  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77141",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: Puerto Rico, AK  AL  AR  AZ  CA  CO  CT  FL  GA  HI  IA  ID  IL  IN  KS  KY  LA  MA  MD  ME  MI  MN  MO  MS  MT  NC  ND  NE  NH  NJ  NM  NV  NY  OH  OK  OR  PA  RI  SC  SD  TN  TX  UT  VA  VT  WA  WI  WV  WY    Foreign: CANADA, ALGERIA                    ARGENTINA  AUSTRALIA                  BELGIUM  BRAZIL  BRIDGEND WALES  CHINA  DUBAI UAE                  EGYPT                      EL SALVADOR                ENGLAND  ESCHBACH GERMANY  EXPORT DORAL FL  FRANCE                     GERMANY  GUATEMALA                  HONDURAS  INDIA  ITALY  JAMAICA  JAPAN  JORDAN  KOREA  LEBANON                    DELIVERY SERVICE (LYNBROOK  NY)  MALAYSIA  MEXICO  MIAMI FL  MOROCCO  NETHERLANDS                NEW ZEALAND  NICARAGUA                  PANAMA  PERU  PHARR TEXAS  POLAND  PUERTO RICO  RIO DE JANEIRO  SAUDI ARABIA               SINGAPORE  SPAIN  SWITZERLAND  TAIWAN  THAILAND  THE NETHERLANDS            TRIPOLI  UK  VIRGIN ISLANDS  VA/DOD: Both see addresses below",
      "recall_number": "Z-2596-2017",
      "product_description": "FEM IM NAIL 15MMDX30CM,  FEM IM NAIL 16MMDX30CM  FEM IM NAIL 14MMDX32CM  FEM IM NAIL 15MMDX32CM  FEM IM NAIL 16MMDX32CM  FEM IM NAIL 14MMDX34CM  FEM IM NAIL 15MMDX34CM  FEM IM NAIL 16MMDX34CM  FEM IM NAIL 14MMDX36CM  FEM IM NAIL 15MMDX36CM  FEM IM NAIL 14MMDX38CM  FEM IM NAIL 15MMDX38CM  FEM IM NAIL 11MMDX40CM  FEM IM NAIL 13MMDX40CM  FEM IM NAIL 14MMDX40CM  FEM IM NAIL 15MMDX40CM  FEM IM NAIL 11MMDX42CM  FEM IM NAIL 12MMDX42CM  FEM IM NAIL 13MMDX42CM  FEM IM NAIL 14MMDX42CM  FEM IM NAIL 15MMDX42CM  FEM IM NAIL 11MMDX44CM  FEM IM NAIL 12MMDX44CM  FEM IM NAIL 13MMDX44CM  FEM IM NAIL 14MMDX44CM  FEM IM NAIL 10MMDX46CM  FEM IM NAIL 11MMDX46CM  FEM IM NAIL 12MMDX46CM  FEM IM NAIL 13MMDX46CM  FEM IM NAIL 14MMDX46CM  FEM IM NAIL 10MMDX48CM  FEM IM NAIL 11MMDX48CM  FEM IM NAIL 12MMDX48CM  FEM IM NAIL 13MMDX48CM  FEM IM NAIL 14MMDX48CM  FEM IM NAIL 13MMDX50CM  FEM IM NAIL 14MMDX50CM",
      "product_quantity": "27279",
      "reason_for_recall": "Zimmer Biomet is initiating a field action for sterile-packaged implants packaged in a single package configuration due to packaging design verification test failures.",
      "recall_initiation_date": "20170428",
      "center_classification_date": "20170615",
      "termination_date": "20190515",
      "report_date": "20170621",
      "code_info": "Item Numbers: 00225230015  00225230016  00225232014  00225232015  00225232016  00225234014  00225234015  00225234016  00225236014  00225236015  00225238014  00225238015  00225240011  00225240013  00225240014  00225240015  00225242011  00225242012  00225242013  00225242014  00225242015  00225244011  00225244012  00225244013  00225244014  00225246010  00225246011  00225246012  00225246013  00225246014  00225248010  00225248011  00225248012  00225248013  00225248014  00225250013  00225250014  Lot Numbers: 60684923  60680624  60680627  60680627R  60680623  60680623R  60680629  60680629R  60680615  60680628  60680615R  60680628R  60680625  60680619  60680619R  60691710  60691708  60691713  60691708R  60732923  60732923R  60691711  60732925  60745751  60691711R  60745751R  60732924  60747393  60747393R  367355  60753357  367355R  60753357R  60721966  60747408  60721966R  367348  60751917  60751917R  60753355  60753356  60758098  60753355R  60715104  60758099  60735584  60735585  60735585R  367352  60789431  60789432  60789431R  367354  60747405  60797046  60797046R  60789429  60789429R  60691709  60789426  60789428  60691709R  60807362  60807365  60807366  60807376  60807376R  60807359  60817729  60817730  60817731  60817733  60817730R  60817733R  