  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77141",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: Puerto Rico, AK  AL  AR  AZ  CA  CO  CT  FL  GA  HI  IA  ID  IL  IN  KS  KY  LA  MA  MD  ME  MI  MN  MO  MS  MT  NC  ND  NE  NH  NJ  NM  NV  NY  OH  OK  OR  PA  RI  SC  SD  TN  TX  UT  VA  VT  WA  WI  WV  WY    Foreign: CANADA, ALGERIA                    ARGENTINA  AUSTRALIA                  BELGIUM  BRAZIL  BRIDGEND WALES  CHINA  DUBAI UAE                  EGYPT                      EL SALVADOR                ENGLAND  ESCHBACH GERMANY  EXPORT DORAL FL  FRANCE                     GERMANY  GUATEMALA                  HONDURAS  INDIA  ITALY  JAMAICA  JAPAN  JORDAN  KOREA  LEBANON                    DELIVERY SERVICE (LYNBROOK  NY)  MALAYSIA  MEXICO  MIAMI FL  MOROCCO  NETHERLANDS                NEW ZEALAND  NICARAGUA                  PANAMA  PERU  PHARR TEXAS  POLAND  PUERTO RICO  RIO DE JANEIRO  SAUDI ARABIA               SINGAPORE  SPAIN  SWITZERLAND  TAIWAN  THAILAND  THE NETHERLANDS            TRIPOLI  UK  VIRGIN ISLANDS  VA/DOD: Both see addresses below",
      "recall_number": "Z-2598-2017",
      "product_description": "FEM IM NAIL 14MMDX30CM LEFT, FEM IM NAIL 14MMDX30CM RIGHT  FEM IM NAIL 13MMDX32CM LEFT  FEM IM NAIL 14MMDX32CM LEFT  FEM IM NAIL 13MMDX32CM RIGHT  FEM IM NAIL 14MMDX32CM RIGHT  FEM IM NAIL 11MMDX34CM LEFT  FEM IM NAIL 13MMDX34CM LEFT  FEM IM NAIL 14MMDX34CM LEFT  FEM IM NAIL 11MMDX34CM RIGHT  FEM IM NAIL 13MMDX34CM RIGHT  FEM IM NAIL 14MMDX34CM RIGHT  FEM IM NAIL 11MMDX36CM LEFT  FEM IM NAIL 12MMDX36CM LEFT  FEM IM NAIL 13MMDX36CM LEFT  FEM IM NAIL 14MMDX36CM LEFT  FEM IM NAIL 11MMDX36CM RIGHT  FEM IM NAIL 12MMDX36CM RIGHT  FEM IM NAIL 13MMDX36CM RIGHT  FEM IM NAIL 14MMDX36CM RIGHT  FEM IM NAIL 10MMDX38CM LEFT  FEM IM NAIL 11MMDX38CM LEFT  FEM IM NAIL 12MMDX38CM LEFT  FEM IM NAIL 13MMDX38CM LEFT  FEM IM NAIL 14MMDX38CM LEFT  FEM IM NAIL 10MMDX38CM RIGHT  FEM IM NAIL 11MMDX38CM RIGHT  FEM IM NAIL 12MMDX38CM RIGHT  FEM IM NAIL 13MMDX38CM RIGHT  FEM IM NAIL 14MMDX38CM RIGHT  FEM IM NAIL 10MMDX40CM LEFT  FEM IM NAIL 11MMDX40CM LEFT  FEM IM NAIL 12MMDX40CM LEFT  FEM IM NAIL 13MMDX40CM LEFT  FEM IM NAIL 14MMDX40CM LEFT  FEM IM NAIL 10MMDX40CM RIGHT  FEM IM NAIL 11MMDX40CM RIGHT  FEM IM NAIL 12MMDX40CM RIGHT  FEM IM NAIL 13MMDX40CM RIGHT  FEM IM NAIL 14MMDX40CM RIGHT  FEM IM NAIL 10MMDX42CM LEFT  FEM IM NAIL 11MMDX42CM LEFT  FEM IM NAIL 12MMDX42CM LEFT  FEM IM NAIL 13MMDX42CM LEFT  FEM IM NAIL 14MMDX42CM LEFT  FEM IM NAIL 10MMDX42CM RIGHT  FEM IM NAIL 11MMDX42CM RIGHT  FEM IM NAIL 12MMDX42CM RIGHT  FEM IM NAIL 13MMDX42CM RIGHT  FEM IM NAIL 14MMDX42CM RIGHT  FEM IM NAIL 10MMDX44CM LEFT  FEM IM NAIL 11MMDX44CM LEFT  FEM IM NAIL 12MMDX44CM LEFT  FEM IM NAIL 13MMDX44CM LEFT  FEM IM NAIL 14MMDX44CM LEFT  FEM IM NAIL 10MMDX44CM RIGHT  FEM IM NAIL 11MMDX44CM RIGHT  FEM IM NAIL 12MMDX44CM RIGHT  FEM IM NAIL 13MMDX44CM RIGHT  FEM IM NAIL 14MMDX44CM RIGHT  FEM IM NAIL 10MMDX46CM LEFT  FEM IM NAIL 11MMDX46CM LEFT  FEM IM NAIL 12MMDX46CM LEFT  FEM IM NAIL 13MMDX46CM LEFT  FEM IM NAIL 14MMDX46CM LEFT  FEM IM NAIL 10MMDX46CM RIGHT  FEM IM NAIL 11MMDX46CM RIGHT  FEM IM NAIL 12MMDX46CM RIGHT  FEM IM NAIL 13MMDX46CM RIGHT  FEM IM NAIL 14MMDX46CM RIGHT  FEM IM NAIL 10MMDX48CM LEFT  FEM IM NAIL 11MMDX48CM LEFT  FEM IM NAIL 12MMDX48CM LEFT  FEM IM NAIL 13MMDX48CM LEFT  FEM IM NAIL 10MMDX48CM RIGHT  FEM IM NAIL 11MMDX48CM RIGHT  FEM IM NAIL 12MMDX48CM RIGHT  FEM IM NAIL 13MMDX48CM RIGHT  FEM IM NAIL 10MMDX50CM LEFT  FEM IM NAIL 11MMDX50CM LEFT  FEM IM NAIL 12MMDX50CM LEFT  FEM IM NAIL 13MMDX50CM LEFT  FEM IM NAIL 10MMDX50CM RIGHT  FEM IM NAIL 11MMDX50CM RIGHT  FEM IM NAIL 12MMDX50CM RIGHT  FEM IM NAIL 13MMDX50CM RIGHT",
