003350 111009781 111018131 810021674 111011908 810008527 810019917 111008583 810008575 111004282 810016339 810027099 810015815 810016102 111002437 111011229 111018270 810021707 810013068 810013216 810015840 810018976 810021355 810011459 111003500 111019459 810003430 111009834 111010787 111000447 111004193 111009841 111013996 111011233 111011254 111021837 111019423 810011451 111018786 810016113 810016649 810013446 810013250 810013442 810027039 810026943 810026944 810027020 810027165 111009759 111006801 111021440 111010437 111019431 810011457 810016665 810016449 810019012 810021618 810008573 810021560 810011535 111009788 810013454 810008650 810015824 111006778 810005319 111024379 810013349 810004013 111009809 810008519 111024368 810003553 111024378 111006784 810016140 810008486 111006797 810013266 111018109 111018795 111022789 111011910 810021606 810007463 810007421 810003677 111011216 810027514 111018640 810019556 810008612 111018784 111024402 810003437 810011343 810008526 111004721 111021995 810027548 810018995 810016138 810019427 810005706 111022585 810011387 111018291 111011963 111009708 111010450 111017239 111022090 111024180 111011906 111021723 111021436 810022090 810015784 111009794 111021433 810003343 810013241 111010980 810027103 111011915 111002410 810027161 111001995 111009741 111006541 111002011 810016121 111004259 111003960 810021559 111024204 810013232 111016740 810027168 810015796 111001027 111012290 111002434 111010962 111001969 810015822 111016745 111024202 111012258 111010952 111024184 111013804 111022788 111010465 810019529 810011303 111013808 810011549 810011538 111021812 810005366 810007473 111023732 111014948 111013807 111001583 111023720 111022587 111011938 810026979 111012462 111012307 111003869 111024186 810016277 111004242 810005286 111022781 111021845 810004846 810015783 111019446 810022064 111010799 111011214 810022363 111009828 810008597 111021790 111003559 810016560 111011185 810013225 111013993 810013206 111011922 111003276 810018982 111003858 810013077 810011449 810011550 810016303 111011237 111019415 111003865 111006785 111003863 111004182 111021466 810004849 111012256 111006775 810022137 810019296 810019297 111018120 111018679 810003615 810008594 111003065 111003037 810022370 810022129 810019042 111004513 111003074 111022971 111003957 810005356 111004518 111023531 111003371 111014919 111009837 111003539 111002743 810019250 810011523 810019254 111021846 111022077 810021666 810022122 810021671 810019272 810019301 810019290 810019278 810019065 810019325 810019408 810019310 810019293 810019009 111009737 111011215 111009766 111009777 111003521 111010981 111011968 111009261 111003686 111017247 111017245 111003567 111003505 810015772 810015773 111004497 111017234 810013185 810019050 810011339 810022145 810022376 810022079 810022078 810021530 810021687 810021563 810018946 810019017 810016151 810027033 810013203 810011349 810011369 810008563 810004696 111021010 111024381 111024199 111020974 111019443 111022093 111022955 810016344 810016656 810016653 810016532 111011363 111008706 111003365 111003283 111000464 111003963 810019260 111004523 111004267 810013197 810022087 810011461 810022126 810016310 111003691 111003350 810007437 111021445 111003695 810013245 111000473 111011969 111012312 810013196 810011305 810015962 111017246 810016105 810011298 810027536 810011487 111012295 810027150 111012279 810016644 111024214 111018263 111018111 111000666 111000560 111011981 111000658 810027036 111020994 111009262 810005321 111011380 111003688 111003510 810021545 810005705 810005253 810005296 111022085 111024394 810003454 111018643 810018983 111011603 111011367 111000462 111021429 810013477 810022094 810019059 111010780 810013451 111021843 810013458 810019282 111000541 111021815 810011470 810008640 111000757 810011337 810016363 810021520 111000792 111003514 111000441 111021811 810011358 810027159 810016146 810019063 810019539 810022082 111021717 810011295 111024398 111000440 810015797 810015800 810013448 810016145 810016322 810019268 810021602 810021664 111001076 111004164 810007406 111011247 111011227 111011226 111010454 111010436 111010773 111010448 810022110 111024549 111024380 111021824 810022096 810008474 810007403 810007413 810007434 111021000 810026542 810016270 810018953 810026999 810007427 810008536 810008637 810019443 810021702 111008602 810005709 111021724 810005279 111012296 111000803 111003508 111004172 111000996 111020952 111014955 810018966 111001082 810016155 810018955 810018996 111003680 810019057 810016524 810016334 810019019 810018980 111009773 810021369 810019524 810019525 810021376 111022089 111011375 111024197 810019441 810019442 810007414 810008607 810021578 810008595 810022141 810008591 810008603 810021616 810007416 111000150 810019010 810011503 810022107 810007410 810016323 810013062 111020968 111024569 810019007 810016321 111019441 111009829 810019550 810011314 810008632 810022091 810027144 111021438 111023543 111023723 111023725 810011347 810018958 111022778 810022359 810021356 810015787 111011957 111011365 111011377 810019332 111014945 810019531 111000268 111024367 111003325 810016307 111000159 111010469 810015779 111003548 810016384 810022140 111013983 111003326 111008599 810026973 810011492 810011506 810027535 111005533 111011212 810011317 810003947 111024205 810021505 810011377 810016306 111014012 810027029 810016539 111000674 111000608 111001084 810016650 111001585 111001034 111011432 111011612 810016543 810011394 810003434 111018288 111018261 111012467 810022371 810019324 111014938 111001997 810027128 111011905 810027016 810027000 111011974 810026968 111009789 111013980 111021842 111022088 111010776 111014009 111021982 111021850 810011368 111024192 810011524 111021537 111024181 810027025 810027163 111016837 111024082 111022073 111022977 810016148 810016305 810004758 111022762 810021537 810007405 111022589 810011505 111011360 111024193 810016356 111018785 810019410 810016534 810016547 111003083 810013336 810013094 810011548 810007432 810008525 111018648 111018276 111022080 810004816 111006552 111018278 810015821 810021541 111021318 810011322 810005324 111020982 111010785 111009753 111000214 111000657 111011453 810015961 810008615 111009749 810015958 810003344 111015637 111024213 810007430 111002399 810026950 111018676 111002401 111001026 810016648 810011496 111022963 111008597 810011334 810011531 111011245 810019421 810011468 111009767 111003968 111006551 111009248 810027526 111008605 111008579 111024194 810021585 111006546 810019323 111016728 111022979 111000213 810019920 810016326 810016373 111022590 810013183 810004568 810015827 810011381 810022065 810015802 810026988 810016330 810027030 810027043 810004847 810016357 810015786 111022596 810015832 111019435 111016733 111021795 111004270 810011365 111019408 111003677 111011221 111010439 111011912 111014932 810018997 810011508 810013069 810013200 810019425 810022383 111016738 111013998 111023545 111014003 111011633 111003335 111003042 111011973 111009744 111003328 111003348 111004643 111004281 810016355 810007417 810021620 810016150 810016657 810015943 111021714 810007441 111003554 810008609 810027160 111011190 111018117 111003502 810026937 810013076 111003287 111009723 111009345 111018639 111009772 111011206 111021711 111012266 810021503 111011207 111020970 111015650 810021605 810019062 810019308 111008590 111009731 111003512 810013047 810021722 111008609 111002397 810018952 111003052 111009270 111024382 810016378 810021733 810021717 810018963 810005282 810016555 111021821 810008490 111021984 810003459 111013969 810019011 810013227 111003674 810013231 111011900 810019048 111018652 111012261 810003734 111003278 111012475 111021016 810011320 810021729 810011306 810021597 111003682 810011513 810013078 111008598 111018660 810019258 111011954 810013239 810019435 111021803 111004713 810019319 111023736 810026952 810026987 111011990 810019068 111002741 111005530 111011246 810013348 111011626 111004522 111004514 111018659 810007426 111004716 810018988 810008599 111003136 810022071 810019033 810013240 810003428 810008518 810019330 111011255 810016165 111014928 111011442 111003354 810016438 111009814 111000261 810016157 111016830 111003290 111012305 810007423 111003488 111004508 111021825 111003363 111011934 810008638 810016458 810016457 111014930 111003347 111021462 111006548 111008604 810013091 111024390 111001880 111024176 111009819 810021354 810021710 810021697 810019291 810018977 810016521 810016638 810015788 810013224 810027013 111003072 810016098 111009725 111019447 810011499 111012281 111011236 111016726 111019462 111023519 810011441 111021697 111009732 810015814 810013198 111020956 111024547 810011301 111009768 111021029 111002404 111011200 111022777 111006774 111003689 111021987 111021431 111021538 111020959 810013287 810021612 111013972 810026953 111014005 111017224 111018806 810026928 810011373 111003651 810005298 810021518 810015789 810019528 810005248 810013205 810013447 810013188 111021849 111000247 111006554 111017243 810022073 111016725 111014010 111003652 111022582 111022772 111022581 810011356 810007458 810021373 810026962 810011378 111010790 111002392 810016116 111024393 111021979 111021991 111022595 111024376 810026992 810003674 111021696 810016550 111003359 111010443 111002417 111021720 111023542 810011522 111009696 111009697 810016568 111009796 111009246 111009807 810027023 810011448 810021589 111010464 810016337 810021538 111018800 810027155 111021003 810003427 111020958 111006787 111010772 810019438 111022945 810005307 111022959 810008556 810013430 111001024 111001580 111012464 810022092 810021699 810004815 111023551 111011444 111018136 111001014 111011224 111003345 111001080 111002453 111011985 111001577 111003289 111003954 810007433 111011217 111012302 810005245 111009260 810027545 111003285 810016579 810022136 111009779 810027142 111009780 810022118 111009771 810003342 810005290 810022127 111011627 810019280 111008699 111018796 111008705 111008611 111003566 810005237 810022386 111010442 111002378 111014917 111021441 111022901 111021458 111010435 111023562 111014929 810019429 111011225 810026989 810021601 810016328 111020979 111022076 111020948 810004828 810003370 810003348 111003134 111004192 810027547 810013478 111012457 111011929 111021464 810005349 111021802 810026997 810011276 111009798 810019423 111010441 810011504 810003452 810016647 111013985 111009726 111009810 111011947 810021712 810027553 111008577 810013463 111011346 810019000 810016124 810016132 810005292 810016313 111017233 111001962 810021519 810008592 810012608 111023739 810013223 111011920 810016663 810022115 810011316 111017231 810011302 810013361 111017226 810005323 810015804 111022083 810013456 111011614 111009240 810026938 111000991 111009267 111004190 111012321 111019422 111021783 810013272 810011530 111005540 111004250 810026942 111020986 111006792 810011283 810011296 111021434 111011211 810013075 810013357 111009740 111024198 111009751 111011179 810005291 810019522 111009268 810007470 111024083 111003654 111003653 810003458 810021510 111002420 111011955 810003466 810013433 810003461 810013099 111004245 111011618 111004272 810003456 810016578 111021708 111009743 111009718 810027024 111002400 111002393 810026957 111001884 111011253 111022961 810008604 111012470 111018133 111019407 111004178 111019421 111004718 810011477 111004720 111000151 111000995 101000034 111001539 111011248 810016361 111012317 111001961 111000446 810003617 810012603 810013338 111010781 111019456 810019527 111021986 101000038 810013067 111001994 111001540 810019013 111005531 810003374 111015615 810013074 810019041 111000535 810019043 810027519 111018798 810019322 111018801 111013970 810026956 111000575 810021592 111013975 810021586 810021610 111011916 111001023 810018993 111021001 111001581 111023734 111001553 810015826 810021348 810021352 111010961 111010812 111018293 111024392 810012607 111018127 111018268 111018119 111018128 111001857 810016274 111018130 111024377 111000315 810018943 810021719 111020987 111001554 810021737 111018804 111001007 810021374 810021349 810021564 810021593 810021583 810021607 810021555 111016769 111016841 810019289 111021006 111020963 111022966 111022936 111022770 111002414 111002390 111003965 810013053 810016557 111022071 111021841 810019044 810021567 810008553 111008589 810019016 111018810 111003870 111022975 810021580 810027522 810016153 810022385 111021983 810027531 810026545 810026965 810027006 810021591 810019922 810021667 810021595 111000805 810027046 810022128 810021362 810021570 810021367 810021571 810021357 810021380 810021368 810021378 111015645 111019450 111008610 810019264 810019305 810021551 111005537 810003886 810021574 111021831 111003176 810013339 101000031 111024720 111010969 810027120 111013810 810022103 810015828 111018790 111005527 111001859 111003272 111001072 101000029 810021539 111003090 810018991 810016324 111006556 810016327 810027534 810027149 810027162 810019318 810022381 810007452 810007449 810007411 810007468 810007447 111000764 111003073 111011610 111021974 810019037 810021594 810021608 810021691 810021542 810021725 810021721 810018939 810019558 810015811 111004501 111000662 111024548 111003513 111000438 111001081 111003538 111000552 111019451 810016137 810015775 111000557 111003687 111003678 111000451 111000708 810013070 810013045 111000449 111000455 111000263 111000154 111000527 111000675 111001566 111000548 111018791 111000562 111000531 111000141 111000458 111000988 111001016 111000547 101000030 101000036 111003321 111001542 111000554 101000042 111000319 111001866 111000576 111000581 111000476 111000316 111000470 111000538 111000566 111000452 111009830 111009754 111009811 111000474 111003668 810018985 111000525 810019535 111021797 111013987 111013979 111008701 111006782 810027104 810011293 810011478 810013218 111011925 111003515 111024088 111014914 810022364 810019255 810016444 810003616 810003338 111010974 111015638 111017228 810003672 111011201 111003257 810013335 111003048 111000672 810013269 810027146 810021661 810021575 810026955 810026954 111011230 111011228 810021549 810018969 111021699 111013809 810022106 111021853 810016135 111008603 111021314 111002389 111002398 111021716 810011326 810011462 111022937 111003536 111016731 111016730 111016727 810016445 111023561 111023547 810019407 810021658 810005238 810016666 810016346 810016122 810021516 810013346 810008559 111019436 810022086 810026982 810021523 810021718 810022114 810016641 810019304 810026546 810008587 810007444 810007415 111000456 810018972 810027157 810021735 810027166 810027044 810019439 810027119 810018986 810019004 810016267 810015945 810005325 810026931 810027004 810019271 810027097 810021681 810011385 111011381 111018115 111011374 111024399 810027169 111003673 810016662 810016542 111002730 111010957 810026961 810007442 111009799 810016317 810016097 810016354 810019038 111011970 111018110 111021820 111022767 111020993 810022113 810027034 810011338 111011935 111023540 810021379 810013083 810016573 810007460 810026964 810003678 810019515 111024719 810011290 810016529 810015816 810016156 810015836 810027529 810027530 810013230 810013190 810013432 810013445 810011319 810011533 810011519 810011329 810019284 810019331 810022116 810022100 810013073 810021532 810021531 810021527 810021535 810021522 810021621 810027100 810027026 810027037 810027027 810016563 810016642 810018989 810019252 810013238 810011318 810007448 810016659 111024084 810013464 810011547 810003340 111000461 810011297 111009813 111024185 111011946 810027047 111021828 810022083 111004161 111023564 111003336 111020962 111020977 111021013 810005313 810026966 810011315 810011527 810026969 810011475 111021008 111021009 111018292 111003277 111000147 111000158 111001548 111001889 111001541 111012293 111011628 111003039 101000041 111003533 810021613 111021806 111006780 111001966 810027135 111000272 111000105 111000139 111000260 111000160 111000246 111000579 810016280 111011948 111011197 111016833 111024178 111017237 810011308 111009734 810016463 111009822 111024171 810016301 111012294 810015812 810016163 111023534 111002419 810021526 810011310 810016111 810022088 810003614 810013341 111000673 810011331 810016676 810016533 810011327 810027140 111023520 810019018 810027035 111004255 810021689 810011287 810016460 111004173 810003346 810007474 111002436 111000990 111004180 810013340 810011383 810011372 111024208 810015837 810027170 810026990 810011526 111004714 810016123 111016716 111003519 111024395 111002465 111003523 111022771 111009795 810011355 810018942 810019925 810019256 111003967 111000610 111000702 111024388 111004175 111003284 111001854 111016839 810022120 111002446 111014918 111000763 111021976 111015618 111015634 111015640 111015611 111015616 111015613 111036294 111036299 111036292 111014921 111014926 111014936 111014942 111036288 111015646 111015643 111015631 111015627 111015626 111015623 111015641 111015620 111021990 111022094 111021985 111015649 111015639 111021782 111021817 111021792 111021818 111021852 111023718 111024089 111014925 111015630 111021856 111021719 111001584 111001578 111000986 111001571 111000989 111001010 111001570 111000984 111001003 111000987 111001017 111000985 111001576 111001886 111001867 111001860 111001885 111001868 111001865 111001562 111009778 111016825 111016828 111017229 111016823 111016826 111016838 111016831 111016842 111016834 111017223 111017232 111017248 111016822 111016761 111018134 111004504 111003059 111018295 111018123 111009755 111018107 111000671 111000709 111000797 111018126 111000704 111018284 111000775 111000753 111000723 111036304 111000777 111000752 111000782 111000768 111000703 111000762 111000784 111000705 111000766 111000799 111000767 111000780 111000809 111000795 111000783 111000754 111036424 111013990 111013989 111013971 111013995 111013966 111013960 111013964 111013957 111024170 111036429 111036426 111036425 111014954 111014920 111014941 111014939 111003648 111003525 111003546 111009703 111009757 810013261 810022382 810013364 810027131 810027129 810013354 810027118 810019053 810015838 810013208 810013044 810013032 810022125 810022132 111000442 111001550 111000269 810013443 810013429 111001090 111001582 111001568 111004509 111004517 111001419 111001543 111001018 810011367 810011485 810011332 111000454 111001418 111001569 111000524 810016465 810016367 810016558 111009801 810021678 810021553 810021677 810021659 810021683 810011325 810011352 810011311 810011345 810011512 810019025 810019049 810018944 810019052 810021528 810021514 810021582 810021728 111021822 111018281 111018259 111003268 810016154 111021992 810019926 810016273 810021732 810019002 810015835 810007402 810019055 111021819 111009838 810027541 810003381 810013079 111024373 111020988 810019540 111011977 111011914 810013213 111022956 810022357 810021599 111018809 810011400 111010760 810027005 810016340 810027127 810004569 111024383 810027096 810022384 111013811 810016643 810027045 111011219 810019541 810003537 810008480 111013954 810003612 111021030 111013955 111013962 111013959 111004260 810016645 111019439 111018685 111009730 810013262 810021554 111015632 810005699 810011536 810008500 810018974 810019034 111022953 111016764 810019262 810018970 810019412 810013441 810015817 810015801 810022109 810022379 810013226 810027158 111022938 810016101 810016269 111002010 111024404 111024371 810021680 111024189 810016382 810013465 810013089 810019031 111022588 111014004 111004249 810021709 111004268 111022967 810005236 111011440 810027521 111009739 810005701 111021844 111014002 810022139 111004247 111009742 111021978 810011397 810011480 810015944 111009770 111009265 111009256 111009251 111009264 810003949 810003950 810003951 810004012 111021018 111021014 111021015 111020980 111012301 111011971 111011455 111011909 111011993 810027017 111011352 111011261 111011172 810003504 111020964 111021031 111020950 111020999 111009722 111009802 111009746 111009758 111009716 111001083 111001068 111001579 111000471 111000578 111001855 111000469 111001882 111001005 111000208 111000580 111000466 111000209 111013953 810013087 111011353 111024207 111022075 111000677 111000533 111000728 111000725 111000720 111001088 810015948 810011507 111018793 111019467 810013233 810026945 111011608 810013187 810005247 111015642 810027125 810007401 810019015 810005294 111018271 810005235 810013186 810008479 810005316 111016835 111011956 111010972 111010795 111014946 810003620 111010794 111011182 111010958 111010821 111010809 111010973 111010968 111011348 111011371 111011609 111011454 111011258 810004813 111012313 111011641 111011637 111011949 111011630 111011621 111010811 111012268 810013029 810022104 111000148 810003511 111022079 111011356 111001887 111001078 111001069 111001012 111001011 111000994 111000998 111000997 810016464 111015636 810013452 111012278 810021736 111000755 111000155 111003063 111003035 111018661 111009842 810021600 810016446 111003351 810005357 111016723 111013984 810016272 111024187 810007422 111003374 111005526 111003370 810018981 810022102 111003147 111002721 810016347 810016637 111009843 810008602 111003649 810016679 810016279 810015808 810016368 111003368 111009252 111002732 111011450 111004256 111021705 111004197 111004507 810004821 810003369 810019251 810013035 810027022 810013058 810019026 111011364 111004500 810019326 111011357 810022066 810016664 111000164 810027552 810019294 111022766 111001590 111010819 111016732 111016743 111016715 111009815 111009728 111009825 810008516 111021452 810007450 810008562 810008487 810003619 810003613 111009804 111009806 810008610 810008524 810027090 810027093 810027107 810003560 810003507 810003554 810003508 111006794 810027092 810027102 810027098 810026948 810027095 810013251 810021730 810021511 111016766 111018789 111018787 111018662 111018682 111018267 810011495 810011541 810011542 810011456 810011444 111019457 111019432 810011363 111022594 111022591 111022900 111022946 111022962 111018282 111018112 111018290 810004817 810004823 810004810 111018269 111018277 111021428 810011471 810011279 111003188 111022082 111003175 111003352 111006795 810022362 111010805 111009745 111011451 111018649 111018665 810011466 111010793 810022380 810021534 810022067 810022063 810026984 810027001 111022933 810013050 810013258 810013222 810013257 810013259 810013252 810013228 810013234 810013260 111014927 111014916 111014931 111010806 810011515 810011286 111022763 810013280 810026977 810026933 111022593 111011354 111011243 810018941 810016266 810019314 810021603 810022377 810016567 810016308 810008598 810019916 810027527 810016129 111017249 111005521 810019411 111012461 111009709 111004530 111011937 111021463 810008584 810021695 810019061 111009705 111012297 111010758 810022080 810018994 810013435 810015809 810008572 111005528 810005246 810019536 111020947 111006544 810021670 810015952 810008641 111009761 810015957 111011959 111002736 111001077 111001000 810013243 810021676 810005707 810019417 111021830 810008601 111024200 111002438 111004712 810019263 810019549 111003958 810015805 810005305 810019534 111009253 111011355 810022081 111011220 111018122 111018108 111009258 810021584 810012605 810013437 810013211 111024169 810016576 810016677 111022973 810011392 810011537 810016114 810019265 810012606 810012602 111000450 810016680 810005283 111022086 111021316 810005284 111021700 810015777 111013958 111013963 810004825 111021439 111002381 810015776 810016451 810008568 810019414 810019542 111018264 111022764 810027009 810007445 111021312 810016667 810019517 810027021 810013182 810013469 810021614 810003371 810019521 111022583 810013363 810013470 111006786 111011921 111000532 810003431 111002441 111023530 111011267 111010817 111021467 810003465 810011284 111009785 111006543 111006536 810018964 111011605 111011256 111011897 111010959 111012264 111009797 810026983 810021693 810011488 111022785 810015818 111012469 111017242 111011980 111003675 111011625 111003655 810022134 810013286 810004814 111013803 810022097 810016343 810013366 810013273 810021507 810013263 111003191 111008593 810003425 810003432 111000146 810021366 111000761 111000606 111000101 111012298 101000035 111012319 111004277 111000463 810016312 810021350 111024174 810013281 810019554 111024190 111024173 111024172 810021359 111000663 810013353 810003436 810016544 111002386 111000140 111024401 111010762 810016358 111003541 111009823 810007464 810011374 810021384 810008600 810013194 111023549 810016134 111022584 810013191 111018667 111003266 810007451 111024203 810016133 111024568 111023529 810027549 111001085 810019912 111022950 810013040 810013041 111009765 810016437 111003322 810013475 111021442 111003498 111003270 810003559 111011240 111001032 810003618 111020975 810021572 810005704 810013246 810021726 810008623 810019064 111019455 810003679 111012311 111014949 111018124 810027110 111012280 810008616 810008560 111011193 111020951 111011975 111018788 111011234 111010820 810021529 810018960 810013249 810022130 810007466 810008577 810016349 111024403 111002740 111016734 111023524 810013082 810008482 111013991 810018979 111004186 810005320 810005368 810005250 111000103 111024384 111003658 111016843 111003494 810019036 111003564 810013365 810011348 111013967 810018959 111022760 111022773 810013081 111024372 111024400 111011441 111013978 810027147 810019051 111018672 810021731 810004809 810005702 810021363 111022087 111016767 810019046 810027546 810013034 810013059 810027148 810027012 111004198 810027007 111000143 810027550 810016553 111003341 111002405 111023548 111010456 810019295 111024201 111021011 111013994 810003349 810021590 810015781 111000211 111000212 810019253 810004567 810013483 810019523 810013450 810004566 810008580 810015823 810007431 111009255 111013974 810018940 810007404 111019414 810011509 111013812 810011497 810011309 111014011 111021813 111000710 111001075 111003053 810003558 810021569 810015947 111011953 111009717 111003959 111000676 810015955 111000607 111014008 111024188 111011382 111009839 810016125 810005365 810005306 111001028 111021446 810011393 111002394 810013334 111001004 111011950 810016348 111018647 111009808 810005367 810019431 111021799 810027108 111021977 810011342 810021375 810013201 810013482 810011362 810011384 111021788 111023538 810004752 111019427 111010774 111011349 810011292 810011517 810005280 810015774 810021517 111022578 111018294 111021315 111021443 810016144 111003127 111011178 111012459 810027525 810013209 810015792 111002402 810016319 111016844 111019463 111006555 111022940 810021509 810008583 111000665 111036453 111036561 810008405 111028682 810028576 810028578 810028556 111042323 810022327 810022123 810021704 810022124 810011498 810011521 810011446 111000271 111018132 111003366 111011595 810007438 810013350 111024182 111024196 111021981 810021734 810019022 810013473 810021562 810016100 810016107 810016128 810016106 810021692 810021694 810016362 810016142 810022085 810011376 810013480 810019267 810019024 810011321 810013037 810013039 810013063 810005352 111021805 111021833 810021588 810011354 111024397 810013096 810022072 111021435 810011366 810021696 810011455 111010978 810011472 810007408 111003862 111003662 111023552 111018280 111002746 810011370 111001863 111003263 111001030 111002449 111012479 111012270 111002738 111002734 810018957 111011196 111016736 111011988 111003342 111020985 810011467 111019417 111024396 810011494 810011336 111011257 111010434 111016719 810019914 111024085 810019259 810019919 810021552 111020978 111012257 111010455 111010788 111003281 810003367 111011239 810016455 111011186 111021787 111021789 810008576 810005281 111012456 111011199 111003038 111002006 810019286 111003962 111003381 810021544 810021365 111011251 111011962 111022761 111015635 111011218 810016141 810011278 111015614 810027122 111011176 810027540 810018990 111021801 810007446 111013806 810011324 810011330 111021809 810021573 810021617 111002705 111003355 810003341 111002747 111024389 111003045 111003146 111003183 111004168 111024086 111008703 810027164 810027151 810027523 111023532 111018642 810011445 111004286 111004177 111000701 810027008 810018951 111022776 111000157 111018113 111018807 111020957 810004643 111020953 111018797 810005241 810005315 810005289 810013036 810015778 810019533 810019316 810019281 810015791 810015829 810015810 810018956 111018808 810027019 810013103 810013056 810015825 111003128 111009763 111016722 111008702 111011592 810016574 111009787 111018803 810021716 810016314 810027533 111003323 111004515 111021808 810005285 810016377 810008629 810016551 111009752 111009711 111001070 111004527 810003438 111018812 810027042 111004520 810021619 111018683 111012269 111022579 111021835 810015793 810003351 810022144 810003552 111009805 810021515 111010451 111015624 111012480 810027543 111009706 111009816 111021786 111003952 111014933 810011540 810016152 810016099 810011389 810019520 111001021 111020965 810008586 111010767 111021005 810021550 810027538 810016127 111024217 111000106 810016375 111023525 810019045 810021675 111012282 810015953 810008492 111010768 111019458 810008551 111004176 810019283 810016448 810026996 111009827 111003951 810027532 111012306 810013055 111031417 810026998 111003353 810011518 111011952 111011904 111022939 810005252 810021525 111012320 111004187 810021598 810022369 810027528 111010766 111012304 111010963 111024366 111010791 810003368 810019006 810018968 111004524 111000568 111010756 111004640 111004715 111000153 111000448 810005703 111008586 810003337 111000526 111003180 111003657 111000561 111011594 111002742 111011992 111000145 111000324 810015960 810015813 810019405 810016569 810013449 111021798 111022775 810004807 810016376 810019269 810019311 810016545 810026929 810026939 810021708 111017235 111003174 810003433 810027516 810016441 810019005 111009762 111009239 111000321 810016268 810027152 111023541 111003144 111020971 111020989 111024216 111024206 810007457 810007424 810026936 810013444 810016443 111018668 810011452 810019279 810003372 111012291 810016143 810027520 810027517 111013986 111011361 111013997 111012253 810003464 111002403 111011264 810016658 810011299 810019333 111023568 111010792 810019915 810018978 111010784 111011965 810021358 810011282 810013031 111018273 111018286 111018279 111015858 111000259 111015612 810016352 810021685 810003735 111015857 810013283 111012265 111012465 810013220 111014915 111016762 111016765 111021796 810027018 111016829 111018680 111010765 810027123 810022108 810021377 111021695 111021785 810016130 810004571 810013284 111004642 111042196 810019553 810027011 810016160 810003380 810019924 111020997 111021784 111022765 111021827 111021536 810013184 81",
      "more_code_info": "0026941 111021704 111000660 810016139 810021604 111016737 810008491 810022387 111009724 111009702 111020954 111006790 111020991 810016452 810016096 810004562 810016562 810003460 111011939 810004563 111021444 111016845 111000668 111006777 810022098 111006800 111020981 810027101 810016117 111000251 111013961 111008596 111002411 810027153 111016717 810027091 111016824 111009700 111010803 810008555 111000670 810027121 111022580 810027139 810027134 111020969 810008529 111011945 111000569 111001963 810008643 111018666 111022095 111023546 111018655 111018283 810008495 810013471 810021506 810008552 810016271 810016335 111011266 111023537 101000037 111002430 111000609 111009803 111002415 101000040 111002416 111001022 111009710 111000215 101000033 111009242 111001551 111003501 810022360 810003383 810015798 111010955 810008626 111011431 810015771 111023733 111041996 810006776 810006770 111042319 111039840 111021451 810008513 810003373 111003491 111010468 111022782 810019303 111018799 111024079 810011514 810007419 111003187 111003337 810026932 810026986 810019923 111003375 111003343 111003121 111003078 111021840 111011972 111014934 111011443 810013355 111014944 111002720 810008511 111010808 810005326 111003499 810007412 810016385 810003885 111022783 810026940 111009775 810013214 111003056 810008624 111014953 111003669 111018645 111003261 111003145 810008569 810003621 111003694 111012254 111011238 810008574 111012277 111003522 111003139 111022072 111002707 810008645 810013264 111003338 111021975 111003529 111011638 111011362 111021807 111010956 111010789 810013100 111021832 111003557 111019444 810008628 111022577 111018811 111001992 111003143 810026974 810016284 111003697 111003274 111003511 810008530 111011192 111010798 810005242 810003557 111003132 111011373 111018653 810026975 111003087 810003502 810007439 111003057 111003313 810007428 810003555 810007465 111009748 810007440 810013267 810003510 111003273 111003047 111003327 111014924 111003075 810021684 810013253 810013229 810022095 111010971 810013360 810021660 810008520 111003271 111011913 810005308 111003088 111003068 111003044 810016325 810013048 111002693 810013030 111003066 111003663 111003378 111003058 810007456 111011366 810003883 810013033 111003124 810008514 111003070 810013202 810013462 810008475 111009736 810013358 810012609 810003882 111022081 111019440 810022093 810019434 111003280 111003373 111010810 111011205 810016162 810016380 810013235 111009836 111003262 111003049 111010965 111019442 111003182 111003282 111002406 810021548 810008515 111009263 111012286 111003357 810008471 810013080 111003666 111021430 111001987 111008585 111003288 111010967 810005358 111021432 810005314 111011991 810011545 111003258 810003736 810016158 111003526 111010804 111002715 111003527 810003884 111001861 111021993 111012468 111001087 111003135 810016283 810018971 111019466 111021989 111011923 111020955 111021535 111021980 810016161 810019404 810013105 810008485 111021836 810013254 810016126 111011617 111003062 111003646 111003507 111019420 111002698 111003140 111003264 810019008 810007409 111002714 111012275 111003324 810013342 111011629 810013352 810008523 111022947 111020998 111003650 111002425 111009729 111003664 810007407 810016120 111011984 111011368 111003192 111002451 810005355 111022078 810003506 810026927 810016315 111003184 810003733 111009259 111002700 810021556 810013189 810019032 810013042 810022378 111021465 111003346 111003358 111003178 810011390 111002749 111011189 810008539 810008564 111009776 810008630 111003079 111001971 810019918 810008611 810008494 810016159 810021504 810013337 810021561 810007471 111003671 810013345 810021686 111009238 810016108 111011613 111002713 810008642 111006788 111003524 111003492 111031402 810019287 810026893 810013043 111003556 111018674 111001988 111001982 810026543 810011502 810021668 810016671 810011291 810013221 810005287 111014952 111012267 810003681 111011926 810011532 111023523 810011277 810008540 810008631 111023528 810011360 111023563 810016147 810019298 810008617 810022375 810019039 810016304 111042193 111042308 111014279 111024721 111024087 810011460 111006553 810007455 111001998 111010813 111011262 810021353 810021371 810021521 810019419 111002000 810011484 810013195 810013282 810027518 810015830 111019445 111000477 111000444 111003041 111003186 111003279 111003149 111023527 111020972 810011288 810011443 810011375 810011344 810013049 111005534 111005532 111021854 111021721 111021715 111021460 111009250 111009719 111009721 111009727 111004512 111004641 810028232 810004811 810004819 111003660 111010467 111010463 111001996 111008581 111004169 111004251 111003545 111010764 810004564 111002445 111002427 111003171 111003369 111003329 111003490 111003330 111003349 810008512 111003676 111003553 111003681 111003670 810004827 810003680 810004822 810004642 111009774 810004701 111003683 111003661 111003856 111004529 111004278 810008589 111004254 111004279 111003966 111003656 111003696 111003690 111003532 111023560 111023518 111023536 111023569 111023556 111011195 111011222 111011268 111011188 111002385 111002431 111002423 111002428 111002424 111002377 111002413 111011606 111011978 111011602 111011624 111011903 111011438 111011623 111001977 111013982 111011250 111011175 810003673 810003463 810003675 111002703 111023727 111023744 111011452 111011902 111024209 111003319 111023737 111011445 111016739 111011927 111014951 111005525 111023722 810016359 810019537 810019538 810011357 810011469 810021370 810016341 810004761 810021361 810011474 810011380 810004754 810004829 810004757 810003732 810003556 111001972 810004697 810004695 810004756 810013440 810013468 810013434 810013481 810016577 111010459 111010460 111002009 111002008 111010783 111006537 111011943 111003190 111003360 111004257 111004269 810019543 111003549 111022942 111022964 111023741 111011928 111024195 111004719 810016282 111021313 111021702 810005351 111004289 111011379 810005304 111004273 810005295 810005249 810005258 810005364 810005312 810005255 810005322 810005310 810005251 111005536 111004502 111010960 810008571 810013057 111002387 810018961 810018984 810005311 810005350 111004280 111004196 111004516 111004639 111004248 111004258 111004644 111009247 111004510 111004521 111008591 111008700 111016744 810008588 810022358 111003077 111003076 111003071 111003060 111004276 810016561 810026946 111002456 111002460 111002384 810008649 111002750 111002737 810016538 111009243 111009266 111003372 810004812 810004820 810004808 810027105 810026930 810027130 810019021 111011940 111011917 810026949 810026971 810019003 810019023 810016651 810013064 810013086 810004694 111006547 111003547 111010775 111012288 111002407 111004243 111004194 111004195 810008625 111003860 111011930 111011171 111011183 111011184 111003361 111003340 111003317 111003356 111003331 111022949 111022981 810027010 810008590 810008579 810019547 810011351 810011516 810019035 810005309 810016575 810007436 111016821 810022121 810021724 810021622 810007462 810011379 810011486 810021508 810026947 111011622 111012471 111011932 111012274 810011382 810027015 111024391 111011270 111012262 111006532 111011249 111016742 111010964 111003379 111015633 111010777 111010449 111021457 111021829 111021317 810021565 111027683 111027658 111027654 111011966 111024081 810004845 111021800 111021794 111024375 111019468 111019433 111019425 111019411 111009241 111022092 111019460 111021847 111023735 111024210 810013279 810013242 810015951 810015954 810021665 810021557 810011544 810008538 810008554 810013255 810013215 810021615 810021673 810019409 810019519 810027031 810026958 810027156 810027112 810016540 810016554 111010458 111016832 111014001 111012473 810022061 810022074 810016338 810018975 810027002 810016571 810016537 111003141 111012315 111004717 810013088 111004284 810003429 111017241 810011489 810011289 810011395 111024385 810015965 810016302 111000804 111014940 111016735 111016729 111016718 111004184 111009818 111009833 111003693 111015622 810021587 810022372 810021536 810021543 810021524 810013343 111011960 111004505 810021581 810003676 810022142 111003489 810008566 810027539 111022951 111010807 810003539 810016678 111024179 111003558 111003177 810008522 111020984 111018265 111012303 111011599 810003339 111010452 111011931 111014000 810004565 810003455 810021360 111015617 810011546 810027551 810011473 810011525 111013968 111024405 111024212 810011313 810013457 810007461 810021703 111019419 111021004 111011187 111024080 111003069 111004181 810008647 111002379 810019430 810019029 810026991 111020976 810013461 810008478 810022366 810022062 810011450 810016374 810011479 810013359 810013268 810011391 810016275 810016278 111009713 810013256 810013247 810013347 111015628 810021727 111011370 111000320 111000107 810021713 111011202 111011232 810011501 810011482 810011476 810008481 810008473 111000544 111000573 810021669 810021720 810021715 810012610 111000534 111000769 810016350 111018116 111005522 111000457 111006540 111018650 111009756 111019461 111019409 111019413 111019428 111019438 111019426 111019410 111019429 810008558 810008619 810008517 111017238 111019469 810019415 810018947 111000539 111000564 111000798 111011987 810013459 810013455 111018266 111000104 111010815 111019452 111000765 810013098 810013051 810013052 810022069 810022089 810016450 810016149 111000443 111000273 111000549 111000258 810008488 810011464 810011340 810004848 810004656 810004760 810004824 810004826 810019406 810019273 810013431 810013474 810013476 111001989 111013956 111013999 810019060 810019321 810019276 810019270 111020995 111020983 810008537 810008622 810016456 810016331 810016281 810016454 810019261 810019302 111022931 111011615 111023554 111002435 111002463 111002452 111004506 111009824 111001980 111010975 111023544 111011982 111004253 111036287 111016840 111005539 111023721 111009720 111011351 111003956 111036298 111036428 111008587 111008608 111020967 111018285 111018260 111000727 111004165 111002692 111012300 111011446 111011958 111016827 111016724 111001558 111004174 111003376 111012308 111018272 111018121 111011213 111036414 111036440 111036420 111041043 111022780 111036433 111001565 111036437 111036303 111009786 111009704 111023535 111003866 111011611 111036416 111018663 111003259 111041044 111009249 111023521 111041047 111014922 111010770 111041038 111004722 111011378 111036422 111022592 111001586 111001557 111023728 111036430 111036295 111009715 111011174 111041045 111036290 111036285 111016720 111001564 111001573 111036293 111001563 111041039 111036431 111011635 111036296 111036435 111023557 111009783 111004288 111000111 111001567 111002702 111001015 111000110 111036284 111036438 111036305 111010970 111036413 111011208 111011203 111011636 111011986 111011447 111003086 111016721 111018677 111023730 111001545 111036436 111001978 111036434 111021816 111001546 111009750 111036415 111036439 111041040 111022957 111002726 111011359 111021447 111010466 111024370 111001544 111024211 111011235 111008607 111011604 111018289 111003061 111023724 111036286 111041042 111021534 111024215 111001572 111003961 111002447 111022978 810021386 111018669 111018678 111022970 810007420 111016846 111013976 111019465 810019913 810003462 111013977 111022096 810003509 111036442 111011439 111011979 111012314 111002391 111001970 111004170 111003067 111003080 810013101 810011453 810027109 111018805 810019056 810018967 111011593 810018948 111010763 810022131 810022119 810022388 810015833 111010438 111011634 810026951 810015964 111019453 810015834 810027513 111006796 810015799 111011607 810013244 810008606 810011483 111013973 111022976 810027032 810027028 111003148 111003509 111018657 111003561 111004283 810015946 111004188 111011896 111009845 810013236 810013193 111011223 111018297 111018125 111018274 111018287 810011341 810016104 810013199 111024365 810026970 111006538 810004692 111003955 810016131 810011454 810005354 810005239 810021714 810013351 810011396 810011491 810016670 810013362 111002745 111021706 111017240 111021810 810011294 810027124 111021793 111001086 111003520 810005240 810026967 810027524 111019424 111004526 111003503 111011173 810011534 111021453 810016639 111019448 111015647 111018671 111019412 111024369 111022586 111001983 111019454 111000542 111000468 111000999 111011941 111011964 810003435 111000993 111002711 111001560 111001556 111021012 111020973 111003528 111000724 111000142 111011907 111011898 111000726 111000729 111000152 111019430 111003542 810011500 810013102 810021568 111018794 111021713 111002012 111001555 810016115 810018950 810008627 111022779 810021723 111003181 810026963 810022365 111004503 810011371 111021718 111020990 111010977 111001074 111003131 810005254 111004162 810013219 111003126 810015950 111001033 810008636 111005523 111011259 810016119 111003043 111011448 111002466 111014007 111024177 111001574 111022774 810016118 111017244 810011307 111004265 111001002 810003541 810005257 810013072 111004723 810021609 810027542 810022077 101000032 111014006 111016836 111002461 111012299 111011449 111012273 111011601 111001547 111001589 111036302 111003089 111012455 111011269 111021448 111001858 111008595 111008600 111018135 111036419 810015785 111008704 111020949 810005244 111005524 111014935 111009769 111018114 111000108 111001031 111000112 111000149 111000322 111000248 810011333 810011312 810013237 810013270 810011447 810026985 810015795 111009738 810011323 111018646 111010796 810021682 810016365 810016276 111020966 111013805 111019418 810011543 810011300 111011944 111011435 810022133 111002723 111006781 810008472 810013217 810018987 810019420 111003867 111021848 111003275 111009791 111002744 111005538 810016640 810008618 111011265 810003540 810007435 810019058 810016364 810007475 111004241 810015807 810011398 810016351 810004818 111004525 111022965 111023719 810022075 810008528 111003859 810008593 810004700 810003457 810004698 810004699 111009699 111009760 111009712 111004499 810008578 810008489 810008581 111010440 810011280 810021558 810021547 810013095 810018949 810019328 810019285 810019292 810019327 810019329 810019288 810019436 810019418 810019432 111001559 810011490 810019266 111004290 111000806 810019315 810019257 810019249 810011465 111002455 810016462 810008521 810008634 111012289 810004693 111003868 810004652 810013204 810004755 111011244 111022958 111012474 111011632 810019544 810019545 111022960 111022952 111022948 111003269 111002709 111021455 111007520 111007035 111007955 111006550 111036469 111006545 111019437 810016646 810019312 111002014 810016528 111004266 810021381 810012604 111003964 111011619 810016535 111006549 111012466 111003550 111036432 111001862 111009792 111036423 111005529 111036307 111036306 111001073 111036427 111036300 111002396 111018296 111018681 810022325 111001879 111002383 810008561 111023539 111012310 111003179 111012287 111009707 111004285 111002001 111001009 111001864 101000046 810008633 111010786 111007025 111004167 111003692 810016332 111010802 111009812 111010954 810008493 810005353 111022944 810011346 111006531 810018998 111023742 111023731 111023526 111023550 111023555 111023565 810019274 111023743 111023567 810008605 810027141 111004191 111016741 111008697 810019320 111008580 810015956 810022135 810022101 810013212 111010797 111002439 111001991 111002716 111003534 111003537 111003667 111003495 111011961 810016548 810026978 111003684 111010446 111010782 810013479 810004753 810011529 810016442 111002724 111002718 111002697 810011493 810021387 111009831 111011600 810005700 111024191 111008594 111015619 111003133 111011358 810003439 810021533 810008565 810003384 810003426 810021711 810013356 111011210 111021838 111011369 111002382 810011442 111001856 111009698 810016109 810013026 111022934 111022935 810026994 810026972 810011386 810011528 810022105 810021663 810013046 810011350 810011399 810019067 810026544 810008596 810016661 810027132 810008608 810016527 810016366 810016546 810016531 810016570 810008644 810008582 111004528 810016672 810019413 810008646 810016556 810021577 810008496 810013028 810019416 111006534 810016333 810013027 810016530 810008484 111015629 810016674 810019921 111012263 810016660 810008477 810016371 810016360 111011631 810016164 810015831 810026976 810013060 810011361 810005243 810005293 111009826 111009817 111010445 111015621 111012276 111011639 810003347 810019514 810019555 111022786 111008601 111001881 111011376 111014923 111023558 111021709 111011177 111011433 111001979 111004189 810021706 111003953 810015803 111000163 111000249 810015941 810015942 810016381 810016309 810015949 810021385 111023726 810021679 810026960 810018973 111024386 111009701 810019309 810026995 111004519 111003647 111000445 101000014 101000044 810016342 810027038 810013084 810016552 111024387 810026934 111024374 111010444 810013453 