      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74010",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, \"Palestinian Territory, Occupied\", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.",
      "recall_number": "Z-2610-2016",
      "product_description": "Beacon Tip Torcon NB Advantage Catheter  Catheter, Intravascular, Diagnostic",
      "product_quantity": "2,662,505",
      "reason_for_recall": "Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation.  degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.",
      "recall_initiation_date": "20160415",
      "center_classification_date": "20160822",
      "termination_date": "20170804",
      "report_date": "20160831",
      "code_info": "All Lots.  Catalog prefixes HNBR5.0 and HNBR6.0    Catalog/product number; Global/order number  HNBR5.0-35-100-P-NS-0 G09916  HNBR5.0-35-100-P-NS-C1 G08623  HNBR5.0-35-100-P-NS-C2 G08796  HNBR5.0-35-100-P-NS-CANADA-041480 G12570  HNBR5.0-35-100-P-NS-DAV G08699  HNBR5.0-35-100-P-NS-FC3 G11204  HNBR5.0-35-100-P-NS-H1 G08421  HNBR5.0-35-100-P-NS-H3 G08918  HNBR5.0-35-100-P-NS-HN1 G09741  HNBR5.0-35-100-P-NS-HN4 G08917  HNBR5.0-35-100-P-NS-HNK1 G27811  HNBR5.0-35-100-P-NS-JB1 G08472  HNBR5.0-35-100-P-NS-JB2 G08809  HNBR5.0-35-100-P-NS-JB3 G08577  HNBR5.0-35-100-P-NS-MAN G08511  HNBR5.0-35-100-P-NS-MIK G11652  HNBR5.0-35-100-P-NS-MPA G11468  HNBR5.0-35-100-P-NS-SIM1 G08471  HNBR5.0-35-100-P-NS-SIM2 G08422  HNBR5.0-35-100-P-NS-SIM3 G08597  HNBR5.0-35-100-P-NS-VERT G11483  HNBR5.0-35-100-P-NS-WAIKATO G13803  HNBR5.0-35-100-P-NS-WNBG G08805  HNBR5.0-35-125-P-NS-C2 G12899  HNBR5.0-35-125-P-NS-C-PS G31970  HNBR5.0-35-125-P-NS-DAV G11375  HNBR5.0-35-125-P-NS-H1 G13258  HNBR5.0-35-125-P-NS-SIM2 G11658  HNBR5.0-35-125-P-NS-VANSCHIE1 G35733  HNBR5.0-35-125-P-NS-VANSCHIE2 G35734  HNBR5.0-35-125-P-NS-VANSCHIE3 G35735  HNBR5.0-35-125-P-NS-VANSCHIE4 G35736  HNBR5.0-35-125-P-NS-VANSCHIE5 G35737  HNBR5.0-35-125-P-NS-VS1 G35974  HNBR5.0-35-125-P-NS-WAIKATO G27813  HNBR5.0-35-40-P-NS-KMP G09751  HNBR5.0-35-40-P-NS-TEGT G13804  HNBR5.0-35-40-P-NS-VANSCHIE1 G31410  HNBR5.0-35-40-P-NS-VANSCHIE2 G31409  HNBR5.0-35-40-P-NS-VANSCHIE3 