367739  367739R  367737  60817725  60817727  367737R  60817727R  60822101  60822112  60822112R  60821407  60822113  60821407R  367749  367749R  60822105  60822106  60822107  60822106R  60822107R  60822102  60822102R  60821408  60827839  60827840  60827839R  60827840R  60836675  367751  60828364  367750  367752  60836673  367750R  367766  367767  367779  367766R  60836664  60836664R  60828363  60833614  60833960  60836676  60838315  60838317  60843576  60833614R  60838315R  60843576R  60833622  60843575  60827843  60833609  60836659  60836670  60836679  60836680  60838316  60827843R  60836659R  367753  60827844  60828362  60830568  60843574  60844203  60844204  60848502  367753R  60827844R  60844203R  60844204R  60848502R  60829849  60838314  60848496  60848497  60848504  60848497R  60836677  60848508  60854060  60836677R  60848508R  60854049  367781  60854056  60854057  60854056R  60854057R  60854050  60854061  60854061R  60854051  60854058  60854058R  367881  60854052  60854071  60854072  60859070  60859073  60854052R  60854071R  60854072R  60859070R  60859073R  367878  367878R  60844202  60854055  60874456  60854055R  60874456R  367882  367882R  60859067  60859067R  367877  368034  367983  368221  368224  368230  368341  368224R  368069  368069R  368223  60885947  60885949  60885947R  60885949R  60885951  60885951R  368231  60893397  367985  367985R  367984  368032  368033  368042  368208  368213  368225  368477  368225R  368041  368070  368206  367986  367986R  60919007  60919007R  60859072  60919010  60836674  60925352  60836674R  60925352R  368872  60931434  60931435  60931437  60931438  60937694  368854  368854R  368869  368869R  368843  368853  368900  368900R  368811  368839  368811R  368839R  368851  368859  368867  368873  368867R  368873R  368838  368846  368858  368846R  368832  368840  368852  368866  368868  368832R  368852R  368868R  368812  368830  368844  368847  368862  368865  368847R  368862R  368861  61018709  61023630  60984608  61023622  61023623  61023624  61023631  61023623R  369074  60984611  60984611R  369061  369071  369067  369072  61018710  61049120  61049121  61049122  369062  369070  61049135  61049136  61049119  61075885  61075885R  61069827  369066  368845  61069825  61085256  61085258  61085259  61069825R  61085255  61049654  61049654R  369182  369179  61114064  61117739  61117742  61114065  61114063  61114072  61114073  61120894  61133012  61092440  61147965  61147978  61114071  61147973  61147975  61114071R  61147970  61143357  61143358  61164612  61164612R  61117741  61147971  61147974  61147976  61147977  61151161  61151162  61159192  61159193  61164613  61147971R  61147977R  61151162R  61159192R  61175871  61175871R  61175872  61184353  61175872R  61178284  61178285  61184352  61178284R  61198489  61198489R  61191182  61191182R  369263  61191195  61191195R  61172881  61191183  61199929  61172884  61203580  61217316  61217316R  61220901  61220898  61198493  61211713  61220905  61217317  61199934  61243933  61243933R  61234879  61237551  61237551R  61234876  61199935  61243932  61244762  61263718  61244766  61240299  61240299R  61276454  61276456  61276454R  61276456R  61266210  61276463  61284073  61276463R  61284073R  61194288  61228310  61194287  61276459  61276459R  61184354  61234880  61272881  61284074  61276466  61285983  61294507  61306349  61294507R  61306349R  61244765  61244765R  61285980  61307908  61307911  61259381  61276464  61307910  61324284  61293416  61251809  61307898  61307909  61307905  61290156  61313655  61313655R  61328157  61120897  61293414  61317589  61328154  61328156  369461  61313666  61344231  61351675  61338969  61348460  61348460R  61220902  61338984  61338985  61344230  61351681  