111036664 111025584 111031122 111031153 810016136 111009832 111001958 111001888 111001552 111002450 111000459 111018792 111003857 810016320 111001869 810008613 810008635 111008698 111023740 111011951 111022980 111022943 111003864 111002015 111010816 111011198 810004654 810004653 810004655 810004644 111010759 810022389 810022390 111004275 111004263 111022972 111002443 810004646 810003948 810004759 810026980 111011180 810003505 810003503 111022968 810016654 111023533 810011388 810008614 810003561 810003538 111023553 111002005 111010769 810021700 810021698 111011942 111003064 810003453 810021546 111012292 111000465 111004511 810005288 810007429 810013460 111019449 111021855 111003665 111008606 111021804 111011347 111021449 810021701 810013179 111004261 111003185 810015819 111004264 810008585 810027154 810019277 810019551 810019548 810022068 810022138 810026981 810019028 810026926 111021839 810003382 111022084 111021701 111013992 810011510 111002706 810016383 111004252 810013472 810015806 810027143 810027126 810027167 111010470 111000257 111001588 111009821 111004244 111001575 111020961 111014943 111018651 111018637 111018675 111018638 810021688 111011983 111018118 111036289 111036441 111011598 111011252 111011596 111006539 111010757 810016436 111017227 111011437 810027133 810019317 111023559 111041041 111004262 111036301 810026993 810027094 111009714 111016768 810007472 810007418 810015782 111023566 810004645 111036297 111023729 111041046 111022784 111022769 111022787 111022941 111036418 111036291 810022374 810022070 810022361 810022367 810013265 810013288 810013285 810016668 810016655 111001959 810022111 810027138 810027145 810027136 810016369 810027137 810027113 810016370 810016353 810013438 810013054 810021513 810013104 810013092 810013180 810013181 810013207 810021690 810021596 810021540 810026925 810026935 810016675 810016318 111021791 810016522 111022954 111023738 111011616 111011242 111009244 111013100 810018945 810019518 810013271 810019440 810016565 810019428 810008648 111036421 810013085 810019001 810019516 810013467 810013344 810008567 810011281 111003054 111010778 111018298 111011933 810021512 810011520 111009790 810011463 111011263 111012260 111000759 111000760 111003367 111003334 810016525 810016439 111000210 111000250 111000793 111011924 111011989 810011539 810012601 111010453 111000540 111005535 111009784 810007467 810013210 810007453 810008476 810008570 810008639 111027706 111025628 111027779 111024842 111015230 810026903 111027921 810002218 111013077 111027874 111013561 111022690 810019569 111024445 111036660 810006624 111023328 111027922 810026859 111027778 810026789 810006596 810000849 111027608 111027766 111018773 810021642 111040028 810021755 111007209 111034929 111027655 111036478 111042324 111034930 111026476 111036467 111033466 111022729 810001209 111025740 810028539 111027713 111036513 111033668 810019583 810001233 111031180 810028185 111005272 111009017 111013790 111022701 111042149 111007574 111025468 810001341 111030972 810000561 810001594 111013069 111016930 111027817 111036497 810026786 111027826 810000694 111033838 111025764 111026268 810019620 810019342 810026846 111036495 111013564 810019946 810000695 111023347 111013110 810028500 111027812 111033763 111042161 111007719 111007964 111007043 810008164 111007958 111031540 111026702 810021811 111036532 111036466 111015163 111031410 810008931 810008464 111026713 810006758 111031480 111015209 111014699 810021873 111027900 810026838 111007933 111015183 111015165 111036910 810019334 111024877 111039841 111005198 111016712 810000962 810000584 111013748 111007082 111013904 810001320 111007715 810000582 111026467 111024826 111027552 810028526 111042207 810001581 111040162 111033867 111033865 111036746 810019366 810002214 111024447 810028258 111031554 111025501 111007037 810008463 111036941 111025762 111034904 111033626 111033877 111036589 810005816 111040175 810005779 111025102 111025734 810002057 111036356 111033982 111007542 111036392 111025120 111013896 111009027 111015591 111020730 810008414 111030966 111036535 111020729 111025080 111013118 111012716 810002635 111007090 810021972 810001935 111027794 111039967 111018774 111024428 111015171 111014665 810006607 111036675 810028326 111033532 810002071 111033784 111024014 111016515 810003285 810000888 111007682 111024011 111036919 111040033 111031407 810003284 111031118 810026738 111025646 111025593 111033657 810001583 810028238 810028230 111022921 111033898 810021436 111033727 810028255 810021769 111005346 111007728 111023309 111013132 810005811 810000963 810026706 111027615 111027586 111031452 111040018 111031419 111036327 111042335 111040031 810026735 111027767 810021485 810005783 810005797 810019388 810002240 810028542 111027738 111025536 111023338 111026721 810028265 111015193 111042249 111036935 111025657 810027083 111034933 111036687 111027666 111031484 111028686 111031398 111013580 111025590 111025587 111031130 111013065 810008386 111025527 111013559 111013087 111018564 111027605 111027756 111034926 111027858 111007576 810021417 111036352 111036459 111036686 111031223 810028206 111036596 810028170 111036558 111036461 111031216 111036320 111036624 111036636 810028225 810028199 810000628 111007057 810021400 810000966 111040167 111028517 810028210 111026273 111027760 111025749 111005348 111036590 111027924 111031503 111027810 111027696 810028256 111024717 111025131 810006606 810006599 810026892 111022423 111033907 810002219 111026468 111015241 810021395 111013697 111022435 111015153 810011573 111034055 111025481 111009019 111033754 111027939 111027832 810000795 810019612 111013593 810022329 111027730 810000504 810022013 810021804 111034948 111033618 111036537 111033640 111012705 111015474 810008426 810000957 810005798 111013112 810008864 810002978 111015488 111018556 810008456 810026731 810000959 111015234 111015497 111012693 810026778 810008871 810005815 810008939 111014648 810026768 810008934 111014644 111018565 111015500 111013693 810008403 810026736 810008465 810008402 810005821 111015491 810008798 810026777 111022401 111022409 111025525 111018555 810008937 111013084 111020742 810002221 111026716 111005277 111031413 810026747 111007735 810008865 810021989 810000479 111025735 111013113 810007343 111013910 111027729 111031416 111027852 111017068 810021763 810002483 111024544 111007688 810002738 810028183 111033660 111015147 810021952 111042209 111034913 111007947 111034054 810026813 111023053 111031170 111031376 111006334 810006628 111040168 111031479 111027681 111014667 810006616 111025541 111007699 810013119 111036488 111024863 111005200 111026705 111033484 810028229 810005788 810000283 111042247 111018576 111036514 810026833 111033718 810026845 111036745 810022153 111040076 810007369 111005212 111040157 810001100 111026696 111007672 111042250 810001925 111031169 111033603 810002478 810026791 810002968 111025478 810002997 810001944 810006645 111013689 111036755 810002567 111025673 111027749 111007030 810002006 810011556 111024524 810011551 810006642 111027910 111027892 810002058 111007074 810021750 111020761 111031396 111025696 810002813 810022022 111022412 111022398 111025517 111033809 111007080 111015833 810002778 810021870 810008423 111013760 111007525 111022716 111036905 810028577 810019560 111022724 111018560 810001322 111033542 111042259 111013572 111014666 810008450 810006769 810028251 810021777 810021944 111025486 111042273 810001023 111018553 810002162 111027685 810021947 111016708 810006727 111016938 111027813 111013543 111025524 111033737 810000935 111042302 111007533 111024537 111013545 111036949 111016532 111027561 810001588 111028474 111007939 111036646 111027788 111007711 111036555 111036741 810000282 111014285 111034043 111016695 810002014 810002001 111036708 111036582 111007521 810026869 111036757 111028479 111034894 111027784 810002101 810028574 111005206 810026714 111034890 111041999 810028545 810021644 810028334 810002166 810005820 111020736 111024012 810026717 111040165 111026701 111033612 810000251 810007356 111007208 810021990 810028524 810001928 810002012 810000284 111033861 111040034 111018575 111040037 810001998 111026577 111026695 111036652 810000249 111025533 810000106 111007942 111042187 111016697 810028560 810002171 111013781 111027828 111040016 111025084 810021936 111007213 111015204 810006620 111007084 111033691 111027665 111005330 111036580 111036517 810019603 810006635 810021959 810006653 111017079 111028697 111031457 111025638 810002376 810026839 810027062 111007051 111025759 810001032 810019396 810028247 111027854 111005227 111024818 111023348 111012718 111012692 810002520 111012726 111013563 810001986 810001937 810001940 810002047 111024836 111014658 111013571 810000697 810000853 111027768 111020720 111027835 111027935 111023349 810028346 111031421 111023317 111033526 111027928 111027720 111022919 111031465 111033525 111014655 810019372 111015442 111015603 111012717 111027761 810021928 810001465 810021451 111025601 111022920 111041998 111036552 810019361 111031206 111036637 111026691 111026261 111026238 111036909 111039876 111036742 111013777 111014643 111015203 111014292 111018776 111015492 810001225 810028565 111015440 810028180 111042288 810028220 810026822 810002317 111005241 810006747 111027669 111026472 111027719 111027741 810002973 810006754 111027541 810006771 111033599 111026583 810006714 810006720 810008466 810006715 810006786 810006756 810006601 111024838 111024860 111025130 810021915 111024718 810021933 810000548 810021793 111025690 111013590 111025691 111024830 111024709 111013550 111025750 111024837 111040047 111024849 810021955 111031196 810019613 810022017 111036671 810019353 810019350 111042159 111013547 111013609 810021998 810021640 810021490 111025532 111025559 810000596 810008398 810008461 111042188 111042174 810001129 111033721 111009021 111025630 810000508 111034881 111025115 111036728 810001533 111033686 111028510 810026757 810006611 111025614 111028507 810001037 111036588 111027717 111022691 810002775 111013555 111033850 810022328 111007943 810002772 111028476 111005226 111042186 111005201 810028584 810006625 111026251 111013130 810019615 111007042 111031143 111015845 111027763 810021970 810002995 810022334 810026750 111027860 111030983 111025752 111012733 111007089 111036648 111016924 111039842 810021925 111013738 111025742 111036927 111026446 111036929 111036958 111025571 111031224 810026721 111033863 810022290 810000626 810001162 111031217 111026258 111027925 810026732 810001125 111026611 810001036 111025491 810006626 111025602 810026809 111033534 111025492 111033499 111030984 810026785 810028562 810019574 111042163 111028515 810000863 111015222 111023334 810002532 810001939 111033586 111033834 111007022 111030971 111042173 111033629 111026574 810006726 111036939 810022321 111027687 111027558 810022320 111027653 111014264 111027800 111028489 111031211 111034884 111039887 111028480 111031204 111017073 111034857 810000280 111014288 111031441 111034882 810002073 111027785 111005222 111036519 111022439 111007049 111031553 111013737 111036481 111027814 111026568 810001921 111036401 810021934 810021478 810028510 111022416 111022696 111034921 810002773 111025100 111022437 810002808 111033719 810001291 111033892 810026861 810001216 111027752 810028522 111024840 810000505 111024833 810002741 111013594 111031472 111027733 810027073 111007935 111036618 111007937 111007941 111013952 111025664 111025627 111013577 111025683 111025702 111026607 111013889 111028509 810002481 111028483 810008163 111024714 111028475 111036578 111031478 111013761 111020763 810026749 111039838 111013900 111014695 111015832 111014679 810019402 111027692 810000864 111016922 111013753 111028676 111016689 111015465 810022295 111025504 111030978 111027861 111026254 111023997 111036593 111020744 111022930 111020768 111015495 111033673 810021433 111013727 810001211 111013749 111026256 810002216 111027604 111031475 111033480 810019605 111023996 810026814 111036761 810026797 111031542 111007684 111036688 810005775 111013768 111027808 111028692 111033591 111023992 810001128 111013702 111036391 810021465 111026709 111028677 111031259 111036719 810006641 111026252 111025562 111013109 111027919 111025576 111025685 111033764 810001025 111024829 111036736 111022417 810021971 810028339 111022406 810001229 810003273 810002777 111030974 810006657 810006640 111018557 111007708 111039843 810001340 111013605 810021861 111034065 111013719 111025556 810021775 810002174 810019354 111005338 111018550 111013089 810000758 111024528 810006673 111040030 111026710 111034874 111033620 111026593 810026871 111026612 111031202 111033636 810022326 111026422 810000796 111022932 810028195 111025554 111025658 810002327 111033580 810002536 810002480 111039971 111036318 810008467 111033545 111042292 111027908 810000700 111031450 111039884 810000250 810000167 111013613 111028688 810000243 111028679 111036315 810001213 111036349 810001484 111013050 111012691 111031512 810021404 111033681 111027577 810026807 111028496 111040000 810019389 111007661 810026772 111036731 810019578 111013773 111042169 111042303 810000464 111023058 810002980 111007731 810026798 810019340 111007956 111027856 111014269 111025591 810001284 111007673 111033606 111012721 810019565 111042339 111022699 111027881 810000830 111020728 111042265 810000851 111005341 111034905 111034895 111034898 111033752 111005216 810021773 111036344 810002271 111024841 810001020 111014633 111036698 111031378 111023329 810019378 810007376 810000923 111034071 111042260 810028218 111027705 111033621 810028205 111033905 810002570 111028691 111026687 111026694 111013735 111033819 810026702 111031125 111007024 111005229 111013083 111033689 810026904 111013115 111015182 810000047 810000049 111015156 111015239 810000065 111023062 111023060 111036696 810006651 810002972 810028540 810019636 810002385 111033471 111041979 111031168 810000686 111031404 111031165 111020759 111040069 810001071 111027791 810026843 111014256 111022694 810019568 111024716 111027704 111031383 111036368 810028332 111028502 111024844 810021963 810006662 810028235 810000563 111015232 111025632 810001324 111039829 810013128 810000559 810001934 111007571 111025110 111033888 111027651 111016917 111040171 111031500 111027937 111027672 111012694 111012730 111033536 810021739 111036764 111031252 111031434 111036630 111042315 111036355 810001158 810028209 111022425 111007077 111014696 810001072 111027590 111033760 111031490 111013607 111033862 111026255 810000247 111027901 810021480 810001096 810008436 111018759 111036450 810000578 111006325 111040086 111007206 111007931 810002069 111013604 810000523 111024013 111033613 810008938 111036926 810028325 810021802 111025497 810021740 810028217 111007536 111015455 810001084 111007713 810001022 810000954 111027680 111041983 810002807 111018756 111012722 810026857 810019373 111024533 111040029 111042015 111014277 810008397 111031411 111031505 111024523 111027848 111015187 111027753 111040189 111007534 810026826 111018758 810019927 111018570 810000890 111012732 810003272 810008433 111006324 111033548 810019394 111027872 111022698 810019356 111007210 810028506 111033847 111033729 111042212 111026449 111025480 111013587 810022016 111014630 810001938 111030982 111005235 810021954 111036942 810026868 111034892 810006663 111036460 111017081 111025600 111022926 810001051 111016940 810002486 111024448 111025758 810028511 111040024 111027548 111005344 810028213 810008171 810019343 111033703 111027759 111014249 810002437 111040186 111007671 111015472 810027050 810019371 111036357 111005238 810022296 111014266 111013064 111006333 111013570 810000185 111033554 111031186 810021978 111036706 111036528 111034027 810022298 111025598 111040172 111014709 111015494 111027864 111023335 111005253 111018569 111015604 111023346 111023312 111023341 111027883 810001163 111025078 111013723 111007511 11 1007207 810000625 810000611 111013576 810000323 111015457 111015215 111025739 111041982 810003282 111033740 111033713 111026437 111036339 111007965 111027587 111027863 111013569 111016520 111014713 111033544 111033890 111031133 111013120 111014276 810001276 111020739 111033671 111020740 111033576 111007721 810026771 111023324 111026240 810028204 810002434 111027913 810002533 111033666 111033625 111033835 111027578 111015443 111024411 111031456 111031121 111033651 111036379 111027790 810026695 111031468 111040156 111022710 111015503 111005347 111022719 111015214 111022695 810027076 810021864 810026549 810028262 810019401 810008459 111027537 810001101 111036725 810001044 810001131 111042298 810008443 810008170 810008448 111027557 111027542 810001283 810005781 810006780 111031263 111033776 111031448 111027754 111007945 111013096 810000436 111014700 111007712 810000050 111013709 810001262 810028345 111033694 810000061 810000547 111014711 111013063 111007959 810001242 810001279 111034875 111007736 111007689 111007734 810027074 111027684 810021647 111009015 111015225 111031461 111005207 810000690 111033645 810002443 111016936 810013118 810019638 810002168 810002831 810002564 111033814 111013067 810000850 111024855 111015852 111015854 810000482 111025487 810000585 111026566 810021806 111012723 111007086 810006629 810019940 810026913 111031556 111031134 810001301 810021953 810008424 111016904 111013730 111031171 111013085 111031538 810001255 810001924 810002669 111024874 111027718 810006746 810021924 111007687 810005823 810021749 810000594 111009008 111025651 810022164 810021401 111026604 111015856 111031513 111025568 111015493 111023313 111033642 111028498 810002167 111030969 111033467 111022925 111033747 111031487 111040081 111036709 810028261 111036678 111015501 810002381 111033641 111036659 111015844 810000797 111034949 111020753 810022291 810021862 111025128 111034937 111013101 111040044 111042175 111041993 111033690 111033675 810021646 810007365 111034063 111031476 810019607 810008391 111033700 810021950 111033486 111007737 810026554 810006777 810006757 111022902 111040163 111031481 810001589 111024831 111013537 111014690 111014701 111018561 810001479 111024865 111015198 111014672 111027918 111014706 111014656 111014634 111034906 810001303 111027890 111013579 111033778 111034060 111027619 111039852 810001257 111040154 111039996 111040012 810001343 111033903 111036610 810028617 111036475 111022707 111022693 111036496 111005215 111036595 810019600 810021968 810002048 810028512 810008418 810022025 111014621 111007554 111036647 810000965 810000435 810026899 111033707 111033565 111033688 111007707 111026571 111027598 111025539 111033908 111025599 111025631 111036962 810028169 810021430 810021979 111025597 111025489 111022721 111013759 111022430 111031466 111033894 111031117 111027579 111031256 111007093 810000744 810001525 111036544 111016700 111031249 111033840 810006778 111040063 111022922 810008161 111017077 111025469 111027875 810021865 111018581 111014302 111030957 111025622 810021988 111027559 111036645 111027884 810026830 810026858 111034923 111012706 111036718 111014647 810021483 111024007 111026689 810002323 111025124 810022159 810002000 111034902 111007214 810005814 810021789 810026683 111012695 111033768 111016942 111020727 111039960 111039860 111027702 111042306 810022284 111007020 111022913 111027775 810001474 111013775 810028252 111013887 810001551 810000638 111025133 111031142 111031488 111006323 111031176 111039874 111034856 810022316 810006648 111036388 111022400 111028485 111024871 111013619 810002178 111013596 111028693 111026610 111027606 111040062 111013705 111039995 810021776 111033705 810026860 111013710 111007052 810006718 111016527 111013538 810000634 111023995 111022431 111013076 111018562 111018580 810000955 111016909 111027574 111025107 111014694 111005237 111023322 111025656 810005809 810028534 111027914 810028237 111039848 810028328 111034863 111005224 111016905 810021397 111007541 111017085 111033766 111007053 111013722 810026875 810019567 111039988 810021987 810000451 810021431 810000169 810000736 111026724 810028570 810022287 111033566 111024444 111036729 810026742 111025743 111033560 111013908 111025678 810026795 111009006 810006614 111007953 111033794 111031194 111013731 111033684 810028551 810027063 810026743 810002104 810007375 111025508 111039959 111031166 111036550 810026884 810028186 810019576 111036723 810022028 111024539 810026715 111024869 111016923 111026246 111024711 111034051 810002521 810006643 810026851 810026920 810026862 810028245 111027774 111027830 111036314 111036385 111042290 810026769 111031467 111027649 111040040 111042201 111040022 111007531 111033798 111036533 111025573 111028681 810019597 111036691 111013539 810007353 111027678 111033731 111033652 111007726 111025777 111025502 111015490 810026817 111036733 810022278 111033685 111027556 810021872 810006745 111027792 111034885 810028548 810026724 810002050 810019939 810026781 111034046 111026474 810005780 810021467 810021956 111034033 111033774 111026441 810001034 810001095 810001027 810019370 111015849 111014294 111024443 111036657 810026703 810005817 111040187 810001354 111031137 111022403 810028546 111027846 810000956 111025779 111025679 111005250 111036612 810021435 810021450 111042150 111013558 111025548 810021456 111042340 111007065 111023326 111025537 111036760 111026704 111027947 111033889 111026573 111026275 111015458 810026905 111033897 810028192 111027865 111028477 111031406 111028471 810022308 810002812 111031167 810002103 810027068 111028685 111031264 111016523 111033781 111036619 111036570 810021923 111031158 111042177 111027652 111022908 810005776 810002523 810002075 111013897 111018563 810002003 111033654 111036482 810001234 111040160 111040041 810021926 111007705 810026874 810021914 111033581 111036574 111028491 111033795 111033792 111014259 810021964 810021447 111030964 111033583 111026599 810011562 111026609 111042309 111025503 111015466 111007526 111017078 111031221 810002993 111024446 111033546 810006644 111034067 111034072 111013893 111027793 111036627 111040025 111031534 810019391 810007354 810001099 810003241 111031395 111042314 111025498 810019379 810011565 111031494 111042152 810000767 111025561 810001527 111027735 810008449 111024862 810026770 810019380 810021797 810026753 111042253 111036557 810001988 810021962 810019359 810001215 810000914 810001282 111007690 111013688 111040088 810001082 111024450 111036967 111028519 111024532 111033494 111025684 111031175 111013075 810028340 111031119 810000586 111007556 111014632 111034061 810001587 111026265 111040158 810002173 111007686 111025674 810001102 810000834 810021788 111013073 810021785 810019593 111030960 111036621 111024412 111015176 111024406 111007019 111025605 810001480 810008393 810008420 111042020 111033516 111013733 111034034 111012690 111027853 111036572 810028552 111042342 111034934 810021454 111034935 111025761 111025659 111023318 111014661 810002313 111031382 111025648 111025528 111023339 111026621 111007578 810002373 111013551 111005218 111042252 111027906 810022280 111005337 111042009 111025624 810021474 111033879 111015210 111014257 810026866 111033780 810002009 810000108 111027951 111005290 810002868 810021993 111014268 810002534 111013694 111034868 810001208 810028520 111025471 111033585 111025688 810000964 111036581 111009013 111025594 111014638 810028184 810026827 111027838 111024827 810021795 111007528 810006785 810001041 111027948 111027882 111042274 111027538 111027677 111036700 111036599 111027569 111027894 111022422 111025670 111017069 111027664 111025780 810021781 810000912 810000894 111036548 810006618 111023993 111027815 111027819 810008415 111033816 111027673 111027842 111025538 111031420 111027825 810007358 111013947 810019365 810002238 111033824 810002518 810002215 111031514 111025699 111041988 810019596 111025500 810000609 111022432 111036381 111012713 111033477 111025747 810019368 111025557 111022915 111018768 111020755 111014657 111013595 111033643 810001104 810006647 111033638 810026740 810026834 111006318 111024529 111007073 111013568 810019559 111042194 810001650 111033512 810001653 111005256 111027657 111026618 810002102 810019618 111039863 111024006 810019357 111024819 111027946 111026700 810000124 111036587 810021931 810021453 810002477 111036341 111040002 111031384 810000937 111042337 111015439 111018554 111026617 810001991 810008470 810002005 111013905 810000285 111034073 111018578 111031389 111005329 111027945 810021984 810002435 111030961 810027085 810002827 111005225 111012731 111036547 111025608 810001520 111025610 111031379 111024822 111033788 810002989 111007692 111031138 111013608 111031213 111033687 111025682 111027742 810028171 111027711 810021761 111018764 111042217 810022306 810028234 810022293 111013780 111040166 810001073 111020758 111031257 111013091 111026720 111041986 111042203 111026266 111013712 810028196 111025661 111025756 111013592 810028550 111039859 111039957 111036512 111039857 810021912 111005213 111015459 111031377 111031519 810026804 810003238 111025075 810002436 111030965 111015847 111042004 810022282 111031507 810028547 111036936 810008166 111022722 111022906 111031200 111036511 111025635 111031132 111026245 111036350 111036387 111007055 810028523 111036665 111036730 810022156 111007075 810022161 111031522 111033663 810008862 111036591 810001315 111033906 111033803 111014254 810002066 111025545 111025612 111031510 111027834 111025515 111013690 111040060 111025592 810001206 111036704 111036634 111025616 111031391 111042291 111031250 111027795 810002811 111027593 111040933 111040038 111028695 810019403 810019606 111028683 810022019 810021961 111036351 111023055 111036364 111028694 111036373 111028678 810019352 810019931 111022918 111028703 111023056 111024420 111022702 810000861 111007944 810000960 111014645 810006638 810001263 111015599 810002568 810002067 111039851 111025555 111025676 111025123 111027851 810007370 810002384 810000889 111024432 111030976 111025093 810000602 111013736 810002274 810001472 810000768 810000612 810022152 810021426 111013546 810000862 111031429 810001126 111013588 810028336 111027934 111031422 111033479 111013106 810019635 111022692 111015831 810001277 810001207 111022917 111013585 111026431 111042326 111042287 111027544 111027545 111007738 810001134 111036597 111027807 810006619 111027797 810002096 111013610 111042266 111025640 111026247 810022024 111024413 111026429 111033588 810021439 111031506 111036713 111020737 111027765 111013714 111016937 810022011 810006636 111013757 111024001 810022001 111034865 810027077 111034056 810028175 111024839 810002530 810002046 111027674 810000574 810001127 810026775 111039991 810027067 810021418 810003244 111026257 111014685 111042311 111007938 111012711 111013698 810019339 111033852 111033704 810028189 111040178 111033521 111015846 810000468 810002737 111007060 810000685 111031183 111027676 111025127 111014675 111014281 111013784 111031485 111033909 111034070 810002329 111014670 810002433 111014653 111014625 111033493 810002235 111014275 810021799 111034066 810006655 810021429 810022307 111031222 111027622 111027616 111034052 111031155 111040070 111025620 810026725 111033596 111033553 111007091 810000887 111031390 111013603 111036938 111013779 111016534 111007727 111007099 111040011 111027601 111026478 111025655 111033644 810002739 111042160 111025596 111033600 111036924 111027625 111014629 111034931 810021932 111033475 810026694 111014290 111027942 111018571 111036635 111013911 810000560 111013136 810006609 111017088 111027770 111013798 111027802 810021863 111022911 111013606 111015185 111025604 810002163 111028702 810006603 111036525 810021948 111007674 111016946 810002211 111014637 111034074 111007046 810008452 111025121 111042168 111016535 810002630 111036762 810019586 111025695 111031255 111027671 111028701 111042279 111005332 810001226 111025677 111025662 810001281 111036945 111027867 111033701 111028488 810028224 810019341 810008441 111031412 810006779 810019599 111042185 111033807 111042213 810026690 810019616 111042258 810001132 111027818 111023051 111031401 810001945 810001932 111031539 111031511 111031258 810002998 810000802 111033800 111007730 111033502 810000800 810008422 111033573 810026697 111033722 111013948 810026723 111033855 111024017 111036456 111023330 111024452 111036340 810021930 810003247 810019571 810026727 810001159 810002312 111025745 111024424 111036625 111007031 810001231 111042208 111028675 111036559 810019590 111015447 810028537 111034886 111034925 111015477 810027064 111036361 810000936 810028257 810005784 111039980 810000580 810000939 810026924 111027920 111025744 810005786 810026739 111026264 111015168 111007962 810026700 111036923 810002107 111033853 111033470 810022015 810006661 111016926 111025736 111015173 810011558 810028188 810028544 111040934 111036348 111026711 111015159 111014246 111024407 111013566 111025106 810022029 810021916 111025772 810001947 810021446 111027891 810002054 111015180 810003243 111009023 810002636 111023049 810028535 810001526 111039835 810001990 111036584 111018579 810019386 111033561 810021741 111018549 810000643 810021742 810013129 111013114 111015594 111014707 111023319 111014684 111014620 810000522 111005245 111033555 810000624 810000633 810000754 111027840 111024858 111007948 111030962 111036390 111039844 111025094 810021940 111024543 810019363 111015592 810026716 810006608 810006600 810028182 111007725 810022299 111013119 111042316 111042220 111007702 810008408 810028567 810019345 810019602 111031549 810021994 111036560 111031120 111025588 111031173 111033601 111013763 111033738 810001031 810001014 810026888 810002072 111036527 810001155 111031177 810002874 111025546 111026692 111026248 111040032 810026819 111022434 111042312 111042289 111027688 111024003 111033739 810006787 810013120 810002008 111013704 111036492 810021428 111013097 111005287 111042007 111033805 111039836 810021457 111007952 111008105 111034047 111026686 111036685 111033678 810026752 810005819 111028499 111034041 810002258 111007575 111036346 111033815 111007683 111007044 111020750 111026587 810000630 810007377 111012704 810001045 810003283 810000544 111013134 810001240 111031509 111013776 111026572 111013072 810026873 111031533 111022438 111013092 111033753 111036576 810019945 810001243 111040021 111026722 111017089 810022005 111028492 111015237 111033551 810019932 111036601 111036631 810001473 111025687 111025526 111014664 111013706 111033882 810001531 111033881 810001549 810022303 111042296 810001292 810000465 111026481 810028571 111024005 111031197 111036931 111016714 810021766 111007550 111033683 111016530 810001995 810006719 810001933 111042165 111042024 810001994 111015586 810002015 111014247 810026792 111013602 810013135 111013802 111042333 111042180 111025669 810002383 111025633 810019344 810005822 111024439 810013117 810013116 111027580 111036604 111036667 111025475 111036906 810006656 111024440 111036395 810019637 810026728 810028559 111027787 810026729 111026430 111036504 111007508 810008447 810005792 810007359 111026591 810019573 111028500 810008168 111031515 111031399 111013557 810000892 111024715 810001319 111015460 111040007 810021981 111015224 810026896 111023999 111034062 111042248 111033680 810021767 111027690 810002110 111031440 111027743 810022324 111034069 111042156 111016518 810026820 111025516 111025097 111027689 111027902 810002876 111022709 111036522 111025477 810000741 111027571 111033574 810002320 810008445 111013562 810002633 810002109 111005214 810028503 111013054 111033523 810019348 810019338 810008395 111013098 810001258 111024425 810027081 111039830 111023327 111013611 111028504 810026773 111025637 810001275 111036682 111040084 111031161 111036569 111025113 111033811 810008455 111033828 810008435 111042172 810022319 810002010 111036677 810000545 111031498 810002169 111015470 111013894 810002176 810002223 111016688 810026745 111024847 111031215 810022146 810028259 111033741 810028323 111033859 810022151 810028208 111031253 111033639 111027751 810028243 111033884 111027694 111033822 810026737 111040085 810028508 810006610 111033469 111033755 111036319 111026456 111026274 810019634 810028324 810000737 111026235 111024004 810013134 810002484 111033866 810002061 111031492 111039886 111033483 111027954 111031198 111022688 111007215 810021976 810002055 810026847 810000618 810008454 111012696 111034899 810008173 111036953 111005271 810000913 810008457 111028506 111025092 111007559 111022907 111014659 111013058 111024535 810001546 111033595 810028202 111033507 810026776 111024423 111024856 111033802 810000636 810007371 810001521 111027667 111033750 810000832 111033579 111020735 111033759 111025769 111014293 810002161 111018782 111033790 810026815 810021409 111039866 111039973 111033913 810008392 111016698 810006793 111027564 111036711 111014286 111027855 111031453 810021396 111027607 810003242 111007029 111007940 111007087 810006652 810019928 111031502 111039870 810027088 111014683 111014712 111034889 111033496 111028704 111033837 111039865 111025586 111022413 111018572 111025603 810006775 810001999 111023990 111031210 810001342 111022704 111039883 111024817 111031550 810000951 810028228 810002233 810001923 810001081 111020762 111040013 111026455 111036474 810001524 810008431 111007563 111026606 810001018 111027668 111027699 111042320 111036662 111036622 810026691 111039964 810026921 810006631 111026602 111015192 111039849 111014654 111007696 111036702 111016911 111007967 111031426 111027656 111033813 111027701 111034900 111024453 111033712 111040057 111033746 111013745 111042016 111013785 111033716 111033886 111033635 810021967 111027816 810019767 111005217 111025566 111027721 111033758 111027857 111031126 810002999 111034876 111025076 111005334 810002063 810002060 810000684 810011557 111013909 810028532 111034032 111025519 111033582 810028174 111033870 111031225 111033602 111036396 810000599 111013599 810021779 810021764 111005249 810000276 810022030 111036541 111040006 111039888 810028221 111039875 810021911 111042310 111036479 111040008 111039987 810028527 111013589 111025553 111015226 810021745 111009010 111017071 810001523 111027912 111014291 111036767 111023054 810001074 111015473 810021868 111028501 810021747 810019397 111024527 810000437 810021945 810022163 810026915 111033584 810008437 111023059 111025763 810001554 111025698 810022002 111036382 111040067 111039965 111033804 111026581 810008169 111024430 111024843 111005255 810019938 111016536 111036768 111025741 111036486 111013747 111007040 111007076 111022912 111025660 111024442 111024019 111024426 810022281 111007691 810028176 810002064 111039879 111026570 111031152 111027585 111027724 810000603 111007704 111013127 111005280 111007720 111023315 111036644 111025563 810000635 111025671 111036577 810001471 111033806 810026693 111039962 111039986 111031464 810026711 111016529 111007739 810002098 111007694 810019622 810002074 111036721 111027822 111027777 111042294 111007524 111036407 111009022 111031443 111025768 810001161 111031218 111033616 111013094 111024414 810021866 810000632 111027591 810007374 111012708 810019944 111007662 810019929 810001026 810006623 810002976 111031504 111027617 111007062 111018778 111005230 810022149 111022689 111036673 810026894 111007032 111025552 111023325 111033468 111036322 111013061 111042155 111025760 111025096 810021974 810002983 111025773 111027820 111042286 111042014 111039839 111013891 810001039 111041987 111022700 810001353 111023336 111022905 111033558 111007505 111027716 810002100 810021434 111034068 810022311 111007529 810021939 810008430 810008165 111007723 810008394 111013548 810008417 810008432 111015229 111033490 111036332 111036579 111014301 810021957 810001351 111012709 111020723 111040161 111040078 111040074 111040072 111040180 810022285 810022292 810001948 111033503 111013907 111013886 111007709 111015456 111014300 810022297 810001651 810006639 111022420 111042270 111005243 111027627 111028505 810001212 111022715 810026759 111033504 111023061 810008387 111013703 111025770 810026552 810006668 810021468 111040164 810028181 810001048 111033634 810021753 111027844 810026880 810019570 111007562 111031474 111007700 111036915 111023991 111015464 810002781 111042197 111033508 111014628 111040087 111040068 111036766 111033723 111030973 111036966 810026783 111024851 111027602 111026428 111005284 111014663 111007957 111036462 810001482 111033773 111027737 111031493 111025494 111033547 111034944 111025776 111036468 111012702 111013950 111033765 810021943 111025082 111033589 111036502 111026250 111026270 111007666 111036523 810026914 111015208 111031462 111033791 111033856 111025090 111016941 810026816 111034927 810001345 810001218 810001552 111025484 111007081 111015588 111007539 111007706 111015475 111012712 111025496 111013597 810026898 111005228 111025653 111033762 111015179 810028250 810000466 111013769 111031135 810019934 111013578 111013729 111013762 111018754 111034891 111026460 810002105 810000597 810002004 111033743 111027897 810019575 111015175 111007540 111015221 111027708 111042006 111014671 111024859 810000588 111033556 111015189 111027904 111031541 111033783 111033569 111025087 111006317 111006326 810026837 111007703 111020749 810000241 810021442 111007685 810008407 111014272 111013591 810021420 111036620 810021403 810021792 810001294 810001300 810028269 111005292 111026594 111005342 111016524 111025564 111033843 111031160 810022331 111020733 111020756 111020746 111013135 111027562 111033849 810000952 111031265 810002368 810000064 111020764 111005257 111022404 810026720 810019385 111040082 810026734 810021491 111033771 111030970 810021482 810028621 111022714 111014714 111015479 111022706 111033518 111025485 111033571 111007026 810001244 111025474 111014623 810002631 111018761 810006664 810006659 810001350 810027080 810000921 111033810 111033514 810006666 810026748 111042178 111031139 111027662 111036539 111014289 111023352 111028503 111033844 111013700 111031393 810028519 810019614 810000969 111031392 810002428 111031156 111033659 810000575 111033482 111031214 111022910 810000699 111027880 810013127 810002991 810006772 111007522 111027799 111033851 111027745 810008400 111025551 810008406 111025782 810027066 810028178 111033481 111033530 111036758 810019360 111036360 111033646 111012697 810000520 810003246 111031532 810001318 810026881 111015463 111007555 111005273 111027611 810008469 111033787 810000483 111025642 810028555 111034871 111040042 111034936 111033726 111042305 111042277 111034922 111025623 810000910 810006670 810008172 810002068 111006322 810001590 111034896 111015482 111025634 810001290 111013741 111007512 810000967 111025647 111033789 111025575 111014283 111015590 111006321 111014287 111040185 111014631 111014267 111036684 111013070 810002517 111013787 111016931 111027886 111012736 111034036 111033711 111033858 111040075 111025081 810021805 111036765 111031436 111036563 810021743 111033904 111007514 111036946 111013103 810002177 111040183 111036694 810007357 111015160 111007070 111031529 111025665 810013132 111005205 111022424 111015151 111027868 111031394 111028689 810001217 111033535 111031199 810001346 111026697 111027670 111024438 111031124 111007732 810001323 111024437 810002529 111036769 810019070 111026585 111007960 111033511 810002528 810002565 810002562 111036377 111028699 111039845 111034045 810008413 810000480 111014641 111013898 111014674 111015169 111007565 810028517 111013695 810008427 810008419 810008167 810001038 810021427 111026616 810026825 111022909 810028330 111016690 810002779 810021639 810001157 111033734 111042322 111020748 810000922 111042219 111040052 111024433 810021918 111024422 810026800 111040170 111040035 810021949 810022323 111036376 111020751 111007097 111020718 111040936 111014248 111007564 111016516 810013121 810006613 111034048 810002147 111016913 111042181 111005283 111005340 111031212 111005244 111005236 111005209 111005336 111016519 111022436 810000968 111013107 810021917 810026864 111014255 111022708 810021985 111031545 111023314 111030979 810026709 810002430 111016526 111033912 111031148 111015476 810000595 111033696 810011554 111036603 111033725 111036928 111022427 810006721 810006723 810027057 810022335 111036703 111027592 111027612 111027583 111036540 111027613 111036944 111036920 111025581 111028696 810006761 111013901 810027087 111036367 111020731 111013686 111025737 111026444 111020757 810022314 111033509 111042171 810001164 810026890 111005252 111025522 111036613 111034869 111042304 111027661 111033682 810000837 111024876 111034946 810001280 111027888 810007352 111033796 111031187 810021402 111033745 810028344 111025582 111026483 111025578 111015178 111015236 111025767 111022927 810006671 111022924 111020766 111015839 810006725 810019566 810019629 111022725 111025505 111022697 111022727 111034873 111034897 111034867 111034880 810002062 810019591 810006783 111031439 111027757 111033476 111031552 111007663 111025595 111033878 111036324 111033799 111036756 111033662 111036705 111034059 111013581 810000629 810000576 111031520 810008446 810001098 111007946 111031127 810002442 810002444 111014715 111005281 111005210 111015155 810021999 111033488 810026923 111036594 810001256 111042023 111039853 111007569 111013728 111027773 111025615 111028473 810013126 111026242 111039861 810000953 111024434 111018558 111036449 111028698 111015448 810006672 810002374 111042267 810008933 810026806 810008936 810008932 810008799 111015484 810026685 111025570 810008861 810008940 111022407 111018567 810005813 111036365 111015219 111024872 111015157 111014705 111031181 111015233 810006773 111036515 111039837 111039856 810022014 111026232 111027726 111007039 111007561 111027862 111015441 111024002 111034053 810019631 111031451 111034945 111033609 810000184 111027859 111027871 810021758 111027543 810026910 111022726 111013742 111025506 111017087 111034855 111034887 111034947 810000622 111024427 111031403 111012724 111030956 111013071 111012699 810008468 111013734 111017074 111036498 111033864 111027915 111013726 111036752 111027938 111033885 810019072 111015478 111036654 810000742 810021958 111009012 810022312 111014295 111030958 111039981 111024410 111026482 111026458 810028581 111026452 111024541 111042195 111027905 111007538 111013086 111036952 810019398 810000642 810001467 111030967 111033730 111042013 810022006 810026760 810026710 111033529 111034883 111036476 111027836 111015205 111025119 111016943 111015218 810002099 111033575 111022397 111036542 810019594 111012707 810001992 111027887 111031184 810001054 111014261 810002974 111015200 810002873 111024531 810006763 810019628 810000838 111023311 111036457 111027725 111013584 810001080 111012703 111014252 111027744 810006660 111020719 111027827 810026852 111036406 111039966 111025579 111025111 111005279 810000891 111034903 111024850 111025572 810028179 810021422 111036747 111023344 111033633 111024451 111027798 810027060 810019942 111025495 111014677 111036623 111036735 111015223 810000587 111014676 111013793 111026267 111005223 111007513 810001321 111007083 111007085 111025738 111034040 810021759 111036689 810021463 810001344 111006319 111027553 810001103 111036485 111033857 810021643 111036551 111025547 810000110 810002563 810028168 810002106 111036573 111027736 810001264 111027750 111042344 810026551 810021787 810028513 111033833 810028580 810028569 810028586 810021399 111042295 111015181 111007579 111031387 111031397 810001043 111033710 810022147 111036653 810001230 111027927 111033900 111031497 111036586 810021757 111024880 810019627 810002220 810007372 111042255 111013612 810002979 111036380 810002969 111026234 111018769 810007362 810019595 810000938 111036564 111025105 810021486 810007364 810008458 111042263 111031414 111025109 111005254 810022003 111040043 810008874 111018568 810000620 810022330 111014273 111013105 111005288 111009009 111013082 111007056 111036374 810028236 111042285 111025074 810021980 111013552 111040061 810008872 810001553 111033891 111033543 810005818 810022301 810002328 111026249 111026241 111033478 111030977 111014253 111033495 810028231 111014262 810000924 810005812 810002366 111013751 111014649 111042338 810021738 810002629 810001304 111022414 810021913 810026900 810028554 810000738 111033568 111040054 810002232 111036695 111036472 810019572 111027879 810028200 111016927 111042190 111040079 111016934 111024857 810005787 111027547 111015855 810000281 111036452 111033648 810021444 810022007 810002440 810028536 111026703 111033825 810026885 111024018 111036520 111040064 111005274 111033728 111042214 111015211 111026272 111014270 111024870 810001010 111007059 111039878 810026821 111026233 810002172 111025649 111027697 111027930 111015602 111026584 111014717 810006650 111039854 111033669 810022020 810007363 111027648 111042158 810000181 111014673 111033883 111024009 111007530 111033831 111031191 810027089 111017072 810001993 111013125 111007069 111026448 111022428 111042010 111036592 111039869 111022410 111005333 810001347 810028338 111042211 111027929 111027833 111036326 810000182 810001469 810013123 111041995 111026260 111026723 111031248 111014686 111034042 111034893 111036399 111013796 111026717 111033539 810000766 111031496 111024010 111025569 111025103 810028222 111033714 111034866 111036411 111033777 810028239 111036531 810026733 810005774 810028337 111024449 111031559 111036672 810026713 111015498 810002984 810019358 810022010 810028173 111014274 810008175 111013691 111005204 111042199 111031455 111036963 111031528 111031489 810002970 810026682 810019632 111036679 111026620 111025483 111033709 111040066 111042162 810026705 111036336 810021784 111039864 810028264 111033676 810021424 810028260 111026243 810028521 111027576 810026744 111013586 111005220 111025490 111025488 111027917 810026766 111031193 810002097 111005331 810001052 111018771 111015449 111018775 810021995 111036744 810003286 111034919 111007532 810026741 111027769 111033567 111027594 810022155 810000688 111015243 810002325 111039968 810026877 810003001 810006612 111006320 111025514 810028211 810021469 111026457 111036912 111036943 111025774 111026450 810026761 111036633 810028201 111027695 111036666 111013766 111042184 810028322 810028582 111015195 810026678 111015589 111015468 111013713 111013774 810026765 111033517 111042192 810028215 810021470 810002671 111025775 111025585 111013104 810001927 111031188 111005291 111031428 111027781 111023994 111033868 111016521 810001205 810003237 111005232 111012735 111025567 111005278 111024882 111036918 111027841 111031254 111040020 111033836 111027764 111014636 111027540 111036506 111015451 111015587 111012725 810001586 111013133 111027703 111015504 111027551 111013051 111027550 111006330 111036484 810026854 111022405 810019633 810026891 810022276 810028177 111033770 111013055 111016692 810021412 111036386 111039850 111040005 111026601 111040015 111013789 111009014 810003274 111025636 111016711 810000692 810021489 111036771 111036774 111031201 810021966 111013716 810021641 111036614 111033655 111036477 111024834 111025101 111027896 810027070 111042283 810028223 111034877 111027734 810019933 810001936 111015144 111027923 111033505 810028331 111016691 810008440 810008439 810028533 111042170 111007729 111007212 111033749 111027772 111036748 111031179 810026887 810019564 111025681 810021477 111015145 111042254 111036720 810028557 111028493 810022008 111027806 111022429 111013541 111036717 810003000 111013786 111036754 111027555 111036543 111033797 111031527 111025531 111016525 810026856 810002990 111015585 