G31408  HNBR5.0-35-60-P-NS-BELL-1.5 G27815  HNBR5.0-35-65-P-NS-C1 G08441  HNBR5.0-35-65-P-NS-C2 G08442  HNBR5.0-35-65-P-NS-C3 G08443  HNBR5.0-35-65-P-NS-DAV G11414  HNBR5.0-35-65-P-NS-KMP G11458  HNBR5.0-35-65-P-NS-LEV1 G11459  HNBR5.0-35-65-P-NS-LEV2 G11460  HNBR5.0-35-65-P-NS-RC1 G08469  HNBR5.0-35-65-P-NS-RC2 G08470  HNBR5.0-35-65-P-NS-RDC G08596  HNBR5.0-35-65-P-NS-RIM G08444  HNBR5.0-35-65-P-NS-RLG G11380  HNBR5.0-35-65-P-NS-SHK0.8 G32368  HNBR5.0-35-65-P-NS-SIM1 G09853  HNBR5.0-35-65-P-NS-SIM2 G11478  HNBR5.0-35-65-P-NS-VANSCHIE1 G13795  HNBR5.0-35-65-P-NS-VANSCHIE2 G13794  HNBR5.0-35-65-P-NS-VANSCHIE3 G13792  HNBR5.0-35-65-P-NS-VANSCHIE4 G13783  HNBR5.0-35-65-P-NS-VANSCHIE5 G13793  HNBR5.0-35-70-P-NS-DAV G27820  HNBR5.0-35-80-P-NS-C1 G10063  HNBR5.0-35-80-P-NS-C2 G11202  HNBR5.0-35-80-P-NS-CHG4 G13711  HNBR5.0-35-80-P-NS-JR3.5 G11453  HNBR5.0-35-80-P-NS-MPA G09888  HNBR5.0-35-80-P-NS-RDC G13703  HNBR5.0-35-80-P-NS-RIM G12595  HNBR5.0-38-100-P-NS-C1 G09344  HNBR5.0-38-100-P-NS-C2 G09967  HNBR5.0-38-100-P-NS-CK1 G08839  HNBR5.0-38-100-P-NS-DAV G08581  HNBR5.0-38-100-P-NS-H1 G08419  HNBR5.0-38-100-P-NS-H3 G11199  HNBR5.0-38-100-P-NS-HN4 G11200  HNBR5.0-38-100-P-NS-HN5 G09899  HNBR5.0-38-100-P-NS-JB1 G08583  HNBR5.0-38-100-P-NS-JB2 G09654  HNBR5.0-38-100-P-NS-JB3 G11192  HNBR5.0-38-100-P-NS-JIM G11446  HNBR5.0-38-100-P-NS-JR4 G10874  HNBR5.0-38-100-P-NS-KMP G11457  HNBR5.0-38-100-P-NS-MAN G11201  HNBR5.0-38-100-P-NS-MPA G11205  HNBR5.0-38-100-P-NS-MPB G11206  HNBR5.0-38-100-P-NS-RDC G11379  HNBR5.0-38-100-P-NS-RIM G12870  HNBR5.0-38-100-P-NS-SHETTY-050181 G11475  HNBR5.0-38-100-P-NS-SIM1 G08584  HNBR5.0-38-100-P-NS-SIM2 G08420  HNBR5.0-38-100-P-NS-SIM3 G08585  HNBR5.0-38-100-P-NS-TEGT G11193  HNBR5.0-38-100-P-NS-VERT G11484  HNBR5.0-38-100-P-NS-VS2 G11487  HNBR5.0-38-100-P-NS-VS3 G11488  HNBR5.0-38-100-P-NS-VTK G27829  HNBR5.0-38-100-P-NS-WNBG G08582  HNBR5.0-38-110-P-NS-VERT G12128  HNBR5.0-38-120-P-NS-DAV G27830  HNBR5.0-38-125-P-NS-H1 G11378  HNBR5.0-38-125-P-NS-HN2 G11523  HNBR5.0-38-125-P-NS-HN5 G13119  HNBR5.0-38-125-P-NS-JB1 G11726  HNBR5.0-38-125-P-NS-JB2 G11883  HNBR5.0-38-125-P-NS-MPA G12154  