61357112  61359840  61359843  369530  369531  369540  61270623  369530R  61375269  61375271  61375270  61379626  61382372  61382373  61389798  61382369  61382374  61379625  61389799  61394801  61394801R  61389801  61199931  61420486  61413327  61413328  61428520  61413336  61428527  61428527R  61413333  61422220  61413335  61446643  61446662  61450097  61428529  61446647  61434765  61446638  61446638R  61428524  61434767  61452760  61452765  61434767R  61452761  61452764  61452763  61458940  369851  61468136  61476315  61476317  61479998  61476320  61484935  61484935R  61490741  61490742  61490855  61484895  61484932  61468140  61468142  61468143  61475202  61476318  61490739  369852  369853  61490730  61490751  61490751R  61490740  61508359  61509894  61490715  61502925  61516763  61533703  61516762  61542199  61542196  61556937  61556938  61562064  61565519  61542203  61562059  61578049  61597113  61602071  61614579  61614581  61602074  61618129  61626798  61626797  61672744  61675992  61674962  61675989  61676004  61693747  61693748  61693749  61693751  61676189  61700386  61700386R  61693750  61725785  61746155  61746152  61746153  61746159  61746154  61749004  61815321  61815331  61815330  61815334  61853654  61853657  61815335  61863108  61863164  61859296  61862960  61863110  61863125  61863171  370497  370498  61815316  61856241  61862961  61862962  61865655  61865656  61865661  61865662  61865663  61865664  61865667  61865657  61865659  61878888  61878932  61878933  61878934  61878935  61878936  61878937  61878938  61878939  61888765  61888773  61888766  61888770  61888777  61893681  61895643  61895646  61895647  61895648  370507  370511  61888775  61893682  61893683  61893684  61893686  61895644  61895645  61895649  61895650  61895651  61895652  61893685  61895654  61895655  61895656  61879902  61888769  61893687  61893689  61888778  61884646  61893688  61884647  370564  370587  370593  370597  370603  61916665  61916666  370520  61941385  61956892  61956893  61958356  61958357  61958359  61979103  61979105  61979108  61935565  61958360  61979106  61990774  61996887  62001460  61993647  370712  370713  62001458  62001461  62012318  62032467  370709  370714  62034074  62050747  62032466  62050732  62050733  62050734  62050735  62050740  62050741  62050743  62050744  62050745  62050746  62050748  62067631  62050739  62067625  62067626  62067634  62067636  62067642  62067650  62074148  62067647  62067651  62067653  62074114  62074142  62074145  62074146  62074147  62074150  370710  62050738  62074115  62074116  62074141  62074143  62074144  62074149  62074151  62050737  62067632  62083787  62083788  62083789  62083790  62083781  62083782  62083783  62083785  62083786  62083791  62083792  62083794  62091238  62091246  62091248  62091249  62091250  62091252  62083780  62083784  62091240  62091243  62091245  62091254  62093276  62093277  62093278  62093407  62093412  62106935  370711  62093416  62106930  62106931  62106934  62111423  62111422  62121736  62121737  62121738  62121739  62122603  62122604  62122606  62122609  62122611  62122613  62124388  62121740  62122596  62122599  62122607  62124387  62124389  62124390  62126624  62126633  370941  62126635  62126636  370942  62050742  62126634  62137462  62140753  62140756  62145209  62140087  62140754  62145210  62145211  62145213  62145214  62145212  62158355  62160907  370715  371032  62158340  62158343  62158359  62158360  62158362  62158364  62158341  62160894  370934  370936  370937  370938  370939  62093419  62112763  62177520  62177521  62177522  62177525  62177533  62177536  62177537  62177539  62177540  62177542  62091241  62170715  62199543  62204414  62204943  62204944  62205060  370940  