  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Dover",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80466",
      "recalling_firm": "Zimmer Surgical Inc",
      "address_1": "200 W Ohio Ave",
      "address_2": "N/A",
      "postal_code": "44622-9642",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Distributors in AL, AZ, CA, FL, IA, IL, MA, MI, MN, MS, NH, NJ, NY, OH, RI, SD, and TX. Foreign distribution to Canada, Australia, China, and Netherlands.",
      "recall_number": "Z-2602-2018",
      "product_description": "Zimmer Dermatone AN, Model No. 88710100",
      "product_quantity": "140",
      "reason_for_recall": "Devices may have a loose control bar, which could compromise the ability to control the thickness of the graft. Devices could also have a loose width plate that if undetected, could result in an imperfect, yet still usable graft. Potential risks include harvesting a skin graft of a thickness much greater than intended, described as patient gouged, minor delay of surgery, and sub-optimal (but usable) graft.",
      "recall_initiation_date": "20180608",
      "center_classification_date": "20180802",
      "termination_date": "20190719",
      "report_date": "20180808",
      "code_info": "Lots 63578134 , 63578135, 63592351, 63618116, 63646910, 63817639, 63898373"
    }
  ]
}