      "product_quantity": "24969",
      "reason_for_recall": "Zimmer Biomet is initiating a field action for sterile-packaged implants packaged in a single package configuration due to packaging design verification test failures.",
      "recall_initiation_date": "20170428",
      "center_classification_date": "20170615",
      "termination_date": "20190515",
      "report_date": "20170621",
      "code_info": "510K: K965098,K032367  Item Numbers: 00225730004  00225730014  00225732003  00225732004  00225732013  00225732014  00225734001  00225734003  00225734004  00225734011  00225734013  00225734014  00225736001  00225736002  00225736003  00225736004  00225736011  00225736012  00225736013  00225736014  00225738000  00225738001  00225738002  00225738003  00225738004  00225738010  00225738011  00225738012  00225738013  00225738014  00225740000  00225740001  00225740002  00225740003  00225740004  00225740010  00225740011  00225740012  00225740013  00225740014  00225742000  00225742001  00225742002  00225742003  00225742004  00225742010  00225742011  00225742012  00225742013  00225742014  00225744000  00225744001  00225744002  00225744003  00225744004  00225744010  00225744011  00225744012  00225744013  00225744014  00225746000  00225746001  00225746002  00225746003  00225746004  00225746010  00225746011  00225746012  00225746013  00225746014  00225748000  00225748001  00225748002  00225748003  00225748010  00225748011  00225748012  00225748013  00225750000  00225750001  00225750002  00225750003  00225750010  00225750011  00225750012  00225750013  Lot Numbers: 60667034  60667034R  60668410  60668410R  60598252  60654244  60671255  60654244R  60671255R  60654236  60684926  60684999  60654236R  60684926R  60684999R  60629337  60629337R  60691903  60691928  60691934  60691903R  60691928R  60691934R  60695640  60684996  60684996R  60671256  60671258  60691915  60695638  60671256R  60671258R  60691915R  60695638R  60691893  60691901  60691909  60691911  60695628  60691901R  60691909R  60691911R  60695628R  367281  367297  60693123  60695646  60711202  60713644  367281R  367297R  60693123R  60695646R  60711202R  60713644R  60691907  60691913  60691935  60715185  60715189  60718446  60719890  60691907R  60691935R  60715185R  60715189R  60718446R  60719890R  60691906  60691910  60695619  60695620  60708944  60691906R  60691910R  60695619R  60695620R  60708944R  60695631  60715178  60715182  60715188  60695631R  60715178R  60715182R  60715188R  60691908  60715186  60691908R  60715186R  60691912  60691914  60691912R  60691914R  60715181  60715183  60718445  60718448  60715181R  60715183R  60718445R  60718448R  60691933  60715187  60718444  60718447  60721968  60725140  60725141  60725142  60691933R  60715187R  60718444R  60718447R  60721968R  60725140R  60725142R  60728324  60728324R  60715180  60715179  60732935  60739212  60732935R  60739212R  60732958  60739211  60732958R  60732953  60744655  60732953R  60744655R  60695634  60732948  60732955  60744662  60744669  60695634R  60732948R  60732955R  60744662R  60744669R  60747498  60747544  60747560  60747498R  60747560R  60715184  60747518  60715184R  60747518R  367304  60732956  60744665  60744667  60744670  60747574  60749830  60732956R  60744665R  60744667R  60744670R  60747574R  60749830R  60747573  60747629  60747634  60754399  60754401  60755628  60747573R  60747634R  60754401R  60747639  60747639R  60747509  60747509R  60750992  60750992R  