62204420  62204939  62204940  62204945  62204946  62204947  62204948  62204949  371058  62204950  62204938  62204941  62204942  62204951  62204952  62205057  62205058  62205059  62142231  62204421  62204422  62204423  62204424  62204425  62204426  62204427  62215729  62215730  62204428  62213571  62213577  62215731  62215732  62215733  62215734  62215735  62216925  62216926  62216927  62215736  62216928  62216929  62216930  62216931  62219449  62216932  62216933  62216934  62219444  62219446  62224807  62219437  62224806  62225036  62225012  62225024  62225032  62225033  62225035  62225034  62225037  62228918  62228927  62225038  62228919  62228924  62228920  62228925  62228926  62228921  62228922  62228923  62234442  62234443  62234444  62234447  62234450  62234456  62234457  62234469  62234475  62234480  62225039  371057  62228917  371060  62250636  62250632  62250634  62250648  62250650  62250637  62268691  62268769  62268770  62250633  62265684  62268771  62268693  62299025  62305253  37107563  62312795  62311311  62366079  62377677  62384159  62384158  62395203  62395195  62395196  62395201  62395202  62403194  62403211  62423514  62429216  62434804  62440008  37108770  37108768  37108771  37108772  37108801  37108675  37108863  62434807  62440012  62466756  37108726  37108773  37108864  37108672  62434805  62466758  62466759  62497614  62497615  62466778  62497616  62497610  62497621  62514985  62516422  62529914  62497622  62466760  62516421  62529903  62524788  62536995  62554161  62554169  62554171  62554174  62554178  62554129  62554182  62559613  62559615  62554143  62559401  62539563  62559400  62559614  62561547  62561546  62561548  62566305  62571440  62571441  62571446  62571438  62571443  62576241  62577996  37211566  62574332  37211565  62584494  62576239  62583239  37211720  62584504  37212809  37212174  37214505  37215061  37215066  37215085  37216470  37218551  37218887  37219940  37211721  37212177  37212403  37213934  37213935  37213936  37214512  37214513  37215091  37215093  37215587  37215590  37216088  37216090  37217163  37217282  37217283  37218044  37218553  37218890  37219165  37219167  37219452  37219946  37219947  37215086  37216473  37217478  62584499  62601468  37213079  37216464  37216471  62598527  37215080  37213083  37215069  37215075  37216467  37216468  37217165  62598530  62598648  62598652  37213082  37215087  37215095  37217480  37215073  37216469  37215067  37215094  37215097  37220512  37211719  37215092  37220525  37212532  37213077  37215065  37215098  37217473  37212531  37220545  37211422  37212175  37215081  37217479  37220508  62621923  37218888  37215078  37220541  37220536  37220540  37220544  37220533  37220542  37220546  37215070  37215099  37220535  62578236  37215071  37215102  62642063  62647719  37219971  62647720  62658487  62661777  62658488  62661778  62679323  62683256  62685325  62685326  62687741  62687753  62690094  37219166  37109099  62728904  62740591  62603174  62727947  62735835  62742189  37109103  37109105  62742159  37109143  62603168  37109266  62802836  62793632  37109252  37109269  37109286  37109287  37109260  37109288  37109289  37109278  37109303  37109312  37109251  37109272  37109273  37109294  62834710  37109311  37109267  37109270  37109284  37109299  37109282  37109309  37109257  37109308  62946922  62946923  62946924  62946925  62963441  37109268  37109298  62834712  63204340  63204336  63204337  63204339  63212103  63212102  63242705  63288977  63291297  63291294  63304969  63304975  63317351  63317352  63324827  63347132  63291296  63304970  63342280  63398625  63317350  63476933  63383475  63503532  63503546  63383473  63537645  63540733  63551241  63583722"
    }
  ]
}