810001532 810002830 111007546 810026681 810028207 111033895 810019598 111033733 810002273 111025733 111031226 810021452 111033594 111042301 111007504 810021809 111027943 810000640 111027804 810019584 810001316 111018559 111025606 111027710 810001530 111031141 810002180 111009007 111025692 111007027 111005343 111036951 111027877 810005790 111031547 111013123 810022021 111014697 810000698 111039977 111015483 810013133 810001033 111007681 810002053 111013794 111036626 810000799 111024867 111039961 810000836 111020743 111033501 810019619 111033541 111014660 111025535 810027054 810028267 810027084 810021942 810022333 111026269 810007347 810026784 111026259 111013754 111017082 111042011 111036470 111031405 111027595 810026708 810022313 111033656 810028507 111033615 111033808 111028486 111031129 810001024 810002668 111016703 810026805 111025549 111018780 111040184 111007545 111015842 810002531 810019943 810001466 111009020 111022923 111013906 111018772 111005285 111036518 111014635 111033899 810026701 111025534 111036370 810026823 111033472 111024853 810026746 810001017 11 1033649 810021472 111040014 111036948 111028516 111034038 111027847 810000639 810019623 810019625 111005234 111014260 810019936 810021415 810019387 111031546 111031128 810008409 810002872 810021481 810026751 810028198 111027597 810026719 810002370 111022419 111031516 810021996 111033628 111039872 111036956 111022730 111007535 810002632 810002321 111016709 111031558 810002482 810002445 810028248 810021800 810019374 111013090 111025544 810000252 111007205 111013949 111017070 111040176 111028470 111033533 810001317 111033540 111027584 111036398 810019346 111007547 111036965 111027624 111036608 111033697 810002315 111036750 111027549 810006724 111024866 111023050 810006722 810006717 810019400 810019610 111031449 111007021 111024409 111027843 111027866 111026708 810006762 111027907 111025629 111040049 810019347 810021791 111013068 111027731 810022027 111013057 111040055 111009024 111031190 111015485 810001130 111013739 111013752 111027693 111027805 111036773 111026236 111023048 111034940 810001949 111014271 111027614 111036501 810002330 810001476 810019604 810028233 111034064 810002535 111016915 111017084 111027682 111025654 810001475 111013121 810000803 111031247 111025694 111039982 810022012 810006604 111015217 111033549 111031159 810002108 810000631 111007515 111016933 111027953 111024823 111013801 111007517 810001591 111026239 810028575 111036384 810005782 111013717 810003240 111031548 810002256 810013122 810008389 810021460 111015597 810006790 810002175 111040173 111039992 111033910 810028343 111024868 810000627 810026799 111028472 111036575 111007509 111033664 111033510 111033592 111013549 111022903 810000691 111024710 111026475 111007669 111012720 810021867 111017086 810002375 810002259 810002522 111025086 810007379 111033793 111014640 810005795 111030963 111036375 111039976 111033647 111036650 810001895 810026692 111025472 111033702 111025613 111036632 111027782 810000481 111030975 111031162 111025122 810002257 810013130 111015467 810013125 810021421 810019611 111007934 111013708 111034860 111014642 810019383 111016704 810026829 111007932 111013554 111034035 810006622 111007701 111040019 111031246 810027079 111013899 810001232 810002013 810022283 810000581 111027789 810002314 810028226 111034924 111007676 111027609 111042017 111027889 810026794 111026231 111040045 810006716 111013128 810028501 810001929 111039880 111027610 111026600 111023350 810000253 111023342 810002372 810022294 111027626 111036655 111033715 810000577 111013797 810006791 111040169 810021946 810000610 810021991 111027748 111024525 111016699 810028563 111027780 810005796 111015167 111039885 111023343 111039882 111025771 111025643 810000450 810002275 111025476 111036343 111033841 111031209 810002828 810027086 111027950 111025118 111027926 810021492 111022399 810011553 810006658 111036669 111033896 810002977 111022916 111014298 810000621 810002971 111031154 111013755 111027952 111015438 111025550 810022315 810019589 111036337 111027786 111034870 111026619 810000829 111014651 111012727 111031491 810022277 111015152 111033473 810003276 810019563 111036701 111014708 810000835 111015593 810002237 810013131 111036508 111036371 111025686 111025781 810002222 111018765 111023320 810000438 111027714 111031219 111027663 111036408 810027056 810026755 111027936 810021935 111014678 810021921 810028212 810006792 810022009 111015231 111033661 111015595 111015605 111005335 111036405 810006755 810006788 111031460 111033557 111036342 810002809 111031131 111026262 111015235 111025639 111016939 810001210 111014280 111027679 111016944 111014284 111031400 111027911 111027771 111030980 111015830 111025529 111027849 810021393 810000687 810002869 111015220 111033672 111026707 810006649 111007722 810008442 111025511 810021922 111028511 111039833 810019351 111025099 810026679 111031151 810027049 111013111 111036321 111013601 111024435 810000277 111025518 111033538 810001593 810026550 111027776 111033826 111027870 111013095 111020721 111034879 810000831 111031251 810021807 111030981 111033706 111028494 111005246 111020747 810021801 111006331 111028518 810026754 111031470 111033871 111033519 111042000 810001049 111039871 810021919 810019609 111033674 111036491 111033772 810027071 111039999 111025126 111040010 111024825 810022288 810000893 111007034 111007071 111028680 111028687 111025693 111015469 111020760 111014652 810021408 111039975 111036932 810021774 111015161 111036605 111036649 111007068 111031136 810001989 810021803 810000452 111031483 111016902 111033593 111033751 810005791 111022713 111007548 111022711 111039862 111025626 810001035 810026762 111005251 111016910 111016517 111027554 111020725 810006632 111042278 111036529 111005276 111039963 111015174 111025580 810026876 810021762 111031473 111015450 810000886 111013074 111036372 111040056 111022904 810002272 111007949 810028193 810022289 111033527 111024846 111013895 111022718 111040059 111031517 111013903 810001050 111042268 810026840 810026878 111007095 111007033 810008404 111007551 810011563 111034932 810026555 111031261 111033724 111015216 111013540 810026842 810026918 810028509 111042336 111025112 111027898 810001652 810002377 810008425 810026812 810026916 111007558 810003275 111020734 810002164 111022728 111005275 111025746 810001931 111016531 810001259 111014669 111014263 111033587 111026253 111031445 111015837 111026470 111031499 810019382 810021812 111007570 810000583 111042154 111024416 111031432 810001348 810007360 111025479 111025565 111014689 810028525 810006615 111034049 111020724 111027600 111015446 810001468 111014710 111033846 810002627 111036571 111007038 111025701 111012729 810000757 111013079 111036493 111034908 111042334 810008428 111033880 111005233 111042275 810001019 810021973 111026229 111023998 111027596 111025473 111014716 810026707 111007064 111034858 810001481 111033624 111031381 111017080 111031430 111040023 111016929 810026684 111042331 111015462 111005282 111024008 111033692 111013912 111034911 810002634 111025072 810006665 111024542 810022157 810021768 111025114 810021871 111036538 111036331 810021808 810026801 111036526 111025129 111023351 111042191 111024536 111016533 111022928 810001185 111039877 810019384 111042012 810026553 111025609 810021937 111031444 111036490 111031438 111036338 111036930 111013117 810002002 810002429 111027931 111027572 111014624 111036937 111023332 810028530 810019617 810000911 810021790 111020722 111041980 111033622 810028327 111033693 810002814 111013725 111013951 111027809 111033667 810002217 810000740 111027824 810008453 111027932 810019580 111027845 111036458 111036961 111009005 111042202 111026228 111036494 111042218 111007050 810007378 111020741 810006654 810007345 810001188 810001029 810022023 111016707 111013129 111026622 111027618 810002667 111013560 111015162 810026556 111013582 810000689 111040046 810001654 810002432 810022304 111024455 810000950 111036772 111005240 111014693 810011572 111013053 111024408 111033699 810007349 810021783 111007537 111013124 111042189 111031164 810026699 810002875 111036710 111023321 810002239 810026687 111005231 111040083 111023323 111024828 111007079 111031116 111027783 111015584 111027837 111040077 111026592 111027829 111033677 111027700 111016914 111015502 810019364 810019069 810006605 111042206 810028573 810008935 111026688 111034872 111040188 810022158 111041991 111015188 111036609 810026793 111031526 111023340 810026756 810019579 111031192 111023310 810006621 810021458 810021405 111042026 111007668 111034907 810021875 810001260 810007380 111031537 810019336 810026780 111036759 111033914 111007041 810002867 111033832 810002371 810028216 111033563 111033559 111033823 111036394 810021416 111031150 810008444 111036699 810028187 111042205 810019071 111036454 810021780 111033590 111031123 111031260 810019349 111027903 111024534 810002316 810001545 111026714 810002056 111026597 111015601 810001261 111023063 111031172 111033736 111036615 111036600 111033619 111040027 810022279 111036499 111036753 810028502 111025663 111027878 111031508 810027058 111025748 111033637 111034915 111024538 111033552 111036330 810000521 810002479 111036393 111025668 111007063 111026227 111024540 810011560 111027941 810002438 810002527 111024852 810006630 810000765 111020767 111025680 111027565 111015583 810028246 111027909 111027570 810021874 111042183 810002318 111025577 111016935 810028518 810021432 111024419 810002379 111015190 810008396 810022317 111018753 810022322 111015201 111024879 810002870 111031459 111028514 111036683 111033817 111027563 111027589 111027573 111027539 810000125 111013902 111024431 111042327 810002519 111033465 111036727 810026917 810021752 810021794 810028172 111025589 111022418 810000048 111018781 111013138 111031536 111013116 111036353 810019769 111025778 111031530 111042261 111042025 111027740 111007954 111016908 111033812 111016928 111033848 111039990 111007710 810028538 810021459 111024832 111026569 111031146 111027869 111034938 111033698 111025073 111031423 111033720 111031495 111036500 810002776 111027940 111015148 111007695 111033767 111024530 111031454 111025085 810006634 111028690 111034941 111007568 111031149 111027839 111007543 111025088 810028191 111036471 810026832 111026435 111025751 810000614 111023355 111025641 111013888 111031521 810002670 810008429 111033820 810019362 111025499 111016932 810000467 111034888 810022150 111015172 810000885 111025104 111036316 111033550 111013131 810001528 810000449 111018770 810002051 810000246 810002967 111013692 111013078 111036400 111006329 111023052 111005239 111005270 810001580 111007066 111013614 111033756 810008451 111006328 810021771 111022720 111013744 810026872 111025697 111034918 111009028 810027055 810022286 810022162 111033572 810026835 111033732 111039828 111027803 111036359 810000623 810002322 111007054 810002996 111013792 111027885 810027065 111015471 810021869 810022165 111033607 111025765 111026271 111036674 810001187 111005202 810021471 111025509 111028490 111024835 810002986 111024545 111033520 111013137 111005221 111036335 111033630 111013758 810026730 111031433 111024441 111036950 111025753 111036959 111042317 111040053 111042176 810021796 111036363 810019588 111024015 111036329 111040051 111034917 111016916 111016920 810021927 111022433 810008412 111023316 111013771 810021746 111033570 810000121 111023333 111028478 111042313 810027069 111040039 111027850 111033562 111036451 810026696 111023345 810028514 111024821 111036553 111041989 111015242 810005777 111040181 111020738 111007679 111027546 810021476 111024421 111031147 111034039 111014622 810002165 810002224 810008460 111015461 810001522 810011571 810000453 111025766 810019355 111033679 810002829 111031195 111030985 111015487 111007693 810021438 111014668 111014626 111039978 810026922 111040155 111014662 111013542 810006669 111033782 111036489 810028268 111031388 810019582 111013687 111026712 111033506 810002260 111028487 111039993 111040080 111027599 111007094 111033827 111036611 111023337 810021977 111033821 111018577 111039989 810001987 810002007 111036556 111033830 111034901 111033876 111034037 810021406 111033829 810021484 111036751 810007351 111036638 810007342 810021394 810021425 111040036 111034914 111026690 810002148 111014627 810002871 111033911 810000739 111031415 810026810 111014250 111014297 111033537 111027709 111036354 810022309 810026763 810019592 111025543 111031185 111041978 111041990 111042282 111039846 810026718 111036568 111036641 111034864 111025618 111014619 810006760 111022712 810002369 810021407 111036565 111027686 111036554 111034943 810022332 810028342 810022300 111033742 111024016 810026867 111042284 111031535 810028266 111036639 111025574 111036907 111036707 810026688 810026726 111031163 111040017 111033735 111042164 111039969 111042198 111040182 111036628 810019375 111028508 810026704 810002810 111025645 810011559 810028583 111031189 810026774 810026808 111033869 111027723 111016705 810007346 111007048 810008421 111016701 810028227 810028240 111026480 111013890 111033497 111031486 111027603 111015186 810005794 810021638 111025617 111026442 810026767 810027072 111036473 111042251 111040935 810027078 111042307 810026855 810019377 111042256 111020716 810006637 111024526 810028505 111034916 810001584 810026853 111016528 111027659 111007527 810026912 810028504 111033818 111015238 810002485 810008462 111015851 111013080 111039970 810001227 810021960 111027801 111013770 111025583 810001941 111013711 111013724 111013544 111015598 111005211 111015170 810000798 111022705 111015838 111016912 111007567 810000600 111024864 111036404 810019392 111031145 810021473 111031555 111033487 111027582 111007072 111034861 111014687 111026693 111013696 810006617 111027916 810026907 810026796 111039998 111015202 810002816 111031418 111013553 810021983 111024861 111031551 111033901 810000109 111013066 111033874 810026686 810013124 111033845 111034057 111007577 111024873 111007697 111024824 111033842 810006667 810006627 111013615 111036922 111022415 111022408 111014692 810002380 111036347 810013065 111002007 111002003 111036448 810002326 111007096 810006646 111015445 810021969 111025625 111027715 810000562 111018766 810007367 810000801 111031203 810021744 111007507 111031386 111025077 111027746 111013052 111033485 810019937 810026908 810001997 111040179 111016687 111039983 111024820 810002992 111016907 111024845 111007553 111027811 810019601 111033872 111033779 810022026 111025482 111025083 810000598 111036510 111015196 810021798 111016696 810021920 810019381 810006597 111042008 810008410 111031531 111027728 111018763 111027707 810000274 111026706 810026712 111031477 111024418 111013778 111033515 111033744 111034031 111028484 810001016 810021411 111007678 111007665 111031523 810000122 810019561 111007523 111013721 810021756 111007078 111007950 810026889 111014265 111020726 111025079 111027588 810026883 111033839 111015836 111020732 111007714 111013062 111013059 810005785 111034044 810006782 111027876 111020745 111024708 111015191 111024875 111034942 810021810 810019585 810022154 810001186 810000519 111013707 810007348 810026758 111036358 810027082 810006728 111003497 111002444 111002418 111011434 111003320 111011194 111002462 111011967 111003530 111003565 111011918 111002448 111003544 111021454 810013061 111010979 111001878 810007443 810019626 810021982 111002013 111003138 111011209 111003050 111012476 111003552 111003496 111002454 111002722 111002708 111011995 111002395 111002459 111002458 111002426 111003517 810019624 810019630 111025675 111025607 111002696 111001883 111003333 111003531 111003551 111003659 111003493 810022112 810006594 810000756 111002412 810019526 111002429 111003189 111002710 111002422 111002717 111002699 111002432 111010461 111004179 111003563 111036323 810028329 111026230 810002319 810008873 810008800 810026680 810000183 111042179 111040065 111014681 111040004 111031557 810026802 810001352 111026453 111031380 111031427 810000170 111031469 810027053 810027059 111042003 111026237 810021464 810002049 111036908 111040058 111034920 810007366 111033614 111036724 111033617 111028497 810026849 810026764 111024881 111033598 111033491 111005247 111006316 810002566 810001214 810000507 810001293 810001228 111025667 111025666 111027944 111034862 810000637 111033531 111026595 111025621 111036716 111013567 810007368 810028497 810028528 111027739 111013102 810028619 810008174 810008434 111007088 810001083 810001278 111040190 111007557 810001548 111013783 111022929 810001042 810001189 810021441 111036642 111015848 111025095 111040174 111041992 111031544 810022305 810000166 111031525 810001070 111013583 111041985 111007519 111013093 810026677 810000123 111027893 111027895 111031157 111033627 111007717 810001478 111027675 111033801 111013108 111014682 111007518 111024848 810001926 111015158 111031435 111025116 810002231 111025098 810028335 111007718 111033887 810026911 111013122 111034028 810000601 810021778 111033854 111020754 111014299 111007675 111015184 111033757 810007341 111014650 111007580 810002740 111007667 111007966 111013765 111036397 111015840 111007677 111007963 111014639 111040001 810021951 111033608 810002382 111015829 111007023 111007680 111013620 111039974 111011241 810001470 111033769 111025512 111007670 111039873 111027722 111014251 810007361 810011552 810011555 810006748 111042300 111036934 810002212 111042328 111042200 810001012 111036583 111042005 111036562 111042262 111042157 111014278 111042182 111042341 111040050 111031409 810001015 111014282 111036692 111015146 111039972 111031437 810028333 111003125 111003123 810022099 111002409 111001985 810018962 111001967 810022143 810019306 111021823 111003260 810019275 111018658 111012255 111012309 810013093 111001974 111003085 111001990 111003036 111002421 111001999 111010966 111002728 111004185 111002704 111003082 111001968 111010953 111012252 111001993 111012259 810022373 111039958 111033474 111002739 111007572 111028495 810028242 810006595 810028197 111031524 810019941 810001592 111034878 111025507 810026844 111042215 111040177 111007552 111013800 111031182 111034912 810000615 111015199 111002467 111002464 111001981 810001582 111036546 111001973 111009269 810021475 111036412 111036505 810001219 111031442 111025125 810026863 111002408 111042027 111033577 111012271 111042276 111036545 111018757 111014258 111018686 111025089 111026469 810006784 810000579 810022004 810000833 111027560 111036738 810001047 810021751 810008388 111014703 111027758 111027873 810026779 111031207 810007344 810027061 111027831 810000619 111036629 111033597 111025510 111003172 111003055 111003129 111036487 111036463 111034928 810028253 111036366 111042022 111033632 810026841 111036740 111031205 111036566 111042321 810001950 810006598 810002975 111042293 111031431 111027949 111018551 111030968 111033873 810021941 810000755 111034029 810007373 111042166 111022426 111025652 810000920 111022914 111031482 111033631 111034030 111027698 810000743 111033605 111033658 111027568 111007936 810000764 111022717 810002525 111033717 111034026 111014698 111005289 111036749 111007733 810000852 810019073 810026850 111013088 111007951 111033578 810000617 111007698 111033786 810022356 111013575 111014691 111024878 111007961 111025523 111026576 810008797 810019040 810028618 111014702 111033489 810022084 111027755 810008401 111024712 111020765 111036770 810026897 810028566 810022000 111039997  Model Number: DR ICD REF CDDRA500Q UDI-DI Code: 05415067032003 Serial Numbers: 111042398 810007540 111030472 111032157 810007571 111035304 111039938 111041830 111032546 111031002 810019188 111014825 111032144 111032535 111042488 810028679 810019191 111016802 111021355 111031038 111029005 810014900 111035533 810018350 810017688 810028759 810028760 810008122 810008133 810014349 111014153 810018126 111011536 111014241 810023659 111035280 810017425 810017430 810017854 810017451 111019471 111035071 111021969 111022986 810017847 810017744 810017863 810017690 810017672 810017702 810017727 810017741 810017661 810017745 810017662 810017653 810028393 810028382 810028366 810028395 810028380 810017853 111043019 810017685 111043269 111043270 111043296 111043099 111043108 111000691 111000412 111002225 111018855 111000425 111001217 111024705 111014240 111014168 111014226 810007891 111018999 111028409 810000230 111015113 810023847 810023930 810023758 810018468 810023864 810023619 810024686 810014712 111012049 810020771 810015692 810004954 111021670 111012045 810005521 810025920 111010275 111002811 111019730 810010860 111010517 111023438 810020838 810020632 810006880 111012081 111012246 111012683 111012608 111020155 810023473 810023958 810009359 810011178 111011152 810025583 810011996 810022227 810020568 111021319 810010110 810003417 111013853 111001499 111008659 111008637 111020604 111019931 810011260 111022322 111008688 810004419 810003499 810003365 111022657 810012503 810017911 111012632 810023965 810024724 810024666 810019723 810010194 111010309 810007718 810025551 810012131 810024842 810014852 810011176 810014984 810011948 810014391 810010121 111002041 810015008 810025421 810017118 810013683 810005568 810010357 111012090 111012095 810014056 111009897 810020719 810023471 810020894 111006569 111002260 810014668 810014994 111002991 810025524 810025599 810003946 810004532 111021372 810003332 810003362 111022280 810005808 111020800 810010156 111002808 810010032 810010139 111020887 111021326 111012622 111011167 810021287 111020807 810012499 810021290 810010383 810012513 111009605 810010271 111002215 111002168 810023394 111009975 810010841 810005804 810012166 810015039 111021333 810020835 111009914 111018106 810025606 111019956 810009729 810023389 810021285 810023995 810024455 810010225 810010017 810014087 810014653 810015931 810017618 810018076 810019721 810020776 810020890 111002982 111021377 810005763 810005081 111008696 111009205 111013836 111024513 111010882 810022258 810003543 111024517 810010353 810022275 810024430 111014206 111013859 111014237 810025540 111011540 810010793 111011997 111022284 111021289 810022405 810003419 810023920 810010006 111020799 810025517 810025544 810007734 810010199 810010062 810010174 810014678 810011149 810003422 111008654 810005180 810022392 111013914 111008658 810004637 111008656 111020564 111019757 111008686 810011179 810003335 111022666 111002137 810010178 810006894 810011128 810010179 810014645 810010348 810014636 810011231 111020862 810025938 810015040 810023400 111003448 810025620 810009327 810010827 111002169 111002138 810021214 810009993 111002069 111010548 810009362 111011160 810006895 111023818 111013857 810018112 810014948 111018635 810014361 111020590 111012615 810024802 111022349 810005438 810025585 810017395 810004387 111010552 810021237 810007639 810014829 111010486 810025593 810010683 810003606 810025922 111010841 810011250 810020960 810024442 810009979 111015825 111011163 810010001 111002307 810010203 111002088 810025470 810005013 810022228 810019722 810017368 111019727 111009600 810025942 810025413 111014176 111014215 810007685 111014196 111024506 111023833 111020588 810023939 810014659 810025564 111018104 810024699 810020574 810022424 810013322 111002503 810017949 810024712 810022397 810022347 810028962 111020599 111019930 810025630 810021221 111002152 810023413 810025513 810019740 810004406 111021374 810010040 810025411 111009619 810026533 810020971 810005550 810024767 810009401 111021282 810025631 810022348 810023459 810010812 111002264 111014221 810014868 810007656 810025923 810012141 810010320 810010200 810005757 810010880 111003921 111002272 810025554 111009596 810010838 111006715 111012682 111024702 111023464 810014828 111013854 810018079 810017336 810009961 810010345 111008671 111002269 111014242 810022407 810022447 810022422 810022412 810022453 111010869 111002193 111010877 111009905 111009590 111010521 810003328 111021340 810009378 810009356 810025901 810021254 810019734 810015759 810015939 810025899 810025594 810025577 810025462 810024723 810024722 810023993 810023485 810023945 810023919 810012170 810012460 810012476 810010379 810010334 810010180 810010185 810010085 810010106 810011232 810009352 810009411 810009692 810009402 810009421 111002275 111009555 111018983 111019008 810005014 111019004 111018393 111006575 111012027 111012630 111012086 111009393 111009900 111010269 111004464 111003001 111002349 111002295 810010181 111018605 111010828 111009617 111011527 111009957 111009942 111012093 810025573 810021193 810020979 810012481 111023472 111018401 111018870 810010215 111009918 111009919 810010089 810010206 111010481 810004529 810010228 810024834 810012542 111018456 810009690 810010053 810012548 810024824 111002980 810010359 111001226 810010167 810010279 111013672 810023396 111020583 111014198 810010363 810010684 810025423 111013845 111002160 810011160 810015037 810014458 111022300 810011171 810022257 111010848 810009418 810025917 810009383 810010045 810005200 810004400 111009917 810004961 810007768 810025793 810020775 810025937 111001517 111010306 810010060 810010147 810024731 111010513 111019773 111011523 111002166 810010318 810025518 810022245 810017237 810025636 111018615 810007676 810012454 810015696 810017247 810017902 810017961 810010221 810010127 810009339 810012520 810012169 810012550 810025447 810017122 810024691 810015758 810006911 810025454 810009382 810017983 810008000 810010700 810010858 810024689 810023494 810021279 810012705 810010000 810012482 111012617 810011084 111021370 810010136 810010240 810010222 810005523 810025789 810025961 810025469 111014179 810014858 111009388 810014857 810011133 810009329 111018099 810019730 111009219 810014855 810015768 810023404 810009324 810009312 810025461 810004947 810025409 810023990 810021267 810009863 810011091 810011187 810010022 810006887 810021199 111023034 111011528 810012515 810017599 810025910 810025940 810012526 810020631 810015763 810009992 111019724 810014970 810011261 810010124 810010117 810010191 810010264 810009386 810012502 810021298 810020630 810022231 810022242 111008647 810025608 810025622 810013945 810010676 810010224 810023405 810007982 810013329 810010803 810011144 111014204 810020573 111018853 111019774 810009736 111018980 111018379 810010857 810012533 810021302 111010858 810017342 810015058 111012058 810014989 810007764 810020638 111023485 810022341 111000699 111020037 111002270 111012064 111000588 111004491 111020881 810011227 810014967 810010142 111002483 810014704 111018996 111018609 111021665 810022441 810022395 810014629 810010155 810023498 810022353 810023491 810010330 810004958 810009317 810009336 111002217 810023616 810023612 111022640 810018075 810017923 111009960 810020996 111000357 810021243 810021035 810012147 111018632 810025792 810014632 810010138 810020580 810009399 810010367 111018608 111002978 810020975 111019761 111003481 810025555 810010129 810012514 810025539 810010795 111009207 111020878 810014874 111020556 810015021 810005764 810010342 810021204 810010365 111018477 111018096 111018616 810007730 810015929 111010541 810009406 810011243 111010508 111009615 810009403 810012462 810006882 111019751 810017113 111008648 810025604 810010043 810010030 810010075 111020059 111022324 810024852 810015005 810005432 810005578 810005092 810004951 111022351 810011088 810025607 111009523 810010171 111000423 111002487 810007649 810009965 111010474 111001228 810020840 810014870 810009968 810011202 810020914 111009945 810004343 111020829 810011999 111010299 810010273 810005560 111022325 111009553 810021226 111010826 810024446 810009725 111002789 810021038 810014468 810005556 111000237 810025548 810025569 810020825 810011153 810004689 810011953 810014665 810025939 810025912 111014190 810010011 111021290 111021275 111022672 111002486 810007736 111002200 810023523 111018251 111020050 111012010 111020053 810007715 111002036 111002286 810011162 810010161 810003729 810003517 111019941 111018438 810012512 810020815 810005086 111019960 810023937 810020814 810017977 810023983 111010852 810014875 111010857 111011531 111023028 111011538 810013315 810014376 810014640 810020986 111021685 810012155 810020912 111021278 111005513 111021675 111002062 810025519 810005522 111019731 810021276 111002299 810010679 111020819 111024073 810025785 111021373 810021241 810011259 810004533 810003944 111009925 810009408 111002198 111002280 810010337 111003103 111018854 111012618 111009558 111022661 810010881 810025616 810010051 111012083 810007657 111020033 111020896 810010693 111009903 111009529 810020965 810020973 810024679 810011943 810011952 810010263 810010233 810011956 111018622 111012053 111002226 810009350 810010084 810005762 810010689 111009573 810011951 810006914 111019719 810006892 111003843 810009438 810005559 810020581 810024849 111009536 810012125 111023031 810012004 111002987 810025626 810005182 810010008 810024664 810012494 810011981 810003330 810020980 810015930 810005011 810013311 810009734 810010009 810007646 810025779 810017241 810014455 810010197 810023963 810025941 810011108 810021258 810022391 111021663 111009933 111009968 111010888 810009966 111009609 111009565 810011971 810014456 111009569 810003670 810011950 810011263 111019736 810007668 810017976 810010811 810010794 810010834 810004398 810009986 111020836 810010103 810012151 810003421 810004627 810010003 810005207 111022353 111020891 810010366 810022263 810022269 810025930 810011224 111012005 810020956 810007694 810020905 111009206 111001054 111023806 101000065 101000047 810014466 810010057 111001058 810011990 810004841 111010880 810017116 810025784 810017381 810010819 111013668 810012056 111013667 810010038 810009975 810007680 111018240 111018250 111018867 111018986 810025633 810024850 810012468 810013324 810015030 810025542 810024806 810023392 810022237 810012497 111010526 810004414 810017919 810019725 111010496 111010562 810012456 810003336 810014086 810005420 111019937 810005495 810003665 810005265 111021358 810004346 111009524 111018988 810017916 111008638 111020048 810004527 810024429 810003492 810014639 810004749 810003327 810025595 810024769 810015003 810015933 810010699 810009976 810010007 111010478 810011946 111010297 810012467 810024439 111018381 111023816 810013310 810011962 111010525 810014644 810010024 810004407 810024847 810017338 111009607 111009971 810007688 810005575 810014661 810003664 810025586 111020834 810010229 810014693 111020570 810010797 810014074 810020955 810025625 810015020 810015004 810024772 810010202 810012461 810009366 810004959 810017327 810010021 810025579 810021311 810020902 111020584 810020822 810024728 111014191 111010866 111009389 810017112 810007693 111009197 111012009 810025407 810021209 810015760 810014702 111020031 810025467 810021283 111006558 111009214 810010371 810003542 111003154 810020811 111010511 111019009 111020868 111011158 111009208 111009198 111013848 111022340 111012087 111023459 111002645 810005489 810012509 810004340 810025581 810010048 111002171 810011968 810004964 810020993 111018398 810010039 111010890 810005767 810022350 111010289 111010476 810011146 810012452 111009920 111012046 810014840 810014822 111009532 810010828 111003024 810025949 810011210 810011228 810017610 810020791 810021046 810021223 810021250 810021213 810023510 810014818 810021316 810025534 810012162 111001524 810006918 111019011 111002074 111013833 810011191 111011155 111008649 111004480 111009196 810004405 111002176 810023991 810010244 810023463 111012092 810014462 810023609 111018990 810012473 111020899 810010370 810020770 810014077 810014675 810015001 810005806 810006904 111010294 111010282 810024827 810003603 810017126 810021242 810025905 810014968 111018095 810011269 111009594 111023800 810010869 111009554 111009527 810017341 111008674 111013834 111019756 111019928 810010686 111001039 810010173 111018470 111000505 810014442 810025514 810024835 111009410 111010302 111020030 810021310 810021002 810021263 810021219 810021260 111018392 810024864 111002261 111001212 111000118 111001066 101000063 111001525 111001040 111008676 810010076 810009860 810014649 111004494 111022270 111013838 111013660 111018469 111009564 111009614 810007663 111010491 810003609 111018399 111018388 810020790 810024826 111022671 111000437 810024678 111021338 810004334 111009526 810009983 111019739 111002835 810009424 810015035 111022333 111001093 111014228 111019938 810023477 111001062 111022327 111006505 111018248 810005263 111000433 810025919 111013837 810023618 111000361 111001244 810009381 810024761 810020841 810020762 810020806 111002035 810009432 810022345 111024068 111021691 810022264 111019746 810021034 810012479 111022334 111020077 810009358 111022344 111001209 810012484 810024717 810024822 810024862 810014674 810020988 111009604 111000362 111003926 111010531 810009360 111021288 111003933 111001507 111001050 111022653 810009307 111024060 810005208 111000374 111021364 111021254 111020039 111001254 810014669 111000400 810009335 111001060 810010393 810014337 111009929 111001097 111000735 810021043 810022354 111022350 111000717 111001067 111022339 810009373 111000416 810012543 810025934 111000743 810023972 810012511 810025457 810013304 810022252 810024729 810021045 810020763 111020585 810025927 111020826 111020815 111021267 111021255 810005564 810005553 810005274 810005275 810009742 810011208 111009540 111010845 111003925 810017566 810017374 810020820 111020835 111020803 810010312 810018099 810025532 810022446 111009563 810009365 810017120 810010806 810010862 810025909 810021274 111023022 111009597 111020151 111020606 111003109 810010335 810010175 810010227 810024694 810010859 810020994 810020842 810021270 810020766 810020826 810009344 810009877 810010262 810005807 810010134 810010193 810010283 810003520 111002590 111011533 111009547 111019704 111009560 11 1002792 810005805 111010874 111022276 810010254 810004528 810004535 810004333 111009622 111011151 810017115 111020072 810024844 810023941 111020817 810015045 810007704 810023499 111014169 111000177 111000253 810024861 810011097 111000178 810025618 810024696 810025463 810025458 810025523 810020717 810022400 810022393 810017931 810017223 810017563 810017591 810010049 111021279 111021268 111002600 111022998 111019972 111019953 810014441 111001535 111000176 810012123 101000048 111002669 111002058 111000818 111000737 810010843 111000367 810013326 810010026 810010100 111001519 111000415 111001202 111001237 101000022 111000382 111000642 111001218 111000510 111000240 810010176 111001240 111001204 111001219 111001484 111000420 111000172 111001528 111001059 810021201 810014992 810010871 111019720 810014972 111019703 810003607 810005089 810023512 111019697 810017562 111012245 111009953 810022450 810014987 111018856 111022656 810010184 810024454 810023506 111020040 111019702 111018479 111018868 111018614 810020627 810014633 810014657 810025531 810005580 111009940 810024690 810024714 810013684 810009423 810010361 111000407 810023520 810004690 810004747 810012472 810012537 810013316 111020563 810024682 810012025 810012180 111010547 111014239 111014243 810011132 810010079 810010238 111022647 810010817 810009737 810009952 810010696 111002830 111003423 111022291 111022297 111022281 810023981 810025952 810025946 810014326 810014328 810010884 810010691 810005502 810005488 810005519 111002660 810025414 111018860 111018858 111018861 111018859 810025416 810025578 810010135 810021309 810017390 810010339 810010671 810009354 810009368 810017595 810024715 810015926 810022250 810010223 810025948 810025546 810012558 810014343 810024825 810024830 810024798 810017256 810017271 810017279 810017244 810012055 810012043 111002591 111022283 111010498 111001233 111019695 111009407 111019759 111022271 810009997 810012545 111002605 111002172 810023406 810024774 810017397 810009435 111010830 810025962 810003415 111018871 810010847 810012535 810012701 810011175 810011211 810010789 810010250 810010697 810021312 810021319 111009521 111018623 810012068 111000436 810004536 810005803 111014186 111001049 810010055 810025521 111002801 111002072 111002842 810003667 810005945 111021368 111006572 810025936 111002158 111002092 111010540 810007720 810019749 111021335 111001245 810023470 810009332 111010530 810014681 810012029 810012089 810012039 810021191 810017937 810017941 810005261 810021255 111010567 810012138 111002057 810011105 810007592 111021249 810028942 111006524 810009999 810025928 810005091 111002635 810009392 810009319 111000363 810023489 111000165 111018486 101000018 111001234 111002489 111009984 111020036 810023611 810023515 111000687 111000383 101000061 810025533 111009956 810017347 810010137 810012478 810010112 111020066 810014991 810011151 810003521 111009200 111020148 810015034 810009874 111009387 810019746 810015695 810020715 810011158 810017119 810024677 810009343 810014075 810017135 810004955 810011230 810014070 810019737 810014832 810021234 111000410 810009440 810004463 111020049 810021264 810012142 810014666 810017613 810023414 810010347 810014841 810004464 810023973 111013846 810024764 810013305 111019010 810011085 810024680 810010195 810010044 810010328 111019920 810014454 810015928 111020876 111010832 111010839 111013680 810020998 810009886 111019784 810015932 810014655 810023994 810025603 111012022 111001227 810011089 810011086 810025790 810023976 810012560 810014830 810014873 810024695 810020778 810022342 810022409 810010172 810010274 810010336 810010063 810010385 810017953 810004632 111010843 111023810 111022267 810010164 111019005 111012611 111009534 810010820 810020824 111003431 810010234 810020572 810010338 810014088 111009199 810024771 810024762 810024663 810023524 111014220 810010169 111010271 810007706 111011534 810020636 810023475 810021000 111019726 810013943 810010213 810011268 111002805 111018253 111001055 111000602 111018613 111018866 810024448 810023980 810011239 810003877 810025511 810014952 810014842 810025509 111023041 111020042 810010281 111018241 810013307 111032377 111032374 111029925 111029926 111037423 111037416 111037419 111037410 111037424 111012623 111009539 111018603 111009932 810014082 810012528 810013318 810010284 810024733 810024760 810017576 810024730 810017567 810017580 810024734 810012007 810011993 810014638 111001481 111000170 810024795 810017129 810017350 810007763 810023946 810021217 810017356 810007675 810012540 810025915 111009588 810012491 810021247 810009688 810025547 111001211 810015698 810009706 810017582 810010373 810025954 810022404 810010146 810022246 111013913 810025580 810010375 810020981 111022639 810005268 810012134 810021257 810010242 810022340 111012023 111019755 111021292 111019721 111013841 810012536 810025637 111009525 810014695 810014673 111009907 810017597 810021307 810010319 111011165 111012687 810005209 111021367 111009961 810003723 111010301 111018451 810010081 810012063 810014701 111006562 810010230 810010091 810023416 111020605 810010372 810009987 111009587 810019739 111023820 810014851 111012077 111012016 810009427 810010313 111006571 810009302 810010352 810009405 810020782 111009974 111009977 810021224 810004638 111009602 111009922 810011225 810024829 111011529 111020597 810010071 111011522 810004383 810009668 111023009 810020906 810017964 810014949 810010064 111021350 810014658 810020891 810025553 810017234 111019968 810006879 810024763 111009913 111009613 111009542 810012050 810009330 810014676 810010863 810014837 810014839 810014820 810006915 111002619 111012012 111014201 810017932 810017940 111018610 111018436 111009921 111018992 810005201 111021693 810012165 111018849 111020139 810007740 810015073 810020716 810010099 810025543 810009875 810007665 810007766 810012704 810009328 810005424 810005425 810005270 810025545 810025473 810025536 810005577 810005490 111009585 810005572 810005576 810005497 810005503 810005563 111011525 111009939 111009936 111009557 111009621 111009571 111001095 111001216 111001044 111000509 111001200 111000427 111000421 111001495 111001509 810010080 111000600 111019782 810006907 810010866 810010019 810014679 810017216 810028960 810025563 810025780 810025522 810025505 111023007 111000169 111019765 111021273 111020063 111005511 111008662 111013843 810014860 810011944 111002205 111000222 111012082 111002267 111010856 111005520 111001099 111006561 111014177 111000515 111020820 111012028 111003439 111019796 810007700 810007729 810004342 111003486 111000585 111000698 111001213 111000503 111000740 111009218 810005085 810025953 111012673 810010013 810009857 111022296 111009531 810004950 111019925 111020076 111014230 111018238 810017383 810023387 810023969 111002802 810010150 810010168 810009984 810010236 111012071 111010479 111010529 111010503 111019748 111010492 810010152 810010360 810011992 810022344 101000028 810022398 111002984 111000516 810005017 111000587 101000020 111020822 111012609 111004465 810010166 111001210 111001203 111014200 810021196 810010029 810006889 810010389 810010087 810014329 810009428 810009393 810004420 810025529 111020821 111019940 111000408 111000718 810024700 810012014 810014352 810007761 810007786 810020641 810007782 810025516 111018630 111010872 111009562 111003408 111003247 810004530 810007590 111002831 810004341 810005581 810011195 810010837 810009733 810010278 810010231 810017333 810020813 810019731 810022233 111009941 111010879 111005515 111018449 111009986 111002491 810011266 810006890 810005167 810025894 111002357 111018487 111003469 810009686 810015693 810006905 810003549 111002353 111014171 810019726 810003547 111002359 810009871 111009976 810015761 810023613 111003021 810003498 111000356 810004963 810025420 810012152 810024443 810014637 111006525 111018485 111010863 810014648 111012075 810024427 810005758 810021304 111000819 111024076 111009969 111019699 810017337 810023986 810004751 810011986 810015015 810024447 810011194 111013684 111021688 810003416 810014321 810003878 111023462 111023450 111023467 111023458 810013314 810007771 111021246 111021229 810025613 810025632 810025628 810025634 111019738 111009958 111009980 810003602 111002053 111002054 810011235 111020569 810025781 111002499 810013312 810025770 810003945 810023497 810017349 810017394 810017218 810014999 810014854 810017385 810009669 810024431 810009699 111018440 111019715 111020577 810010208 810010259 810009363 810023410 810022396 810022449 810010391 810009700 810010204 810014964 810014966 810024683 111000748 810004466 810004471 810015697 810015699 810010035 111009593 810012463 810005562 810005799 810014062 810014443 810022414 810025537 810022456 810020968 810017373 810011963 111006520 111019935 111019722 810009967 111002191 810017283 810010145 111012014 810007774 810004842 810014691 810004397 111000815 810010808 111024511 111000719 111000175 810014642 111001047 810011188 111012248 810009969 101000025 810014643 111021320 810011156 810022235 810017245 810005566 810025786 810010349 810025947 810025565 810022432 810010207 810010189 810010016 810011978 810023495 810009377 111021250 111020877 810023943 810007981 810014965 810024848 810021315 810022234 810003500 810011219 810013681 111019694 111001041 111010506 111009944 111008694 111010824 111012004 111012074 111022326 111000360 111018384 111000419 111000364 111000377 111002258 111002077 111000413 111000376 111000255 111009411 111023826 111020832 810009955 111012024 810007986 111006723 111012089 810004410 111019921 111009581 111011524 111018446 810010358 111006574 810021321 810021041 111022302 111010544 810021231 111020150 111012676 810010386 111006722 111024521 810025552 810007781 810022260 810025950 111019763 810010065 810004386 111009577 111024067 810017217 810010394 810009687 111018447 810011970 111018604 810017236 810015212 810023386 810023938 111000366 111000243 810009364 810023418 810023382 810023956 810011185 810023984 810009374 810024833 810009361 111001038 111001048 810012453 810011139 810023949 810020963 810020961 810012464 810003361 810017959 111000373 111001496 111001497 810011240 810011165 111020558 810023970 810023954 810017972 810024818 810011253 111001063 810011249 810023953 810012470 810010673 810023419 810023979 810025572 111012621 810005080 810004945 810023936 810017605 810025422 810003493 810025434 810006885 111022268 111014229 810012017 810014692 111014175 810011964 111021354 810011248 111002686 810004411 111021252 810011256 111005504 810005183 810006902 810009988 810011989 810024705 810014684 810019728 111020044 810010041 111019754 810018092 111019768 810015936 810007743 111002820 111002807 810019756 810019744 810019761 810024441 810024432 111012051 111002817 810023482 810009658 810005517 810005514 111009898 810006917 111023396 111023827 810009705 810009656 810007666 810015031 810014990 810012556 810010369 810010316 111018476 111014185 111006560 111010846 111009583 810007664 111018395 111010285 810011272 111002143 810024840 111013849 810010350 810019732 810020821 810022240 111010884 111021322 111013670 810004537 111018863 111009952 111006503 111002146 810013308 810004469 111019783 810014315 810020634 111024564 111002654 111024554 111000252 810020570 810014377 810010092 810011142 810011201 111020052 111019973 810014318 810017393 111002060 810003551 810009407 111002078 810017908 111002290 111009397 810011192 111009391 111009908 111002833 810006881 810009977 111023042 810017581 111002025 810017232 810017583 111018994 810011954 810009889 810005168 111008646 810005204 810025510 810025904 810017335 111024507 810012489 810010785 111020068 810004388 810009745 810010692 111003033 810010144 111009901 810017131 111002293 810003418 111019742 810025592 111020057 111002181 111006526 810014696 810015687 810025911 111020890 111019712 810010791 810009693 810010056 111020055 810021273 810007595 810024692 810017905 810020635 810012160 111019776 810025619 810003497 111000379 111000370 810005770 111002075 810014464 810014485 810025774 810005760 810017943 810018080 810022270 810010378 810017622 810015048 111021245 810022427 111021371 810025615 810011122 810005950 810004538 810025460 111023027 810010833 810021278 810015937 810025771 111023002 810023605 111019963 111021284 111013655 810009888 810010245 810023486 810003943 810011209 111018847 810017242 111009979 810017607 810007776 810025776 810023610 810019748 810019720 810018105 810021299 810024718 810004750 810025587 810010324 810004748 111020874 810010068 810020784 111010559 810005273 810024438 810012178 111020830 810015002 810010695 810007717 810007669 810005802 810023467 810023487 810014667 810017353 810010196 810007710 111008664 810004332 810012459 111018465 810025903 810007747 810011258 810024820 111018452 810004402 810012530 810003546 810010118 810017280 810004957 810009439 810022273 111003440 810009398 111009591 111002156 810012531 810010201 810017331 810010251 810003605 810003357 810005409 810023960 810010170 810011941 810003519 810023479 810023502 810007641 810012538 810022230 810022394 111003464 810017560 810004534 810025468 810023988 810024451 111019939 810009994 111022355 810007724 810024721 111023812 111009915 810014869 810003494 111019769 810018127 810024697 111012048 111018483 810003663 111020156 810006903 810025559 810009703 810010094 810005206 111018978 810007987 810006897 111023040 111021323 111021243 810023950 810021308 810004688 810017440 810017433 810017422 810017428 810017450 810017459 810017426 111043094 810010816 810011116 810017274 810017277 810024836 810025945 111018247 810010073 810009395 810012510 810004956 810009434 111021287 111024074 810011273 111024613 810024706 810021036 810021300 810023989 810023390 810024863 810026532 810026534 