60732954  60732954R  60747645  60747650  60747506  60747568  60747570  60747637  60750993  60764756  60766461  60747506R  60747568R  60747570R  60747637R  60750993R  60764756R  60766461R  60755623  60771519  60772793  60755623R  60771519R  60772793R  60776457  60776457R  60629186  60629189  60629190  60629191  60629192  60630476  60747521  60755624  60629186R  60629189R  60629191R  60747521R  60750991  60750991R  60776458  60776458R  60755621  60755622  60776463  60777926  60777972  60777977  60777989  60777991  60781793  60755621R  60755622R  60776463R  60777972R  60777977R  60777989R  60777991R  60772794  60777974  60777992  60783535  60784485  60787246  60772794R  60777974R  60783535R  60784485R  60777987  60777990  60783536  60787257  60777987R  60777990R  60783536R  60787257R  60777973  60777975  60777980  60777984  60777988  60783534  60787205  60787228  60792396  60777973R  60777975R  60777980R  60777984R  60777988R  60783534R  60787205R  60787228R  60792396R  60777981  60777985  60787221  60787229  60787240  60777981R  60777985R  60787221R  60777968  60777976  60777986  60777968R  60777976R  60777986R  60777978  60777983  60784486  60787204  60787253  60777978R  60777983R  60784486R  367291  367580  367582  367583  367584  367585  60777979  60777982  60787248  60787255  60787256  60792386  60792390  60792397  60797057  367580R  367582R  367584R  60777979R  60787248R  60787255R  60792386R  60792390R  60792397R  60797057R  60777928  60787237  60792391  60792394  60797056  60792391R  60792394R  367581  60787254  60792392  60792393  60792392R  60792393R  60792395  60792395R  60781794  60792398  60809876  60781794R  60792398R  60809862  60809873  60809878  60809879  60809880  60809882  60809884  60809862R  60809878R  60809879R  60809884R  60809875  60809883  60809875R  60809883R  60809877  60812656  60812669  60812674  60812675  60812678  60812656R  60812669R  60812675R  60809881  60812661  60812671  60812680  60812681  60812682  60809881R  60812661R  60812671R  60812680R  60812682R  60812668  60812659  60812662  60812665  60812670  60819944  60812662R  60812665R  60812670R  60819944R  60812660  60812663  60819947  60820073  60812660R  60819947R  60812677  60819962  60819962R  60812658  60812658R  60819977  60819977R  60819946  60819946R  60812654  60819968  60820004  60820068  60820069  60820072  60825062  60825065  60825068  60825078  60825079  60812654R  60825062R  60825068R  60825078R  60825079R  60812676  60819956  60819974  60819976  60829839  60819956R  60819976R  60819999  60820067  60829841  60819999R  60819960  60819973  60819975  60825074  60829840  60819960R  60819975R  60825074R  60829840R  60819972  60820070  60825055  60819972R  60820070R  60825055R  60812673  60819978  60812673R  60819978R  60819961  60836549  60819961R  60820001  60836551  60836551R  60812655  60836553  60812655R  60836553R  60812667  60812667R  60825071  60836550  60825071R  60836550R  60839981  367909  368003  368014  60820074  60839978  60847909  60851221  60851368  60851412  367909R  368014R  60820074R  60851368R  60851412R  367907  367913  367914  368015  60825056  60839992  60847916  60847917  60851384  367907R  60825056R  60847916R  60851384R  367906  367911  367917  368004  368005  368008  368009  368012  368016  60819958  60851376  60851380  60819958R  60851376R  60851380R  367910  368001  368006  368007  368011  368177  368189  60819969  60851272  60851323  60851418  368177R  368189R  60819969R  60851272R  60851323R  367915  367916  368017  368017R  368013  60858240  60858240R  368276  60858243  60858246  