  "results": [
    {
      "country": "United States",
      "city": "Abbott Park",
      "address_1": "100 Abbott Park Rd",
      "reason_for_recall": "Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with  serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.",
      "address_2": "",
      "product_quantity": "7981 kits",
      "code_info": "List number: 8K26-20, Lot Numbers: 01916E000, 00316F000, 01816G000, 00916I000, 07316I000, 00916L000 ;     List Number: 8K26-25, Lot Numbers: 01816E000, 00216F000, 01716G000, 00816I000, 07216I000",
      "center_classification_date": "20170619",
      "distribution_pattern": "Worldwide Distribution -  USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MONTENEGRO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UNITED KINGDOM, URUGUAY, VIETNAM",
      "state": "IL",
      "product_description": "ARCHITECT SHBG Reagent    The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.",
      "report_date": "20170628",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Abbott Laboratories",
      "recall_number": "Z-2613-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77070",
      "termination_date": "20190603",
      "more_code_info": "",
      "recall_initiation_date": "20170418",
      "postal_code": "60064-3502",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}