810023401 810014359 810012546 111002827 810025918 810019753 810011136 111009215 111021669 111019012 810012060 111018997 810025433 111023471 810015690 111018627 111008679 111002220 810010317 111009906 111009530 810009388 111000751 810010058 810014378 111001061 111002081 810014835 810009349 810011254 111002852 111002132 111010850 810005090 111010557 111019792 810024720 810024710 111019701 111019971 810018072 111009599 810024846 111003166 111018606 111018599 111018864 810005166 111018629 111020833 810010853 810018082 111023025 810010252 810010829 810003518 810010315 810005338 810007780 810010798 111018628 111019706 810014344 810024832 111002992 111002030 111003455 810024854 810010802 810011117 111005508 810025441 810017386 810009985 810014342 810005414 810004953 810004412 111021277 810010258 810017132 810017396 810011174 111018633 810005769 810022266 810017954 111002174 810014327 101000062 111000113 111001199 111009394 810013940 810006910 111012008 810022440 111002839 111018458 810010042 810014048 810012702 111019707 810009678 810012534 111020580 810004421 111006501 111008689 111014192 111009983 111019778 810025560 111021664 111008678 810005012 810010368 810010090 810017628 810023944 111010859 111009572 111012039 111013652 111018396 810010253 111012015 810017579 111002505 111008639 810010355 111009570 111010822 810011275 810006916 111019007 810009422 111009618 111001221 111002274 111019957 810025425 810011199 111002179 810005018 810011154 810011955 111003102 111012631 810009697 810007652 111002145 810006912 111023813 810017267 111006689 111023443 810025455 810012172 111008652 810011980 810017910 810003329 111019000 111002665 810010856 111003106 111018400 810007789 810007753 810006886 810024711 810023955 810021288 810010822 111023832 810017133 111019965 111020572 810010059 810010382 111001521 111022354 810017389 810006891 810017958 111018390 810010388 810004336 810005909 111027021 810010362 111010507 810023500 111013840 810009431 111013657 810011222 111019711 810017960 810025605 810009954 810010810 810017951 810010815 810024445 111023442 810020987 810010356 810003545 810022439 111019745 111020067 111019713 111019772 111019775 810020712 810020834 810021346 810020807 810020823 810020830 111021227 111024069 111010550 111010522 111010560 111010539 111024078 111024075 810017573 810025464 810025471 810015213 810025621 111001239 111020600 111010281 111010303 111010292 111000518 111000430 111000244 111000513 111000217 111018385 111018383 111018380 111000385 111001230 111002348 111020945 111020810 111000435 111000498 111000403 111012637 810003724 111020560 111020601 111020875 111020866 111000426 111000411 810006898 810015052 810015056 810015087 810015083 810019741 810019717 810019727 810014663 810004382 810012517 810012549 810012523 810012519 111000429 111000418 111001064 810009397 810009345 810015762 810006909 810007784 111001493 810004470 111000812 111000521 111000359 111000432 111000683 111000693 810010067 810010025 810010327 810003424 111019944 810014856 111010273 810017257 810017957 810020849 111011170 810011205 111019787 810012067 111018848 111035475 111026171 810018616 111012018 111020047 111020034 810010826 810010889 810010854 810010046 111021274 810023522 810023509 810005267 111003014 810020909 111011162 810024703 111012038 810005093 810012516 810014345 810006883 810006900 810005518 810022448 111021264 111011999 810023385 810006908 810009351 810014486 810005801 810005941 810012544 111002488 810009347 810009310 810010247 810003333 810017281 810014335 810023947 810017285 810024855 810012507 810020976 810022244 810022226 111001096 810005212 810025557 810018107 111019764 810005520 810009384 111014214 810005952 111019967 810024768 810025787 810025598 810025575 810025412 810025601 111012681 810003731 810003331 810004335 810014364 810004467 111011164 111002059 111009927 810007775 810011127 111003030 111008684 810017286 111012030 111010283 810024452 111001220 810014319 810010387 810011124 810023957 810004399 111000692 810005169 810005094 111022287 810020795 111013674 810005019 810012455 810025943 810022420 810020970 810022272 810025602 810023403 810023518 810017369 810021249 810017564 810019752 810017602 810023391 810026531 810026537 810019724 810024685 810019754 810025773 810023490 810025520 810025775 810022401 810019763 810022459 810024857 111010889 810009744 810009683 111024071 111014222 111009396 810022403 810009410 810010235 111022648 111002855 810010086 810009429 810025527 810003496 111018467 810003363 810017915 810015756 111018977 810021343 810010380 810017917 810003420 810003671 810005083 111021324 810010322 111008651 111008691 111008669 810025525 810017332 810025589 810010037 810017268 810025430 111018382 111009209 111008687 111021684 810024858 111014199 810003611 810005010 111019729 111020884 810021229 810024821 810009414 810012087 810017620 810003423 111021689 810005498 810021238 810025507 810024436 111017211 810017278 111001205 111001052 101000027 111000499 111001051 101000067 101000023 111000218 810015022 111020944 810010888 810010248 810014083 810003501 111023444 810020829 111019714 810005448 111023008 111012610 111023437 111021363 810009665 111011161 810017904 111003170 810021305 810024428 111012063 810025437 810011081 810011198 111004481 111010308 111014217 810004633 111010564 810024435 810020969 810005271 810005436 810014067 810010855 810018093 810023402 111023804 810005496 111021344 810010813 810010012 810010678 111002622 810025432 810023474 810007702 810014996 810019742 111002067 810005429 810011137 810022271 111013677 111002026 111018448 111010865 111002637 111018255 111006577 111013683 810004843 111019791 111003162 810005430 111021690 111021233 111019794 111003441 810004391 810005210 111022664 111019786 111019797 111022676 810011183 111021334 111020942 810009851 111002997 111006513 111023435 810015075 810004390 111010475 810007996 810020783 810009853 810010323 111003110 810010095 111020582 810020901 810017939 111020808 810014446 111023835 810018101 111011168 810012547 810017382 111018481 810011223 111022310 810020910 810017947 111002080 810015199 111012249 111020886 111012689 810017596 810012062 810010272 111003007 111022358 111009402 810009878 111023026 810012031 810005485 810007772 810009325 810014333 810018111 810014362 111003474 111001091 810010329 810010018 111002664 810015049 810021317 111009946 111002084 810010848 111002089 810004635 810020903 111002612 810010255 810012033 810014444 810020836 810017568 810010849 810021005 810024859 810009981 810011083 810018088 810011131 810017906 810017261 810018089 810020978 810010883 810018120 810015076 810011972 810014993 111004478 810010675 810012143 111021353 810012153 111020562 810010872 810010823 111020883 810020720 111003105 810010788 810014703 810013320 111002194 810020989 810013941 810010131 810012184 810010036 111002044 810021190 810012521 111002196 810003668 810021189 111021222 111003927 810014316 810010109 810005417 810010685 111019945 810023397 810017980 810023393 810012073 111002598 111020831 810005264 111003833 810017984 111014184 810017269 810023407 810017214 810021265 111023465 810005939 810021225 810018094 810021216 810011982 810009370 810005413 810012048 810021252 810003669 810023411 810005561 810005440 810018130 111002139 810010873 810017215 111023836 810011242 111023814 810023408 111002642 111003160 111002033 810004074 810010116 810020768 810010123 810022243 810021200 810020792 810009974 810017903 810010824 111010554 111013916 111013847 810015215 810012040 111010304 111023030 111023802 810017575 111022312 111002688 810010082 111002019 810024837 111023005 111018478 111002086 111021256 810018103 111009403 111002594 111022318 810017380 810012059 810005416 810009998 810012034 810015085 111008663 810011271 810010878 111021357 810017221 810025405 111002031 111002647 810010104 111010568 111023512 810012557 810004416 810014317 810014065 810018087 810011125 810014040 111002028 810017914 810011220 810015080 111002836 111002256 111014173 111008685 810010275 810019733 810010783 810005447 810010249 810005911 810023458 810011214 111002045 810011958 111020859 111019966 111021686 111011998 111012084 810007773 810005202 810014382 810004076 810011229 111003410 111010532 111002821 810014985 810005574 810005493 111002998 810005571 111023809 810020907 810017379 111022295 810020779 810020799 810012136 111022646 810013323 111006530 111004471 111019918 111002024 810005953 111020143 810017986 111002816 111003834 810017233 111004495 810009675 111020043 810023468 810009990 810011965 111002305 810023409 111002157 810010047 810012158 111004490 810014085 810018110 810024810 111014232 111024584 810020967 810020995 810015757 810021245 111002052 810017344 810017348 810012126 111020074 810014051 810020839 810021253 111022338 810007995 111019795 810019760 111019735 111005519 810010672 810011190 111010527 111006565 111022289 810010241 810021261 810014706 810005554 111002289 111021366 810012529 111002983 810015071 810004010 810009735 810021215 111009935 810009391 810017935 111002592 111021247 111010499 111021331 810017377 810018074 111002180 111021348 810011947 111002167 111002090 810005947 810017359 810022419 810019747 810010326 810015059 810012508 810015694 111014172 810017354 810014066 810009854 810004401 111006502 111022659 111020595 810004337 810010052 810010083 111005503 111023029 810020892 111004472 111023508 111002150 810014373 111002206 810009862 810009962 810021194 810021303 810005431 810012157 111013860 111023815 810007997 810017952 111018484 111020566 810007769 810014071 111003118 111018239 111003838 810005942 810011212 111009963 810005567 111008693 111023803 111002262 111005516 810017558 111014219 810022416 810022351 810014046 810025440 810018091 810015766 111006563 810007684 810005343 810017965 111010543 111002482 810017284 111010545 810003726 810011109 111021351 810012492 810005491 111002049 810018077 810017572 111012031 810004831 810012044 111023036 810012074 111002223 810005181 810012016 111003922 810010096 810015769 810004687 810005337 810010792 810011121 810011186 810011148 111004240 810014076 111018482 810011138 810010256 810007703 111023455 111018489 810009866 111012250 810012148 111001037 111006504 810017926 810023511 111010483 810010840 111021356 810014963 111013666 810004634 111010276 810010809 111019771 810014332 810020896 810005336 810015046 810017928 810011218 810015043 810012154 111011535 810014324 810014330 810013327 111002177 810014045 810012001 111012007 111019934 810005412 810012496 810023617 810017330 810018121 111022357 810005443 810014041 111012052 810005768 111013671 810014323 111018088 810015011 810014688 810017367 111023824 111021280 111003454 111003836 111002148 810014358 111002671 810005492 111002153 111022337 810010886 810010182 810007994 810021322 111013676 111020870 111002641 111002071 810009740 810025445 810015086 810012471 810014375 111002040 810012145 111019961 810004072 810014054 111003837 810017948 810013332 810005494 111003418 111002034 111012686 810014049 810018117 810010314 111002051 111020811 111020046 810021232 111002263 111021692 810021301 810017238 810021320 111002210 810005415 111020869 810010257 111019978 810014461 111020858 111023509 810020833 810017615 810020760 810025444 111020824 810010364 111012042 111010571 810011107 810014043 810020900 810005433 810010108 111012043 810017963 810017213 111021258 111004476 810009671 810012032 810005435 810015042 810004389 810025406 810017362 810018119 810015006 810007721 111018090 111002593 111002070 810014347 111023511 810017252 810009326 111004489 111002201 111023003 111002822 810010130 810017355 810015198 111010875 111002265 111006576 111019977 810007711 111012678 810009660 810024838 810017357 810004540 810021228 810010093 111020078 111018091 810011215 810024765 111020056 810021197 111002079 111018246 810017224 810024725 111002803 810010384 111009973 810017933 111019943 810015084 810010054 810010122 810005486 810014393 810024807 111006506 810015081 810020911 111023460 810012080 111002848 810010105 810021271 111002806 111019758 810009380 810005761 810007690 111002018 810018113 810015700 810017246 810009982 810025600 111015834 810020803 810005772 810011237 810010842 810010787 810004075 810025427 111012021 111018394 111010510 111012070 111024063 111002857 111002066 111003254 111002055 810009677 810010074 111013856 111002616 111004484 810005339 111014212 810017364 810012085 810014064 111023513 111020818 111006519 111021683 810009306 111002691 810020817 810017254 111010570 810005418 810009304 111019949 810014371 111003453 111002209 111002658 810011193 810010183 810014662 810024449 810022338 810014708 111013673 111023024 810005262 810007793 111008667 111009949 111004493 111002144 810021313 810019718 111013651 111018455 810017251 111002602 810010804 810011979 810017574 810024453 810010128 111020867 111021228 810020797 111010518 111003845 111003846 810011987 111018377 111006527 810014044 810023488 810014387 810011994 111002631 810012132 111002358 111012076 111010533 111018480 111021376 111020029 810020844 810023614 111010873 810007726 810010280 810015009 111021341 810023472 810017329 810010688 111004492 111012035 111012037 111014236 810011247 810017227 810025448 810017982 810009980 810012042 810009348 111002304 111003152 810014440 810007795 111010293 810009865 810005341 111003005 810020800 111013858 810007739 810025408 810017944 111002083 111002287 810004636 810012161 810023460 111010484 810011152 111010489 810017974 810017371 810010392 810017955 810014995 810009885 111004468 111020805 810012012 810021198 810023464 111003924 111008683 111009552 810018100 810012504 810017934 111002140 810012480 810012027 810020827 810020805 111020045 810011143 111014225 810004833 810020843 111013663 111002610 810009876 111012685 810004011 810007673 810014073 810017924 111014205 111003929 810010376 111012068 810018071 111002850 111002853 111013852 810011130 111022305 810005516 111018442 111011996 111006578 810015072 111002985 111006568 111012044 810017907 810013325 111018244 111023023 810003662 111010286 111014218 810012124 810017376 810004403 810005411 111020802 111019798 810012163 810011252 810010163 111003151 810025415 810011102 111002638 810017559 810010876 810018109 810012015 810017962 111002161 810017592 810023420 810024811 111008653 111010538 810019764 810005551 810012159 810011217 810011172 810009316 111021365 111010536 111014235 810004631 111006579 810004007 810005912 810010354 810012487 111002837 111002192 810021347 810025623 810015028 111002022 810009701 810009739 810011164 111024563 111002027 111004483 111020145 111002087 111002824 111002628 810024434 111023819 810024727 111002136 111006516 810010097 810009346 111018378 111010488 111002306 111003468 111010493 111010854 810020845 810017971 810005269 810017326 111010279 810005446 810014080 810020801 810014385 111023020 810019755 810009311 810009710 810018116 810014463 810004140 111012020 111013678 111021257 111006564 111002649 810010276 810017577 810021188 810003942 111002076 810004640 810010265 810004073 111009943 810009696 111010494 111003450 111024562 810021266 810017570 111024617 111004482 810010830 810011161 810017927 810018128 810017114 810014322 111010535 810012518 810005800 810014079 111003443 111014193 111002604 810009337 810014392 810018125 810014355 810011998 111002988 810009679 810005500 810020789 111002134 111002589 810017621 810010677 111013862 111021239 810014448 810017925 111010886 111022677 111013656 111014174 810005766 810005260 111010473 810005345 810005487 111011526 111023032 111002268 810012168 810015025 111018097 810021275 111022663 111002068 111004477 111021260 810020959 810012501 810024799 810015765 810018122 810017556 810004630 810009708 810017222 111012003 111023822 810018073 111006566 111020812 111014197 111020863 111020814 810012164 810023417 111023828 810025410 810010341 810009960 111023456 810005499 111014244 111014216 111021332 111014183 810021039 810009667 810018114 111023469 810005419 810004838 810005437 111002504 111002050 810005342 810023521 810010879 111002186 111021259 810005427 810004835 810005410 810012000 111004479 111021220 810023383 810009376 810009314 111010500 111002308 810018102 111004487 111021347 810017598 810004641 810018081 111003923 111019946 111006507 810022421 810011147 810014969 810017946 810013331 111002814 111023825 810014348 111006528 111012032 810024809 810017240 111010885 810010377 810010260 810020713 810023469 810017922 810012156 810020957 111021673 111002135 810009400 810014451 810011213 810010786 810004639 810009662 111002266 810009713 810017578 810009964 111020893 111014170 810019759 810011106 111012057 111003256 810017130 810012135 810017593 111002164 810020718 111002490 111002481 810020804 111002643 810004345 810005346 111002219 810014068 810004137 111004488 111008661 810013330 810020897 810017384 111002690 810012010 810017372 810005272 810017255 810003666 111021359 111002278 111002043 810020846 111004473 810010835 810007792 111023439 810022274 810010850 810005910 810004138 111004237 810014057 810005759 810010266 810012026 111006508 810015691 111020940 810012505 810010698 111006523 111012671 111019789 810011255 810021268 111002082 111021360 810009856 810009707 111020871 111023461 810015688 111002063 111006518 810010177 810025924 810010381 810011114 111002037 111004470 810005344 810020964 111014231 810017375 810011233 810017909 111009982 810015064 810015069 810009852 810024770 810021233 810005445 111022665 810010148 810018078 810012045 111002493 111010287 111018242 810017225 810010033 810005259 810010088 810003727 810009712 111005507 810009394 111023451 111003010 810015057 810018115 810018083 111018103 810018095 810017228 111019717 111003034 810022239 111003847 810005266 810018124 111020879 111012624 810020898 111003919 810017589 111010553 810011094 810017121 810021203 111002587 810012129 810004836 111010504 810020990 111002184 810005423 810012174 111012636 810024773 810024796 810010101 111021244 810017601 810009855 111018443 810005348 810004141 810010343 810012122 111012605 111006559 111020071 111002211 111002615 810011221 111003250 111023831 810004008 810011145 111021270 810005569 810017366 111021352 810023462 111020894 111006514 111002501 810015044 810005277 111021661 810010243 810015200 810005442 810024800 111002652 810021244 810010269 810014460 810014383 810005951 810017561 810015767 810007985 111024058 111021662 111021666 810004139 810004404 810017229 810020761 810017399 810017378 810010344 810017985 810011975 111022637 111002208 111008695 111003163 111024059 111020070 111002221 810009313 111023515 111002790 111002224 111019779 810004009 810010102 810012128 111003930 810005773 111002303 810017266 810009973 111010534 111018386 810015066 810007777 810005771 810025438 810022435 111003157 111004469 810010801 810005558 810017231 111010290 111010284 810021297 810013306 111013664 810017585 111010878 810017134 810005082 810018085 111002046 810007588 810012054 810017987 111003230 111010272 810018108 111012019 111020825 111018461 11 1022292 111002094 111004496 810004628 810017973 810010874 810012144 111010295 111009203 810012133 111009924 111002085 111031374 810015935 810025908 810009978 111009551 111003482 810014871 810004962 111018620 111019923 111009550 111002195 810010845 111020035 810020787 810021282 111019936 810010198 810024816 111003000 810024684 810010877 810003358 810003364 111018624 111019734 111002190 810023968 810022268 111019750 810005340 810010211 111003092 810024437 111018450 810011274 810020772 810003610 810014697 810003359 810005549 810005441 810011264 111023440 810017604 810010246 810010034 810015940 810011957 810017273 810023484 111001094 810003366 111020885 111001480 810009379 810009419 810010220 810010268 810025913 111002273 810025582 111022307 810025783 111003239 810020816 810017600 810004629 111002056 810011244 810024814 810007983 810021207 111022347 810006896 810012167 810021192 111020895 810017212 810017220 111020061 111021346 111021225 111021362 111020041 810023496 810023457 810007752 810007760 111019728 111019753 810005946 111018387 111018453 111012607 810012495 111024516 111020897 810015010 810009659 810015027 810024865 810014950 810014951 810015033 111020065 810006899 810014078 810005211 810005756 810004344 810004339 111023801 111023811 810009338 810009331 810021314 111004239 111004467 111021033 111019752 111008680 111008692 111024514 111010838 111010823 111002183 111002142 111002207 111002202 111006573 111006570 810011200 810009367 111024515 810007988 111012619 111012251 111009589 111008641 111008645 111011530 111009395 111019926 111020064 111019947 111019948 111010847 111004486 111004238 111004485 111012670 810014061 810014388 111002689 810020966 810021345 111022293 111009937 111001527 111001515 111001502 111001520 111022662 810007999 810007586 810007636 810007643 111023039 111022317 111022314 111022298 810009887 810009957 810009959 111002298 111002291 111002204 111002199 111001510 111002846 111002799 111014181 111014210 810009963 810009971 111002672 111002687 111002629 111002279 111009528 111009522 111009398 111009601 810007644 810007648 810007741 810007708 810007705 111019929 111002634 111002650 111012677 111012613 111012017 111012054 111012635 111012065 810007689 810007686 810007698 111012080 111012026 111012078 810007714 810011092 810011098 111010565 111010849 111010537 111011157 810007709 810007654 111023397 111010274 111010472 111010477 111010495 111021329 810007723 111001531 111010836 111010523 111001488 111001503 111023006 810011090 810011110 810010825 810011123 111009950 111009948 810012022 111009392 810007650 810007672 810010890 810010687 810011270 810010784 810011197 810010682 810010839 810012071 810012140 111009947 810012176 810012030 810012179 111009966 111008682 111008665 111003099 111003006 111003018 111011532 111002617 810012037 810011984 810012177 111022285 111022680 111022682 111022686 111022679 111022304 111022658 111022272 111022649 111022645 111022641 111012606 111012626 111019790 111020157 111019743 111002596 111002608 111002588 111002613 111012620 111024567 111018488 111013863 111009909 111035327 111024556 111018472 111023448 111037432 111024553 111023466 111024062 111001053 111023446 111020872 111018437 111018631 111035348 111021694 111021286 111019933 111021265 111009537 111014223 111037431 111035332 810025931 111001243 111001231 111001224 810020721 810022452 810012013 810012052 810017571 810022429 810022339 810014634 810014635 111014224 111013850 810007751 810007737 111009962 111014194 111010561 111018850 111009578 111021231 111023015 111024077 111037426 111021226 111009204 810012083 111002091 111010524 111002065 810025466 111009216 111008681 111009938 810007591 111009201 810018098 810018096 810022346 810020780 810009682 111009959 111021361 810010277 810010219 810010267 111022660 111022644 111022651 111001486 111001511 111001526 111001206 111001100 111001197 111001235 810012485 810012488 810012477 810012465 111005518 111001512 111001498 810009702 810009664 810022249 810022248 810022267 810022444 810022247 810022236 810007745 810007990 810007759 810007993 111008690 111010842 111020565 111010861 111010837 111020594 111020149 111020154 111020581 810014971 111002656 111005502 111002484 111002606 111002502 111002500 810012002 111012627 111012675 111010834 810013942 810014059 810014069 810014058 810014354 810013944 111013679 810023985 111009586 810012008 111010881 111010853 810007594 810007758 810025526 111022678 111022674 111022670 111012679 810017370 810017398 810007671 810007658 810007589 810007637 810007670 810007659 810007638 810007640 810007674 810007660 810014682 111023046 111023045 111002302 111012079 111023018 111023001 111002659 111023000 111023012 810017352 810017391 111013658 111013654 111013681 810014628 810014630 810014654 810021218 810021272 810025449 810025439 810005555 810009436 810015024 810009372 111009549 810020781 810025568 810014647 111009212 111010300 111009606 810009738 810010114 810007699 810024839 810024851 810024831 810023971 810023964 810023483 810007783 810007790 111010558 111012680 111009579 810006893 111009954 111009556 810014876 111018092 111018612 111018634 810007642 810007762 810007779 111019001 810009318 810011945 810005203 810011977 810020832 810024444 810010690 111018600 810003334 111010485 111018852 111020140 810011135 810010014 111021230 111021343 111021272 810022402 810022410 111024512 111024519 810015026 810014986 810025933 810025893 810023514 810023476 810024702 810024688 810024856 810024817 810024815 810021289 810021280 810009340 111024510 111024509 111023514 111010876 111018466 111018459 111024508 810020714 810020629 111015134 810024665 810022413 810009691 810017264 810014370 810015689 810025562 810017608 810021293 810014819 810014838 810014831 810023975 810005016 810015000 810014320 810007687 111012025 810020983 810015201 810011238 810005944 810005948 810017253 810004338 111024072 810005579 810014694 810009425 810015012 810010061 810025617 810024698 810024804 111011166 810012506 111023436 111024520 810010340 810011966 810026541 810026536 111023830 111019716 111019709 111019710 810009731 810004465 810019738 810009308 810013686 111000371 111000511 111000504 111000500 111000372 111012060 111000381 111000174 111000633 111009972 810011203 810010864 810011236 810011216 810011155 810011119 810010796 810011126 810011095 111009566 111009574 111009595 111009603 111009535 111022328 111022309 810011196 111012072 111012033 111012091 111012013 111012085 111012041 111009955 111002347 810009357 810009333 810009309 810009404 810009375 111009584 810022417 810022406 810022418 111009541 810010261 111019749 111018474 810009721 111019740 111008675 111009582 111009616 111009559 111012062 810010165 111003009 810009958 111012604 810010332 111018252 810014834 111018444 810010374 111003119 810025452 111012011 111009543 111002301 810013317 111011537 810024845 111014227 111018991 111023043 111018602 111018611 111019924 810007707 810009323 810009305 111023517 810020837 810014850 810009320 810026535 810023942 810010010 810020567 810024704 810020640 810020999 810020577 810017127 111020062 111019951 810007755 810014836 810017938 810009741 810020895 810017275 111010502 810014988 810025456 810025459 810004539 111009931 810011111 810010875 810010209 810010865 810010143 810025424 111020946 111009902 810025612 111010519 810022443 810012457 810010805 810010821 810010852 810011189 111011521 810023615 810009711 810009714 810009676 810010870 810009717 810009720 810011988 810011976 810011960 810012077 810012079 810015055 810015053 810015051 810015050 810014862 810023962 810023415 810023388 111009568 111010270 111010277 111008642 111020152 111020575 111002292 111002038 111001057 810004948 810004949 111003471 810004837 111004474 810004952 810023974 810004960 810025957 810025955 810024726 810009416 111019762 810017124 810017125 810017111 810012023 810012082 810012081 111000744 111022684 111000219 111000254 111020827 810026539 810010832 111020898 810025538 810024843 810023517 810024866 111022642 810026540 810011118 810021281 111021660 810020991 111021674 810011157 111020141 810021286 111002151 810014360 810011087 810011140 810010818 111019976 111019958 810015019 810014826 810013680 810013685 111018457 111000117 111000241 810022337 810022262 810013679 810019735 810019758 810019766 111018094 111018100 111023017 810014672 810014687 810014671 111019741 111020060 810009334 810009437 810009369 810009390 810009389 810009396 810009322 810009417 810022336 111019744 810023513 810023466 111009598 111009910 111009576 111014234 810011093 810011129 810011099 810023478 810014457 810014459 810014452 111021241 111021281 810006913 810022251 810022355 810022352 810023381 810009695 810009680 810009743 111022306 111021345 111021375 810012075 810012006 810023456 810023384 810025958 810025956 810025418 111018998 810012500 810009387 810012474 810012524 111001229 810010162 810010188 810010210 101000019 111021682 111018981 111019003 111000695 101000017 111000584 810012064 810020773 810020893 101000026 810020904 111021679 111018473 111021291 111021349 810020578 810020637 810020962 111018987 111001533 810021222 810021251 810021246 810025472 810025627 810025611 810025629 810009689 810010141 810010187 810012057 810017260 810017259 810017239 810021001 810010132 810014449 810010814 810010285 810003608 810003604 810015938 810015927 810017250 810017249 810010226 111009611 810009409 810009371 810009385 810009301 810009433 810005084 810005088 810011163 810012559 810017587 111018397 111024560 111020804 111018098 111040644 111002047 111003460 111002048 111040648 111032411 111029108 111040654 111010862 111006567 111002603 111001208 111020038 111001232 111001247 111002271 111029915 111029107 111023817 111018439 111032402 111032367 111018851 111040667 111032393 111018102 111029924 111001530 111029114 111002346 111040656 111005514 111035330 111023021 111032376 111032381 111021676 111020576 111020861 111002996 111020864 111032380 111037411 111005506 111002163 111035339 111032400 111029105 111012625 111040653 111037422 111003485 111009386 111035336 111037417 111005510 111018989 111002218 111040665 111035329 111021238 111020865 111003108 111037430 111032404 111035338 111035333 111018626 111001252 111040645 111005512 111029103 111029921 111029922 111032366 111029919 111029920 111018089 111006521 111013685 111001222 111009928 111009545 111001215 111035340 111003231 111003412 111001246 111035342 111037414 111023463 111035326 111001501 111001492 111001514 111010840 111020158 111024057 111035346 111035334 111018454 111009923 111010867 111020557 111006726 111032401 111032379 111032375 111035347 111002283 111011539 111020801 111032371 111014178 111029095 111029109 111040673 810014055 810014072 810007651 810011207 111014238 111037421 111029098 111002216 111029110 111029111 111029113 111019964 111002185 111002599 111002175 111021667 111024518 111024061 111035328 111001483 111001489 111032397 111010831 111040651 111022303 111032365 111037425 111032370 111032408 111032403 111029930 111029928 111029106 111009620 111002155 111021339 111001536 111037429 111040666 111021678 111001516 111040658 111032406 111032372 111013839 111040659 111019696 111009580 111032396 111035331 111009567 111040649 111029100 111029927 111021223 810018129 810017272 810015764 810014389 810010887 810025584 111022277 111020943 111010520 111020873 810004840 810003360 810004408 111024561 111022997 111022313 810021227 810005422 111018257 810012005 810005276 111003153 810005334 810011180 111004475 810009657 810017913 111023807 111023019 810014705 810015041 810009415 111003019 810014042 810012541 810020828 111010278 810010125 810017345 111021687 810017219 111037412 111010860 111006725 111035337 111032407 111037418 111037427 111029933 111002170 111029112 111009220 111029929 111018475 111022342 810025431 111013682 111012029 111012073 111003920 810010050 111020032 111020889 111008657 111002828 111023823 111013675 810022458 111002154 810024701 111023805 810012146 111002986 810012498 810012469 111022675 810013328 810012130 111010497 111010280 810024713 810011100 810003728 810012053 810010005 810014664 810013319 810020767 111009548 810011103 810009709 810025419 810012139 810023398 810025512 810014453 810014465 810007645 810014084 810017617 111010512 111003480 810010154 810017270 810025556 810025574 810025530 810023465 810010214 810010218 810017616 810022442 810024813 810005765 810011961 810014060 111010482 810023461 810021044 810005913 810012051 111013665 111013662 111010480 810023492 810023508 810010867 111003931 111012000 810010212 111003928 810014339 810005439 111003116 111023454 810018090 810012046 810014997 810021256 111019955 810023603 111018256 810005444 810023604 111019970 810014833 810025429 810015054 111002841 111010549 810017211 111002804 810023399 111012614 111010864 810024687 810017901 810009716 810012490 111021261 810007749 810025932 810024433 810006888 810006906 810024803 111000354 111019770 111001043 111010307 111002188 111002282 810010694 111018625 810014053 810021306 810015007 810009996 111002495 810025960 111009390 810014469 111002133 111002212 810011267 810004385 810022454 810010190 810005501 810003544 810003880 810009663 111009951 111002655 810009694 810005015 810011134 810005484 810015686 810005570 810010790 810015016 810024681 111002178 810010027 111002020 810007765 810005087 111010868 810021295 810025782 810021262 810018106 810012458 810025609 810025772 810014450 810025515 111012629 111012674 810023516 810023606 810020982 810021004 810020958 810023503 810025902 810009873 810009870 810009872 810009867 810010107 810010120 810010158 810010325 111005505 111014203 111013915 111002651 111002646 111040650 111018254 111018862 111020596 111020589 111021328 111002670 111035349 111001242 111020809 111035344 111018460 111014180 111019927 111040661 810026538 810019736 810010015 111021263 810009730 810011104 111009978 810010782 810010159 810010192 810010031 810012703 810010882 810010020 810025567 111020073 810015014 810012078 111018982 111018245 111012247 111023016 111012612 111022336 810012066 810025443 810012018 810025435 810023493 810012019 810012171 810021202 810025465 111023014 111002213 810011969 111008668 111013861 111023468 111001534 810015018 111020882 810021235 810024440 810023480 810025778 810025944 810025959 810009728 810014467 810010205 810009430 810010115 810009342 810009412 810014656 810014338 810010157 810009420 810021047 810014445 810010674 810010232 111021330 810014050 111010827 111002648 111010833 810012003 111001513 111022294 111023445 111020506 111001482 810020579 810020628 810012009 111011169 111012684 810011967 810011141 111002621 810017334 810017339 810022255 810022445 111001532 111012002 810020576 810012061 810013313 810017590 810017360 810012038 810012084 111022638 810025907 810022426 810022455 810020774 111014207 111014211 111014233 810020831 810017930 810024693 810024708 810017594 810020786 810021248 111032368 810005347 111003932 810012181 111004466 810009869 111012616 810015060 810015065 810014824 810020764 810020913 810020765 810020808 111022319 111003120 111003026 810022423 810022436 111002159 810020819 810021291 111012001 111002630 810007728 810011082 810011974 810012475 111019954 810012486 111009964 111023044 810010846 810024853 810009970 810025426 111018076 111042468 111032395 111009970 111032369 111023004 111032378 111006724 111001523 810014827 111023011 111010835 111021672 810007757 810005573 111009533 810014325 111013661 111022685 111022668 111009896 111022687 810011096 810003881 810025588 810010333 111009610 111023037 111032405 111020561 111032373 111001538 111029931 111001056 111037420 111037433 111023441 111019781 111037428 111032412 111009934 111040660 111019760 111021262 111002285 810014081 111002093 111009210 111020941 810018501 111038303 810028678 111019785 111002189 810003548 111018468 111022288 111010528 111014208 810004418 111002257 111014209 810024808 810023940 810025528 810025597 810025614 810025916 810025951 810005557 810019765 810019757 810021033 810015077 810014372 111021285 810017625 810014652 111018105 111006522 111021276 810010680 111009965 810015082 810015078 810020785 111023013 111005517 111008640 810009685 111022999 810024805 111012047 111018445 111019952 810019729 810023987 810019750 810015934 111009409 111009401 810010028 111003100 810011166 111018619 810017603 111021378 810005335 810015036 111020058 810022241 810014381 111018101 810009991 810014367 111018993 810005184 111009405 810023959 111023035 111023829 111003004 810017243 810005421 810017262 810023395 810010023 810010800 111012036 810005205 111013842 810014346 111013835 111022356 111012034 810010836 810021040 810017626 111018636 111002197 111003462 810017400 810017627 810005949 810012088 111018389 810017343 810007991 111014182 111008650 810023948 111002657 111002668 111010501 111010310 111002496 810007785 810011262 810012036 810010282 111022681 810007732 111022673 810021210 810021211 810022430 810025777 810022253 111019718 111010855 810019745 111014245 810011265 810024662 810025417 111019950 810019743 810004834 810004839 810006901 810023978 810025914 810021284 810013682 810004844 810010844 810011150 810014487 810004531 810004830 810012070 111002300 111002999 810023507 810025624 810010004 810010270 810005943 810011226 810012539 810022415 810017557 111022655 810009724 111009930 111008644 111009546 111010844 111018865 111008670 111018872 810028680 810017282 810015074 111005501 111009213 111022650 810021269 810021239 810021236 810021259 810021220 810009315 810009303 810003722 810003730 810003725 810003879 810014384 810019751 111009904 111022282 111022316 810007787 810017588 810007791 810007989 810014651 111020813 810009413 810009353 810014680 810014047 810017981 810014685 810004384 810012149 810024841 810014700 810005426 810013321 810013309 111008655 810014334 810015032 810007992 810022229 111003839 810013333 810009859 810009953 810024456 111021336 810004413 111011153 810014089 111011154 810014650 810007593 111020051 810005165 111009399 810009868 810012086 810025921 810012049 810025900 111021369 810017912 810014447 111014202 111012050 111006515 810010807 810011206 810017365 810020908 111002800 111020806 810024709 111002214 111003934 810015029 111002073 810017351 810017929 111021224 810020899 810020985 810009426 810012065 810012058 810020992 810021037 810005428 810005278 111001508 111010516 111003466 111029097 111008677 111032394 111012633 111029932 111029918 810022261 810022238 810010149 810014368 810011245 810024732 111018621 111014188 111014187 810024828 810005552 810017936 810024860 111032410 810023501 111002990 111020147 810010346 810025446 810017276 111019969 111010514 810010781 810010861 111010490 111000220 111020069 111018391 810023505 111001253 810015023 111009538 111039921 810011173 810023519 810014877 810007662 810014350 111023038 810003550 810025561 810017446 810018620 810018502 810018634 810018433 810018518 810018611 810018655 810028700 810028699 810017416 810012649 810017851 810017681 810017747 810017808 810017689 810017684 810017718 111013851 111013653 810025450 810025570 810010681 111019705 111032398 111023449 111009217 111020823 111020568 111000114 111010871 111003242 111029101 810024707 111009916 810009674 810025926 810025788 111021321 810011113 111022669 810007647 111002632 810007667 810007725 810007727 111022667 111022654 810007696 810007744 810007716 810007750 810007701 810012028 810012175 810012127 111019974 111019942 810014646 111022335 810011942 810010239 810011973 111001529 810022425 810022437 810022438 810022265 810022460 111009544 111009899 111001238 810014052 810014063 810007742 111012688 810012021 810012090 810012020 810012069 111022683 111022652 111012006 810023952 111012061 111012056 810023977 810023961 111020603 111008643 111010296 810022451 810023481 810021240 111020054 111010870 111002620 111023834 810005515 111014195 810024797 111002284 111000684 810012137 810024716 810004468 111010829 111019932 111020554 810022254 111002064 111014189 111014213 111000522 111002819 111002149 111018601 810009880 810021206 810022343 810018118 810014390 810006884 111002352 111002644 111018441 111012069 111009912 111019732 111009592 111010883 810010390 810023966 111003251 810023951 810012076 810021003 810005565 111002979 111009575 111018607 111022332 810022408 111022321 111022323 111008673 111003237 810014983 810004691 810004832 111001248 111009926 111012672 111020573 810004415 810004417 111019793 810009341 810010885 810012121 111009911 810025596 810025558 810025506 810020847 810018097 810017975 810023607 111021240 111003422 111003427 111029104 111029102 111040663 111040646 111021271 111020892 111022286 111040655 111010851 111035325 111035343 111013669 111013855 111013659 111023821 111012628 111024066 810012532 810012527 111023453 111020888 111019698 810005434 111022301 111018243 111012040 810024450 810017263 810012522 810010851 111023452 111009211 810022434 810004462 810024366 810022259 810014686 810025474 810025508 810014660 111029099 111022274 111019780 111012066 111012088 111032399 111040664 111029916 810023412 810025453 810025451 111020138 810010126 810010311 111021283 810012525 810017226 810017258 810022349 111023047 810020793 810022457 810022433 810022428 810022411 810022431 810024812 810024766 810024801 111001487 810017128 810017388 810017392 810017387 810015038 810014631 810021318 810022232 810014369 810017328 810017117 810021277 810007754 810009727 810014825 810010002 810011170 810009723 810009681 810009321 810015202 810024719 111024070 111035345 111021266 111001036 111035335 810014821 810014823 810021205 810020809 810014336 810009972 810017942 111022290 810017439 810017230 111020592 810017361 111037298 810018084 111042394 111010291 111003236 810014853 810014848 810023982 810014859 810011234 111010891 111010305 111019919 810023504 810007746 810010237 810023992 810023967 810007587 810025436 810021042 810007735 111024064 111012634 810020848 810021344 810021230 810014340 810014341 111000378 111000738 810014353 810025549 810025591 810012150 810012483 111008666 111023457 111023516 111002818 111002492 111002485 111002497 111001046 111003159 111000179 111000168 111010887 111001223 111001201 111003225 810003495 111000734 111000733 111000216 810011246 810011101 111000166 111000750 111000746 111000749 810021296 810021294 111011156 810009672 810009661 810009704 810009673 810009726 810015047 810015013 810015017 810014998 810014690 810014689 111000690 111006517 111000741 111000398 111000747 111012059 111040657 111040671 111040662 111040672 111037413 810007733 111037415 810007713 810007682 810007767 111040647 111040670 111040669 111037434 111016772 810020615 111026887 111013029 810027824 111041654 111022754 111030067 111017217 810006682 111006293 111025387 111016779 111037289 111016635 810027508 111040978 810002021 111025426 111017935 810011052 111007913 111037286 111030748 111041944 810009063 810023934 810018547 810003030 810000848 810006515 810002419 810008149 111042380 111014057 810010911 111041864 111021957 111006291 810008078 810000819 810003016 111022534 810023729 810001692 111026897 111005193 810027497 111043074 111043096 111041219 810018201 111030799 111041019 810011046 111041966 111041206 810009111 810014610 111004875 111023405 111032237 111022848 111041502 111007357 111028235 111025385 810001266 810021080 810019203 111032647 111038066 111029081 810010995 810018629 111026166 810018388 810008887 111028730 810014518 111032849 810014507 111030786 111041820 810018523 810011043 810014933 111030493 111032604 810015160 810012237 111023644 810012249 810018358 111032863 810015140 810018305 810022502 111029944 111041496 111041569 111039785 111023867 810018015 810018397 111042403 111035297 111034954 810017417 111035090 111035187 111041726 810022531 810022496 111040966 111042591 810024362 111041899 810015175 111037237 111035324 810018273 810017831 111026304 810008056 810002020 810014816 111041684 810025694 810002766 810002732 111028002 111041226 810023785 111028134 111037986 111005717 111041854 111041032 111022574 810002155 810028635 810001462 111038003 111038097 810001152 810028010 111038383 111028938 111038200 111028974 810021083 810023423 810002153 111007871 810000730 111006954 111027128 111027968 810008127 111026100 111029877 111028289 810018458 810001896 111038487 111042367 111013264 810023823 111035228 810014511 810023902 111016391 810025279 810025257 810018250 810007821 810010946 810006709 111038045 111034994 810006813 111030611 111043027 810027466 810000665 810001915 111041567 810000257 810001619 111042423 111042553 111030994 810017673 111035033 810011029 111032854 111041149 111028389 111038421 810001331 810000786 810000783 111026282 111029810 810017872 111041581 810006871 111038399 111030554 111031104 111023430 111022858 111013298 111006304 111030180 810007849 810025069 111032490 111028371 111038437 111030531 810001459 111023414 810007894 111042483 810002035 810001701 810025167 810028373 111042493 111023664 111030478 810001463 111042437 810020671 111025455 111042534 810006571 111042540 111008214 111035056 111030845 111008088 111039791 111028765 111024629 810018498 810018171 810006487 111032874 810002823 810002136 810019210 111023415 111008242 810019247 810009122 810019194 111033219 111013244 810007544 810003064 810007527 111041956 111032543 111033244 111023075 111037263 810007562 111035463 111023413 810009216 111004897 111015138 111015063 111032150 111037493 111037278 810009117 111038371 111015117 111015107 810008121 810003220 810007561 111032152 111038260 111030854 111041724 111007329 810018160 111030185 111037316 111028903 111028907 810017420 111043081 111028195 111032793 111028837 111028136 111032598 810002197 111032248 810020692 111035125 111030868 111041833 111038139 111038144 111032289 810014537 111038111 111035413 111026164 111013263 111008299 111007657 111038392 111041211 810019175 111023410 111005912 111008771 810002666 111023093 111035037 810000649 810001270 111029829 111028343 810008830 111015550 111029822 111028314 810003131 111008754 111023090 111006961 111022846 111038077 111041575 111039732 111008750 111041679 111015131 810006299 111015524 111029817 111015056 111026942 111008272 111038380 111008240 111038482 111028229 111013648 111016653 111008045 810025296 810019198 111007858 111042428 810018271 810002040 111026971 111032512 810024305 111020674 111035540 111030944 810012234 111032249 111016189 810024314 111015537 111041511 810023719 111039778 810025717 810018315 111041815 810025649 111040192 810018399 111029964 810010935 810017820 111030072 111029978 111037233 111026298 810023818 111037925 810017875 111028078 810017748 111028205 111028079 111028088 111037321 111037299 111028160 111021921 111035287 111035441 111035132 111005985 111008076 111016812 111015576 111007629 111038161 810018379 111028939 111038061 810014906 111013288 111007314 111006698 810014890 111038228 111016788 810001972 111006088 111028960 111013259 810014907 111035534 111028240 111028308 111026331 111032658 111027010 810014895 111032319 111032531 810018427 111017165 810000720 111008770 111008782 111026136 111040979 810023748 810018404 111038446 111038324 111029898 111035191 111041693 111039758 111026404 810001337 810010917 111037508 111038174 111031348 111039764 111026412 810001086 810023437 810000782 111032148 810001636 111006976 111008086 111038461 111038209 111032172 810020682 810028142 111023943 810022501 111032359 111040247 810023776 111004918 810027809 111037495 810002229 810002228 111019604 810027857 111024597 111016806 111028802 111037223 810027960 810023858 111030541 111005685 810001115 111032610 111012982 810002846 111030914 810011050 810023794 111031277 111042353 111039701 111022549 810025302 810002842 810001700 810001704 111026414 810001920 111030469 111013258 111041022 111026364 810014753 111042425 111035017 111005696 111038198 810010918 111024681 111032724 810006585 810018247 111032678 810001624 111028138 810009112 111007910 111041744 111042360 111030196 810008045 111032167 111022756 810008136 810000875 111038015 810014627 111022822 810025158 810002559 111027076 111007646 111024696 810011015 810017678 810014782 111039784 810003129 111030785 810003215 111015112 111020672 111024699 810001635 111023073 810012349 111008276 810023791 111023651 111031285 810007833 111030546 111005729 810025035 111025067 111030946 111038440 111004862 111038250 810003065 111013629 111005694 111023477 111005693 111032619 111038073 111013632 111013621 111038201 810001192 111029078 810023711 111017145 111032442 111039900 111041033 810001309 111023871 