368276R  60858243R  368325  60847919  60858242  60858244  60864518  368325R  60858242R  60858244R  60864518R  367912  368186  368190  368191  60836554  60868008  368191R  60868008R  367908  368002  368185  368188  60819991  60858239  60864519  60868002  60868866  367908R  368185R  368188R  60819991R  60858239R  60868866R  368010  368184  60847918  60847918R  368187  60880785  368187R  368379  368381  60868010  368379R  60868010R  368180  368383  60885923  60885926  60829842  368444  368382  60868009  60883670  60868009R  60883670R  368388  368483  368563  368483R  368564  368594  60851184  368594R  60851184R  368484  368600  368622  368624  368637  368638  368629  368648  368650  368654  368660  368661  368728  368733  368725  368604  368610  368613  368615  368617  368619  368620  368626  368645  368646  368619R  368735  368748  368623  368723  368759  368780  368763  60956527  60956527R  368653  368729  368734  60956526  60964123  60964124  60956526R  60964123R  368642  368700  368719  368752  368760  368786  368641  368649  368657  368671  368761  368762  368800  368647  368651  368656  368721  368724  368738  368710  368627  368672  368713  368713R  368628  368636  368628R  368665  368667  368782  60851372  60851372R  368708  368737  368790  368798  368608  368659  368789  60975078  60975081  60975102  60975081R  368736  368803  368736R  368803R  61003177  61003178  368743  368794  61003179  368658  368705  368757  368791  61003728  61009010  61009011  61009010R  368640  368688  368744  368771  368955  368956  61009012  368640R  368771R  61009012R  61008986  61008986R  368692  368732  61016568  368655  368715  368731  368773  368781  368805  61009014  61016587  61021142  368731R  61009014R  61009008  61009008R  368717  368727  368785  368799  368727R  368652  368726  368741  368742  368787  368793  368795  368804  368952  368954  368726R  368741R  61020169  61020169R  368722  61009009  61035126  368722R  61009009R  61035126R  368712  368783  61016589  61037701  61016589R  61037701R  61020173  61020173R  368777  368953  61009013  61016588  61016588R  368643  368774  368784  368796  368801  368806  61020172  61026005  61020172R  61037789  61040632  61040632R  61030963  61030964  61030963R  61030964R  61030960  61037697  61030960R  61037697R  61030959  61030962  61030962R  368668  61037793  61037795  61037768  61037769  61037768R  61037769R  61037740  61040594  61040595  61037740R  61040594R  61040595R  61037788  61040591  61040625  61037788R  61037763  61037763R  369117  61037797  61040586  61079840  61040586R  61037733  61040622  61076411  61037733R  61040622R  368380  61040596  61040596R  61076409  61076415  61076415R  61076412  61040628  61040628R  61040629  61076419  61040534  61076416  61076418  61076416R  61076418R  61111457  61111457R  61117693  61119027  61120862  61117693R  61119027R  61117694  61120859  61120859R  61120861  61120861R  61120860  61129136  61129136R  61120857  61120865  61128236  61137201  61120857R  61137201R  61117695  61120863  61129139  61137194  61137204  61137205  61137206  61137216  61137217  61137205R  61137216R  61137217R  61137195  61137202  61137207  61137215  61137202R  61137207R  61140838  61143406  61140839  61140840  61143402  61143402R  61160658  61160657  61160659  61160660  61160661  61160663  61160665  61160657R  61160663R  61160665R  61160664  61143404  61164636  61164701  61166694  61140837  61143405  61140837R  61143407  61166702  61166703  61172823  61172824  61172825  61178916  61166704  61166708  61172819  61178914  61178915  61172819R  61172821  61172821R  