  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66448",
      "recalling_firm": "Hospira Inc.",
      "address_1": "275 N Field Dr",
      "address_2": "N/A",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "*** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa",
      "recall_number": "Z-2593-2016",
      "product_description": "Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL)  Canister, ATS, 2000, with Valve     1) List Numbers 43449-01 and 43449-11; LIST NO. 43449 RECEPTAL CANISTER; CANISTER FOR REUSE, THROW AWAY LINER ONLY; DO NOT FILL 900 ABOVE LINE  2) List Numbers 43423-01 and 43423-21; LIST NO. 43423; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; 1400.   3) List Numbers 43445-01 and 43445-11; LIST NO. 43445; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; DO NOT FILL 1900 ABOVE THIS LINE.  4) List Number 43496-01.",
      "product_quantity": "67,410,861 liners and 67,410,861 lids/canisters in total",
      "reason_for_recall": "Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.",
      "recall_initiation_date": "20130910",
      "center_classification_date": "20160818",
      "termination_date": "20161215",
      "report_date": "20160824",
      "code_info": "43423-04-01  43423-04-11  43423-04-21  43445-04-01  43445-04-11  43449-04-01  43449-04-11  43496-04-01"
    }
  ]
}