810000760 810006816 111043135 111032431 111026854 111028200 111026877 111031058 111028716 810008886 111038455 111038239 111037240 111030828 111031297 111028148 111039821 111030952 111024657 111038148 111038363 111038280 111038008 810022539 111035457 111038064 111031276 111032518 111038088 810022525 111029136 810015191 111033221 810002412 111007895 111008744 111014025 111041809 111025447 111019890 111025384 111032662 111041923 810024670 810000326 810008893 111037245 810001452 111023907 111013289 810018439 810008907 111023840 810027841 810012360 111031036 111028330 810000327 111038055 111041503 111041772 810014555 111007916 810020669 810008079 111028731 111032586 111030901 810006216 810003171 810002561 111032876 111013254 810028420 810017683 111031270 111031373 111031360 111031032 810017632 111031290 111028190 111028760 111028271 111028369 111031342 111025427 111028219 111028381 111025459 111025451 111028285 111028378 810000930 810000396 810000229 111030750 810011020 111030761 111008229 111026360 111027052 810017427 111028127 810017434 111028017 810017647 111028751 111030948 810017666 810017722 810017844 111028992 111031089 810002759 111014156 111029141 111015054 810006586 111031293 111015580 810014879 810007525 810002026 111007362 111032880 111032195 810028396 111038113 111012650 810025017 111028747 111030791 810017656 111028172 111028100 810017437 810017447 810006266 810014625 810014576 810017429 810017415 810017424 810017081 111040134 810025003 111038381 810017995 111035245 810024669 111017221 810006811 111026319 810010961 810006511 810018489 810014784 111023912 810006690 111042583 111035229 111016644 810010982 111037962 810023731 111040965 810018344 111032358 111038397 111028395 810024993 810012594 111019475 810018035 111022509 111040105 111035298 111030005 810023802 810027832 111035197 111038309 111027105 111037912 810025726 810023739 111029088 111032767 111041720 111032761 111032841 810002556 810006176 111038539 111008073 111018221 111032484 810002987 111038053 111004870 111032267 810006799 111014849 810008909 810021336 111013260 111007372 810001965 810006199 111012639 111042467 810000302 111015100 111014854 111014023 111012644 111026901 111021966 810025285 111007339 810025299 111008286 111026889 111007312 810000114 111020652 111015582 810014911 810014935 111015578 111016649 111038330 111031061 810012354 810028656 111028323 111030557 810008123 111026902 810014936 111026964 111026802 111031321 111038019 111037243 111037257 111038103 111028789 111016751 111026152 111031274 810017445 810001296 810010927 111024029 111015562 111028090 111030777 111040962 111029878 111020665 810025297 111032154 810018356 111029846 111029850 810024999 111035145 111032138 111028033 810012258 111030755 111038415 111035043 111034961 810018377 111005987 111015136 111035254 810001985 111038060 111026191 111026126 111006709 111038169 111006079 111026989 111028989 810012229 111005989 111021938 810014899 810027961 810010922 810014937 810027819 111016758 810027812 111016805 810014901 111038206 810018437 111026184 111008780 111013003 810002198 111006696 810019177 111035044 111035541 111024612 111028908 111030174 810008908 111030797 111035274 111022832 810025693 810002799 810009062 111032183 111026155 111041761 810028160 111043025 810025148 111040246 810018345 111026141 810007888 810028766 111008269 111022500 111032528 111037956 111007492 810003164 111041011 111031304 111019474 111022836 111017181 810002080 111026919 111026822 111026893 810023868 810018522 810023681 810028723 111023119 810022523 810022528 810007971 111015527 810001254 810002553 810014544 810024992 810025256 810018357 810023773 810022511 111024676 111006682 111030178 810006263 111037485 810009055 810000998 111028074 810000650 111038129 810010960 810001899 810006848 111029889 810014558 111017205 810009207 111018083 111005915 810007585 111021904 810006319 111029036 111030194 111005710 111030912 111031106 810007518 111025320 810012212 111015103 111025332 810001801 810000448 111035271 810020606 111038542 111038197 810015179 810015192 810027882 111006481 111040203 810003179 111035299 810008022 111042399 111022982 111035206 810019226 810027876 810025040 111030943 11 1042620 810009047 111004952 810006545 111008061 810019229 810006809 810014939 111028912 111026207 810002832 810010912 111042586 111038353 810006324 111038431 810028741 111041729 111041653 111030821 810018373 810024365 810015193 810006275 111031015 111029023 111031310 810014509 111041611 111031043 810014812 111030510 810017836 111031034 810001308 111038343 111038026 111016794 111041651 111041708 111033229 111024028 810007524 111032428 810007576 111037977 810010991 111041871 111041500 111028995 111024630 810001567 111022544 810018256 111039825 111031080 111026199 111041523 111029053 111037259 810025045 810002269 111031330 111041713 111038527 810012264 810027864 111035226 111028217 111038290 810019164 111031026 810018243 810024309 111032495 111013252 111038181 111041626 111038050 111008797 111026846 111032709 111032599 111026280 810024395 810001784 111030195 111038241 810001797 111016760 810007862 111026353 810007930 111030740 810002612 810020653 111026972 111041620 111037970 810006497 810003133 810003074 111026875 810025669 111028975 111038131 111005993 111041800 111012668 810008029 810020625 111029035 810024416 810006227 111035114 111007906 111035427 111015094 111029038 810025697 111023130 810011010 111028966 111030544 111032725 111024671 111012799 111041780 111008244 111028721 111043032 810018467 111007927 810028639 111042385 111032296 111006492 111007266 810002833 810002502 111033231 111035000 810028691 111017206 810023884 810028666 111041008 111025310 111018258 810007914 111039731 111007900 810027997 111026172 810001005 810020684 111017186 111043130 111028801 810002019 111025438 111030856 111007335 111007486 111028223 111028095 111029993 810008059 111032856 111030765 111030933 810018264 810006877 111022758 810007564 810023904 111032641 111008260 111026112 810007887 111030808 111014888 111038232 111035034 111041208 111039932 810012247 111006973 111042424 111030470 111030198 111041951 111028988 810011036 810018206 111015547 111032682 111023395 111014826 111015081 111026339 111031047 111032162 810011080 111026217 111026159 810024406 111041580 111043018 810002947 810007877 111026146 810001573 810024385 810017731 810001310 111030911 810025656 111032795 111014862 111028741 810002353 111026789 111032201 810017703 810006514 810018656 111037302 810019172 111028021 111017195 810006730 810006750 810028717 810002348 111026828 111028167 810018534 111031309 111013276 111028130 810028673 111031084 810018519 810018635 111035136 111031333 111027990 111028367 810001691 111028276 111028347 810018621 810007558 810000927 111007867 810008125 810001248 111026132 810018636 810018488 111042390 111005893 111028781 111022568 111022559 810006214 810023824 111026915 111041179 111041165 810023790 111026939 810001643 111041155 111026394 111026959 810019174 111038424 111039820 810025163 810025166 111040215 111039772 111041926 111006078 111007654 111016394 111035321 111041002 111035115 810028630 810028623 111027957 810018440 111004899 111030537 111038259 111035296 111031086 111030992 111032733 111028786 810017841 111031081 111038285 810017654 810017676 111041673 111006719 810000648 111037937 111007648 810025055 810001568 111014058 111014046 111032471 810008137 111007311 111031322 111026977 810011025 810001006 810010976 111028298 111023870 810000721 111026930 810002506 111030840 111016810 810001193 111028069 810014607 111029156 111028108 111034982 810018024 810014613 111006090 111030526 111041891 111030543 111038481 810023733 810002200 111024649 111005891 111022548 111041741 111041627 111005712 111006980 111035516 810008096 810020675 111035156 111020670 111013267 810001960 111017149 810002157 111017167 111037329 111032657 810002140 111028120 810002087 111021919 111026205 810002043 111020676 111021973 810006743 810002763 810006751 810006734 810000651 111004883 810001802 810001450 810001121 810006231 111038211 111028785 111025343 111026411 111028393 810028139 111041875 810000536 111038439 111030859 810002093 810000556 810009065 810000235 111012646 810000319 111028281 810001626 810002045 111026791 111041595 111026095 810017458 810027962 111028331 810020677 111035469 111027123 111008775 810006868 810008067 111032635 111038101 111028072 111032785 810002038 111006968 111016399 111026963 111008221 111016214 111008219 111032230 810006702 810028151 810028368 810002031 111028102 111028150 111026909 810028742 810002300 111023500 111037261 111030999 111038189 810028731 111005898 111031048 111038054 810024401 810001453 810003291 111008755 111030440 111041533 810017824 111043129 111026313 810018382 111032452 111014018 111035052 111014140 111032142 810027999 111035142 111038369 810000475 111031355 810003079 111028344 810023438 810023422 111026359 111026405 111026290 111028816 810002355 111004945 111028112 111038349 111028920 111037503 810017700 111014844 111022748 111019897 111018074 111039735 111037304 810018643 111031362 111026409 111032741 111041173 810006817 111022541 810024307 810024988 810020691 111038018 111032727 111032190 111038276 810018513 111040096 111041925 111032522 111041941 810002301 810020681 810023447 111041934 810023753 111035086 111040244 111038348 111028865 111040250 111040118 111035021 111041706 810002354 810023787 810023625 810023454 111026980 111041940 111040241 111028806 810022541 111029139 111040974 111038087 111024701 111035083 111032542 111007282 111025071 810006825 111028115 111035127 810028762 810002209 810010999 111037285 810008919 111030502 810028663 810001149 810023433 810014802 111016400 111006967 810002287 111023434 111023098 111023401 111026413 810003290 810018308 810006563 111027099 810020663 111025347 111030517 111015554 810006929 810007799 111026170 111017292 111026844 111022561 810014800 810002617 111014895 810008155 111026853 810008915 111030813 810000668 111014039 111035089 111035008 810025154 111041494 111039929 111037513 810003174 111017187 810023722 111033227 111019889 111039895 111029945 111016397 111024621 111035064 810007922 810007897 111022868 810025129 810006178 111039897 111008743 810019178 111041860 111037254 111043103 111040220 810021339 111032337 810027468 111032563 111032548 111030177 810001454 111007342 810025716 111040984 111004924 111038233 810018466 111030909 810023804 810002142 810008900 111027024 111015521 810020690 111032779 111039893 111021961 810006676 111017185 810017814 111032715 810017734 810017648 810014776 111040217 111029046 111030991 111028800 810017838 111030894 111038484 810012584 111037228 111031112 810017751 111030936 810017832 111030433 111037480 810017811 111035522 111031110 810017641 810017815 111038217 111029962 111032817 810028668 111038109 111041859 111021912 810002143 810003235 810025016 111026932 810014560 810023771 111029803 111041535 111038042 810002296 111029965 111028307 111028296 111038507 111038136 111042621 810006936 111031097 111029956 810014567 111028915 111026300 111035444 111030780 810006919 810014551 111025406 111029957 810017871 111025449 810008060 111004921 111037945 111037519 810022520 111024050 111031088 111008783 810024667 111042577 111008745 111035458 111035234 810000759 810018666 111028707 111014147 810017714 111038023 111016388 111028003 111027129 810023852 810020698 111031275 111028210 810019154 111028819 810019183 111028319 111031095 111031292 810018043 111023079 810018196 810018462 111035286 111041768 111037917 810001449 810002307 111024022 111038116 111029910 111041154 111019618 810027481 111035009 810002362 111014021 111014150 111042572 810000907 810023777 111030778 810017694 810017085 810017680 111028734 810017719 111028215 111028181 111030738 111031109 111028728 810006182 111028737 111008757 111041669 810018277 810022507 111030525 111028709 111028711 810003047 810003069 111017153 111040982 111026976 111032719 810012261 111038196 111026811 111038476 111035447 810025015 111026187 111023955 111026988 810018472 810023813 111030904 810008095 810006842 111028179 111040191 111030811 810011079 111017151 111017194 111040129 810024995 111005686 810006742 810025712 810007867 810001333 810008020 810012643 111022503 111041803 111022555 111028874 111041770 810020569 810008151 111041621 111042547 111035195 111016752 111028755 111023637 810007574 810015183 810000261 111023141 111023086 111035509 111032687 111032799 111023085 111024023 111023901 810014572 111041516 111037229 111032827 111012649 810006509 810003062 810009098 810003046 111004929 810006841 810028412 111007377 810003031 810008062 111029992 111005692 111004895 111032757 111028341 111022564 111030508 810023654 111027088 111027958 810000905 111005921 810014504 111030795 111023067 111041680 810008152 111028745 111029073 111032810 111021909 111032451 810000657 111007898 111007890 111018232 111018213 810006274 111007658 111005916 810023911 111032254 810018495 111032258 810019196 111023850 111026346 810003216 810012638 111008225 111026150 111031040 810010994 810006283 810008014 810008148 111012658 111032233 111015575 111035443 111030745 111024627 810017879 810018268 810022504 810021852 810011019 111038401 111032565 810011007 111032616 111038263 111041838 810000188 810012587 810001118 810001180 111017101 111015553 111029043 111035220 111006684 111035317 111028779 111029892 111027107 810002508 111030603 111026223 111029842 111035412 111015822 810018456 111028273 810002288 111014884 111013640 810017728 111028332 810006498 810023624 111007639 111007649 810007979 810023740 111032538 810015132 111032866 111032855 111037282 111028384 111035094 111035110 810002018 111038117 810027474 111032710 810018175 810024372 111030011 111029906 111008241 111023872 810018385 810023658 810028714 810022521 810006234 810028387 111041718 810028670 111008273 810008845 111035446 810001446 111007625 810002746 810022515 111012647 111028853 111026114 111029966 810008126 111029891 111042551 810015155 810007837 111026967 111038313 111038333 111026790 111028710 111035169 111031343 111030621 111026093 111038253 111026142 810001181 111041521 111041583 111027038 111023945 810022540 111007864 810023735 111028237 810000996 111008277 111007636 810002156 111016781 810002752 810003052 111030858 111032145 810019147 810011066 810003123 810001222 810015134 810018033 111026410 111013283 111041035 810015142 111038367 111032802 111014032 810008047 111025342 111035002 111024602 810008053 810006543 810018422 111038393 810018482 111028290 810027958 111021940 810006590 111016392 111024031 111032351 111013306 111014041 111013239 810000463 810001971 111032840 810024320 810018543 111012979 111013274 111031025 810028126 111023956 111035277 111026925 810022538 810014505 111017159 810002757 111032214 111035405 111032772 810025024 111021949 111032732 810003035 111042617 111032605 810009086 111032493 111038269 111007280 810006744 111021950 111023066 810014891 111030759 111015104 810017671 111028382 810017649 111043073 111028183 111032170 111031282 111029002 810017862 111026128 810019184 810008917 810003026 111030460 810006226 111041577 111006094 810000068 810019230 111026386 111035507 810002358 111042379 111029145 111028283 111035285 111014883 111031323 111042465 111028402 810000883 111025315 111014842 111035316 810006253 810006193 810000880 810006186 111042573 111028262 111025428 111025053 111014911 111037484 111041642 810019206 810002416 810002555 111025379 111023475 111028735 810006233 111023502 111023480 111041608 111039806 810009090 111038470 111038207 111030203 111022820 111030457 111008048 810021342 111039743 810027484 111032818 810010950 111035073 111039889 111039726 810012644 111028872 810027834 111028793 810007819 810025272 810025132 111028932 810019246 111035104 111032791 111042375 810007974 810027972 810018197 111042411 111035419 111005920 111029085 111023880 111006942 810006823 111028815 111027979 111027003 810025131 111042351 810011032 810007532 111038308 111041796 810015163 111039937 111014056 111027044 111030450 111038027 810001618 111026941 111007502 111026882 111040131 111013292 111040113 111028752 111038231 111041827 810006679 810014579 810027964 810012257 810025006 111026926 810002513 111018235 111026367 111035140 111029830 111030199 111032851 810001447 111015046 810024402 810006832 810003217 111038432 810000173 810018368 111017144 810006308 111032694 810018640 111022505 111032771 111035433 810020686 111016645 111030788 810001904 111041810 111040193 111032853 111030013 111038418 111015526 111031296 111028294 111028158 111027992 111042362 111032850 810006521 111041715 810027460 111005990 810006534 111043098 810023751 810006329 111024651 111023839 810023843 810006256 111028888 111028304 810002547 810000753 810002554 810002552 111006311 111005983 810000761 111028364 111006073 810002474 111023116 810006200 111022759 810006260 111023101 111023095 810006259 111023113 111023122 111023105 810002088 111038205 111038316 111038106 810006691 111027972 810027811 111028152 111043133 111032875 111020653 810001914 810003022 111028724 111034953 810025022 111034986 111037973 111038021 111027970 810006207 111023072 810003018 111014040 111030535 111030521 111030533 111039801 111019615 111026366 111022990 810006738 111040208 111005902 810002943 111005730 111012666 810009024 111042433 810002764 111028131 810018650 111028051 810028736 111030825 111032614 111026108 111027967 111028155 810028734 111026107 810018457 810009208 111035430 111014838 810018638 810028720 810028654 810009119 810001966 810023830 111023851 111023420 111026946 111028211 111031033 111026885 111038020 111038138 111022821 111035303 111027047 111041955 111042386 111039924 111007643 111039759 111027998 111038411 111014913 810023665 810007886 111019605 810006241 810000496 810020608 810020687 111024054 111032297 111041785 810001455 810008030 111035243 111043016 111038158 810006829 810025260 810017738 111032730 810018179 111040956 111041612 810009135 111007331 810000361 810003233 810007919 111008767 810023455 111030458 810019162 111042417 111041501 810000255 111041158 111030201 111034972 111007868 111019886 111030766 810001538 111035138 810007969 810001200 810025640 111014054 111014062 111016793 111017193 810002028 810018028 111032208 810008024 111015135 810018658 111017179 111013627 810017692 111038041 111038115 111041682 810023770 810024369 810010899 810018390 810018380 810014564 810023817 111030485 111027074 810019165 111038149 810007797 111026827 111032467 810002032 111017169 810000660 111012230 111008307 111029153 111012240 810007898 111041629 111026924 111006706 810002092 810028688 111041703 810020667 111023878 111015072 810008835 810024410 810006244 111026868 810001089 810002347 111027103 111027091 111013266 111021928 111035528 111021923 111035519 111029008 111035166 111007293 111014821 111038118 111026969 111030550 111026330 810014615 111008212 111008220 111007488 810025011 111035171 111037988 111018084 111014880 810001628 111014897 111008793 111024039 810006255 111026123 111015125 111043134 810008066 810001202 111013004 111015813 111007912 810018535 111035039 111035180 111019614 111032869 111041006 810028015 111007861 111038094 810025140 111032705 111043131 810025164 810008089 111014042 810010895 111039718 810001576 111041015 111014865 810027871 111007291 111035252 810006562 111014893 810007517 111022829 810003221 111015808 111015821 111039708 111015809 810009039 810009049 810015159 111038519 111030816 111025398 111025466 111037251 111028146 111035054 111032523 111040240 111032660 810008154 111042354 111023426 810018010 111030628 111008230 810025048 810000130 810023708 111040146 810011031 111039935 111038213 111023979 111013249 111007301 810014815 111014861 810007556 111008310 111005996 810000055 810006710 810000305 111017201 111030520 111040995 810000445 111015122 111019620 111008069 111023970 111041594 111022753 810014926 111041221 111008267 111023474 810012353 111028419 111007650 111032168 111041639 111032487 810012351 111037985 111035208 810012588 111028981 111042387 111024666 810012664 111038190 111043014 111015806 111013011 810010896 111038493 111020660 810009131 810008104 111028873 111038429 111006721 111008071 111007855 810002826 111041557 111026913 810025145 111026864 111026332 111038009 111026156 111007349 810012268 111019898 111026301 111026285 111032243 810023642 111038385 111032722 111022566 810002753 111023859 111041960 810001146 111035128 111041588 810008807 111040957 111041633 111028005 810002796 111039733 111005190 810001516 111007271 810018446 810006484 111005184 111015817 111037488 810023897 111042589 111017214 111026866 111028066 111029844 111038126 111040108 111023074 111035076 810025265 111013287 111007296 810018469 111039912 111035172 810007907 111025421 810006499 810023839 111029146 111038400 111006712 111030443 810025264 111026140 810011018 111028810 810017845 111028387 111040197 810024316 810008890 810007906 810024360 111030002 111029976 810001544 111035225 810014912 810001622 111023908 111027122 111028996 810006519 810002621 111023977 111030879 111042568 111031332 111028833 810025683 111032527 111038025 810007860 810023664 111005900 810001236 810023801 810023747 111022859 111028842 111035261 810014530 111027073 111028813 111029909 111030900 111022552 111031269 111014051 111032353 111029951 111025391 810006184 111025458 111025415 111027025 111022856 111027027 111027118 810018372 810018359 111027120 111025382 810018541 111026862 111025324 810006264 810003093 111025311 111022851 810018455 810006236 810012214 111038305 111004873 810008043 810025682 810023721 810001334 111037969 111016803 111023102 810001145 810007913 111012642 810003067 111039918 810006805 810002768 810010938 111041731 810006495 111023942 810021860 111024572 111013037 810010916 111008295 111026832 111008289 111008281 810020697 810027873 111030449 111026899 111025435 111025333 111026931 810012343 111035255 111035151 111008082 111015566 810003006 111015518 111015810 111015558 111013040 810002785 111015142 810002619 111013269 111012665 111015564 111041161 111026904 111005880 810001539 810001575 111025326 111041027 810027818 111032649 111007345 111038268 111038005 810003126 810002560 111023433 810024419 111042492 111038240 810006581 810018480 111021954 810025695 111007360 111013240 111023124 111006095 111005885 810000664 810014878 810014917 810017720 111027993 111008275 810008076 111016398 810006213 111013039 111007891 111033223 810009092 111035210 111030917 810009033 111032717 111024625 111023121 810027868 810010988 810011045 111005711 111041822 111022744 111023849 810018411 111028415 111032300 111040960 810008911 810008902 810024420 810024370 810014920 111040111 810024396 810011034 810028391 111028990 111032320 810024404 111035203 111037303 111030749 111035521 810014770 111035162 111038085 111039907 111032559 111040196 111024030 111019619 810011049 810012653 111032272 111005875 810006706 810018365 111039771 111032287 810006556 810003232 810018004 810006257 111023930 810006317 111030187 111007298 111023861 111030436 810000826 111037281 111006687 111037527 111029867 111005719 810007537 810018327 111028251 111038509 810009116 111017146 810028123 111031018 111037528 111038405 111029001 111027114 810028347 810011051 111029079 111032322 111038465 111030479 111032500 111028861 111029040 111026870 111007490 111042395 111030542 111024598 810001330 111028302 810018245 111026884 111005908 810020612 810001314 111041791 111017111 810018533 810010913 111035143 111008768 111040114 111016650 111007660 810021858 810025041 810006938 111028766 810018539 810001799 111035273 810027480 111035479 111032570 810018448 810012238 111038203 111041945 810018194 810003167 810001627 810001906 111017155 810000825 111028178 111030954 111023846 111008248 810017883 111041789 810001183 111022745 111005903 111004902 111032675 111022572 111040119 810007557 810027496 111028269 111026325 111028778 810014808 111028812 111026363 111029082 111027022 810022545 111030802 111041730 111031295 810023808 111035411 111023660 111016637 810007535 810024675 111019607 810023840 111028226 111041512 111041643 111040998 810019144 810015129 111035159 810014894 810020711 111026900 810006211 111030853 111032263 111032692 810000846 810002292 111008223 111032624 810008100 111035077 810001007 111029782 810018527 810023886 111018223 810011061 810002507 111029812 111026333 111039777 810008891 111038202 810025068 810023442 810017857 111038377 111017215 810001178 810018475 111006075 810001235 810018145 111030924 810024994 810009110 111016186 111026399 111032257 111041528 810025258 111026888 810011062 810023643 810008085 810027462 111029051 111016387 810006501 810002791 111026843 111012243 111042416 810003096 111024585 810002509 810002468 810006531 810006517 111028388 111028303 111039818 810007804 810014531 111029936 111038520 810008057 111029080 111041547 111035536 810002090 111025445 111042356 810023434 810023743 810025290 810002964 810002624 111023894 111026276 111029015 810011076 111032714 111028708 111029895 111035103 111029012 810018492 810012636 111028164 810018483 810018508 111022536 111038425 111026995 111013010 810008052 111026420 111012999 810018540 111005077 111004931 111015137 111030915 810000731 111027083 111030814 111027016 111031042 810009134 111041012 111014843 810024398 111030499 810015190 810017732 111023862 810002944 810002657 111041000 810014556 111029161 810020705 810022514 810009100 810023772 111031114 111025346 810010931 810017704 111028083 810014794 810014807 111028043 810018417 810025063 111038334 810023829 111042430 111041648 810018151 111033226 111028355 111024056 810018038 810000311 111008063 111007352 810006674 111022546 111022842 810006698 810017817 111008925 810009210 810009072 810002962 810018286 810023841 810001901 810001536 810001897 111007495 810000842 111013031 111004858 111040149 111032828 111043080 810024991 810007975 810008048 111029902 111041901 111023987 111005876 810019161 111013232 111005883 111042523 810002160 810001579 111013262 810002205 111015540 111015532 810002613 111038443 111007300 111030481 111032466 111017197 111028052 111016780 810003045 111031072 810027858 810000308 810001631 111006708 111008245 111041932 810002497 111026981 810000788 810028665 111042466 111007872 111032545 111029011 111024600 111024605 111028124 810012222 810008837 810008916 111004907 810028692 111039817 810014574 810020709 111032806 810019221 111032504 111027028 810003076 810002945 810007882 111012992 111041872 111038264 111026863 111035186 810027852 111032554 111023227 810006794 111032592 111015572 810018510 111037921 810012598 111026839 111021929 810014609 111026395 111026137 810000655 111028093 111007289 111030916 810018396 111033232 111029833 111032417 111020678 111020667 810000270 111028326 111032285 111013633 111008776 111006691 810020703 810020674 111039768 810028143 111041878 810019204 810025676 810024321 111035091 810027978 111028070 810023691 111024046 111028784 111037933 111030186 810010997 111030945 111023071 111026092 111027955 111023104 111035465 111025436 111037295 111028896 111022570 111028107 111018066 111023114 810015185 111028248 111039923 810002359 111022869 111018067 810001903 111030539 810014921 111014014 111016178 111006948 111030616 810014904 111031115 111041873 810027965 111042402 111041889 111028349 810020605 111005991 111032796 111006944 111019623 111013041 111013625 810009080 111019871 111041632 111040092 810012269 810006705 111023103 111039739 111024690 810001295 111030171 810017888 111038224 111042478 810000288 810007822 111014885 810023676 111029937 111026391 111041529 111038192 111038494 810000495 111024658 111008231 111035085 810002609 111008224 111024672 111018065 810001574 810006218 111016778 810000174 111014114 111024673 810006265 810009083 111032436 810000264 810006855 111015541 111030608 111024663 111017213 810023767 810003075 111035025 111028847 810000997 111041531 111041723 810018289 111039914 111035428 810002955 111032777 111041781 111035294 111024631 810001508 810023915 111015083 810017859 111026983 111015127 111042366 111038036 111041710 810021341 111042622 111041520 111028056 111028034 111030165 810002743 810017713 810017670 810002954 111043026 111026840 810017746 111028225 111043125 111043126 810007869 810017638 111031093 810017645 810018041 111026322 810017655 810017826 111031331 111030445 810014754 810017750 810002965 810023833 111040245 111038416 111026387 810009051 810007567 111032665 111015569 111041739 111032606 111005697 111013273 810000877 111017168 111025424 111017182 111006681 111026306 810018330 111035026 111014043 111028006 810017414 111039819 111022567 111041532 111032301 810023755 810002022 810027506 111027982 810018311 810008065 111007338 111040963 810022499 810003043 111032602 111006943 111032783 111023974 111007321 810006536 111023975 111016785 111023961 111042404 810020700 810028705 810000515 810000557 111020656 111039800 810002956 111040224 810009132 111041573 810002620 111040987 810018157 111014115 810027989 111029999 810025061 111024686 111032179 111028949 810023706 111020657 111022826 111019875 111015545 111007494 111026318 111023868 111032569 111028203 111027040 111030512 111023431 111035508 111008278 111042388 111043132 111015064 111028126 810018546 111026356 810028675 111028305 111037938 111008290 111038486 111035176 111035163 810006504 810001570 810001514 810025696 111043010 810028686 111035265 111038453 810007808 810007847 111028050 810011075 810006313 111025321 810002084 111024670 111028987 111035291 111023496 111027059 111042520 810008051 111042432 111030434 810028732 111026188 111021962 111028841 810007838 111028942 810023788 111038407 810006806 111026110 111023899 810012224 810015157 111014130 111037923 111012640 111014116 810009023 111042349 810011011 111008079 810023685 810018654 111007632 810014620 111041877 111006958 111037280 810028704 810006191 111031268 111031291 111028220 111028114 810021081 111035030 111031271 810019158 810017441 111015108 810008899 111028213 111018219 111038430 111039781 111015560 111012801 111005914 111015804 810007873 111007626 111016386 111038374 111024691 111032617 111024695 111022515 111025369 111038001 810014526 810020654 810028128 810023451 111037910 810006217 810000749 111023135 111012970 111023402 810018143 111030065 111014162 111026303 810000991 111035514 111013016 810018612 111031315 111018222 111028329 111025057 111005994 810000069 810023909 810023925 111029039 111040235 111032839 810023922 810014534 111037305 111041786 810018451 810023856 810027836 111023139 111035311 810015148 111022823 810019224 810008141 810025691 810008107 810017812 810008082 810008130 810006570 111041590 810006523 111031019 810025298 111028169 111032540 810023694 810003127 111014133 111014829 111038006 810000823 810028119 810017754 111017112 111008293 111042539 810027509 111042473 810027492 810027465 810027458 111042504 810002623 810027842 111041515 111041765 111018224 111022573 810006867 111041853 111030804 111041777 111004876 810000876 111040242 810000750 111041811 810006248 111007336 111008053 111041699 111022553 810008922 810001251 810001307 111004867 111014899 111008779 111032852 111032744 111028919 810028006 810028004 810000540 111032173 810008805 111028009 111039762 111026375 810000516 111005919 111021942 111027961 810018400 810007541 111029802 111021926 111032642 111018220 111005698 111028928 111029907 810010892 810011030 111028045 810008832 111028322 111035116 111014139 111027020 111026938 111022563 111022524 111041657 111032674 810028744 810008901 810014799 111032229 810018623 111038248 111037955 111041020 111040230 111041930 810017736 810017695 111014868 810017693 111017219 111006303 111019609 111016642 111026315 111022996 810008144 111013303 111032882 111040140 111008758 111005073 810001782 810006579 111018230 810018185 111026324 111026101 810006555 111039702 111006979 810006522 111032457 111029144 111007869 111038512 111040252 810006276 111025069 111028366 810006252 111040101 810006876 111035232 111031009 810007863 111035126 111004934 810008833 111034988 111042369 111038162 111030787 810007857 810024996 810001191 111021913 111027097 111035200 111039727 111035223 111030815 810006542 810028367 111014837 810017743 111041760 111032318 810025036 111028197 111028732 810028646 111034984 111023971 111032740 111012656 111041751 111022518 810011012 111019874 111035150 810020670 810025057 111029839 111028386 810000267 111007907 810002298 111038188 111024687 111038306 111041645 810018444 111029831 810028001 111031346 111029852 810027987 111025344 810022530 810010959 111026795 111027094 810025652 810014580 111042440 810027464 810014550 111032651 111026393 810017856 111035268 111034999 111030906 111041974 810006821 111025456 111025318 111039940 111029888 111007622 111026861 810011026 111018072 111030631 111028993 111030760 111041617 810000434 810010932 111004877 111008263 111025431 810001964 111035263 810008099 111041801 111005978 810002364 111035194 111026220 810001644 111035537 810003083 111034963 111026204 111013237 111029137 111007903 111038498 111005731 111026216 810020707 111006091 810018304 810018294 111041928 111042427 810012332 111014015 810002294 111032278 111026803 111028340 111004894 111006692 810002146 810018198 111030489 111038479 111007896 111006086 111024620 810027815 111028295 111026336 111007322 111021917 810025672 111016791 111038299 810002952 111031067 111037494 111005720 111041950 810006836 111033225 111026986 810008143 111041001 111026418 810008892 111039824 111029912 111041506 810010928 810020658 111032136 111039706 810023763 111015118 111032529 111008788 810023718 111023489 111039749 810018434 810007563 810008888 810008103 810007900 111013027 111037504 810006767 810015158 111034978 810017675 111026830 810017816 111032362 810006752 810001693 111008308 810001518 111041231 111012981 111018070 111030437 810012245 810000227 111019616 111022520 111030435 111007893 810009046 810002551 810001268 810028109 111007279 111042481 111029059 111021959 810027502 111042552 111035406 111028846 810002961 111028845 111037996 111007327 111022522 810012346 810002356 111026096 111027985 111026094 810001252 810006541 111028031 111035270 810018432 111038294 111030736 111026312 111007866 111029006 111007353 111005890 111030805 810000171 810000325 810020655 810023783 810000129 810007921 810007800 111035013 810006592 111028957 111029893 111035123 810018316 111023985 111035011 111007500 810002421 111016648 810014886 810002361 810019148 111041023 810022519 111015111 810022512 111028791 111029983 810008904 111033239 111024626 111028898 111030817 111034967 810000933 810008929 111026842 111022988 111029904 111029987 111023106 111040969 111007897 810023697 810025644 111029876 111035221 111006092 111021925 111032670 111038133 810027820 810018301 111012645 111026286 111016800 111021930 111021911 111023481 111027981 111037914 111041844 810027872 111027956 810024409 111026320 810002352 810028356 810023752 810008120 111032299 810027874 111031055 111026288 111026801 111028370 810002836 810003181 111014906 111041659 111007368 810018295 810003228 111008081 111039954 111008087 810023917 111035181 111038436 111039943 810028127 111038067 111006705 810025673 111039719 111019884 111016775 810021087 111038225 111025399 111042571 810009036 111030452 810006822 111037241 111024668 111041029 111042409 810001272 111041004 111024645 111041190 111024610 810006582 111014126 111041016 111014030 810001616 111035213 111041152 111028397 111038536 111041685 111030625 111030524 810024393 111014900 810018317 111042347 810020706 810025273 810025657 810001640 111043123 810025709 810018548 111035269 810020619 111035178 111015812 111005700 810006240 111042407 111041858 111027991 810006307 810006303 810003039 111030181 111042501 810000502 111032175 810008101 111026175 111041186 810001979 810007572 111040958 111030918 111041794 810006525 111032328 111015044 111023855 111013622 810015196 111023980 111023946 111027997 111028380 111024623 111038119 810002754 810023441 810012593 810007851 111021965 111026350 111019612 111007318 111015552 810023935 810025030 111021971 111030882 111041749 111030529 111012663 111029160 111014892 111040195 111030877 111028978 111030789 111007363 111025305 810007820 111007371 111034990 810001092 111039903 111030834 111026831 111038295 111017176 111019872 111005194 111008068 111016188 111015525 111007637 111015507 111041797 810023695 111012652 111029979 810007827 810007839 810003160 810008808 111032153 111017177 810014581 810020666 810002503 810018167 810023863 111029950 111014033 111014020 111007901 111042536 111026416 810010943 111030989 810001780 810025147 810001868 111041555 111028808 810000503 111035464 111026388 111027121 810022503 111031337 111041498 111031352 111039756 810027482 111040142 111032304 111014834 810021854 810006553 111029042 810006505 810010956 111026124 810006280 111029942 111037296 111028159 111013033 111030439 111032448 11 1032588 111032820 111029009 810009209 111039790 111039783 111039955 111029863 111032209 111032185 810008906 810007910 810007852 111037242 111038193 111038220 111038229 111035259 810020614 111030456 111021907 111014022 111014874 111008044 111008234 111041774 810006229 111007651 111023498 111030947 810028689 111028202 810008075 111028398 111029896 111024655 111038072 111030559 111032661 111034991 111030169 111032166 111038034 111015511 111032169 111032199 111019624 111027117 810018275 810000320 111025448 111041014 810023929 810019235 111038540 810000228 810025007 111040232 810028676 111028718 111030615 111032477 111030483 810014522 111023873 111032756 810010936 111032356 111024026 111007498 810014885 111041417 111031029 810006826 111023128 810003073 111029151 810000286 810000296 810024363 111013013 111038533 111029063 810006578 810006712 810006693 111012967 810018542 111014061 810018665 111020671 810001866 111026335 111032867 111004869 111040210 810007818 111022844 810018176 111017934 111008773 111032831 111032481 111038182 810012595 810018009 111035059 111015116 810015166 111026163 810006735 810001197 111007310 111041225 111035414 111032309 111041522 111028239 111028362 111028275 111037499 111029025 810002618 111022543 810012232 810017436 111028722 111028161 810017084 111038541 111037520 111028282 111028274 810025710 810010898 810025670 810025706 810025668 111037266 111037252 111040199 111028945 111028809 111037265 111028943 111028877 111038466 111038422 111031004 810017696 111035256 111035239 111030884 111030920 111040117 111038534 111038029 111008247 810018013 111028406 810002946 111029019 810006849 810017867 111035092 810003033 810024391 810006874 111026984 111013006 111035302 111024055 111019606 111035472 111039939 810008831 111038387 111032197 111007497 111026116 111027004 810017646 810017633 810023445 111028972 810000787 111015530 111008072 111008274 111032871 111008288 810005844 810001639 111023099 111038366 810001980 111024659 810007530 111026179 111023646 111040954 111038414 111023084 111028723 111032699 111038287 111041769 111042529 111042361 111042359 111017936 810008119 111026806 111014038 810019167 810019145 111022838 111040093 111031075 111031082 111008298 111015126 810011063 111016801 111016660 810018187 111026131 810023448 111017288 111023989 111024603 111041842 810025266 111041938 810014934 111005895 111029824 111013014 111006716 111027127 111026850 111027132 111030455 111023903 810000052 111035309 810025662 810014528 111015078 810007872 810001221 111020663 111030193 810006866 111035032 111039894 111029140 810024417 810028136 111031052 111026165 111026224 111029959 111026183 810006277 111013285 810020696 111024697 111038156 810019237 111013229 810000840 111015581 810000494 810006326 810002798 111007655 111027124 810006306 111023902 111026160 810002793 111014024 810014587 111026906 810003066 111035155 810019233 111035100 111026990 111041737 810018252 111042592 111042543 111028060 111032680 111028206 810018029 111038403 810007870 810006731 111030995 111027984 810002849 111042524 810007528 810007520 810007859 810010930 810002804 111038242 810012216 810006566 111043031 111040122 111008762 111025371 810009104 810027963 111040201 810028125 111043033 810018537 111016218 111040094 111024590 810001194 810001289 111031003 111031063 111031016 810017637 111031286 111031068 111042472 111042559 810012350 810012230 111028301 111032720 810018478 810012580 810000237 111026809 111026283 111021931 810003023 810006210 810009082 111032558 810015172 111026823 111017938 111015565 810018279 111038502 111032532 810007570 111033238 111041183 111042371 111029064 111029958 111038183 111031021 810007533 111026225 810000986 111016656 111043011 111041034 111030480 810017412 810007893 111005191 111035209 111016184 810017710 111016180 111027013 111028141 810010894 111019617 111004908 111025045 111037307 111016196 111006082 111008049 111028999 111027133 111019613 111007383 111029075 810008088 810008146 111043095 111028859 810008109 111032479 111028186 111030507 810009067 111035421 111028357 111037246 810006222 111027960 111032327 111043117 111043072 810023828 111014859 111038352 810017850 111035247 111038212 111035099 810018274 810007811 111038175 111041904 810008106 111007374 111032239 111023656 111032242 111026344 111035167 111030955 111038282 111038398 111024587 810003028 810008912 111027031 111041887 810008802 111007268 810023725 111032838 111015556 111013038 111041508 810017833 810023778 111038215 810025023 810024318 111037490 111004889 810002085 111041536 111024624 111007323 810001981 810001919 111042372 111026916 810006302 810002264 111038095 111032339 810000724 111014824 111025410 810001968 810018328 810024418 111018234 111016651 111004917 810028637 111032811 111042549 111029148 111027102 810018452 810027487 111024647 111018236 111038357 810008071 111041913 111015066 111035060 111014128 810009031 810018386 810001456 111012998 111016652 810023633 111038185 111025377 111028151 111037319 111019603 111039917 111039945 111028068 111037997 111040097 810028014 111023111 111039926 810010962 111038402 111028193 810002770 111025312 810018191 810008037 111037328 810017752 111029961 810017664 810009133 810003013 810003156 810009048 810019205 111035453 810007582 111032823 111015818 810001288 111032502 810027867 111029166 810023865 111007623 810006549 111030876 810009139 111026121 111015080 810014597 111023504 810009105 810014590 111023643 111029823 111028893 810008914 111024577 111026337 111039724 810006185 810003021 111026342 810025165 810018364 111042485 111037220 810018391 810023621 111028403 810025065 111041613 810018648 810008881 111032147 111026209 111014154 111032769 111038517 111030463 111035466 111031053 111038444 810025641 810000993 810018278 111041579 111029044 111028346 810006507 810028394 810022513 111030812 810001649 111026115 111037272 111035451 810003182 111041559 111042439 810014535 810018503 111026345 810017444 111017192 111028404 810006228 111014152 111035177 111008246 810028137 111006693 111042352 111041153 111005683 810000662 111014882 111039934 111007884 111008075 111007874 111008085 111028983 810014760 810028712 111030935 111031101 111028759 111032480 111026201 111007640 111032344 810002663 111028317 111014835 111023125 111007631 111038373 111041037 810000735 111028320 810010969 810008117 111042596 111038329 111041666 810006560 111023493 810023699 111042487 111042486 810011056 111042511 111039767 810023836 810008028 810017443 111022742 111035153 111038024 111035170 111038125 810000321 111042508 810020656 810025727 111041518 111024591 111026293 810028708 111026198 810009123 810028758 810018494 111027975 111028234 810018341 810018614 810025052 111042556 810010921 111029027 111028092 111030990 810018207 810028419 111008054 810025159 111012231 111038147 111042422 810002303 111038208 810028660 111040239 810001565 111022538 810020650 810028756 111017940 111043087 111007299 810007917 810014798 111042435 810028000 810010902 810027829 810014540 111028955 810018284 111041867 111032734 111005074 111016774 111007297 810002512 111029018 111006972 111013268 111033228 810022535 810017866 810023629 111015128 111024035 111038321 111012638 810027990 111015522 810018459 810006833 111035422 810007798 810023709 111006072 111032684 111026311 111035144 111026886 810002848 810002027 810020659 111027106 810002291 111015523 810002304 111005076 111006308 810007916 111032269 810007915 810003086 111027062 111007647 810001085 111013293 111032683 111030878 111017204 111030839 111014122 810006692 111008746 111008268 111006307 810027859 111040964 810023855 810006493 810002208 111030831 111014889 111042368 810002091 810001506 111032458 111038063 810008094 111037253 111022540 810018202 111032456 111028854 810007826 111023065 810010986 810022506 111038325 111014049 810010919 111032204 810023653 111004948 111032215 810009120 111032163 810008055 111038092 810018550 111007489 111031087 810027986 111014832 111006685 111026954 111028278 111039816 810018627 111037946 111032341 111031341 111007501 810001973 810008924 810010958 810003155 111032847 111042470 111042566 810008910 111027006 810006558 111031329 111025389 111038471 111038271 810007569 810001984 111007328 111025441 111037313 111037310 810001091 810007884 111041879 111028921 111026805 111035517 111014123 810002415 810027472 111041719 810028694 111022542 111007634 111008051 810023620 111032787 111033224 111007273 111029990 810014603 111041182 111005887 111005910 810012270 810011064 111005195 111013643 810000784 810022532 810019153 111028917 111041846 810025690 111038288 111041882 111026202 810028706 810028005 111026194 810000269 111017164 111004874 810009041 111026937 111016817 810001963 111033237 111026214 111017175 111013233 111040194 810001956 111027966 810002966 810017874 810000461 810002958 810008903 111028916 810028012 810008857 810019190 111031354 810006588 810014542 111028011 111026824 111030522 111037976 111037264 810023848 111029143 810008895 111007347 111021945 810001239 810002840 111035420 810001571 111032837 111022834 111015513 111041160 111023905 810025651 810000291 111007315 111038079 111038467 810001220 111026991 111027039 810018378 111028414 810001287 111008058 111007621 111014127 111006676 810003122 810025283 810028626 111031326 111042419 810017711 111035015 111039792 810003032 810003224 111027101 111038341 111041637 111016784 810023899 111026895 111041517 111026200 810008042 111038007 810006852 810009071 111029808 111023132 810017658 810018002 111030496 111028761 111027987 111028268 111026958 111028311 810017652 111028350 111031113 111041601 810018536 810018496 810023775 111026385 111027087 111021970 810014797 111006292 111023965 810018505 810000297 810007908 111022843 810028120 111039698 111038084 111037258 111038274 111038323 810002471 111029836 111031056 111030530 810002662 111006493 810025262 810025700 810003234 111025380 111030734 810017639 111006965 111038134 810023688 810025708 810002786 810007815 111030987 111028259 810018653 810003152 111041902 810006678 111028929 810002201 810014909 111029952 111038245 810023842 111037952 810023933 810023932 810008842 111022527 111028843 111025054 111032252 111024638 111032273 810018307 810002658 810014783 810017840 810017839 810027463 111043012 810027473 810017691 810014765 810017822 810019245 111038035 810023872 111027049 810002733 810019213 810017869 810017701 810020613 810018497 810023887 810006567 111032648 810002543 111017096 111035417 111029954 810000474 810002950 111028394 111032333 111014873 111037293 111030929 111038135 111032295 810000429 111043115 810027880 111028363 810002420 111032307 810018428 