HNBR5.0-38-125-P-NS-SIM1 G12072  HNBR5.0-38-125-P-NS-SIM2 G11517  HNBR5.0-38-125-P-NS-VERT G12170  HNBR5.0-38-125-P-NS-VTK G11642  HNBR5.0-38-40-P-NS-0 G24825  HNBR5.0-38-40-P-NS-KMP G09548  HNBR5.0-38-65-P-NS-C1 G08446  HNBR5.0-38-65-P-NS-C2 G08447  HNBR5.0-38-65-P-NS-C3 G08448  HNBR5.0-38-65-P-NS-DAV G11435  HNBR5.0-38-65-P-NS-FINCK-101979 G11437  HNBR5.0-38-65-P-NS-H1 G11395  HNBR5.0-38-65-P-NS-JB1 G11208  HNBR5.0-38-65-P-NS-KMP G09469  HNBR5.0-38-65-P-NS-LEV1 G11194  HNBR5.0-38-65-P-NS-LEV2 G11207  HNBR5.0-38-65-P-NS-LEV3 G11461  HNBR5.0-38-65-P-NS-MPA G11195  HNBR5.0-38-65-P-NS-MPB G11470  HNBR5.0-38-65-P-NS-RC1 G08641  HNBR5.0-38-65-P-NS-RC2 G08642  HNBR5.0-38-65-P-NS-RDC G08578  HNBR5.0-38-65-P-NS-RH G10080  HNBR5.0-38-65-P-NS-RIM G10082  HNBR5.0-38-65-P-NS-RLG G09371  HNBR5.0-38-65-P-NS-SHK1.0 G11476  HNBR5.0-38-65-P-NS-SIM1 G08677  HNBR5.0-38-65-P-NS-SIM2 G08625  HNBR5.0-38-65-P-NS-SIM3 G08849  HNBR5.0-38-65-P-NS-TEGT G11377  HNBR5.0-38-65-P-NS-VERT G11485  HNBR5.0-38-65-P-NS-VS1 G11486  HNBR5.0-38-65-P-NS-VS2 G12715  HNBR5.0-38-65-P-NS-VSC1 G20791  HNBR5.0-38-65-P-NS-VSC2 G20038  HNBR5.0-38-65-P-NS-VSC3 G21042  HNBR5.0-38-75-P-NS-CHG2.5 G27848  HNBR5.0-38-75-P-NS-CHG3 G30040  HNBR5.0-38-75-P-NS-CHG-D G27846  HNBR5.0-38-80-P-NS-C1 G11203  HNBR5.0-38-80-P-NS-C2 G08586  HNBR5.0-38-80-P-NS-CHG2.5 G11430  HNBR5.0-38-80-P-NS-CHG3 G13573  HNBR5.0-38-80-P-NS-CHG-A G13559  HNBR5.0-38-80-P-NS-CHG-B G13560  HNBR5.0-38-80-P-NS-CHG-C G13561  HNBR5.0-38-80-P-NS-CHG-D G13562  HNBR5.0-38-80-P-NS-CHG-E G13563  HNBR5.0-38-80-P-NS-JR1 G13370  HNBR5.0-38-80-P-NS-MIK G09312  HNBR5.0-38-80-P-NS-MK1A G11465  HNBR5.0-38-80-P-NS-MPA G11467  HNBR5.0-38-80-P-NS-MPB G11469  HNBR5.0-38-80-P-NS-RC1 G11530  HNBR5.0-38-80-P-NS-RC2 G11471  HNBR5.0-38-80-P-NS-RDC G12293  HNBR5.0-38-80-P-NS-RH G27852  HNBR5.0-38-80-P-NS-RIM G10028  HNBR5.0-38-80-P-NS-RLG G27853  HNBR5.0-38-80-P-NS-SIM2 G11886  HNBR5.0-38-80-P-NS-VS G11209  HNBR5.0-38-80-P-NS-VS1 G11189  HNBR5.0-38-80-P-NS-VS2 G11190  HNBR5.0-38-80-P-NS-VS3 G11191  HNBR6.0-38-40-P-NS-KMP G27856  HNBR6.0-38-65-P-NS-0 G27858  HNBR6.0-38-65-P-NS-KMP G27860",
      "more_code_info": ""
    }
  ]
}