  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80388",
      "recalling_firm": "Becton, Dickinson and Company, BD Biosciences",
      "address_1": "2350 Qume Dr",
      "address_2": "N/A",
      "postal_code": "95131-1812",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "US Distribution was made to FL, KS, KY, MA, NJ, and SD.",
      "recall_number": "Z-2598-2018",
      "product_description": "BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only.  The firm name on the label is Becton, Dickinson and Company, BD Biosciences, San Jose, CA.     The Loader offers various settings to resuspend and mix samples. It can draw from 12 x 75-mm tubes in 30- and 40-tube racks. A barcode reader verifies the ID on tube racks and individual tubes in 30- tube racks. A built-in imaging system provides safety checks, such as verifying the correct rack type and tube layout, and ensures the tubes were loaded correctly.",
      "product_quantity": "8 devices",
      "reason_for_recall": "The solenoid plunger tip length is too short  resulting in insufficient engagement to the Universal Loader enclosure to lock during use.",
      "recall_initiation_date": "20180516",
      "center_classification_date": "20180802",
      "termination_date": "20180813",
      "report_date": "20180808",
      "code_info": "Serial #2662879000001, UDI (01)00382906628794(11)170915(21)Z662879000001;     Serial #2662879000002, UDI (01)00382906628794(11)170915(21)Z662879000002;     Serial #2662879000003, UDI (01)00382906628794(11)170915(21)Z662879000003;     Serial #2662879000004, UDI (01)00382906628794(11)170915(21)Z662879000004;     Serial #2662879000009, UDI (01)00382906628794(11)170915(21)Z662879000009;     Serial #2662879000010; UDI (01)00382906628794(11 )170915(21)Z662879000010;     Serial #2662879000011; UDI (01)00382906628794(11)170915(21)Z662879000011; and     Serial #2662879000012, UDI (01)00382906628794(11)170915(21)Z662879000012."
    }
  ]
}