111030881 810024412 111038384 111031000 810000908 810024352 111040993 111038180 810001335 111041672 810000172 810002546 111032807 111033236 111035407 111027999 111025457 111013305 810014919 111038283 810024400 111032698 111041959 810023644 111038122 810028164 111032702 810025725 111035306 111030898 810025277 111013256 810003169 810025291 810028365 111038410 810025025 810007825 810007920 111032212 810027968 111032291 111035012 111041563 810018363 810023878 810017810 111037967 111031057 111035066 111038395 111042441 111042535 810014622 111030607 111038278 810018177 111032414 810023646 111029167 810018524 111041615 111006484 111007278 111024036 810028007 111042569 111037297 111032240 810010998 111007332 111015070 111007343 810002625 810008927 111023897 111006702 810006331 810000262 810001641 111032764 810022526 111032235 111026222 111007305 810006486 810012219 810003225 111035113 111015140 111041589 111006096 111015143 810018392 810023705 111017143 111023126 111017110 111023082 111023966 111028782 810018525 111016796 111035275 810018192 810018254 111038338 111008763 111032311 111015803 111035439 111041855 111026908 810003025 111015534 111037984 111038155 810025134 111012972 111043083 111030810 111023904 111032672 111039746 111035538 810018258 111035478 810023427 111043136 111038214 111016749 111016396 111005918 810023890 111007496 810002407 111026190 111014823 111031288 810001505 810007834 810010948 810025288 111027988 810007579 810002089 810009130 111013623 111025452 111018229 111041212 111031302 111008741 810008023 111043100 111025434 111028153 111028036 810018490 111032690 810024990 111035122 111008239 111028997 111032478 111032627 111040228 111015129 810019242 111033243 810027494 810025714 111007355 111041687 111035057 111042581 810000659 810012586 810023667 111014013 810011003 111004951 810018036 111030930 810018208 810008882 810008111 111014158 111015124 810002843 111038454 111007292 810006528 810006492 111041909 111029881 111030007 111026104 810025655 111030818 810009038 111023895 810001783 111035146 111037981 810012657 111008279 810014538 111035505 111037311 810000322 810019181 111032566 111029165 111029083 111035506 111024583 111038265 810012236 810006494 111028075 111037987 111012659 810023870 111041905 810018354 810023670 810000785 111007881 111028959 810014781 810028361 810018189 111028297 111030744 111026317 111025397 111007904 111040141 111040953 111043112 111032418 111043078 111032881 111037941 111027009 810023628 111031284 810007577 810010989 810011033 810018615 111032354 111041222 111041694 111043102 111042606 810028631 810025157 111038057 111032276 111016192 810009114 111006690 810023648 111037958 111039744 810021830 111043119 111039752 111037943 810023891 810023759 810028369 111030772 111037972 111023133 810017432 111028856 111013034 111017173 111028413 810006246 111038442 810012243 111042582 111042554 111038089 111031367 111031046 810006591 111028106 111031071 111028032 111025060 810007805 111028247 111031105 111028292 111030487 111028260 111028328 111031011 810006267 810006201 810014903 111041670 111030829 111026196 810006873 111040116 111031010 111022750 810002226 111022746 111035262 111032286 810001648 810027843 810018003 111041003 810006804 111026370 111014820 111007308 111035152 810018664 111025420 111025417 810025648 111042438 111032437 810022542 111029072 810006589 111017109 810000539 111041799 111041544 111024669 810027966 111029071 111017104 111029827 111030492 111023112 810002660 111023909 810028353 111041968 111022507 810000879 111037932 111023944 111023976 810002755 111021960 111026968 111035307 111026849 111022565 111022547 810023446 111029980 111042618 111026398 810023425 111026355 111028119 111007267 111041939 810023428 111014050 111038346 111028046 111028909 111024650 111014134 111039810 111041946 111035105 111035134 111028844 111041678 111029946 810025124 111041969 111040238 111043075 111038535 111028267 810017886 810008061 810002957 111025329 810008092 111024683 111022508 111013636 810015181 111037963 111007380 810006535 111013248 111008778 810015150 111030211 810018626 111030442 111032697 111037294 111031311 111035308 111038179 111024644 810008920 810007848 810023827 111030494 111024033 111031375 810007968 111038426 111038336 111037982 111038337 111038289 111038537 111030820 111028016 111030850 111038298 810012637 111008046 111028977 111032459 810019234 111038068 111038351 111035147 111005689 810009129 810027475 111032171 810028113 810018506 111038124 111037989 111026798 111015053 810014813 810023866 810018419 111029170 810008157 111026978 111041576 111028834 810028657 810012253 111032325 111022995 111015531 111023134 111041603 111004916 111006969 111016631 111015811 111039920 111030202 810027817 111032631 810006700 111031076 111015802 111024700 111030792 111032711 810012576 111040143 810006502 111042522 810006860 810024997 810019197 111037269 810014553 810014523 111029022 111041681 111035431 810020626 810025143 810025275 111038091 810006565 111017209 111028911 111029173 111028227 111031317 810006687 810017643 111028266 111043273 111028059 111037911 111027131 810003071 111034989 111005682 810018146 111014026 810006513 111035201 111030071 111022560 810018144 111007319 111013243 111013242 111032607 111028221 111041888 810001087 111015067 111004878 111035426 111024640 111007627 111023877 111043084 111035068 111015102 111018073 111023096 111028177 111006093 111013255 810002262 810001958 111014036 111013275 111032432 810007890 810006485 111032774 111039946 111041410 111025454 111004893 111008090 111016195 111008291 111041587 810022529 810023800 111038237 810017837 111031020 810017823 111028004 810012639 111028025 111035157 111039913 111041818 810014774 810006696 810022527 111031356 111012662 810006853 810000646 111038223 810018012 810018017 111028018 810023805 111042533 111028396 111042532 111035529 111033222 810018001 111035081 111040139 111026277 111037507 111037965 111030000 111032688 810008913 111005726 810001794 810000517 810006798 111026934 111030630 111035283 111030822 810008087 111032731 810020702 111025400 810000733 111029855 111031314 111032721 111035264 810007966 111029004 810002210 111029055 810027957 810007866 810018667 810027848 111038123 111026149 111035266 810028644 111042401 111032729 111041640 810001970 810028725 810002286 111037998 111038314 111041961 111032868 810003078 810012331 810025053 111032224 810023912 810027863 111028899 810001088 111016639 810023730 111037949 810025033 111035241 810006689 111038130 111041883 111023898 111024694 111042593 111027085 111039776 810028727 111023409 810003029 810012662 810006322 111041936 111042584 111041900 810018384 810006875 111028839 111023110 111023070 810000444 111041828 810006175 111032308 111043122 111032556 111032290 111032298 111028030 111034964 111032633 111032501 111032450 810006325 111041817 111030837 810000990 810007802 111031096 810006262 810028624 810018282 111035448 111030937 111027092 111028233 810011044 111035434 111015514 810028374 111028089 111030191 111023499 111040106 111026923 111041851 810002794 111022545 810023821 810003223 111007905 111024601 111026910 111035518 111030069 111028980 111039815 810002268 111037993 111007652 810003014 111037273 111006711 810018651 810010966 111026788 111022499 111026848 111027068 810006321 111035098 111027048 111027119 111022824 810006219 810006278 111034962 111034960 111026883 111027041 111027045 111026985 111035061 111015542 111035096 810006320 111035067 111006957 111035016 111026928 111034976 111035051 111035027 111035035 111012664 111014903 111035055 111022992 111019885 111021935 111022837 810000501 111023131 111035097 111022830 111005882 111026994 111035004 111034985 111008777 111008794 111013230 111034973 111035095 111013024 111026859 111026814 111034959 810027881 810027828 810027861 111026178 111034981 111035050 111016747 111035045 111035003 810002266 111035020 111035088 111026851 111023087 111027109 111022993 810006258 810027844 810027856 810027822 810027813 810027878 111022845 111042611 810017642 810009035 810012651 111037318 810008894 111030769 810012597 810000821 111032313 810001905 810008064 810025031 810008132 810008128 111029949 111041582 111037971 810000845 111013641 111029886 810006305 810007858 111032814 111016773 810028384 810023672 810027471 810018322 810024384 111040089 810014896 810027477 111008232 111024634 810001090 111030610 111041716 810020673 111032701 111030073 111032788 810025049 810017657 111028208 810002227 111041916 810008070 810023714 810023918 810023651 810023894 810023921 810023888 810023910 810023424 810023443 810018416 810023640 111031344 810023809 810018430 810023716 810023429 810018406 111031279 810018039 810023634 111031325 810018032 810023421 810019160 111031280 810023627 810018155 810019156 810023666 111028925 111031357 111028758 111040219 810018476 810023850 810009085 111029988 810018022 810025026 111030633 810010920 111034955 810008147 111032804 111027978 111015551 111027072 111032666 111014875 111037323 111030617 810019218 111029045 111037322 111042590 111025444 111029168 111023429 810007813 111035107 111042393 111041967 810011022 111035456 111032645 810009021 111031001 111030889 810023435 111032292 810009096 111031022 810025042 810018276 810012647 111029869 111039712 111005872 111028870 810001299 810001312 111006956 810002499 111028081 810014773 111032176 111025409 111032780 810003095 810011067 111030827 810007521 111032503 810002789 111032469 810001909 111022835 111038114 810009081 111005709 111005899 111013304 111032164 111028736 810000315 810019227 111023649 111005705 111012973 111007276 111042405 111005704 111030204 810003134 111038322 111008262 111029003 111040218 810006861 810025286 810012641 111039715 111006982 810025156 111030179 111040112 111033235 111030172 810006576 810006796 111028149 111035058 111035165 111029077 810018515 810020704 111035065 111038031 111040147 810027835 111030459 111037954 111030886 111032241 810006568 810000292 810006242 111041568 810014766 111032826 810002761 111040976 111028383 111035117 810018034 810007911 810006187 810018310 111026874 111008243 111014019 111007378 810007552 810025032 111032161 810006921 810008159 810007904 810001800 111014045 111016787 111006959 111023642 111007316 810000992 111023497 111022827 111012228 810000982 111041676 111041524 111007307 111008261 111041194 111004884 810007899 810014545 111016385 810001907 111026197 810006559 111024678 111028022 111027980 810025149 810015147 111030609 810027488 810005845 810002270 111032141 810007823 111016789 111016179 111029033 111025336 111032331 810001982 111026369 810018320 111005708 810018370 111007364 810002351 810000514 810028141 810003037 810006808 111028035 111028998 810006551 111038191 111029859 111040148 111037935 810012226 111005913 111007330 111013017 810019173 810006843 810023803 810019214 111007659 810010907 111023403 810025704 111006490 810018333 810018241 111030561 810028351 111028768 810018142 810014804 810014789 111028121 111043038 810018321 810018474 111042479 810018169 111042436 810008090 810027511 111013235 810009068 111013278 111023911 111007309 111014910 111013630 111017199 111014905 111030836 111007350 810017884 111014120 810012225 111026210 810001238 810026547 810006859 810025005 810006741 111023957 111038513 111008236 111028965 810001460 111030871 810018339 810002144 111029031 810009103 111032638 111032836 111016655 810002614 810018019 111022840 111014016 810006574 810001184 111032293 810023893 111007883 111021927 111007917 111041898 111031094 111041918 111030538 111028787 810006703 111028073 111004887 111008066 810027875 810012260 810000430 810024383 810000995 810001457 810000666 810001803 810010996 810000359 810020620 111028775 810000259 111032425 810003173 810011074 810002548 111025373 111029940 810018303 111025378 111029969 111032259 810007976 111032234 111023949 111014822 810010973 111041823 111041701 111032247 111006297 111006947 111026419 111032187 810023679 111042570 111041592 810025639 111041707 810021857 810018154 810009042 111043020 111039910 111018081 111017188 111017172 111024025 111026879 810001687 111022539 111016193 111035432 111006951 810000112 111031035 111038132 111014901 810007560 111008766 111008304 111028390 810007584 111028123 111042350 810002036 111032608 111008739 111023931 111005188 810012220 810028110 810014803 810014608 111037308 111041624 111026173 810001515 810012572 111006300 810025293 810000190 111008270 810012342 111041010 810000847 111043121 111008254 111008255 810012591 111008228 810014764 810024413 810003214 810006328 810002835 810003218 111017099 810008040 810008050 810008080 810000932 111029884 111017290 810002610 111029806 810007980 810001776 810001779 111015517 111015087 111006495 111013296 111028717 111024051 810008828 810003212 111028405 111026957 810000518 810001912 810014584 111005695 111008060 111004953 111013231 111013020 810014775 810003226 810014546 111013005 111013634 111004915 111026815 810023867 810014618 111032591 111006683 111005722 111014145 111039931 810020665 111041691 810025027 111041840 111029848 111027026 810014503 111041539 810021337 111006488 111028727 111015512 111032465 111012239 111032485 810014562 810018516 111006700 111040202 810007972 810014532 111013637 111013628 111006077 810014536 111008280 810002425 111042426 111026120 810002290 810006208 111013019 810008041 810001617 810006677 111038286 810003077 111038069 810020701 111040100 111030170 111028750 810008073 111015049 810023797 810007824 111023507 810001791 810023857 810011016 111007862 111004919 810011070 111014848 111026348 111026390 111026295 810007864 111008789 111041917 810018461 111023089 810015164 111028214 111028712 111040214 111032812 111025443 111030798 111040098 111019876 111025345 111025381 111012236 111041550 111028902 111006799 111030743 111007366 111024652 111038065 111023659 111023117 111028769 111038226 111023473 111027108 111027125 111029169 111017941 111037230 111025301 111028306 111041600 810000906 111028391 810002557 111027056 111037511 111018217 111027077 810018624 810019220 111014850 111014890 810006223 111030629 810019187 111023476 111028866 111014819 111041619 111028264 111026371 810002498 111026955 810008896 810002308 111017200 111004882 111032149 810018298 810023757 111028956 810018507 810001621 810023779 810023789 111041942 111041971 810023696 810024356 111031039 111032703 111042574 810018203 111042408 810000266 111041976 111006083 810003055 111027037 111029865 111032603 111023842 111026927 111026865 111042515 111042541 810006865 111022554 111018071 111029172 111041223 111013228 111007924 810005846 810005848 810005847 111005079 111041972 111031266 111030213 111012229 111030890 810023761 111007313 810008110 111007373 810017830 111026896 810008836 111028733 810002267 111008740 111023393 111020677 111032193 810023832 111018079 111026147 810007902 111032762 111028316 111030905 810028669 810008925 111032454 111031368 111032498 111029835 111039952 111038379 111029804 111035217 111037317 111032789 810024392 111038391 810018499 111029090 810012655 810024308 111029020 810002659 810020680 810001961 111004930 111030448 810025728 810024376 111032446 111040227 810025638 111026143 111026289 111032200 111035164 111005877 111027043 810012355 111013626 810006593 810024373 810023744 810018210 111022825 810018429 111032573 810017994 810023877 111040233 111035544 810023796 111026834 111012991 810025280 810002409 810018163 111006480 111041915 111015819 111041166 810023649 111035190 111015533 111042376 111008796 111032135 810010984 111015509 111006703 111034980 810023768 111016748 111031062 810003162 810006587 810023756 810023853 111038460 111024047 810028134 111034997 111038501 111039948 111031100 111031012 810023638 111038359 111004944 111004922 111026948 111030860 111012238 810003227 111014146 111016770 111005197 111014867 810002086 810024310 111038370 111012990 111032667 111039944 810003080 810006812 111024044 111007479 111029908 111023427 111026122 111042431 111039723 810028726 111038204 111028358 810002297 810001253 111038152 810001195 810015165 111030863 810019240 111042528 810001148 111030907 810007853 111038492 111023900 111030604 810023647 111031111 111014047 810006224 810001507 810001512 111041026 111023078 111027034 810000667 111032878 810015176 111015557 111031267 810000647 111041495 111041597 111026966 111041510 810017887 111029998 810000926 111023852 810024987 111015141 111026960 810001175 810028685 111034979 810006768 111019610 111012660 111028192 111035438 810019241 111014896 810002406 111007356 810025037 111023940 111007375 111004938 810023674 810002516 810023764 111014879 111015105 111026892 111030471 111038093 111005906 111023963 111032691 111023487 810002346 111015059 810018657 111035314 111032482 111033234 810001297 111008302 810008074 111023838 810025058 111013631 111032191 111038469 111041821 810010933 111012241 111028991 810028348 111043015 810000751 111008784 111004943 111041513 810008134 810006864 111012233 111025411 111025368 111016757 111037492 111007481 111032739 111028293 111028795 111015519 111038524 111028139 111034971 810008921 111007641 810028162 111038100 111027078 111035525 810003222 810001246 810024985 111041641 111008256 810028360 111032677 111022872 111038076 810002615 111028931 810014612 111024698 810000290 810018453 810011024 810028743 111030068 810006683 111041847 111030477 810014525 810024415 111016206 810025718 111029871 810000067 810009059 111025439 111021906 810001306 111032246 111005701 810007875 111030189 111029854 111028772 111025334 111025068 111029868 111008756 111041884 810001326 810009058 111041935 810001448 111041229 111035461 111023657 111023647 810018401 111005186 111041189 810027503 111017108 111028373 810001510 111017091 111041409 111032158 111038473 810028698 111030447 111032601 111028944 111029093 111038368 810000779 111032434 111035289 111013307 810000881 111018215 111029955 111040126 111023982 111027115 111024038 111023953 810006845 111032763 111041728 111008057 111029024 111038302 810028375 810014565 810001124 810014563 111026374 111042517 810028364 111038030 111032521 810000301 810027959 810000843 111006290 111042496 111018228 111004898 810027980 111023083 810023871 810001119 810025711 111014028 111005979 111006496 111006305 810002426 111016630 810010929 111007370 111032676 810028166 111041157 111026812 111032612 111041702 111029900 810014529 810011073 111032444 111035179 810012575 111028374 810009026 810012596 810018200 810001176 810011000 111029982 810027994 810009136 111026869 810014521 810027974 810011069 111040980 111041167 111039741 111037506 810018485 111032628 111029858 111031372 111039947 111028738 810022497 111038538 810012352 111027081 111024042 810021086 810028011 810018162 111039813 111028055 111035424 111025316 810018195 111039710 111038317 111008213 111041560 111038365 111023136 111031074 111032723 111004892 810006309 111032447 111035305 810028362 810028372 111028097 810018531 810008058 810017864 111005886 111016632 111017105 111042530 111017161 111023638 810002847 111006298 111018075 111017937 810018331 111007487 810014916 111025465 111015814 111008787 810001093 111020668 810009218 810012213 810017421 810018408 111008296 111015520 810010968 111005976 810024364 111021924 810028371 810015177 111035018 111023077 810001265 111007333 810025719 111039956 111039811 111006483 111032832 810008114 810000395 810019149 810002196 111004885 111029007 810001122 111032829 810023762 111008792 111006312 111012996 810010942 810021338 111023641 810018329 810017993 111004871 111027033 111038235 111042477 111026974 810018270 810017819 111043034 111028749 111032571 111008083 810018326 111014887 810012585 111028867 111032822 111037926 111022558 111037957 810003060 810018625 111037517 111026343 810018617 810028655 810028729 111035102 111026154 810028667 111026139 111037279 810028682 111026117 810018644 111031028 810021084 111028013 111031361 810019186 810028701 810018670 810028636 810018412 111038243 810001634 810008018 810018449 111028729 810001630 111029087 810018471 810018008 810000934 111006704 810023908 111041911 111038176 810025681 111012235 111028315 111023662 810011009 810020603 810023675 810012659 810010905 111026211 111026961 111014060 810023838 111016208 111016194 810023746 111032360 111034977 810025004 111038441 810001464 810001336 810018188 111030770 111028725 111025430 111006713 810001696 111007326 111030864 111013644 111030742 111030847 111030851 111030925 111008043 810027969 111035119 111032473 111035436 810015173 810027489 111026807 111026797 111037950 810002199 111032695 810006251 111017107 810025645 810024315 111038252 810002139 111040216 111042513 111032835 810019228 111026109 111039808 111035041 111041412 111041957 111028012 111026296 111037287 111028170 111029061 111026185 111031294 111038483 111017178 810002141 111024020 810001642 111041910 810018334 111039892 111028982 111028979 810018251 111028411 111024611 111039807 111022517 111026826 810023724 111025055 111029029 111015123 111023847 810017431 111030938 810020643 111029813 111028894 111041922 810028733 810028421 810017698 810028377 810028418 810028370 111005075 111032865 111006677 111004925 111032438 111032768 810002138 111037964 810020689 810001697 111035230 810019179 111026837 111008786 111007920 111015077 111041937 810007578 810023754 111021920 111031044 810022536 111038362 111008064 810001689 810007519 810002230 111038177 111032679 810014889 111014136 111032332 111026206 810001796 810018517 810000312 111027075 111030064 810009126 810006537 111027973 111040222 111024688 111024034 111025049 111023856 810007855 111023492 111033218 810023436 810025138 111005699 111017191 111022857 111029092 111025335 111039786 111032824 111030747 111030809 810001519 111005986 111023858 111041585 810008072 810017835 810014806 810018409 810018027 810017699 111024609 810002078 810011023 111041187 810017413 111035440 111025395 111030735 111014863 810000645 111016627 111035512 111014149 111027963 111028026 111026818 810017726 810018389 810020685 810023905 810018021 810018393 111030768 810023700 810018450 810020618 111028280 810018018 111028799 11 1028385 111031287 111031349 111028840 111028948 111031066 810018299 810018421 111028805 810018487 111028904 111028803 111028236 810018159 111028889 111028324 810000433 111026403 810000313 111016190 111028249 111014858 111006695 111034998 111042602 810018037 810006221 111030846 111013028 111037978 111029872 111016191 111016753 111032283 111038350 111016221 810000663 810009032 810018619 111012977 111023422 111006966 111028936 111007485 111040968 810017827 810009107 111021948 810028135 810025255 111023142 810025136 810020651 810023449 111041663 810001969 810010915 111041735 810012600 810017848 111039890 810025679 111041170 810012335 111025046 111007341 111024664 810025671 111026221 111032830 111040115 111034993 810028745 810025647 810002305 111023417 111016216 810002023 111033233 810014570 111030190 111038105 111038347 111041172 111021943 111032427 111032560 111041652 111022757 111037255 111016814 111026307 111005917 810012242 810025699 111013291 810017813 111026933 810023765 111005889 111041184 111041970 111029938 810003015 810000500 111037919 810027850 111035078 810018484 111038081 111030892 810018445 111030870 111022850 111024049 111023857 810023874 111016795 111038419 111020662 810007829 111028241 111026905 111018085 111039919 810020662 111023962 810014601 111016808 111022533 111022501 111022504 111022528 111035022 810024989 111040132 111007284 111008742 111024606 111032821 810007874 111006482 810008898 111015076 111040229 111024024 810006807 111032430 111016207 111016215 810018025 810027827 111024653 111028140 810018205 810027870 111028122 810027838 111028174 111037516 111039780 111004910 111035063 111034975 810018152 810018269 111007491 111038462 111028788 111026151 810025701 810001182 810002365 111024685 111028748 810018371 810009095 111017152 810023769 111030173 111028968 111029174 111037515 111018225 810028661 111017156 111027064 810008054 810023750 111032515 111024675 810025066 111032282 111008253 810024317 810024387 111032618 111025322 111024646 111025376 810012359 111030446 111019883 810025666 810009125 111006074 810001332 810003081 810001199 111023108 111026158 111023960 111023107 111032668 111041021 111043024 111029807 810007896 111026987 111022551 111040133 111013639 111043105 111041671 111035511 111025331 810006315 111014142 810017737 111023398 111029155 810000878 810010954 810024355 810006857 810007903 810017707 111026278 111043090 111035047 810000431 810002306 810012221 111028954 810014938 810027810 111030926 111014143 111006953 810018193 111005884 111023408 111040961 810001285 810009079 810000652 810009091 810006554 111035211 111029815 111041795 111030536 111022865 810017881 111004896 111015084 810009050 810023682 111023881 111024592 810012337 810027996 111041526 111042414 810002025 111028924 111041782 111038164 111043092 111032736 111030540 810011013 810003229 111024582 810018668 810003176 810003124 111037226 810028640 810006837 111039765 111032615 810003038 111041591 111023068 810007918 810020693 111029162 111013030 810027862 810003009 810027849 810023879 111022516 111022513 111022571 111022556 111019473 111039711 111008215 111008790 111022521 111030753 111030941 111041209 111032544 111016212 810006518 111031090 810006314 810025721 810017659 111027036 111032577 111037496 111032551 810009084 810011041 111027082 810007865 111040103 111032685 810018170 810018302 810019169 810023432 810023931 111029985 810023622 810023631 810023892 810023439 810023914 810023907 810019189 810023898 810023431 810023641 810018045 810023895 810023927 111031328 810018335 111028798 111041159 810021085 810018362 111028245 810019166 111028950 111028827 111031370 810018661 810000239 810025146 111028914 810002076 810018204 111038011 810018253 810021079 810019217 111019894 111028851 111038490 810018470 810023819 111032587 111040996 111029840 111024575 111026340 111012798 111038022 111032797 111030532 111026213 111032846 810006180 810017083 111004926 111037270 810001959 810006923 810002845 111029984 111030899 810006573 111041736 810014533 810027967 810006732 810023430 111028780 111037217 111028790 111034974 111028797 111030934 810010985 111025059 111025317 111023655 111018068 111028879 111038168 111018064 111022576 810010909 111028416 111028010 111025413 111023865 810017665 810014624 111006955 111008772 111030793 810015162 111025403 111037325 111017939 111017933 810010980 810018374 810002788 111017931 111032555 111027976 810002756 810022534 111041614 810023749 810025664 111039748 810023846 810018293 111028963 810019243 111035442 111038500 810003130 810028349 810006496 111043111 111038451 111038254 111041865 111006310 810017829 810017852 810014547 810014549 810028711 111026329 810018183 111006981 111014857 111025450 810019168 810007516 810023826 111038143 111021908 111024639 111030624 111028242 111004906 810027976 810020642 810018249 111037999 111033125 111034969 111032580 111032578 111038310 111038004 810025715 111035476 810008039 111041893 810014757 810017724 810018283 111023958 111026793 111026841 111024604 810001804 810014923 810001775 111028952 810018173 111024636 810001790 810003003 810028693 810028687 111013007 111032165 111017203 810000644 111015563 810014888 111026326 111041210 111026377 111041683 111026215 111028353 111028392 810001645 810023834 111006983 111007911 111008078 810007928 111014902 810002824 810023807 810002626 810000304 111041605 111008238 111026338 111038267 810012661 810010993 111008751 111031049 810023673 810001913 111041931 111015539 810027970 810008093 111029967 111032453 111029971 111025300 111038499 810000824 810018423 810007801 111028257 111021955 111022526 111028906 111028967 111007348 810008930 111021951 810003132 111014907 111027002 111029870 810003059 810011035 810017880 111027054 111026833 111028379 111039740 810001541 810028118 810000397 111015516 111022529 111004861 810002963 810025713 810012250 810001224 111020673 111028015 810023837 111004905 810001513 111012995 111004923 111007274 111016786 810006932 111021947 111032589 111021932 111021918 810002837 810008015 810002838 810006530 111005728 810001250 111028027 810023811 111005687 111041009 111038300 111028243 810028645 111042384 111037930 810014767 111032800 810006330 111029157 111029021 810006524 111032634 810018042 810019239 810017873 111035284 111041874 810018366 111028287 111014118 810006279 111038013 111038279 111035527 810009078 810009106 111041664 111038102 111030497 111026327 111007880 111034992 810006516 111032205 810027499 810020622 111029963 810020611 810018161 111039742 111016210 810002081 111015060 111032862 810027504 111038511 111038543 111038080 111030503 810025663 111032174 111038039 111015132 111030794 111034983 111015045 111026825 810001517 111023968 810006529 111019608 810001976 111015085 111025302 111005072 111019891 810001983 111032728 111038459 111006071 111016654 810028003 810014586 111016390 810007551 111039930 111035031 810019225 111017154 810024673 810000538 111039775 810024407 810018287 111037521 810009108 810009109 111028829 111028145 111041890 810018403 111030842 111029152 111014881 111025442 111032288 111032329 810023426 810001778 111043082 111032324 111032595 111043076 111031030 111038433 111006952 111020666 111041868 111035272 810018662 111031069 111030751 111032786 111042608 111042497 111031283 111030453 111031335 111028222 111028333 810019238 111035513 111028171 111028336 111041218 111041912 810018168 810008923 810023684 111041527 810014883 111032194 810014811 111041574 810017723 111007340 111028773 111004863 111007645 111013645 810000271 810000254 810025680 111040990 111028922 111025404 111028162 111025463 111032255 111015544 111016181 111038389 810001116 111012661 111041625 810018528 810010957 111030872 111015086 810003070 111030910 810001313 111012986 111017100 111029060 810000309 111028984 111035231 810017849 111028348 810015161 111030737 111035224 810025289 810025303 111013238 111035435 111029034 111035074 810000497 111023118 111004912 810006323 111006294 111024692 111006680 111037929 111035526 810003041 111006701 111028418 111006085 111035276 810014930 111005879 111013236 111021963 111026316 111035481 111030775 111035249 111030527 111013302 111035320 111007919 111013250 111038199 111021937 111026970 111016819 111026943 111016807 810025263 111041525 111042579 111038052 111042615 111038335 810028633 111015058 111037927 810027860 810009056 111015508 111024693 810007909 111040204 111023845 111030865 111028142 111028256 111026804 810001190 111027032 810017730 111028325 111027986 111031273 111025330 111028024 810006753 111028973 111029150 111030491 810002959 111042576 810007559 810018641 810017674 111014852 111025308 111032630 111029026 810006926 810001638 111038128 810024313 111026180 810006818 111004927 810009070 111037222 111022510 111039814 810028147 111037300 810001009 111027983 810009088 810009099 111005714 111039796 111017932 111027974 111037920 111038477 810014527 810020646 810006740 111026098 810006935 111008252 810008838 111032845 111016809 111013624 810006205 111029939 111041920 810023650 111023954 111030509 111030553 111030632 810006301 810007850 111025065 810001247 810006209 810009043 111030482 810001196 111007894 111026402 111007304 111030551 111030486 111038142 111041616 111040251 111040209 810012578 810028107 111015823 810027512 810006814 810001204 111030921 111035023 810018410 111040221 810028703 810028751 111029873 111031307 810011058 810017418 111038326 810028124 111041863 810018259 111015548 810006801 810028750 111028053 111042400 111026890 810024668 111032639 111042391 810025276 810003061 111030862 111030556 111032653 810002363 111013272 111027051 111041787 111013247 111029138 111029014 810025171 111030754 810023881 111030498 111027057 111038356 810023661 810024671 111030614 810025059 111035049 111031316 810008153 111015528 111023425 810018442 810025678 810025707 810006332 111038071 111041175 111032582 810025001 111041677 111025070 111041841 111032312 111029154 111028757 810007924 810022524 111041688 810002044 111023423 111028884 810023885 111008265 111030833 111035237 111028940 111041499 810010903 111004946 111005973 111005975 111028753 111026936 111012653 111042560 111030516 111025051 810007523 111023137 810023742 810007814 111008282 111032686 111023645 111039769 810028108 111037944 111038521 111032784 111021946 111031318 111007295 111034957 111015088 111041005 810002202 111042604 111023875 111030500 111041826 111040152 111032596 111043097 111032520 111032137 810012656 111035418 111032511 111032508 810010992 111035455 111030451 810000752 111015573 111028882 111035425 111030175 111035006 111024633 810006834 111028144 111035139 111035293 111042545 810009121 111035168 111029901 810018376 810014801 810000115 111005716 111032160 111024588 810001625 810028738 111004859 111032251 111007870 111023866 810025002 111007483 111007480 111023969 111026898 111016759 810022498 111018231 810022544 111025433 111042357 810025137 810022517 111028783 111025386 111007918 111014160 111014871 111028901 111008047 810025038 111025453 111013022 111004864 810003236 810003231 810028664 111023491 810011028 810010944 111023952 111041636 111025402 810006851 111032524 111041927 810018646 111042382 111035196 810002154 810002760 111026876 111030891 810002608 810019201 111038382 810002787 810027816 111028868 111032553 111028832 111035242 111032244 810025144 810008021 111024622 111032271 810007580 111026105 111030003 810001458 111028254 111004914 810001685 810001537 111038210 810024353 810000882 111025372 810006215 810006237 111028971 111023505 111031327 111031363 111035189 111037250 111037277 111037224 111014846 111031023 810019163 111031281 111028970 111042491 810002500 810002473 111014833 111032486 111031308 111028157 111027113 810002467 111029801 111016203 111035219 810009066 810018647 810028111 111038098 111030619 111038375 111015069 111038058 810023880 810014763 111035019 111032693 111026381 810006844 810018337 810023703 111008294 111014159 111041415 810002295 111038360 111028080 810012579 111035062 810009052 111023127 111026193 810001338 111005192 111016790 810006802 111016798 111042567 111025340 111037234 111037510 111026396 111041191 111028400 111030515 810000653 111028811 111027014 111041783 111028986 111021968 111016211 111005189 111039911 810008135 810011004 111025467 111028058 111028185 111042514 810017729 111030988 111027977 111042503 111032262 111032825 111030931 111042527 111042506 111028110 111023879 111028351 111025052 111025323 810017668 810017679 111030620 111028156 111028313 111013245 111028218 810018300 111038074 111038420 810006736 111037953 810028662 810002505 111023484 810003168 111007303 111028933 111028399 111025401 810014928 111028116 111037931 111037301 810002042 810025152 810024368 111038150 111023088 111032584 111028754 810027983 111032669 810002152 810012259 810023678 111039737 111029960 111007892 111022987 111016658 111028309 810028392 111026857 111028103 810019202 111032696 111014141 111029091 111032264 111038545 810001445 810023786 111038340 810002360 810025141 111041656 810001917 111005980 111026208 810006920 111004860 810010908 111008292 111027007 810010940 111042516 111032819 810027979 111030903 111024660 810017682 810018199 810028748 111022743 111039916 810012599 111030466 810008840 111041775 111030464 810000929 810007892 111040975 810006795 111038159 810008031 111031079 810023710 810024397 111038448 111038355 111032726 111008089 111004936 111029981 810018006 111039950 111013253 111026817 111038428 111037915 810023798 810023810 810001781 111006087 111040099 111038059 810006488 810000477 111022852 111007285 111032636 810002742 111024045 111035278 810007901 111029030 111028014 810003024 111026992 111026860 810003008 810008815 111005688 810014932 810027467 111038070 111026397 111007351 111024041 111024667 810012652 111022841 111028037 810003178 111029834 810014931 111038311 111027053 111006945 111016197 111024661 111004880 111027079 111032435 111032334 111014031 111032363 810014596 810024411 111035319 111016638 810028130 111042548 111018218 111027000 810010983 111008250 111013642 810001305 111043030 810017709 111032713 111032488 111029054 810024361 810001916 810001977 111035258 111016777 111032549 111018080 111006964 111042521 111013650 810025684 111032228 111008258 111037231 111040999 810014942 111032815 810018631 111015535 111038002 111043128 111025061 111007269 111035111 111022985 810008806 810023704 111038284 111028143 111030885 111031324 810003135 810025010 810000728 111028071 111030511 111028720 810012211 111038376 111041808 810007905 111024632 111037514 111038458 111038195 810024676 810025050 810006247 111005684 810027869 111028265 111041832 810027883 111004954 111024641 111030919 111013270 111035313 111022828 810006190 111022855 111030558 111033217 810028163 111007473 810025043 111004891 810025013 111030940 111041411 111037276 810025261 111030206 111032755 111028794 111030873 111028253 810012333 111032600 810003082 810012665 810006878 111023864 111027005 810006533 810003004 810002145 810014516 810018464 111032216 111026372 111038166 111041217 111026195 810014902 111032873 111040090 810017733 810017721 810003128 810006316 111028763 111030819 111032843 111032314 111037239 111030922 111004947 810018491 810006695 111032475 810014508 810025054 111041805 111041897 111041705 810014568 111039747 111040212 810028354 111023951 810017739 810014814 111029853 111021952 111016662 111027130 111016776 810002784 111022530 111033240 111028875 810025021 111025429 111032803 810001690 810020602 810007832 810018172 111039798 810003154 810008086 111033220 111028719 111023848 111028746 111026157 111039909 810002616 810012262 111004886 810011006 111032279 810018291 111029825 111027061 810008829 111039949 111035108 111029975 111012654 111023983 111028857 111028838 810019244 111035410 111028001 111016763 810006557 810003072 810006675 111007381 810001967 111005974 810001908 111021936 810028416 111026907 810003068 810014600 111038452 111028869 111039927 810011071 810023906 111035246 810028161 111032681 111026799 810002293 111030826 810014623 111028890 810021340 111028041 810000729 810017861 111017289 810028721 111014853 111005981 111015571 111023640 111015570 111019621 810027837 111026872 111028113 810018659 111028777 810028628 111038249 810000984 111032516 111037922 111007638 810006561 111030192 111041644 111026145 111004872 111013281 111012975 111014909 810002558 810006188 111006497 810023723 111017196 810018280 111008092 810008804 111017094 111013299 111014877 111007382 111013284 111013280 111020655 111026361 111041195 111040225 111039750 111017210 111005715 111030528 111041176 111015079 111015101 810002515 810025133 111026389 810025661 111005874 111014872 810006575 111041727 111041628 111040249 111030767 111039755 111042364 111026952 111029089 111042370 111038247 111038275 810024390 111038141 810002734 810007977 810002736 810010975 111040985 111025392 111023412 111032499 111006720 111023391 810025653 111030182 810014790 111032537 111028082 111035112 111030830 810006520 111041571 810006824 810006225 810000727 810006713 111015047 111038104 111014886 111022994 111022863 111004911 810012660 111041862 111038342 810014583 111035471 810009220 111042598 111028252 111029037 111022870 810000232 111024574 810009206 111040970 111029135 111041030 111037502 810025130 810001688 111018069 810008142 810012239 810002079 111024580 810008083 111017222 111025437 111025423 810006803 810018398 810025720 111022747 111016641 111007325 810003010 111035148 810007843 111039951 810022516 111024027 111037928 810007881 111026379 111007914 810001686 810007878 111026820 111007630 111021944 111015549 111022502 810000446 111007633 810025294 810025686 810002771 111014155 111023959 111041908 810010972 111032884 810006333 111028994 810018426 111035188 111030163 810015149 111028962 111025432 111037483 810007536 810010906 111041762 111030534 111026203 111041181 810014519 111035109 111032622 810015178 111026138 810006510 111023988 111023973 810003040 111028291 810006538 810023663 810009102 111030006 111041572 111038165 111037256 810019146 111013297 111032178 111023140 111013035 111025446 111031005 810018361 810027991 111023981 111032302 810007830 810010964 810003019 810003042 810006550 810018141 810012645 111028923 111038408 111040994 810025126 111032491 111025306 111029147 111037947 111013234 810001273 810023873 111037501 111008062 111029805 810007554 810001975 810014795 810018186 111030467 111028201 111026102 111038491 111039933 810023781 111024578 111032476 111038218 111041025 111030549 111042588 111025303 111035133 810006204 111025338 111023064 111026287 810002423 111038273 810006544 810006846 111028774 810021078 810028413 810006686 111029159 810010891 810018460 810012227 810023854 810012235 810018544 111037291 111041534 111027969 810018639 810009030 111035240 111038266 111007928 111039942 111008266 111023418 111030986 111004928 111040120 111018237 810028122 111042561 111026167 810028359 810027807 111030514 810025269 810018265 810024399 111037924 111039782 810018383 111038246 111030807 111029049 111038194 810006733 810002951 111038394 810023793 111029086 111032310 810014761 111026881 111037505 111041607 810002037 111021972 810006281 111014864 111029076 111025394 810014585 111032507 810025008 111041885 111014135 111017190 111016634 111032650 111014137 810007812 111005718 111032861 111016389 111032513 111007860 810002790 111028807 111031059 111025058 111006714 810000459 810002408 810009087 111035233 810000447 810008102 111035250 810020678 111030849 111023947 111032870 111023964 810000762 810023774 111037983 111028937 111031358 111029974 111041416 111023984 810006212 810000839 111030875 111025339 810019152 111035198 810002501 810000844 111028352 810002544 810006203 111042498 111027084 810006181 111040972 810018332 111035121 111017198 111038075 111007885 111014055 111015121 810027459 111025405 111028417 111014048 111017158 810020672 111031278 111031305 810028352 810027498 810025009 111035070 810001274 810025160 810017716 111028897 111042476 111031054 810008858 111026858 111030949 810003219 810011001 810011059 810006856 810003157 111028764 111008222 810001793 810000909 111030913 111035484 111032643 810019231 810007912 111032472 111022741 111007361 111013025 810001328 810009124 111022989 111032159 111035075 810002664 111035312 111041658 111041921 810002261 111018077 111028101 111024677 111032189 111030932 111041869 810015128 810002549 111016393 111041906 111039773 810009219 810025643 111023972 810023656 111026996 111028881 111041790 111028061 111040102 111035193 810014513 810025642 810007550 111037942 111029913 111035214 111030164 810006238 111014017 111028828 111024579 111012667 810011005 111016636 111015115 810011048 810018530 111016811 111041947 111032808 810003175 810006577 810018014 111041880 111030771 810002469 111041798 810009027 810007568 111025460 111042484 111030784 810000287 111007359 111023837 111014856 810002851 810006174 111032203 111038028 111008056 111041593 111018063 111032760 111006679 111016746 111037512 111014839 810003180 810015156 111042373 810027510 111026894 111021916 111007875 111025412 810025284 111035539 810001703 810001647 111030998 111005977 810008809 111029977 810006707 810025062 810012650 111041214 810012663 810025060 810014915 111032742 111029816 111033242 810009025 810023639 810006850 810009028 810018246 810027831 111032621 111025440 810012646 810012251 111029819 810014591 810009060 810014575 111006487 111039705 111035467 111008283 810007816 111026176 111032280 111007923 111038296 810018473 810000240 810018309 111041866 111024589 111041894 111041013 111015567 111015092 111037992 810011072 111032489 111015095 810025162 111038049 111030861 810018165 111032223 111042510 111027055 810028156 810025274 111035282 111037248 810001957 111007886 111038107 111032883 810017855 810015137 810002782 111032877 111042475 810000542 111038485 810008105 111025050 111039899 111032238 111027030 111035135 810001325 111037961 111032494 111041530 111028726 111026973 111037960 111006718 111004933 810020710 111035477 810028112 111040207 111037274 810018020 810025170 810025161 810000994 810008883 111017106 111023488 111041192 111017162 111040138 111040234 111026845 810028681 810027479 810019209 111040226 810028363 111014117 111040986 810000828 111038048 111023896 111027017 111026417 111032143 810015152 111007503 111012994 111022749 111032792 111031031 111028864 111038096 810009044 810002841 111019892 111023486 111023400 810014515 810020694 111040200 810012345 810012640 111028231 111032671 810014924 111035462 810027998 111016640 111023503 111027135 111028892 111038108 111008765 111024573 111026328 810018342 810028153 810003007 111024021 111038258 810017423 111042421 111041814 810011039 810011054 111030462 111016797 111040988 810017876 111012980 111012974 111028111 111004866 810002422 111035279 111006798 111023129 111008791 810011068 111029832 111035101 111041806 810001177 111028338 111038361 810015153 111038396 111026308 111028913 111026118 810001798 810002034 111026880 111026829 111026999 111025319 111026415 111026383 111026852 111028188 111030908 111028209 111025325 810017753 111026182 111028084 111028173 111028163 111042546 111042555 111042558 111028207 111042500 111028168 111028064 111042603 810017635 111042474 111028047 111028019 111042544 111041895 111028109 810017636 111031359 111031065 111042507 111042495 111028028 111031306 111042595 111042420 111030997 810017640 111042518 810017660 111042612 111021956 111039779 111035292 810025304 810019195 111017160 111022562 111005894 111022535 111041168 810008138 111040243 810015184 111037235 810018266 810001004 111035470 111038037 810023726 111038515 810025150 111014121 111007475 111022867 810019176 111030505 111022864 111039953 111028335 111029883 810020657 810006827 810008036 111027046 810012218 810018369 810006202 810014778 810008118 111025062 810027469 111026134 810006839 810023632 111032547 111026821 810018375 111022506 111042434 111033230 810008077 810023668 111041813 111042406 111043021 810027830 111029809 111029973 111041542 810027992 111038508 111006710 111008309 111038463 111041750 111041767 810028117 111032706 810028378 111031319 111040205 111032330 111025066 111015133 111015538 111007344 810001540 111032497 111030950 810001117 810000258 111007281 111008067 111028376 810014506 111024599 111041661 111014891 111023482 111030473 111007888 111013032 111017220 111040124 810003051 810018267 111032533 111035106 111031312 111038154 810003084 111035515 111035323 111035014 111035257 111042564 111038409 111037327 111027065 111024689 111012987 810028155 111037936 810008889 111029996 111027110 111012804 111032654 111042509 810015174 111029142 111042471 111035185 111042525 810007555 810006206 810018343 810007553 111037974 810006235 810002795 111016756 111031351 810001795 111019611 810008019 111028029 111005183 111015099 810001298 111019878 810000535 111022751 111017166 111015110 810027846 111039730 111027015 111026835 111027089 111022871 111023094 111016198 111014132 111023091 111012643 111014113 111023138 111038445 111028739 810017686 111031103 810017651 111035520 810012267 810023677 111028321 111031339 111041792 810025687 111041753 111006984 111028821 111028166 111016755 111040253 111021953 810001902 810001900 810001286 810024319 111038082 810000988 810012246 111013002 111005909 111032539 810007543 111040150 111028825 111008259 810018663 111028132 810018669 810018454 111025307 810018652 111037247 810018340 810018431 111028230 111023653 