  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Komorniki",
      "state": "N/A",
      "country": "Poland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88469",
      "recalling_firm": "ARJOHUNTLEIGH POLSKA Sp. z.o.o.",
      "address_1": "Ul. Ks. Piotra Wawrzyniaka 2",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, FL, IL, KY, MD, NV, NY, PA, and SD.",
      "recall_number": "Z-2588-2021",
      "product_description": "Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo modules (retrofit kits) that can be installed on Citadel (model no. INDIDAA) or Enterprise (model no. INDIJAA) beds manufactured without IndiGo modules assembled at the time of manufacturing and subsequently retrofitted on the beds.",
      "product_quantity": "367 units",
      "reason_for_recall": "The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.",
      "recall_initiation_date": "20210610",
      "center_classification_date": "20210929",
      "termination_date": "20260203",
      "report_date": "20211006",
      "code_info": "Manufacturing dates 5/25/2018-2/15/2021 (printed on both the bed and IndiGo module if the module was assembled on the bed at the time of manufacture, or printed on the module if it was later installed on the bed."
    }
  ]
}