61185730  61185730R  61172826  61179444  61185813  61188443  61179444R  61184397  61193884  61184399  61187520  61187542  61187550  61185812  61187519  61187548  61187549  61198797  61191348  61198471  61198467  61198469  61198470  61198467R  61198469R  61191346  61198468  61063875  61063878  61191347  61210330  61214452  61206017  61210335  61210338  61210339  61210341  61221598  61224619  61206017R  61210335R  61228314  61198472  61221416  61234053  61234054  61221416R  61210331  61219281  61219286  61221415  61238840  61219286R  61221415R  61221410  61221410R  61219284  61229628  61233399  61233401  61248168  61248169  61251811  61219287  61210333  61210336  61219282  61247089  61210333R  61237544  61248166  61253369  61244974  61266597  61240131  61266610  61266615  61266625  61266615R  61237542  61242685  61244975  61253368  61266602  61229623  61233353  61266596  61271213  61271214  61271213R  61237543  61249762  61266601  61266601R  61285963  61229630  61242708  61256538  61040507  61229620  61229624  61249761  61256536  61272866  61229624R  61271206  61233371  61256537  61261768  61279145  61285956  61279145R  61229627  61258825  61279153  61285952  61285964  61290767  61290769  61266623  61219285  61229622  61258826  61279144  61279146  61279146R  61210337  61233391  61253370  61266626  61279150  61279151  61279154  61279155  61233391R  61271205  61233402  61279152  61295040  61285955  61285960  61285962  61272865  61295041  61295042  61303203  61303204  61303203R  61295037  61303201  61312956  61295045  61314562  61307043  61312953  61312955  61312954  61317568  61317569  61347111  61063877  61360005  61087376  61087383  61395148  61395148R  61428554  61428554R  369833  61440391  61468130  61468129  61492632  61475219  61279143  61492631  61492633  61516788  61589628  61597257  61597271  61585013  61597255  61611834  61622046  61622047  61597272  61622041  61622048  61635838  61634097  61635836  61635839  61639612  61622045  61639611  61759088  61759090  62137496  62148988  62158959  62259848  62259834  62266697  62285471  62294778  62294786  62294792  62299159  62284876  62280715  62285470  62330722  62330723  62330724  62333354  62355628  62355622  62355663  62390948  62355621  62355626  62390947  62345776  62355623  62355624  62355667  62390950  62394958  62390986  62390989  62390992  62390983  62390979  62406151  62406153  62413068  62413078  62413076  62390974  62413072  62413077  62421239  62421238  62427390  45002929  45003084  62430038  62430042  62430043  62430044  62430045  62437538  62437541  62430046  45002928  62437532  62437517  62437527  62437552  37108416  37108791  62446166  37108523  37108402  37108493  62446175  62453729  37108392  37108410  37108489  37108522  37108525  37108528  37108529  37108547  37108577  37108604  37108641  37108654  37108656  37108658  37108776  37108850  62430039  62435341  62437537  62446169  62453730  37108411  37108413  37108490  37108491  37108527  37108532  37108539  37108655  37108732  37108805  37108853  37108866  37108867  37108869  62430041  37108530  37108391  37108412  37108417  37108606  37108640  37108851  37108420  37108545  37108602  37108603  37108632  37108811  37108847  37108852  62450840  62450861  37108865  37108492  37108526  37108533  37108542  37108543  37108576  37108581  37108774  62437515  37108722  37108807  62446172  62446177  37108405  37108670  37108733  37108803  37108849  62446173  62446176  62450671  62450881  62459657  37108868  62446167  62446170  62450675  62459655  37108414  37108415  37108660  62432239  62453738  