{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Batesville",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88548",
      "recalling_firm": "Hill-Rom, Inc.",
      "address_1": "1069 State Road 46 E",
      "address_2": "N/A",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL,  ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria,  Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan,  Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal,  Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand,  Turkey, United Arab Emirates.",
      "recall_number": "Z-2597-2021",
      "product_description": "OctoStretch with Stretch Leveller - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.  Model Number: 3156056",
      "product_quantity": "440 units",
      "reason_for_recall": "Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion.  When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.",
      "recall_initiation_date": "20210811",
      "center_classification_date": "20210930",
      "report_date": "20211006",
      "code_info": "N/A"
    }
  ]
}