810000314 111028354 111005182 810018532 111038256 810018297 111037320 810017442 111038301 810028672 111028337 810018481 810012256 111028086 810008011 810009089 111032758 111041164 111017218 810001629 111035084 111023853 111017095 111028368 111023941 111034968 111028044 111027995 111028129 111023950 111008091 111028375 810007534 111005724 111041964 810007538 111015579 810025047 111031366 111030888 810028697 810015171 810023792 810018511 111038234 810009093 810019155 810012228 810014594 810014573 111021964 111032156 111032629 111008264 111041743 111028049 810001000 111012976 810000980 111023069 111012983 111006699 810012583 111017090 810006931 111015073 810006928 111030506 111008218 111017184 111013042 810019199 810015167 810006569 111035227 111041852 111028165 810021335 810018323 810014882 111041725 111028360 111032585 111015574 111027080 111040123 111041228 111041870 111022833 111019881 111029814 111005995 111028927 810000722 111032213 111013300 810003137 111008080 111032348 810006819 111029997 111020649 111020659 111020654 810007565 810024403 111026357 111037959 111029052 111014840 810028381 111035542 111032737 810018447 810008131 810000460 810001632 111022847 810002758 111008050 111041213 111041635 111032716 111041171 111016217 111007365 111018227 111012242 810002203 111014131 111004920 111032303 111029986 111035260 810010945 111032620 810012217 810010979 111028312 111032514 111028279 810012357 810003057 810001898 810006766 810018628 111023421 111039901 810014602 810025698 111030848 111008760 810009061 111038529 810022537 810002424 111008055 111026873 111016219 111041024 810003085 810008156 111041638 111028762 111035205 111007909 111037268 810006312 111030927 810027470 111015807 111025043 810020668 810000734 111032506 111026813 111032775 111035048 111029028 111040153 111028818 111023661 810023780 111016628 111041793 810014626 810014559 111022854 111022853 810017890 810014616 111038040 111013277 111030547 810012254 810000298 111024656 111007283 810001566 111038307 810000537 111038503 111005892 111038172 810002797 111008303 111005897 111032180 111038230 111035130 111032575 810006310 111006678 111004950 111030773 111030183 111037236 111006949 111013279 111017207 810008097 810009127 111035408 810000256 111029069 111024642 810000236 111040248 111037916 111038510 810017806 810023882 810009053 111037481 111032470 111032809 111032151 111022739 111031077 111030838 810018346 111031024 111041738 810006232 111026376 810006249 111026382 810006282 111016201 810023851 111039904 111035069 111042396 111041779 111038078 111040959 111038236 810023692 111032790 111032139 111039700 111032222 111038423 111039725 111038327 111038017 111038043 111031014 810012577 111031313 111037994 111030887 111038251 810000543 111004949 111035482 810002083 810007880 111016646 111039797 111027098 810018262 810023680 810018181 111038112 111032708 810006711 810018178 111014034 111015561 111032186 810001203 810011008 111041655 810006503 111008074 111005707 810018618 810003044 111041556 111038062 810015154 111035244 111032652 111007653 111037275 810015169 111005703 111035531 810000983 111038056 111035183 111038146 111029935 111032644 810002792 111007306 111026169 111006070 111005988 810000985 810000234 111025408 111016782 810000999 111028094 111025388 810001201 810002017 810022500 111030518 810021829 111030513 810014910 111032656 111035235 111016804 111023874 810007835 111032664 111040236 810006318 111027069 111030928 111029856 810008027 111035535 111025048 111042600 111023479 111016185 111023869 810000656 111028117 810009138 810028148 111041606 111032859 810019157 810006311 111027090 111041973 111026129 111015091 810023707 111015120 111039716 810010981 111013000 111029062 111030475 111013246 111013012 111012648 111004939 111013647 111013265 111022531 111021939 111025047 111038046 111013001 111015546 111029820 111026111 810009069 810010910 111015529 111035253 810006704 810006681 111006302 111007290 810006540 810003058 111014129 111015515 111041578 111032227 111007926 810000054 111037497 810017712 810015141 111035212 111034965 810018026 111032441 111029843 111026836 810001646 810014913 111004940 810000884 111005196 810002310 111035315 810020623 111023501 111028985 111035161 111028880 111038504 810002476 111040955 810007803 111032700 111032260 810019182 111041963 111035036 111014161 111004888 810023812 111004900 111021905 111026174 111041837 810011057 111026309 111026292 111037486 111031334 810008897 111017163 111038506 111032231 810006508 111029826 810008113 111005713 111026847 111007275 111032140 111041722 810025646 810000289 111038145 111005185 111032226 810001151 810018526 111031300 810012340 810009073 111030501 111027126 111032536 810025703 111041586 111015130 111028224 810023745 111029903 810018319 111032534 810006749 111026922 810014554 111037518 111038121 111042502 810028140 810014887 810023903 111038438 111030008 810006830 111037487 810007889 111037238 810018255 111032568 810014605 111026800 810003011 810014758 810027985 810025123 810014884 111029811 810000111 111028408 111038530 810025034 111035474 111032673 810020649 810000981 810017865 111041150 111030806 111017216 111035238 111035459 810025300 810025168 810019185 810028707 111030739 810003172 810003170 111023483 111018078 810024312 111022860 810012233 111032611 810015139 810025675 810011014 111028118 111028199 111028105 111031272 810008918 111024571 810002735 810017807 111028098 810017725 111035318 810020708 810008069 111030764 810020648 810018338 111032613 111038272 111028008 111030874 111041662 111038372 810018182 111030626 810028634 111041665 810025018 111041007 111031050 810018529 111030613 111028212 111032345 810028414 111008285 111035202 810011078 810000558 111012978 111038044 111028261 810020661 111043124 111035460 111035222 810014599 810011065 111041552 111014124 111028835 111006697 810025658 810010987 111031369 111031320 111031037 810017825 810028116 111029066 810024408 111030184 111007642 111040967 111023120 810000534 111037934 810001509 810019223 810010977 111007863 111029866 111027060 111039745 810000541 810017650 810018016 111041151 111043113 111029057 810000328 111006306 111038086 810022510 111039729 810025287 111032211 111039766 111022532 810024674 111035029 111028377 111032449 111039714 111040991 111030776 111041700 810028629 810014512 111027096 810014578 111024637 111014151 810000931 111016213 111005901 111029948 810028627 111038457 810027486 810018325 111023399 111023407 111035072 810025271 810008032 810027977 111039760 810002661 111029050 111031301 810017842 810028724 810011027 111038277 810002410 111028091 810007967 111038167 111041509 810018306 810025705 810019192 111035409 810018520 810007926 111028196 810000478 111020651 111031107 810023698 810003289 810002413 111041216 810000822 111024643 111037260 810012338 111042469 810020645 111007358 111027136 111026189 111023081 111006717 810000189 810001698 111007272 111038047 810002417 111029032 111030781 810003056 111037980 111039728 111023424 111005888 810006872 810025292 111032416 111030504 810006539 810023901 111041825 111028969 111027042 111028194 810015138 810007828 810003230 111028300 810002765 810001615 810001910 810003158 111041816 111029875 111039826 111041507 810023835 810023820 111038496 111015559 810007871 810007844 11 1026373 810003125 111026354 111026192 111017092 111035160 111027095 111035149 810002960 810012571 810012658 111030783 111035118 111038238 810008068 111016200 810008116 111030774 810023923 810007973 810015189 810006765 111043023 810018290 111040981 111040104 810025020 810017705 810002030 111026186 111035423 810020683 810007879 810006828 111030545 111035093 111013021 111008764 111031289 111041538 810028376 810028417 810008885 111026310 111038140 111032632 810028390 111028277 810006800 111026982 111032352 111031092 111015506 111032277 810025281 111021934 111032462 111028067 111042443 810024367 810020644 111023844 111040997 111041764 810023652 111026299 111039707 111008052 810028129 810018313 810023825 111041919 810003027 810006854 111028087 111023910 111031340 810023453 111022831 810002299 111028175 111041660 810006584 111031045 810002762 810006548 111031098 111008059 111014912 810000989 111032207 810020604 111038090 111039793 810022518 111028946 810011040 111030454 810006937 111042480 111028255 111022755 810002988 111025327 111008287 111030176 111032220 810002204 111041566 810001633 810000781 111038262 111032879 810008084 810020679 111028792 111017170 111027996 111023860 111031298 111035322 810018613 810011047 111032707 810018355 111015065 111014855 810000175 810000186 111015068 111015114 111023406 111004935 111024037 111024040 810000658 810003094 810000324 111023411 111012802 111012800 111016659 111037968 111035087 111030993 111028863 111028814 111022984 111032335 810006527 810018248 810001327 111035452 111035510 810002744 810008033 111028836 111030210 111004941 111032625 111029995 111038447 111023404 111030857 111006489 111026406 810018147 111041668 111040137 111032439 810001620 111015577 111038151 810024421 810024388 111026855 111038222 111013015 111014037 111017102 810028150 111006301 111006299 111032517 810001153 111015510 810014771 111042565 111006081 810027984 111038390 111035024 111014125 810023713 810023926 810024359 111041215 810021334 810028158 111032196 810006847 111035300 111032712 810008035 810002470 111028401 810023889 111028953 111029841 111029897 810006564 810009118 111030484 111029989 111029070 810001271 111032316 111032343 810028747 810000233 111014148 111006491 810002834 111014827 111028042 111031371 111012669 111041650 111029899 810017082 810006250 111026291 111028048 810006239 111030867 111013635 111029821 810015197 810006862 810006701 810002158 810009077 111041618 111004890 810003017 111028104 111032735 111025328 111032250 111024052 111023658 111030606 111030208 111026408 111038033 111028742 810008046 111029000 111027019 111006688 111041755 810020617 810028728 111041784 810011017 111037526 810014614 810014772 111024586 810009037 810006927 810002472 810002542 111040128 111008305 111030461 111039928 810014756 111027971 111028900 810020699 810008814 111030490 111026321 111028147 810014566 810023822 810020660 111040107 111041031 810028138 111004901 111039823 111028125 810017870 810012348 111030741 810017885 810017889 111037524 111041565 111037218 111016820 111029013 810022509 111005690 810014880 111035251 111037975 810007841 111022861 111020669 111042363 111038255 810018149 810018030 810008158 810002953 111040983 111026302 111032848 810000303 111028334 810007573 111015568 111005078 111032232 111041196 111032483 111042490 111041207 810028159 111041413 810023690 111032562 810001249 111015109 111015093 810002159 111032350 111041903 111028935 111027029 810028149 111032294 111025309 111030939 111026979 111031041 111012993 810018242 111023876 810024414 111028796 111014845 810014561 111043110 111028216 111024053 111014830 111037324 810018443 111028359 810000719 111022550 111038525 111043109 810001867 810014517 111026997 111023495 111041886 111007656 111039905 111040198 111028910 111038522 810002024 111043086 111032443 111032281 111026993 111022511 111026352 810019171 111026281 111040110 111026891 810027825 111008211 111012803 810028757 111006946 810003048 111039802 111008781 810006699 111027001 111038099 111028976 111012805 111006971 111032583 810019151 810018642 111028412 111030835 111004865 810028716 810028696 111005723 111032646 810027866 111032738 810010963 111006978 111006686 111042378 111038526 810017834 111032860 810023924 111016182 111043088 111035216 111034958 111041929 111035137 111042594 810006192 111041646 810014593 111019873 810000299 111008785 111026148 111007320 111023478 810020610 111037249 810025028 810025723 111028804 810011053 111032181 810012336 111026358 111028895 111035080 111029882 810007817 810000820 111008769 810014927 810007806 111023639 111041695 810000187 111028076 810010893 810006546 111016815 111028020 111035192 810023738 810028671 810018288 810014918 810025169 810012241 810014805 111040145 111029862 111039753 111029164 810019236 810010937 810008139 111024682 810024672 810006245 810017457 111026380 111034956 111026407 111028299 810006189 810017449 111026365 111017180 111023419 111042348 111041697 810017697 111038083 111038127 810028746 111030757 111041933 810027485 111032609 810027975 111030523 810012266 810022533 111039799 111029058 111026284 111012989 111026161 810012215 810025270 111029991 111029947 810014908 111026341 810018257 111026384 111026334 810017809 810014539 810017717 111043116 111028228 111032445 810007840 111025304 111032530 111032270 111032355 111030605 111028858 810000987 111014847 111017147 111007302 111004942 111014059 810008826 111030441 111035184 810008115 810002206 111035207 111029048 810006764 810014510 111032306 111027134 810007549 111038216 810018538 810018622 810000827 810007583 810007539 111014027 810025301 810028386 810018031 111037990 111007484 111007482 111032509 111028135 810018425 111023863 810017882 111027086 810027476 111028743 111041232 111040231 111013295 111032526 810007842 111023893 111026314 810010965 810002550 810010951 810028165 810006737 810002783 810014755 810024304 111037948 111041850 111040973 111041742 111019882 810028152 111041881 810014793 810014817 810014940 810014787 810014788 810017740 810017828 111038495 810018504 810008129 111040091 111032424 810014810 810012654 810014619 111008748 810006858 111026125 810002302 810010974 111041558 111014052 111030752 111005721 810012231 111012997 111028137 111024570 111028250 810027823 111031085 111041949 111031345 111030012 111037221 111006970 111037292 810018414 810023815 111032637 810014759 111038434 810006820 810027879 810009115 111029880 810006729 810010923 810010941 810008049 111037913 810001695 111030548 111028850 810018263 810025667 810002839 810025012 111041758 111037306 111004868 111041773 111038297 111041696 111032718 111024608 111025414 111004913 111041907 111038153 810028131 111026965 810003177 111031007 111028887 111026914 111030444 111030612 111042413 111035010 111032245 111041017 111042505 810018415 111006295 810001451 111017148 810002309 111017103 810001777 111015061 810002033 810028389 810028350 810028379 111037283 810007927 111032842 111006499 810017631 111032315 111042482 111042381 111042531 111014138 810017858 111038427 111014851 111038292 810014617 111025390 810018348 111008249 111025393 810002039 810014611 111022569 111038468 111014119 111035483 111032155 111007270 810018418 810014621 111030438 111028905 111030866 111030167 111023506 111032704 810018512 111042374 810009057 111032463 111006485 111007644 111008306 810006580 111026945 111029010 111023394 111019823 111026103 111038531 810023669 810008013 810018500 111030166 111028826 111041975 810025650 810018184 111043101 111041958 111026349 111024595 810012341 111041188 111026305 810014548 810010952 111043106 111031008 111035154 810019216 111041690 111031108 810014777 111035028 111028947 810015130 111028133 111035046 111039770 111019476 810009140 810019232 111039906 810001578 810006708 111032218 111026810 810014892 111028007 111034970 810008145 111015805 810023637 111035248 810018413 810025019 111035131 111032834 111038489 810025677 111032597 810012240 810014941 111041584 810014929 111027959 111041824 111027116 111041504 111041734 810021828 810027982 111042526 810018493 111042619 111034987 810028684 111032857 111037314 111038523 810028761 810023717 111041631 111032225 810028385 111032655 111028057 810007531 810011055 810028144 810007529 111030562 111038345 111037315 111041036 111041230 810019222 810023862 810025056 111041163 810023636 810019248 111037940 810019212 111032429 810028114 810025665 810014592 111032346 810006220 111032177 111035429 111035042 111029149 810024375 810019208 810024306 111035310 810018272 810021853 810014595 111041018 810018514 810027877 111032326 111032426 111037326 810012344 111024665 111043007 111028038 111043114 111028339 111038110 111038412 111041752 810010939 111032188 810007868 810018436 810008928 111032782 111041848 810006685 810003213 111042610 810001543 111028182 111040992 111039898 111035001 111028063 810025139 810018244 810028659 810024986 810018166 810014520 111028930 111030832 111032323 810008811 111032593 810025724 810017877 810027839 111032440 111040127 810028632 111029968 111035005 111026951 111030896 810014524 810024998 810023816 810023734 810023876 810023844 111039809 810023660 111032455 111023841 810018285 111029818 111032689 111042418 810012347 810019150 111026392 111039761 810025125 111041561 111031051 111031347 810018549 111041604 810024389 810024394 111032206 111035129 111039896 810010947 810014897 111035175 111042537 810009128 111041551 810018318 111040135 111037232 111030622 810023630 810017846 111016629 810001147 810014914 810015133 810015186 111032765 810006506 810015187 111042377 111042345 810023737 810024382 810019193 810023782 111038528 111037489 111042587 810025278 810027493 111037523 111038312 111040211 810023626 810014768 111028964 111029972 810007970 111039789 810010971 111030801 810018164 810028713 810028710 111038184 810028754 810019219 111041704 810012263 111041514 111039822 111038221 111041733 810018405 111037525 111039709 111038474 810008926 810015194 111029911 111028852 810008124 810024386 810022505 111038244 111016626 111032460 810001569 810001623 111023097 111024593 111006694 111007369 111018233 111028085 111025064 111024684 810019170 111035182 111028407 810002611 111016205 111026294 810008112 111028187 810020676 810023450 810023727 111041545 810018435 111040095 111038488 111032192 810002041 810002016 111007354 111007877 111007902 111037939 111032184 111029885 810014769 810012642 810027973 111022537 810015144 111043036 810027853 111023432 111035290 111030902 810001974 111040121 111028189 810028677 111042392 810017749 810014569 111043091 111043276 111043290 111043104 111043089 111039721 810021855 810006491 810011038 111041778 111026816 810002414 111032492 111030782 810001154 111035445 111016401 810027840 111029041 111035082 111042346 111032461 111006960 810018261 111028776 810008843 810007895 810018153 111014866 111038320 111020658 111015543 111041686 810002622 111015090 111041647 810018420 810006254 810008859 111030476 111026113 111035295 111004937 111030474 111021967 810012244 111027071 810001150 810001237 810000925 111019622 111016220 810002745 111007899 111023986 111024048 111007878 111008774 810007548 810006268 810014606 111040977 810023657 111032413 111028176 111038339 111039751 111028951 111006950 111019472 111041754 111022983 810002942 810008044 810008098 111027023 111022866 111029849 111013026 111025337 111030756 810007978 111030205 111038480 111019887 111019888 111014144 810002800 111035053 111026119 111040151 111039704 111016187 111026918 111026917 111019877 111026421 111026950 111029851 810024405 810007929 810018349 111032640 111032663 111024581 111024628 111008233 111039804 810002475 810023861 111007337 810018352 810001535 111016799 111024674 111038157 810008091 810014792 810006490 111030880 111026177 810001339 810017742 810018292 111035124 810001179 111004909 111038456 111038010 111008752 111029094 810008803 810006512 810025268 810012358 810008063 810014557 111030852 111025422 810023655 810006572 810025067 810018260 111003415 810007681 111024704 111010569 111003226 111003421 111039925 111037262 111037491 111033241 810009075 111034966 111030951 111029065 111041711 111030762 111008753 810020624 111028410 111015106 810025044 810001699 810025064 810020621 810001198 810000360 810017735 111032510 111024662 111030010 111019879 111026940 111038012 810006694 111020664 810027971 810027821 810008860 810007542 810012573 111014904 111031006 111012244 111006486 111014870 111040125 111015820 111042607 111027964 111029158 111029887 111040237 111032347 111035468 810018649 111023226 810002265 111038388 111032594 111013271 810018477 111038281 111022557 111012234 111039717 111032581 111030897 111024043 111027008 810002137 111023967 111041634 111042542 810025172 111038544 111041224 810025282 810006526 111006975 111016657 111007474 111038404 111038497 810008108 111014841 111008300 111035532 111039902 810023859 111041977 810017708 111030070 810028358 111027011 810008841 810023693 111026911 810001245 810001008 111041570 810000725 810018381 111039699 810000307 111014029 810002349 111015815 111032805 111023092 111026871 111032182 810012589 111032146 810007925 111026878 810008813 111032801 810023671 810018367 810012648 111041227 810001865 111026106 810009137 111007887 111030001 810023806 111029894 111032464 810003049 111025383 111035079 111035473 810023741 111041028 111038464 810001311 111035040 810001911 111024635 810012255 810007546 111030207 111030618 111008065 810015151 111022514 810000661 810002803 810018158 111022525 111030869 111030803 111030468 111035524 111026808 111028862 111028342 111028822 810008905 810001542 111003841 111003461 111003438 810020794 810015770 111003032 810020997 111002147 111003425 111003112 111003478 111022330 810017606 810009861 810009718 111003233 810020984 810020575 111003479 111003429 111024565 111022273 111003113 111003101 810014351 111021680 111002834 111020591 111003011 111003025 810010799 111021379 111003840 111002796 810010160 111003456 111024618 111002788 111001042 111020144 111001065 810017611 111027093 810006684 810015131 111026130 810007885 111028286 111041162 810028115 111024703 810025535 810011120 111022266 111003107 111002640 111002843 810014374 111003027 111003093 810025550 810012185 111003031 111018985 111022348 111020555 111003832 111003008 111003002 111003253 111002823 810025794 111038450 111038358 111038344 810000723 111003104 810009882 111021269 111003096 111003229 111001236 111003844 111010551 810011184 111018995 810007692 111003467 810025898 111023010 810010395 810020798 111021337 111002597 111003240 111002810 111003029 111001098 111020142 111003013 111020559 810020633 810020571 111002849 111002281 810020777 111023808 111022341 111002994 111002276 111001504 111002350 810017619 810017584 810025635 111020567 810017609 111002981 810014365 810007677 810007984 111001485 810007998 111001500 111003433 111003245 810007655 111003424 111010515 111019723 111032858 111028365 810009670 111003458 810010140 111002259 810012183 810009732 111003017 810025906 111021659 810010072 111020075 111003117 810020769 111003244 111003437 111003165 111003091 111003227 111023100 810018336 810007695 111022278 111001505 810015070 111024614 111003473 810007712 111003459 810017340 810025541 810017614 810007788 810015067 111003449 111010556 111001250 111003465 111009612 810025896 810009666 810015079 810009989 111001491 810010098 810023869 111035141 111038257 810002350 111041965 111030560 810007836 111038328 111025044 111032564 810000231 111030746 810008017 111032266 111016633 111041896 810020616 111023080 810014762 111023416 111022523 111030893 111038417 810001120 810009054 810006810 111035416 111016395 810010897 111027100 111005873 810019159 111035449 810018521 810009022 111028876 111028744 111042563 111039722 810018351 111032794 111027112 111032305 111035437 111025370 111007889 111042512 810006230 111025374 810028622 111032552 810023845 111027111 111032468 810028643 111032626 111026962 810009034 810003163 111032574 810018281 810006547 111031027 111039788 111028714 111029890 111016661 111013301 111038261 810017669 810023799 111021910 111013286 810001978 111025407 111021914 111032342 111017150 810006583 111022991 111035301 810001702 810018296 810025135 810023662 111041505 111032336 111043077 810025155 111029860 111030519 111005691 111005907 111005982 111028288 111026162 111023392 111039941 810002844 111042519 111028926 810014786 810006327 810023883 810006797 111041541 111037244 810028013 111017093 111039720 111022519 111015096 111007876 111041839 111014035 810023452 111030758 111041892 111026168 111038514 111027018 810021082 111007334 111038516 111037979 111027063 111028855 111040213 810025729 810027995 111027050 111024032 111012985 810012248 111027962 111041849 111041856 810025029 111035204 111029074 111027994 111016813 810009029 111038354 111029067 810023689 810012582 111005911 810002510 810018180 810008810 810006863 810000554 810025000 810028146 810008844 111014869 111028824 810007923 810008839 810007831 111015057 810000318 111042355 111032778 810012334 111008217 111038472 810007856 111008301 111026218 111038518 810014582 111042585 111032217 111026099 111032864 810017448 810018005 111026133 810007845 111042429 810024311 111032557 111032833 810018660 111035415 111029879 111028356 111032268 111026347 111042499 111031338 111031083 810010970 111039787 111023648 111030843 810010904 111032759 810015168 111014831 810018633 810010900 810025654 810025660 111026794 111024680 810028641 810025722 810011077 111024594 111032590 111006494 111015050 111026920 810014598 111032816 111030552 111035215 810014588 111043035 810009074 111031099 111030623 111015074 111013036 111015048 810009094 111018082 111042383 111037991 111008297 111042580 810001223 111014878 111017291 810009076 810020652 810008884 111028023 810017435 111007857 810007807 111040971 111028198 810010955 111038435 111023470 111003164 810010351 810007794 111009608 810010070 111003451 111021325 111024555 111022320 810011995 810025791 810012041 111003483 810012173 810010077 111019708 810011991 111021668 111003463 810010186 810009879 111022299 810020810 810009883 111002856 810011959 810015068 810015062 111003446 810011949 810014331 111022279 810017921 111003420 111003158 111003016 111019777 111003484 111002595 111002653 810010868 810007731 111003012 111003444 810009890 111003435 810010153 111002829 810020812 111003470 810011112 810017287 111003228 810017979 111003169 810009881 111002798 810012072 111003472 810009719 111003097 810007678 810010066 810009864 111002989 810017978 111003430 111020153 111002813 810025571 810011181 111002609 111003094 810010078 111003023 111024557 111008672 810020802 111020587 111003442 111002854 111021234 111023447 111003476 111003249 810009995 810009884 111003022 111022331 810011182 111003409 810010111 111002851 111002995 111020571 111002826 111002812 810020639 111010505 810010321 810007679 111003223 111003842 111021242 111022308 111003114 111003222 111024558 111022275 111002636 111009202 810007697 810011204 810010151 111022346 111020837 111021253 810025925 111002845 810012047 810009951 810017401 111021342 111011159 111003432 810014366 810014386 810014363 111020602 111020574 810020788 111021232 810007756 810017956 111002838 111009561 111001249 111002661 111022343 111001251 111001241 810014380 111020586 111001518 111003416 111001479 111001494 111002666 810012182 111020593 810009956 111003487 810011167 111001522 111003161 810025566 810025935 111003156 111002993 810011257 810015061 111021327 111003248 810009040 810010216 111024616 810007653 111019959 111024566 111024615 810017358 111022329 111022311 111019767 111018984 810010113 111021251 111002141 111003234 810011985 111038170 111027058 111042575 111039703 111032236 111030855 810027993 810009217 111037290 810000113 111017171 111005732 111040206 810011997 111024559 111002847 810007661 111003436 810010069 810012024 111026856 111028713 111016816 111032561 810000732 810002607 810024354 810018324 111013257 810010925 111003428 111002815 810017565 111003028 111003477 111022269 810008140 111005905 111042605 111042412 810019215 111042578 111038364 111031017 111032743 111038318 111003003 111028077 111026127 111041861 810009715 111024706 111003095 810006934 111029068 810023732 810015063 810009722 111002042 111003241 810010924 111014044 810027988 111030844 810027847 111041812 810027495 111029016 810006933 111028961 810011251 111003255 810011159 111024065 111032776 111041180 810007581 111029017 111041602 111038137 111025416 111030883 111003235 111020578 810002077 111028771 111024654 111025419 111041519 810000066 111030823 111007288 810014541 111029800 111003426 111002793 810007778 111002797 111003312 111001198 810003063 810018360 111035480 111030996 810027501 810010926 810015143 111026362 810001003 810006680 810006697 810018040 111041834 810023766 810021859 111029994 111041857 111019896 810007566 810019762 810007876 111014860 111002844 111002832 111020579 111003246 111010563 111022352 810017945 111002662 111003452 111020860 810007738 111002791 111020598 111022315 111026219 111026903 111014828 111013008 111028941 111038000 810027981 810006930 810008834 810000726 111039734 111026998 111017189 810020647 810027505 111026351 111002809 111003445 111003457 111002203 111001214 111003224 111003168 111002614 111003417 111003232 111003413 111024619 111040109 111035218 111016202 111028039 111003447 111003015 111032798 111030555 111023494 111023076 111026867 111037951 111001490 111024552 111002840 111002601 111021681 111025425 810006688 111030942 810014785 111015082 111013294 810010934 111013261 111008226 111015824 810001792 810009045 810015180 810025046 810017623 810020796 111014894 810023720 810009101 111032813 810025692 810023623 810011983 111008271 810023760 111021671 111019006 111041948 810021856 111003111 111020553 111003098 111003411 111003155 111015119 111007624 810018150 111028817 111012968 111032221 810008827 111041156 111016647 810007810 111026144 810017612 810010119 111003115 111028770 810001869 810008025 810000127 810002822 810002802 111038038 810003153 810002801 111039915 810008012 111023123 810018190 111032256 810008150 810002949 810017843 111007277 111027067 810001461 111015816 810008026 810018156 810002805 111030800 810006869 111032340 111032572 111029874 111038187 810018395 810023814 111032659 810020818 111010509 111043118 111037288 111040223 810020607 111032550 111040130 111029838 111041740 111041819 111042538 810011002 810022543 810006552 810012011 111003020 111022345 111020816 810007770 111030923 111043093 810017586 111028756 810017706 111007477 111035543 810017624  Neutrino: Model Number: NxT VR ICD REF CDVRA600Q UDI-DI Code: 05415067034373 Serial Numbers: 810016068 111017554 111017845 111017564 111017591 111017561 111017385 810021495 810005696 810012620 111017359 111017396 111019396 810004125 111019405 810021649 810013177 111017378 810013413 111019039 810026573 111019273 111018939 111018525 810013163 111017597 111017384 810005688 810019078 810016247 810019134 810019136 111019270 111019374 810012615 111019373 810013372 810021497 810013298 810013418 111019403 111017391 111017388 111017398 810022040 810021653 810016239 810013373 810021496 111018523 111017365 111017362 111022359 810019074 810019076 111017538 810013422 111019271 810021493 111019570 810021502 111017386 111019599 111019592 111019600 111020521 111020639 111020504 111020545 111021590 111021584 111021582 111019028 111019034 111019038 111018914 111018952 111021588 111021072 111022367 111022396 111022370 111022387 111022375 111022377 111022361 111021593 111021591 111021575 111021585 111021578 111019369 810005697 810005678 111019368 111018529 111019389 810019502 810021891 111019375 111017583 810005692 111017400 810021500 111019250 810013302 111018940 810021494 810013160 111017578 111017560 111019035 111019029 111019365 111018521 810013175 810019140 111017555 111017542 111018922 810005680 111021057 810013153 810013428 111017405 111017601 810005682 111019252 111019362 111018949 810004253 810004305 810019494 111022374 810004254 810003595 810016252 810022047 810013524 111019386 810003856 810003916 810019774 111019372 111020634 111019587 810004169 810016067 810019142 810019782 810019785 810019506 810016420 810022036 810026563 810022043 810019772 810016435 810026562 111018532 810022044 810019776 810019787 111020512 810019077 810013526 111020647 111018930 111018957 810021889 810022034 810016410 810003802 810016432 111017552 810022031 810026586 111019032 810021879 111017553 810013551 810013540 810019507 111017406 810016427 810013527 810013138 810004249 810005698 810004736 111017368 810004735 810011608 810021894 810004788 810011603 810021881 810027592 810027565 810027596 810027567 810027568 810027595 810011595 810011599 810011596 810003713 810003712 810003716 810003715 810003775 810003655 810016249 810022045 810026588 111021077 810004251 810004056 111017562 111018524 111019572 111019577 111019406 111019274 111018517 810003859 810027580 810016038 810003803 111019584 111019392 111019376 111019378 111019387 810003805 810021878 810005691 810004053 810013276 810011609 810027581 810026575 810004790 111018931 111019595 810021885 810013176 810013525 810013425 810013296 810021892 810026564 111021580 810016411 810021887 810022046 810022052 810016037 810027584 810021498 810013301 810019513 810019512 810013412 810013426 810016253 111017582 810005686 810021880 810013419 810005679 111019573 111017579 111021576 810005685 111021060 111020517 810003654 810013541 810005753 111020532 810005683 810005695 810005674 810022048 111017599 810016090 810004256 810019500 810016043 810011610 111020635 810011598 810011602 111020628 810022039 810005755 810003448 810022051 810003773 810016248 810004316 111017586 810004605 810021884 810003774 111020620 111020619 810022038 810016415 810004250 810016241 810004167 810021882 810013292 810016062 810013531 111017596 111018518 810026583 810003594 810016258 810016071 810004052 810026578 810026587 810013543 111019383 810013528 111022366 810005690 810022056 111018955 810016066 810013157 111021583 111022380 810011594 810016236 810016419 810003804 810003772 810019781 810016237 810016238 810016089 810016243 810016422 810016421 810016244 810016431 810016406 810016425 111017557 810016064 111022138 810004604 111017374 810013294 810004880 810013162 810022055 810021888 810016070 810012613 810026570 810027582 810022042 111017363 111022141 111022146 111017413 810027577 810013161 810013295 810003596 810013534 810004122 810027585 810019505 111020548 810019495 810003857 810019499 810013411 111020552 810003995 810003593 111022137 111021579 810013299 810003652 111020528 111018330 810016058 810016413 810013113 810013303 810013165 810016074 810016429 810022058 111017544 111017546 810005681 111018329 810013530 111017572 810016092 810016040 111019579 810004168 810016091 810016050 111020621 111017371 810004602 810004787 111018324 810022033 810026574 111019398 111022383 111022388 810027575 111020525 810004313 111020529 111020633 810013539 111021658 111020505 111020547 810027569 810026572 810026579 810026593 810026577 810019079 810013290 810016060 810016093 810019141 810019510 810019509 810019143 810013421 810016428 810016407 810004314 810016039 111018951 810013297 810016424 810003912 111018513 810027572 810027570 111019593 810012629 810019135 810004170 111019367 810013409 810005687 810013137 810026566 111019598 111019578 111017548 111017602 810004732 111017547 810003915 111018948 111022381 810019511 111021586 111021592 111021589 111022139 810019080 810019784 810019783 810013529 111022140 111022143 810027583 111022148 111018941 111019371 111019597 810003860 111018935 810003854 111018916 810022032 111019575 810013548 810003653 111017851 810013410 810013154 810027588 810013278 810013300 810026592 810003714 111018934 111018516 111018534 111019031 111021064 810003592 111022376 810019780 111019030 810013158 810013178 810005684 810003858 810026559 810027564 111019382 111017379 111017416 111018921 111018535 111017389 111017382 111017399 810027587 111017392 810019504 810026568 810026585 810019508 810019777 810026598 810027573 111019381 111017373 810013159 810021876 111017569 111019594 111017395 111017850 810016048 810013535 111019393 111022363 111017541 810003861 810016046 111018919 810013547 111020542 111022384 111018932 810013544 810027586 810016408 810027574 111018938 810013115 810026589 111018927 810013151 810016260 111022373 111018522 810027589 810016085 810016069 810016057 810016086 810016075 111017380 111017358 111017381 111017375 111017369 111017539 111019585 111020523 111017387 810016263 810011593 111019602 111017848 111018953 111018936 111018950 810012619 810004304 111020520 810004302 111021075 111017409 111018337 111018317 111021070 810011597 111018945 111017849 111017364 810013533 111018920 810016434 111021594 810016053 810013114 810013274 111022386 111018321 810004303 810016061 810005675 111019568 111019588 111017550 810004055 810004255 810003852 111019582 810005754 810013155 810003994 810016072 111021068 810013536 111018533 810004171 810016259 810016083 810016245 810016076 810013111 111018326 111018915 810016059 810013532 810022054 810005694 810016045 810012618 810016051 111019596 111017401 111017571 810005677 810005673 111020509 111019390 810016044 810016049 810016081 111019027 810019788 810021890 810027571 810027594 810021651 810021648 810027579 111017577 810016416 810027591 810019138 810026569 810022035 111017412 111018918 810013423 810003914 810026591 810013289 810016077 810013427 111018323 810011607 810013416 810019139 111018520 810016426 111018313 810016418 810019075 810019082 810022049 810019137 111017377 111018530 111021573 810013152 111020642 111021596 111018954 111018913 810013550 810005676 111018515 111018511 111017397 810004603 111020627 810016065 111012884 111012886 111012889 111012885 111012807 111012892 111012887 111012894 111012890 111012888 111012893 111012891 111012895 111013043 111012883 111019255 111013864 111020489 810027578 111021581 810019790 810019779 810011604 810011601 810021652 810011600 810003776 810019791 111021587 810016262 111017549 810019778 810004791 810022060 810011592 810016073 111022368 111020630 111017594 810016080 111018917 810004312 111018943 810013293 111019251 111022360 111018956 810027576 810011591 810012617 810027566 810004252 810004734 810019786 810003993 810016417 810013140 810016078 810016088 810011606 810012627 810013417 810016254 810004248 810021895 111022389 810013420 810012622 810013537 111018519 111020538 111022144 111017370 111022394 810003855 810004881 111019590 111020530 111020516 810016430 810016261 810016047 810021893 810013552 810016264 111020648 111018316 111019395 810003806 111017595 111017565 111018325 111017394 810004882 810004123 111017589 111019601 111019569 810013164 810016250 111019591 111022147 810004247 111018928 810013291 810016265 111022145 810013136 111017410 810016036 810016095 810016094 111022390 111021066 111022392 111017584 111017574 810004315 111019581 111018924 810013549 111019364 111018937 810003996 111022393 810013424 810003913 810027593 810026571 810016084 810016052 810013139 810016423 810013546 810013156 810022053 810004606 810026584 810016082 810016054 810016063 810026580 810026596 810026582 810026595 810026594 810026590 810026560 111022395 111022369 111019400 111019402 111019388 111019377 111019404 111018942 111018926 810026567 810026581 810026576 111018944 810016255 810016242 111021067 111017393 111018322 111018925 810012616 810012614 810012612 111021076 810004618 111017390 111017593 810016409 111019391 111020511 111017590 810016433 111018526 111018514 111018332 111017592 111019366 810004733 111017563 111022385 810016055 810019775 111019384 111018527 810004054 810003853 810016056 810012630 111018933 111022378 810016087 810004124 111022362 810016414 810013415 111020550 111019037 111019370 111017567 810016257 810013545 810004126 810016256 810019773 810021501 810012611 810013174 810012621 810019083 810016246 810016412 810005693 810026565 111021595 810003656 810013414 111018331 111018315 111018336 111018528 111019036 810022057 810022037 111020626 810026597 810021655 810021654 810013275 810021657 810019081 810019503 810019498 111019379 111017545 111017600 111017551 810019789 111017842 810005689 111017846 111019363 111018946 111018947 810016251 810021877 810021656 810021883 810021499 810021650 810021886 810019497 810026561 810019496 810016042 810004306 810022050 810004884 810004883 111017575 111017559 810013112 810013553 810022041 111018320 111017540 111017844 111017566 111017568 111017573 111020551 111020514 111020526 111020544 111020549 111020518 111020515 111020534 111022364 111018334 111022372 111022391 111022382 111022365 111019574 111019380 111019401 111019385 11102051 3 111020536 111020638 111020640 111020631 111020507 111020645 111020508 111020644 111019571 111019586 111019589 111020539 111020541 111020632 111020625 111020646 111020636 111020546 111019580 111021061 111021062 111021071 111021063 111021073 111021059 111021065 111021577 111021056 111022371 111022142 111022379 111021055 111021054 111021078 111021069 111020622 111020533 111020540 111020522 111020524 111020637 111020641 111020629 111020643 111020519 111020623 111017403 111017361 111017366 111017383 111017411 111017408 111017576 111017376 111017415 111017367 111017417 111017414 111017372  Model Number: NxT DR ICD REF CDDRA600Q UDI-DI Code: 05415067034335 Serial Numbers: 111019825 111019841 111018417 111018420 111018429 111018419 111018407 111017348 810010300 111020493 810010406 111020492 111019558 111019664 810012271 810021091 810010296 810017785 111019802 111017512 111017355 810014481 111020498 111020496 111020499 111020494 111021040 111019658 111019835 111019826 111019842 111017353 111021138 111017507 111017529 111017501 111017508 111021470 111021160 111019659 111021170 111021136 111021133 111021157 111021142 111022115 111022127 111022129 111022131 111021035 111021559 111021558 111022112 111022101 810005742 810005730 810005744 810005739 810010411 810010402 810010419 810012432 810012389 810012388 810012390 810012305 810012296 810012371 810012374 810012276 810021162 111019654 111018425 810012287 810012283 111018406 111020107 810021174 111018427 111020501 111017838 810017325 111020495 810005748 810025752 810010403 810012375 810012325 111020089 810021169 111020136 111018418 810017305 111021485 810005721 810017304 810010397 111019025 810021170 111020093 810017312 111020500 111018687 810017759 810017297 810017324 810017311 111019804 111018702 111018694 111020102 810005750 810017891 111022109 810021183 111018692 810021128 810021130 810005737 111021135 111021132 810024754 111020123 111018421 810021121 810025744 810025751 810010418 111022098 810021117 111021567 810012286 111018824 111021149 111021034 810021109 810004597 810025868 810014436 810025747 810021154 810025880 810021144 810021146 810010287 111021557 810005171 810021095 810025759 810021161 810021120 810021163 810015113 810025757 810014957 810018229 810018064 810018052 810012445 810017470 111021165 810018058 810021171 810021101 810018223 810017970 810017302 810018211 810017477 810018237 810017895 810017768 810024748 810017300 810017307 810015106 810003988 111021131 810025886 810025874 810025981 111021167 111021468 111018890 111018817 810021173 111018880 810017476 810017462 111022125 810012372 810010420 111022124 810014725 111018813 810014037 810014420 810012280 810015125 111020086 111021051 111018818 810021090 111018831 810024740 111019246 810014945 111017302 111017294 810014745 810004119 810012387 810014956 810021145 111021154 111021052 810005133 810005037 810004784 810004727 111022132 111019653 810018062 810025753 810017969 810021115 810014423 810017298 810021164 810003843 111017317 111019833 810017775 810025976 810004783 810017484 810025689 810014961 810017480 810017460 810017467 810017468 810017474 111019799 810005038 810021107 111022099 111017510 810004120 810004598 810021142 810021137 810025870 810021141 810021182 111020084 810021116 810005039 810024749 810005172 111021162 810024738 810021132 810017483 810018215 810012307 810004601 810025973 111019022 810010399 810010292 810010307 810025982 810018069 810021186 111021563 810021180 810021113 810021187 810018066 111018428 810014428 810014751 111019559 111017530 810005727 111020130 810025876 810025887 810012306 810014723 810004320 111021474 810021127 810018067 810021184 810017762 111022113 111017357 111022120 810012322 810005132 111017303 810010414 810018222 810021108 810012449 111017312 810014496 810014748 810012450 810025967 111018823 111018402 810014489 111021152 111018690 810012383 810021143 810014488 810014424 810018234 111018695 810004599 810021093 810014416 810004057 111021489 810026558 111021049 111022114 810021110 810014415 810004879 810014429 810005720 810004875 810004121 810015116 111018826 810025966 810025985 111018887 810014977 810014955 810014498 810005733 810014736 810025735 810025882 810025881 810010409 810024742 111018879 111018822 810010400 810017771 810012318 810010417 810004728 810021147 111018698 810014417 111019242 810014418 810012310 810012299 810004307 810012278 111017324 810024741 810025748 810017306 810010306 810014427 111020114 810025749 810004311 810005740 810014954 111020100 810005743 111019021 111020110 810021152 111021141 810004877 810014747 111019019 111017520 111017837 810025741 810004785 810004782 111018819 111022123 111017841 810014947 810003850 810012288 111020128 810012281 810012284 810025873 111022133 810010290 810014421 810005741 810026015 810025988 810026011 810014039 810004596 810014495 111018889 810025884 111020087 810021088 111022100 111022105 111019245 111019800 111022122 111020085 111018411 810021179 810005131 810014980 111022134 111018886 810014735 111021044 810005735 810018216 810012279 810021140 111019824 111018703 111019026 111021140 810010416 111019836 111020104 810014732 111017528 111021166 810025869 810025877 111019830 111017319 111019248 810021157 810021105 111021168 111017526 111019838 111017322 111021572 111017341 810005723 111021565 810014431 810018212 810025737 810003711 810015110 111022103 111020105 111020127 810010401 111020131 810012292 810026014 810017783 111020106 111017342 111017308 111021042 111017499 111017320 111019648 111019650 810017296 810017289 111017344 111018814 111022135 111019663 111017519 810010301 810010288 111019652 111019665 111019564 111019566 111019562 111019563 111018367 111017299 111017316 111017525 111017330 111020497 111021490 810024751 111021151 111021053 810021131 810014962 111021488 810010297 810017318 111022104 810010308 810012373 810021114 810018230 810014430 111021568 111019014 810024747 810024736 810005725 810005722 810005724 111019018 810010286 111022110 111021476 810004060 810014438 810010415 810010404 810010294 810014981 810014979 111017329 111017505 111021155 111021483 810014978 810014943 810003851 111017532 111017333 111018825 810017292 810017294 810017315 810017288 810017782 111017305 111017331 111017314 111017340 111017325 810024750 810024752 810024746 810024737 810024745 111021037 111019846 111019845 111019839 111019844 810025730 810021098 111017351 810025980 810004876 111017509 810003446 111017523 111017522 111017833 111017296 111017346 111021139 810014744 111018828 810014741 810012361 111020112 111017498 810014946 810010413 111017518 810021124 111018830 111018877 810012301 810014494 810014959 810025745 111018874 111020115 810005732 111017834 111017332 810004059 111017310 810012295 111017318 111020101 810025983 810018054 810010310 111017345 810021177 810018070 111020502 810010398 111021561 111019561 111020109 111020117 111021041 810010407 810012314 111020132 111018820 111018700 111019647 111020079 111018434 111019657 810012273 810017469 810014497 111022106 111018696 111018693 810025871 111021161 111018832 111018412 111021469 810014475 810014434 111018404 810025889 111017347 810004317 810021172 810021158 810018213 810017765 810017776 810017777 810017772 810017767 111017500 111017502 111017350 111017343 111017513 111017511 111017516 111017304 111017338 810017757 810025888 810025878 111017356 810017779 810021097 810014480 810014738 810012311 810005174 111021556 810025971 111020094 810024753 810017475 810017773 810018228 810021185 810003800 810003921 810004600 