62472003  62471991  62471996  37108421  37108535  37108790  37108848  62354472  62468630  62468631  62472004  37108540  62471995  62471997  62474261  62474268  62446174  62482418  62474269  37108544  62485824  62486578  62421249  62474258  62474259  62474260  62490464  62490481  62490482  62490483  62490484  62490485  62490497  62490501  62472005  62490469  37108653  62472006  62511303  62520433  62524451  62524452  62531671  62531672  62539393  62531673  62539192  62539394  62539195  62539196  62539280  62554053  62539397  62554063  62539395  62539408  62437524  62554074  62559755  62559758  62562415  62562417  62562421  37213165  62520429  62562419  62482422  62554066  62559757  37212020  37212253  37212010  37212017  37212021  37212831  37212832  37211775  37211779  37211780  37212012  37212023  37213159  37212013  37212025  37213168  37218609  37218610  62565725  62565731  37211459  37211777  37211778  37211781  37212009  37212018  37212022  37212026  37212250  37212251  37212252  37212254  37212430  37212433  37212977  37212980  37212983  37213156  37213169  37213616  37213619  37213620  37214581  37214584  37214585  37214586  37214587  37214588  37214589  37214590  37214591  37214592  37214594  37214595  37214597  37214598  37214600  37214970  37214971  37215279  37215282  37215284  37215286  37215287  37215374  37215646  37215651  37215653  37215654  37215655  37215656  37215657  37215894  37215895  37215896  37216135  37216141  37216143  37216144  37216550  37216619  37216622  37216625  37216627  37216629  37216632  37216633  37216634  37216635  37216843  37216844  37216845  37217201  37217202  37217203  37217299  37217300  37218612  37218613  37218614  37218972  37218974  37218977  37218980  37218981  37218982  37218984  37218985  37218986  37218987  37218988  37218990  37218991  37218992  37218994  37218995  37218996  37219229  37219232  37219233  37219234  37219235  37219500  37219501  37108601  37211782  37212014  37213170  37214593  37215643  37216631  62565719  37213148  37213150  37213611  37213154  37213614  37213162  37217931  37217935  37212024  37213153  37213160  37213163  37213164  37215645  37215647  37215650  37216140  37216623  37217933  37217937  37218611  37218989  62602995  62606709  62606719  37212015  37217932  37213155  37213158  62577179  37212982  37213166  37213604  37216626  37216891  37217934  37220699  62606712  37212829  37213157  37215010  37215649  37220678  37220691  37211776  37212019  37212248  37212570  37213603  37213612  37213981  37216136  37217298  62611686  62616452  37212830  37216621  37220669  37220688  37220695  62616449  37213167  37213172  37213617  37216628  37213610  37213618  62602966  62616485  37220667  37220675  37220685  37220687  37220701  37216846  37220659  37220664  37220676  37220680  37220684  37220696  37220703  37220706  62616447  62616453  37217936  37220662  37220671  37220673  37220677  37220679  37220681  37220682  37220689  37220690  37220692  37220694  37220700  37220704  62611685  37220672  37212016  37212981  37213152  37213161  37215648  37216142  37216630  37220693  37220698  37220702  37220705  62554078  62628613  37212432  37215285  37220686  37213613  37213615  37214583  37213171  37213605  37213608  37213980  37214596  37214599  62616479  62616455  37218993  62616444  37215652  37217938  62654107  37109132  37109133  37109164  37216624  37109134  37109137  62734008  37109129  37109136  37109167  62734004  37109165  62353703  62955756  62955755  62959774  62955754  62955786  63035255  62968061  62959484  63242713"
    }
  ]
}