111019269 111012855 111012847 111020614 111012878 111012858 111012838 111012842 111012850 111012876 111018003 111018004 111023892 111012877 111018002 111018001 111017998 111018000 111013049 111023888 111023883 111023891 111023887 111012870 111012840 111012839 111012854 111012867 111012856 111012859 111024227 111012869 111012861 111012871 111012872 111023890 111023884 111023889 111012862 111012863 111018010 111018008 111018009 111012849 111012879 111012852 111013047 111012841 111012848 111012864 111013045 111012860 111012868 111012882 111012873 111012843 111012880 111012846 111012845 111012881 111012865 111012866 111012853 111023885 111018005 111012851 111013048 111013046 111018011 111020612 111013879 111013875 111020617 111013884 111013531 111013878 111013876 111013880 111013881 111013883 111020608 111020607 111020613 111020615 111020326 111016335 111013877 111019268 111024221 111024222 111024224 111024223 111024218 111024226 111025945 111025951 111025950 111025948 111025947 111025944 111025946 111019266 111019265 111017997 111018007 111018006 111013866 111013865 111013870 111013869 111013867 111013868 111013882 111013872 111017999 111013874 111024220 111024225 111013532 111012857 111017831 111017832 111013873 111024219 111019267 111020616 111020610 111020609 111020611 111013871 111025949 111025952 111025943 111035357 810024220 810024218 810024222 810024219 810024216 810024221 810024217 111035370 111035351 111035373 111035371 111035359 111035362 111035374 111035364 111035363 810028647 111035367 111035358 111035361 111035365 111035356 111035350 111035366 810024227 111030430 111028706 810024230 810024229 810024235 810024232 810024233 111030426 111030429 111030428 111030432 111030431 111030424 111030425 111035352 111023886 111032423 111032421 111032422 111032419 111032420 111035372 111012874 111012875 111012844 111032382 810024226 810024228 810024224 810024225 810024223 111035355 111035368 111030427 111035369 111035354 111028705 810028652 810028651 810028648 810028653 810024234 810028650 810028649 111035353 111035360 810010303 810010309 111018875 810012384 810004058 111022107 111021153 810012442 810005173 810018214 810021159 810017899 810024743 111018706 810025892 810025890 810005170 810014499 111018829 111019247 810010293 111019243 111018689 111017524 810014437 111018699 810004731 810014739 810014743 810014750 810015099 810015118 810015103 810015102 810015108 810014473 810010298 810014958 810025987 810025974 810014439 810014435 810025989 810025991 810012320 810015101 810010302 810015124 810015120 810012315 810015114 810015104 810015115 810012309 810012436 810025984 810014479 810026557 810025965 810025992 810025743 810015105 810015119 810015126 810015127 810012363 810012317 810012448 810012323 810012321 810014471 810014953 810017784 810014752 810021099 111021036 810012328 810017316 810025964 810005036 111022121 111021043 111021486 810003917 810017761 810018220 810025742 810012272 111017328 111020137 111020080 810021149 810017968 111020133 810010410 810025990 810021104 810025968 810025746 810010396 111021169 810012435 810014976 810026013 810014472 810025872 810010295 810024735 810021166 810021133 810012438 810012443 810012285 810021175 111021566 111020116 810012277 810015111 810022553 810014491 111022111 810025879 810010289 111021048 810012431 810014502 810018236 810004309 111019827 810018046 810021112 810021178 111021147 111019829 810014944 111020120 111021144 810012282 810014501 810012298 810021119 111019803 810015121 810012300 111020097 810005035 111021473 810025970 810017488 810017780 810018226 111022116 810014422 810018050 111020135 111019801 111022126 111022118 111022119 111022136 810024739 111018408 111018369 111018691 810017758 810012329 810017487 111018688 810005749 111018422 111020095 111018373 111018416 111019834 111018413 111018892 810021148 810014973 810017472 111021164 810014490 810017471 111018701 111018704 111018403 111017327 111021472 810017321 111020126 810003918 810021129 111018409 111017337 810014425 111021159 111018821 111018888 111017836 111022128 111018891 111017515 810014982 810018231 111018405 111020618 810018057 810014426 810015100 810015107 810015122 810014974 810017897 111019805 111020124 111020091 111020090 810017893 810018056 810018239 810018225 810018061 111020129 810018227 111019560 111020125 111020098 810014474 111019024 810025986 810012275 810012289 810012362 810012326 810012294 810004786 111019832 111018424 111022108 111018432 810014975 810014960 810017485 810017482 111021137 111021158 111018376 810025969 111022117 810012434 111020113 810018217 111021038 810014733 111022102 111018893 810012433 111017300 111021143 810012304 810010291 810021153 810015117 111017301 810004319 111019831 810004878 111022097 810004729 111020122 810017303 810012447 810014419 810003590 111021482 810003842 111018410 810021181 111018423 810017894 810018068 111019649 810026012 810014478 810014477 810017778 810017770 810017293 810025891 810017486 810017319 111018371 810025883 810017629 810017966 810021126 111019847 810017489 810012439 111018883 111022130 111020490 111018414 810015112 810015098 111021047 111018415 810012316 111018433 810017314 810021150 810017317 810017630 111021050 810005738 810017774 810025733 810021096 810014500 111019646 810014730 810014729 810014483 111019016 810012324 111019655 810014432 111017349 111018435 111017295 111018375 810021139 111017339 810021125 810017766 111017835 111017506 810017322 111017531 810017896 111017503 810012441 111019565 111019662 111019244 810005747 810017465 810017461 111019651 810017313 111019567 111020134 810021167 810025885 111021146 111021478 111020121 810017763 111019660 810005752 111019661 111019656 111020099 810010299 111017517 810010408 810021156 810018053 810017756 111021046 810010412 810017478 810012319 111021148 111017306 111019013 810017290 111021479 810012364 810004730 810021100 810021123 111021475 111018873 111018816 111021571 810014399 810021094 810021138 810017769 810017967 810021102 111021477 810021103 810014749 810005130 810014731 111018705 111018827 111018815 111018881 810014724 810014737 810014734 810014728 810014727 810014742 810014746 810018048 810018232 111018697 111019015 111019020 810021122 810021135 810021092 810021168 810021134 810021136 810021155 810021151 111018368 111018431 810017320 810017323 810017301 810017309 810017310 810017295 810017291 810018060 810018059 810018065 810018047 810018219 810018221 810018235 810018055 810012313 810012381 810010304 810012444 810012446 810012303 810012291 810012290 810012297 810012437 810018238 810018051 810003650 810003797 810003987 810003746 810004165 810004163 810004061 810003708 810003709 810004118 810003849 810004166 810003844 810005745 810005731 810005728 810005726 810003588 810003710 810003647 810003989 810003707 810003919 810003991 810003920 810003745 810003801 810003744 810003799 810004164 810003848 810004308 810004162 810003845 810004310 810003847 810003846 810004318 810004117 810003798 810003742 111018878 810017898 810014036 810014038 810014433 810014476 810014484 810014493 810014482 810014492 810014470 810005751 810005734 810005746 810005736 810017463 810017464 810017479 810017466 810017473 810018049 810018224 810018233 810018063 810012302 810012365 810012382 810012385 810012330 810012308 810012327 810012274 810012312 810012293 810017760 810005729 810004321 810003648 810003651 810003743 810003649 810003990 810017781 810017900 810017892 810018218 810017299 810017308 810017764 111020082 111021163 111021039 111021134 111021156 111021150 111019017 111019837 111019840 111020111 111020118 111020108 111021481 111021471 111021487 810022555 810022547 810022551 810022554 810022548 810022546 810022550 810022549 810025750 810025738 810025739 810025734 810025758 810025756 810025731 810025688 810025740 810025736 111017298 111017313 111017334 111017321 111017307 111017323 111017297 111017315 111017293 111017335 111017336 111017352 111017354  Entrant: Model Number: VR ICD REF CDVRA300Q UDI-DI Code: 05415067032843 Serial Numbers: 810027235 810016172 111004795 111004775 111033929 111010018 111010037 111022001 111021553 111011580 111002554 111005418 111002524 111022016 111006768 111006759 111006744 111021539 111021550 111010934 111010158 810027691 810027654 111007144 111004783 810011431 810011410 810011436 111009229 111009427 810016404 810015869 810016405 810021390 111002559 111004788 111005432 111010910 111010000 111004658 111004605 810008769 810003697 810005396 810021824 810005901 810005908 810005895 810005841 810004590 111022017 111011720 111011723 810005868 111006729 810011405 810011437 810008545 111002538 810004592 810021633 111003854 810013145 111003946 111013935 111009002 111008328 810021823 111004657 111013942 810019444 810016191 810007476 810011439 111005822 810015895 111004735 810003925 810016225 111009440 810015983 111005441 111013446 111005437 810008541 111002556 810016390 111003939 810005218 810004651 810003924 810005863 810005850 810013515 810016211 810015995 111008334 111005818 111013451 810013395 810005860 810013173 810005834 111011721 111004600 111011735 111011715 111005442 111005157 810003764 810003763 810016025 810015852 810011408 111004597 810015884 111004778 111007159 810005405 810015975 810016205 111002541 111010032 111010030 111002561 111005431 111004629 111004633 111002558 111011727 810004860 810011440 111009999 111006764 810021819 810027701 810021627 111005150 111003940 111010130 111005430 111003937 810016235 111009412 810016170 111004562 810005828 810011402 810013502 810027240 111009235 810027231 810003699 810013150 810015972 810016232 810013493 810021838 111006734 810005483 810016499 810015848 111007135 810003973 810004233 810004228 111013917 111013930 111013920 810027251 810027253 810015886 810016208 810027683 810027700 810027656 111004549 111004763 810021845 810015875 810016014 810016397 810015867 111010132 111004613 111004661 111003942 111010926 111005820 810008547 810008502 810015846 810015868 111005417 111022012 810021820 111004764 111004733 810016204 810027230 111005829 111006741 810015881 810013511 111016668 810013512 810005825 111004561 111004650 810016207 111007174 111019358 111010150 111007142 111004536 111010914 111007131 810013490 810016468 810008550 810005829 810019454 810019469 111016679 810005853 810008532 111007151 810013510 111007187 111007186 111005834 810021637 810015984 810019453 111007130 111016667 111006739 810015888 111042981 810016028 810021813 111004610 111004614 810015878 810027200 810016398 810013507 810027180 111004765 111004798 810016209 810027239 111007133 111010008 111009227 111008989 810013390 111008988 810012987 810027243 810027242 810016020 111006728 111003622 810015966 111036227 810019456 810016032 810011409 111010927 810016210 810016011 111006727 810019458 111005832 810027248 810027214 810027205 810027183 810027208 810027184 810027196 810027223 810027221 810027194 810027201 810027193 810027213 810027185 810027178 810027234 810027220 810027244 111004760 111004773 111004779 111004790 111004740 111004767 111004728 111004557 810027651 810027669 810027686 810027652 810027676 810027655 810027666 810027668 810027688 810027671 810027679 810027670 810027680 810027672 810027675 810027687 810016022 810016168 810016021 810015994 810015890 810013506 810015879 810015850 810015872 810015849 810015856 810019451 810019491 810019476 810027203 810027222 810013518 810013489 810013486 810013487 810013521 810013514 810005859 810016206 111042943 111042993 111042968 111042973 111042990 111042992 111042987 111042980 111042985 111042975 111042961 111042966 111042984 111042969 111042958 111012444 810004861 810015865 111004648 111004607 810003701 111009422 111002531 810003975 111005826 810015889 810011427 111008331 810021905 810021904 111010006 810003580 111005837 810003578 810013391 810013519 111004744 111007171 810005027 111004586 111009435 810019460 810008531 810003640 111008992 111008997 810027215 810008767 111010893 111004596 111007158 810013403 111004635 111005845 111008356 810005481 111002544 810016015 111008815 810004649 810019455 810011429 111004796 111007182 111004794 810005116 810005029 810005115 810003641 810003639 810004778 810004722 111011552 810016234 111009443 810016197 810012993 111010925 111008975 810019484 810021903 810013509 111004627 111008974 111004617 111004724 810015978 111009230 111004585 111009004 111004649 810016492 810013141 810008535 810007479 810003638 810004230 810027188 111011713 111007168 810005831 810027660 111011725 111010929 111010144 810013109 111005434 111008335 111004638 810021901 810005185 810016202 111010949 111005423 111005428 111008814 111010149 111040578 810027252 111009418 111003944 810011433 810021629 810016217 810016230 810021840 810015893 810021900 810013495 111009237 810027236 810021632 810016195 810016200 810015973 810027117 111010123 111011724 810016203 810011426 810011417 810005404 810005402 810005401 810011411 111010157 111007169 111036252 111010038 111010029 111010163 111010023 111021548 111022011 111036232 111036276 111036235 111021549 111021555 111021552 111021547 111002560 111002545 111041265 111008311 111041254 111041279 111041239 111041255 111033916 111033923 111033932 111033920 111033926 111033930 111033922 111033939 111033925 111033931 111033942 111033919 111033921 111033934 111033924 111033918 111033940 111033917 111033933 111033927 111033935 111033938 111033915 111010912 111010915 111010899 111010935 111010928 111036226 111036255 111036239 111010133 111010160 111010152 111012435 111011739 111011585 111011569 111011409 111011559 111011576 111011548 111011719 111012445 111012437 111012439 111021545 111021554 111011417 111036283 111036220 111011751 111011711 111011583 111011712 111040597 111012450 111012436 111012431 111012454 111036218 111036244 111036241 111036251 111036277 111036282 111036219 111036240 111041250 111041237 111041273 111041259 111040588 111036247 111041251 111041244 111041242 111041248 111022018 111022007 111022002 111022021 111022008 111022003 111036229 111036237 111008100 111008097 111008104 111008098 111008801 111008807 111008802 111008805 111036245 111036234 111036216 111036246 111033943 111040617 111036281 111007128 111007199 111007127 111007176 111031232 111031228 111031242 111031235 111031239 111031227 111031238 111031231 111036280 111031236 111009222 111009224 111040607 111011553 111011546 111011406 111033937 111033928 111036279 111036274 111011407 111040619 111040593 111040583 111040585 111041274 111041249 111041264 111041258 111041284 111041263 111041271 111041278 111041269 111041252 111041277 111041253 111041268 111041241 111041238 111022000 111022013 111022020 111021998 111022015 111022009 111022005 111022006 111042945 111042957 111042998 111042977 111043000 111013428 111013454 111013437 111042986 111042988 111042983 111042946 111042978 111042972 111042963 111042974 111013438 111031220 111013944 111013923 111013931 111013928 111031234 111031233 111031243 111033941 111040592 111036250 111036242 111040614 111040609 111040582 111040606 111040605 111040590 111040608 111040589 111040611 111040581 111031229 111031241 111031240 111031237 111031245 111031244 111040623 111031230 111040622 111040574 111041235 111041266 111041240 111033944 111016665 111018060 111018058 111018057 111007146 810003820 111005846 111009421 111002548 111008324 111010920 810003922 111011701 111004784 111007173 810008762 111009415 111011743 810016473 810016166 810016486 810016221 810015855 810013396 810016034 810013397 810004781 810015847 810027229 810003819 810015862 810012991 810016026 810013369 111010904 111006757 111036248 810019452 111010903 810013380 111009444 810005398 810015876 810015851 111010005 810005865 810011401 111011425 111013945 111011728 111009000 111008996 111013433 810004231 810005025 810016220 810021815 111013924 810013370 810015845 111013430 111004616 111008357 111006731 111008338 111008813 810027199 810027190 810027241 810027246 810027233 810027247 810027216 810005836 111010041 810013110 111007178 810027250 111021551 111016684 111004785 111007161 111021546 810027176 111005828 810008508 810003581 810003637 810016215 111019355 111004655 111011414 111011427 810005855 111012433 810027195 111011740 111011413 111008330 111009232 810021906 810019463 810019471 810021834 111009434 810013492 810013374 111004619 111003948 810012984 111010003 810003791 810003818 810005399 111008351 111003949 810013520 810013388 810005837 111013429 111012442 111014956 111005436 111007137 111008347 111008352 111008329 111008333 111004737 111004782 111011577 111011419 111011556 111007194 111011704 111011714 111007201 111002539 111007184 111005840 111005817 111004792 111002535 810015842 111004620 810016480 810019488 810019479 810019466 810004148 111008803 111010021 111009987 111009991 810008543 111005439 111009990 111007162 810004779 810016175 810015891 810016018 111008816 111004537 810016479 810005026 810016484 810004234 111011699 111007203 111004791 111007166 111007177 810005330 810008653 111004556 810013503 810019450 810008510 810016482 810008534 810021848 810027238 810027663 810005219 111004593 810016487 111011571 810008765 111010035 111010010 810003817 810003788 810027226 111004615 111019360 111008350 810015968 111008811 810005406 111008986 810016489 810019486 810027225 111010151 111011543 111011557 111013444 111008993 810011413 111004623 111004628 111004748 111004770 111007147 810005474 111021541 111010155 810004235 810004229 810016171 810004724 810015979 111008809 810016395 810015877 810015873 810008654 810013499 111011408 810016226 810004038 111011550 111011731 111011562 810005408 810008764 810016466 810004863 810016498 810003579 810005838 111009231 810016229 111004599 111007149 111005165 111005836 111002551 810005479 810016403 810019474 810027211 810005717 111008800 810005478 111004772 810013522 810007477 810016214 810016027 810015967 810015988 810015874 810016401 810016402 810016192 810016216 810015866 111003945 111010945 810005830 111004738 810005891 810005547 810004665 111005421 810013496 810005906 111013927 810016471 810027678 111007181 810012986 810004594 111007170 810003821 810005477 810013383 111010924 111013934 810021849 810027659 810003825 111011418 111010947 810012989 810011422 111004786 810027703 810027658 111004747 810005327 111004771 810027116 810021843 810005894 810027175 810003974 111007160 810005186 111003947 111002555 111010892 810005332 810019485 810008770 111010016 111013938 810027177 111010942 810005187 111010142 111002528 111004755 810011430 810013384 111005433 111007198 810003824 810013402 810013485 810005407 810003826 111007164 111009997 111010906 810003700 810004103 111011423 810021897 810005892 810005864 810005852 111008987 810005331 810016493 111008990 111011421 810005861 111011742 111021999 111009416 810013516 810013513 810005220 810019490 810019475 111004588 111004621 810005896 111004612 810005889 810005833 810013144 810013149 810016227 810016231 810013399 810013408 810013142 810016500 810016483 810016477 810016391 111009226 111009433 111002529 810005328 810011438 810019447 810019493 111011564 111021996 810016219 810005824 111011426 111009414 810021844 810004862 810016224 810003787 810011414 810015970 810011428 810008660 810016012 810013497 810019482 111014957 810013148 810015883 111011541 810021827 111006742 111011582 111007165 111006766 111016673 810021839 810008655 810011404 810008768 111010917 111010159 810011407 111000975 111004592 111010936 111016682 810016174 810016389 810011432 810021628 111007189 111008320 111016680 111006758 111007172 111016686 111006675 810011403 111010141 810013504 111004603 111010139 111006746 111008812 111036278 810004780 111004731 810021831 111008991 111008999 111008343 111011737 810004666 810004227 111005161 111010125 111003950 111019361 810027681 810027674 111010907 111010930 111010894 810004663 111013941 810008651 810008659 810015882 810015841 111010940 111004608 111042994 111010012 810015870 810016386 111013452 111011566 810004650 111013432 810008548 810019472 810019448 111007134 810005028 111002562 111011558 111009993 111010007 810003789 810013386 810003823 810003698 810005222 111010939 810005904 111010042 810005480 810008542 111011752 111011554 111013436 111013937 810008549 810011412 111004589 111003941 111001669 111011567 111009989 111010025 111009236 111008316 111006773 111006762 111007140 111004758 111005842 111005425 111004547 111016672 111010154 111017942 111004636 810013171 810004232 810004150 810005118 810005329 810005475 810004593 810003442 810005541 810016400 810016494 810021842 810005548 810005826 810005476 810019487 810021626 111007145 810004591 810004726 810013501 111013929 810015981 111010013 111010901 111010919 111004647 111004601 111004622 111002546 111003626 111002963 111004584 810004039 810003976 810004105 111006755 111004730 111004563 810005856 111007156 810008546 810008544 810013170 111002557 111011561 111013933 111009430 111004590 810013407 111008353 810016213 810015858 810005857 111010033 810004648 810016228 111004598 810004647 111009442 111005154 111005847 810015991 810027179 810016013 810003923 810016467 810011424 810008505 810015885 810027665 111008319 810005400 111006738 810015980 810005866 111003849 111005420 111005440 111005438 810021825 810008656 810008503 111008332 111009225 810005867 810005839 810005858 810005832 810005843 810016399 810021896 810013172 810003762 111005839 810005890 810005898 111011555 111005429 111005435 111008348 810008509 810021898 810021822 810015992 810008506 810005333 810004723 810021908 111004752 111004545 810016491 111011749 111036253 810015993 810016222 111011570 810019468 111011415 810015976 810019445 111010031 810015977 111009998 810021835 810011420 810016496 810004037 810013508 810015864 810015854 810011419 810016495 810004725 810011416 810011421 810003972 810013517 810011434 810015859 111004582 111007163 111007152 810005862 810003765 810013394 111005443 111002518 111005419 111008339 810005543 810027664 111010028 111010136 810013498 810013505 810021631 810008507 810016475 810008761 810015871 111010905 111010138 111002966 810027706 810027209 810013376 810016167 810016396 111007157 810013147 111004659 111003853 810027227 810027228 810005542 810016198 111013440 810021817 111005819 111006740 810004595 111005827 111010923 111016685 810027192 810027219 810027254 810027186 810027202 111010024 111013926 111011579 111011729 111036231 111036228 111036223 111036243 111036275 111006730 111007175 111012438 810019459 111004630 810011435 810004236 810004147 810005117 810005119 111007150 111007179 810027115 810027198 810016488 111006732 111006754 810008658 810008766 810008533 111013453 111013431 111004656 111008976 111006733 810013166 111011718 810016497 111022014 810021392 111010896 810027189 810027212 810027693 810027684 810027699 810027661 810027245 810027224 111003851 111021125 810007398 111021130 810007391 111022019 111002969 810021389 111011588 111011575 810012980 810012995 810013377 810021391 810021623 111010946 810013367 810013392 810013381 111011424 111008354 111011578 111011544 111010898 810027704 810019461 810027705 810027187 111004539 111004606 111009223 111011738 111004725 111004609 111008321 111004652 810013368 111008325 111018301 111006767 810021836 111010134 111002532 111004625 810016016 810019457 810007382 111009423 111013921 111010146 810012973 111008337 111010017 111021127 810012992 810012979 111021126 111010135 810003822 810027662 810027210 810007395 810027218 810027698 810027232 810013385 810027702 810027697 111003645 810007386 810019480 810027682 111021540 111022010 810013484 111003852 111013922 810019467 810019478 111022004 111004651 810012981 111009420 111004602 111009424 111009426 111003633 810021388 111013449 111003850 111008327 111009447 111008342 111008315 111008980 111008978 111008341 810021910 810013108 111011696 111011563 111007167 111012453 111011716 111011422 810007396 111008313 111003936 111011750 111005831 111008982 810027217 111011698 810007394 111004559 810027206 810012975 810012994 111011697 111003641 111005424 111004618 111004594 111011586 810019473 810027673 810027695 810027181 810027692 111011745 111021542 111008344 810012976 111005166 111021543 810027685 111013940 111013946 810012982 111008323 810016392 111018300 111011560 111004653 810007385 810007389 810027653 810021833 111003624 111008326 111011741 111013918 111004646 810013106 111009437 111003636 111003621 111010128 111010918 810007400 810019462 810013404 810013389 111011726 111004645 810021837 111010932 111004632 111007153 810012990 111010895 111002973 111013919 810021624 810027696 810027191 810016387 111003848 111010036 810027237 810012971 810027182 810027677 810027249 810012974 810021909 810012985 810012977 111010002 810012978 810007392 111008314 810012988 111012441 810016169 111021129 810007393 810016394 111013939 111013943 810016485 810016474 111002549 111010127 111004587 810019483 810019464 111010931 810019470 111019356 111004624 810016201 810016223 810021635 111004583 111004626 810013371 810015989 810016218 111008817 810007478 810007399 810007383 810007384 810005221 111008355 111010126 111010145 111011590 810008657 111004637 111008983 810011415 810007390 111002564 111011709 111004793 810004664 810004864 810016393 81000 5715 810005719 810005716 111004595 810004777 111009436 111040615 111041281 111040580 111001660 111002530 111041272 111040600 111041233 111011705 111011547 111009417 111036222 111001664 111042971 111036221 111011717 111040602 111010001 111008981 111042967 111041245 111042989 111042960 111003629 111002526 111002976 111003619 111007138 111041236 111013445 111000966 111000979 111010951 111008317 111011706 111036217 111009001 111042995 111013450 111036225 111033936 111042959 111001670 111010027 111012447 111012452 111011736 111009995 111000982 111009428 111041267 111010143 111042965 111013443 111001657 111002542 111011430 111000971 111001677 111000969 111010937 111010026 111000981 111042982 111011730 111041246 111010921 111010950 111005160 111040618 111008336 111002553 111013427 111007143 111003632 111002550 111011428 111000972 111000962 111042944 111011700 111011568 111040576 111000964 111040586 111001672 111001681 111041283 111042976 111008994 111040575 111012443 111011572 111040621 111040616 111009413 111000973 111013442 111010916 111011549 111040620 111001674 111002972 111002520 111010938 111040599 111010933 111010909 111042991 111012446 111008971 111010009 111010015 111003644 111012448 111007200 111013426 111008099 111001665 111041275 111010034 111036236 111010148 111042997 111011702 111010147 111001663 111009228 810016212 810008652 810015892 810027197 111005823 111005155 810019489 111003623 111010941 111002534 111002974 111040595 111000977 111010922 111041234 111001678 111040603 111000963 111003634 111000968 111041276 111009439 111001667 111040594 111040577 111041280 111003628 111009996 111036249 111041247 111011747 111040579 111002522 111010045 111009445 111011733 111011734 111010014 111011410 111009438 111011581 111011420 111001658 111000980 111003625 111036254 111013435 111000976 111036238 111010129 111008984 111012432 111002519 111007197 111001676 111042962 111007139 111010019 111002533 111002563 111040601 111005825 111003627 111001668 111040596 111009419 111001671 111036224 111040604 111008340 111010004 111042942 810013382 810007397 111010040 111040598 111002540 111003938 111006750 111006747 111000974 111013448 111021124 111008322 111021997 111021128 810021625 111004542 111010948 111011589 111009429 111009432 111010131 111010043 810013523 111005416 810011425 111010022 111008985 111008312 111004739 111013936 810005899 810013494 810004106 810021636 111003631 111003643 111003935 810004102 111010908 810016472 810016476 111016670 810021818 810003926 111016666 810005827 810016029 111008810 810016199 810015853 111004799 111005844 810019492 810007480 810007388 111007190 810013398 810011423 810012983 810021846 111004591 810005482 810013387 810016478 810016470 810019465 810021630 810021899 810021907 810005189 111008977 111005158 111011584 111011542 810005546 810003790 810013143 111011545 111010913 111004634 111003642 810021850 111004631 810005188 111010039 111002523 111002552 810021821 111007195 810013379 111003943 111003635 111002965 111013447 111041257 111041243 111002543 111040610 111041256 810004151 111005427 111011574 111010011 810005902 810003766 111005830 810016469 810005840 111010911 111009988 111009992 111011411 111005821 111001673 111009425 810013488 111004766 111004543 810008501 111011748 810016481 810013393 810021902 810004040 810013500 810005835 111007141 111006745 111011707 111002521 111010900 111008318 810016196 810008504 111007183 111004756 111036230 111040591 111002547 111009233 111003630 111001662 111000983 810004149 810004041 111005164 111042996 111004654 111011551 810015969 810019446 111002537 111004726 111004751 111007192 111011746 111001675 111003620 111003640 111001666 111010161 111003637 111001659 111011429 111003639 111036233 111012449 111010162 111040584 111040613 111042979 111010044 111007196 111008973 111013932 111013925 111001680 111004727 111010140 810021847 810016490 810015843 810015887 810015985 810015974 810015860 810015982 810016031 810016033 111010153 111011587 111000978 111008349 111009221 111013441 111011573 111016669 111005159 111005838 111008979 810013107 111002967 810016173 810007387 810007381 810005851 810005907 810005854 810005403 810016024 810015971 111011703 111011710 111002536 111009446 111007155 111043001 111010944 810003577 111010137 810005849 810013491 111009441 111042970 111042999 810016193 810021634 111040587 810016233 810008763 111041262 810021814 810016388 111009994 111008798 810019449 810015880 810015987 810016035 810016194 810027207 810005842 810005900 111009234 111008998 810013167 810013406 810013400 810013375 810015861 810015894 810016019 810016023 111007129 111007132 111007185 111006760 111006771 111006748 810004662 810013169 810013146 810013168 810011406 810011418 810005903 810005897 810005905 810005893 810005718 810005540 810005539 111004769 111004551 111004754 111004544 111004564 111004757 111004746 111004534 111004787 111004801 111004800 111040950 111040952 111004538 111040951 111005152 111004550 111016683 111004781 111004729 111004789 111004552 111004554 111004743 111004750 111004776 111005835 111004555 111004753 111004546 111040949 111004742 111008345  Model Number: DR ICD REF CDDRA300Q UDI-DI Code: 05415067032836 Serial Numbers: 111041297 111040639 111040940 111041289 111041298 111041300 111041299 111041291 111041303 111041295 111010712 111037464 111037466 111012200 111012202 111012184 111012590 111004224 111037467 810024501 810024477 810024479 810024471 810024468 810024467 810024480 111005128 810020857 111001294 111006923 810017798 810022467 111008619 111009370 111009349 111005480 111005489 111004215 111007812 111007386 111004214 111003794 111001261 111004203 810004763 810012416 111010101 111010681 810014845 810022475 111009365 111002017 111012570 111012572 111008009 111008038 111009360 111008628 111009176 810004661 111010699 111013820 111004403 810004867 810014396 111007779 810012715 111010693 111010112 111003827 111007416 810024503 810020870 111008631 111012591 111001651 810020855 111001258 810005376 111013417 810005392 111003786 111012568 111007611 810020592 810021067 810020869 810020877 111005122 111035375 810022483 111010702 111004410 111007801 111005669 111004415 111001272 111007806 810012407 111008029 111001648 111012580 111010657 111001265 111007591 111010652 111001276 111010116 111010097 111012208 111012181 111012215 111007419 111007810 111007452 111010707 111010679 111010700 111012593 111012589 111012219 111012588 111012573 810021053 810021052 810021065 111006914 111006932 111006934 111006918 111006925 810017408 111004221 111007982 810012399 810008009 810008006 111007835 810012377 111012225 810014722 810012187 111003774 810017404 111008728 111008715 810012551 810012562 810020851 111007793 111007804 810017406 810024509 810022484 810024489 111020904 810012733 111006915 111007420 111007844 111007797 111012205 111010710 111009353 111010703 111004399 810004765 810004589 111007999 111012600 111007969 810017800 111007846 111008013 111006927 810005389 810005388 111010103 111007411 111008032 111008012 111008024 111008018 111004201 111006941 810017990 810017796 810015206 111015860 111008709 111008004 111007830 111005138 111005668 111007387 111006919 810022480 810022469 810022486 810022485 810022482 810022461 810022470 810022481 810022489 810024500 810024498 810024474 810024473 810024463 810024492 810024460 810021048 810021076 810020853 810015205 810015209 810015207 810015211 111042953 111042955 111042952 111042950 111010665 111012176 810022474 111005644 111007464 111008020 111009348 810020860 111001266 111007586 810018000 111007992 810004588 111005648 810020885 111001303 111008030 810017405 810021327 111003761 810012209 111004210 810020866 111010109 111004232 111005475 111005479 810014863 111004406 810020879 810024505 810021051 810012405 810015090 810005176 810017801 810022471 810012409 810020861 111007400 810022494 111012199 111004389 810012429 810012696 111003778 111013407 111009361 111003785 111007968 810018135 111012197 111006935 111004209 111004402 111004411 810020590 111004235 111006929 111001641 810003440 810012553 810020585 111005636 111005640 111007390 111004231 111019351 111001642 111012207 111007600 810012204 111004396 111008614 111008027 111007975 111019335 111001286 111008711 111004404 111012195 111005656 810004721 111003820 810017794 111001649 111004218 810012195 111000941 111003789 111007403 111008724 810021064 111001282 810012203 111012182 111012583 111007990 111003830 111003788 111000946 111001300 111012192 111007384 111007463 111001639 810024507 111010649 810024504 810024508 111005661 810024495 111012214 111000952 111001628 111012223 111042947 111008612 810018134 810020591 810018138 111012586 810014409 111003779 111000956 810012367 111003825 111008040 111008041 111041261 111041304 111041285 111041301 111041292 111041287 111008721 111007401 111007457 111037477 111037460 111037474 111035398 111035381 111035485 111035391 111035393 111035383 111035404 111035400 111035500 111035376 111035388 111035498 111004230 111004412 111004386 111043051 111043046 111043056 111043068 111012185 111012203 111035382 111043040 111035494 111035491 111035493 111035489 111035496 111043059 111043060 111035386 111035402 111035380 111035399 111035403 111035377 111035378 111035385 111035401 111035486 111035387 111035501 111035390 111035384 111035392 111035379 111043062 111043064 111043049 111041286 111041296 111007451 111007418 111007395 111019343 111040941 111040636 111040627 111040625 111043066 111043043 111040631 111040939 111040937 111040938 111043005 111003793 111009172 111043055 111043054 111040643 111040630 111041288 111019329 111019341 111043004 111040638 111043006 111019330 111019323 111007389 111007399 111007405 111007406 111043069 111043052 810017402 810017788 810017787 810021056 111005113 111001295 810008007 111005472 111007832 111009192 111008629 111010096 111012186 111009381 111005112 810012206 111009178 111010694 111007589 810005712 111008028 810021324 111003772 810014405 810014400 111013419 111012595 810022479 111012579 111020902 111005665 810012380 810024481 111012179 111010675 111010120 810005711 111010105 810012566 111005645 111007800 111008003 111007397 810024458 810021073 111006906 810014843 810005022 111006907 810012719 810017992 810017409 810012711 810005369 810024464 111012212 111004394 111003823 111010683 111012188 111013418 111009350 111008737 111007599 111007970 111007414 111008722 111007422 111008001 810003815 111008613 111008633 111008620 111008034 111004226 111012196 810015088 111020905 810005114 810005021 810021054 810021057 810021068 810020876 810021326 810021329 810021072 111008714 111001630 111007436 111019339 111012172 111013832 111006909 111007583 111019347 111012598 111007404 111007971 111007442 810020582 111003821 111000948 810015208 810012564 810012406 111008008 111008021 111009364 111008636 111007818 111008022 111001269 111004212 111004216 111005651 111009378 111007590 111004199 111004222 111001285 111010114 111007783 810003814 810012192 111001280 111003791 810004657 810005024 111010650 810003576 810005215 810004660 111003828 111008716 810017991 111001279 111007977 810005179 111007433 111007828 810017407 810012726 810020598 111009358 810012369 810020872 111007429 111012226 111019327 111009179 111001652 111013826 810014864 111007819 111010669 111010680 810012421 810012423 111007596 111007998 111007417 810005213 111001260 810005386 810022466 810022477 111007826 810024478 810014844 810022478 111012582 810015210 111007465 111010107 810012568 810014401 111003773 810012208 111007845 111007595 111007603 111007973 810003572 111010714 810012202 810005178 810005377 111007995 810012730 810021066 810021328 810020883 111007460 810017805 810015204 111007851 111007833 810022495 810022490 810012189 810012430 810012201 810008010 810008008 111005646 111010695 111007813 111007598 810003816 810018137 810005177 810005394 810004559 810004658 111001650 111004227 810012197 111006912 810012729 810012194 810012379 111008713 810022491 810022464 111013415 111007840 111003776 810012207 111008720 111003784 111000943 810020593 111004228 111005488 111009195 810005370 810012366 111005666 810005385 111007440 810021055 111043048 111007979 810005379 111010708 810012734 810005175 810015089 810012198 111007438 111013411 810020850 810012400 111010659 810020863 810024483 111007972 810024491 111008717 810021074 111007605 111008617 810021331 111007817 111007843 810014402 111007618 810004558 111005674 111010709 111007469 810024497 810020868 810003633 111008627 111013824 111007996 111010715 810020600 810005382 111009363 111005485 111006904 111009180 810004766 810003694 111012224 111012221 111007987 111007424 111006917 810017998 111007978 111008734 111010674 111007447 111009184 810022488 111010121 111010686 810012418 810012635 111012601 111012575 111013828 111007614 111007604 111007803 111007837 810017793 111010690 111007827 111007391 111007388 111007427 111001634 111013818 111010692 111008033 111012189 810017799 810020856 111017533 810005216 111010119 810012567 111000957 810005217 810004868 810004869 111005092 111005081 111005641 111007472 111004388 111004204 810020594 810021063 810020583 111000951 111004400 810012725 810012699 810012731 111003792 810012718 810012717 111004202 111007612 111007796 810014398 810014414 810014406 111008036 111008011 111008019 111003826 111007838 111008002 111009380 111010670 111007798 810012634 810012623 111008730 810014717 111019328 111006910 111006903 111007410 810012403 810012714 111007991 111013416 111007462 111007409 111006922 810012196 810012411 810008004 111007802 111006913 111008632 111006936 111007984 111008719 111008006 111007588 810012554 810017791 111007466 111007834 810012393 810017790 810012413 810014407 810005373 810012392 810012728 810012561 111007795 111007985 810012424 810017989 111007824 111006908 810017786 810012402 111008738 111007836 111008723 111007791 111007788 111008917 111007413 111007585 111007601 111005660 810012712 810012420 111007825 111007809 111043063 111043050 111043058 111043042 111043057 111042956 111043045 111043070 111043053 810012376 810012723 810024484 810024487 111010104 111004408 810014403 111009369 111005633 111012183 111009185 111009175 111012210 111012211 111012569 810012727 111008736 111007847 111005474 111008630 810022465 111006926 111012191 111013420 111013414 810003813 111012174 810024485 810020864 810017988 810024472 111010662 111004405 111004207 810018136 111006930 111007850 111004229 111004531 111007455 111010098 111010108 810004866 810004865 111012581 111019325 111009371 111003824 111009187 111009366 111019331 111010110 111010100 111010672 111010667 111010656 111010647 111019346 810004762 810004718 810004720 810004717 111019321 111019336 111019345 111019350 111019320 111012194 111004213 111013827 111013409 111004532 111019337 111019348 111019324 810015093 810015097 810015095 810015091 810020882 810020884 810020596 810020597 810015094 810015092 810024459 810024502 810024475 810021325 111003762 810003636 111012206 111008025 111010111 810024496 111009354 111007458 111009347 111004223 111012222 810012697 111012574 111007782 111007792 810017997 810005384 111012602 810005023 810012191 111010701 111009346 111008735 810005375 111013819 810005371 810005383 111007581 810024470 810021077 111002016 810012555 111010691 111007997 111004219 111001291 111012173 111007799 111013817 111013825 810012200 111013831 810024469 810017797 810003634 111013830 111007468 810012396 111013815 111007584 810017996 111008718 810005390 810012186 810003573 111005482 810004659 111013813 111013412 111004390 111009362 111007789 111005484 810003695 810012425 810012210 810014408 810020854 111012190 111006940 111001283 111009382 111004414 111007421 810012415 111007385 111007848 111007587 810012199 111010648 810005393 111010676 111007459 111006902 111012227 810017803 810018131 111008007 111006921 111010661 111010704 111010660 111007602 111005481 111005487 111001296 810005113 810005111 810005020 810005110 810005112 810020874 111005477 111005653 810012395 111005136 111005675 810020595 111012603 111001281 111001278 111004407 810017795 810014865 810014866 810005391 111007434 111005133 111005673 111003781 810022476 111001301 810024490 810024493 810024461 810014867 111005476 111007444 111004200 810017410 810020852 111007615 810017411 810024488 111007974 810012722 810014395 810018133 810012698 810021069 111007785 810021071 810012397 810012724 810012720 111007852 810020601 810003575 111007780 111007988 111007593 111007446 111007415 111001647 111004205 111006931 111007423 111008726 111007808 810021061 810012422 111001259 111008712 810005378 111001644 810005214 810003635 111007396 810004764 111001654 111008708 111007619 111005473 111008733 111007607 810014715 810012401 810012721 810012391 810024499 111009190 111003783 111009186 810008005 810012735 810014847 810003632 810022492 810022463 111009352 111010653 111010664 810015096 111012571 111001277 810020587 111018028 111012220 111035487 111035497 111035502 111035394 111035397 111004220 111001275 810020875 810020889 111007989 810012193 111001293 111009173 111009377 111009383 111006924 111013823 111010099 111010117 111012180 111007402 111007453 111007449 111007613 111009355 111020903 111008624 810017403 111004409 810012428 111037468 810018140 111005478 810012700 810012188 111007841 810012378 111009189 111008016 810022462 111005471 810020588 810024476 810022473 810012414 111008037 111019326 111012592 111003822 810014411 111003796 810014404 111012209 810012417 111005490 810014413 111019349 111008618 810021059 111008010 111010688 111008634 111009376 111010697 810012552 111013816 111003819 810020873 810024506 810012408 810024466 810022493 111004211 810012713 810012563 810012625 810012205 810021049 810020880 111010684 111003764 111009351 111001297 111000944 111009177 111007617 810003696 810020584 111001292 111019334 111007786 111008625 111037456 111003771 111004834 111001302 111040634 111037472 111040626 111001631 111010122 111043041 111018304 111041306 111009372 111001271 111005115 111008621 111037469 111008622 111008035 111019340 111040632 111013408 111000950 111008031 111000958 111041307 111007609 111006920 111043061 111040641 111004533 111001646 111001643 111040628 111008017 111008023 111013410 111013425 111019352 111001298 111043002 111001633 111009174 111004401 111035396 111035492 111035499 111042951 111010118 111000955 111018305 111010654 111035490 111010115 111001627 111004225 111043003 111041302 111035389 111043065 111019332 111010685 111037462 111040635 111010678 111043008 111001267 810005387 111005670 810003574 111009193 111005486 810015203 111041290 111004393 111003766 111004413 111008635 111009367 111001273 111012177 111035504 111001263 111000940 111018302 111001257 111041260 111042948 111041293 111000947 111037471 111010671 111010677 111037463 111042954 111008616 111010682 111001632 111010663 111007407 111037459 111009171 111001287 111001640 111042949 111037475 111007993 111001264 111012171 111009356 111040637 111000953 111010705 111043039 111010668 111000949 111012204 111019342 111009183 111001655 111035503 111009385 111040633 111013413 111040629 111037458 111009191 810020862 111003831 111012187 111007854 111012594 111019338 111008626 111008725 810021070 111009188 111003765 111008026 111012599 810005395 810008003 111009373 111001629 111001288 111040624 111003782 810020888 111007471 111004387 111007981 810020871 111007822 810012410 111007435 810024465 111008727 111009194 810017999 111004395 111004233 111004206 810012569 111001637 810008002 810014412 810018139 810022487 810021332 810024494 810005713 810005714 810014714 810008001 111013822 111012577 111009181 111007831 111008039 111007432 111007430 810003812 810012570 810005372 111007983 111000942 111003775 111001636 810014719 111007986 111008000 111041294 111001645 111003768 111010673 111001268 111008015 111010716 111003777 111012175 810021050 111005655 111007408 810014718 810022472 810012419 111004208 810020589 111001653 111006939 111007805 111006928 111007428 111007816 111007811 111037470 111012578 111001290 810020881 111001270 810024462 111035488 111001289 111018303 111012584 111007814 111007437 111007821 111007608 111037461 111037455 111037465 111037478 111037476 111037457 111037473 111005093 111001304 111008014 111013821 111013829 111012201 111007829 111041305 111004391 810012716 810012404 111003767 111001274 111004815 111015861 810012368 111007398 111007450 111007849 111006933 810014716 810014720 111019322 810014721 111013421 111013423 111040640 111005652 111007392 810017792 111005483 810017802 111008707 810012427 111007426 111010651 810017789 111003763 810014713 810020865 111008623 111008042 111003770 111019344 111037479 111010102 111035395 111043047 810004719 810012732 810014394 810012394 111010666 111013424 810020878 810021323 810005380 810012426 111006938 111000954 111012178 111007853 111001256 111012193 111035495 810020599 810022468 111019333 810012412 810014397 111001635 810018132 810012398 810024482 111007820 111009379 810024457 810020859 810020887 810020867 111010655 810014846 111007980 810012370 111009384 111009375 810021075 810021062 810020886 111007431 111007441 111007393 111007610 111004392 111043067 810005374 810005381 111003829 111003790 111012597 111012587 111012585 111010658 111010706 111018306 111010687 111010713 111010698 111010711 111005137 111040948 111032391 111005095 111032754 111005658 111032750 111005667 111005103 111004840 111032751 111004832 111005100 111032389 111040943 111040945 111032747 111004807 111040946 111004824 111005632 111005096 111005106 111032388 111005664 111005654 111007594 111005089 111005679 111007454 111005132 111005080 111032390 111004843 111005140 111004836 111004818 111004854 111005126 111004845 111004837 111005131 111004813 111004829 111004833 111005082 111004852 111032745 111032752 111017534 111040942 111004849 111004805 111005130 111005083 111005647 111004846 111004812 111005657 111005659 111005680 111005109 111005117 111004806 111005120 111005142 111005090 111005088 111007794 111007597 111006937 111007807 111032387 111005135 111005635 111005139 111005637 111005677 111004804 111005108 111032748 111008710 111005084 111005099 111032753 111040944 111017535 111032749 111004826 111017537 111004850 810012565 111012216 111012217 111012218 111012576 111032386 111032392 111032385 111040947"
    }
  ]
}